CoChat

Toward a Run-to-Run Adaptive Artificial Pancreas: In Silico Results

Toffanin, Chiara, Visentin, Roberto, Messori, Mirko, Palma, Federico Di, Magni, Lalo, Cobelli, Claudio · 2018 · IEEE Transactions on Biomedical Engineering
Cited by 107

The DOI matches its Crossref record; 12 of the 12 references it cites were found in a registry or an index.

Abstract

OBJECTIVE: Contemporary and future outpatient long-term artificial pancreas (AP) studies need to cope with the well-known large intra- and interday glucose variability occurring in type 1 diabetic (T1D) subjects. Here, we propose an adaptive model predictive control (MPC) strategy to account for it and test it in silico. METHODS: A run-to-run (R2R) approach adapts the subcutaneous basal insulin delivery during the night and the carbohydrate-to-insulin ratio (CR) during the day, based on some performance indices calculated from subcutaneous continuous glucose sensor data. In particular, R2R aims, first, to reduce the percentage of time in hypoglycemia and, secondarily, to improve the percentage of time in euglycemia and average glucose. In silico simulations are performed by using the University of Virginia/Padova T1D simulator enriched by incorporating three novel features: intra- and interday variability of insulin sensitivity, different distributions of CR at breakfast, lunch, and dinner, and dawn phenomenon. RESULTS: After about two months, using the R2R approach with a scenario characterized by a random 30% variation of the nominal insulin sensitivity the time in range and the time in tight range are increased by 11.39% and 44.87%, respectively, and the time spent above 180 mg/dl is reduced by 48.74%. CONCLUSIONS: An adaptive MPC algorithm based on R2R shows in silico great potential to capture intra- and interday glucose variability by improving both overnight and postprandial glucose control without increasing hypoglycemia. SIGNIFICANCE: Making an AP adaptive is key for long-term real-life outpatient studies. These good in silico results are very encouraging and worth testing in vivo.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1109/tbme.2017.2652062

Cites

12 of 12 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 12 references

Cited by

See all 107

Metadata

Authors
Toffanin, C., Visentin, R., Messori, M., Palma, F. D., Magni, L., & Cobelli, C.
Venue
IEEE Transactions on Biomedical Engineering
Published
2018-03
Volume
65
Issue
3
Pages
479-488
ISSN
  • 0018-9294 (ISSN-L)
  • 1558-2531
OpenAlex
W2572287943
Cited by
107
References
12
FWCI
11.88
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Toward a Run-to-Run Adaptive Artificial Pancreas: In Silico Results. https://app.cochat.ai/papers/10.1109/tbme.2017.2652062
The paper
Toffanin, C., Visentin, R., Messori, M., Palma, F. D., Magni, L., & Cobelli, C. (2018). Toward a Run-to-Run Adaptive Artificial Pancreas: In Silico Results. IEEE Transactions on Biomedical Engineering. https://doi.org/10.1109/tbme.2017.2652062

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.