Design and In Vitro Interference Test of Microwave Noninvasive Blood Glucose Monitoring Sensor
Open accessAbstract
A design of a microwave noninvasive continuous blood glucose monitoring sensor and its interference test results are presented. The novelty of the proposed sensor is that it comprises two spatially separated split-ring resonators, where one interacts with the change in glucose level of a sample under test while the other ring is used as a reference. The reference ring has a slightly different resonant frequency and is desensitized to the sample owing to its location, thus allowing changes in temperature to be calibrated out. From an oral glucose tolerance test with two additional commercially available sensors (blood strip and continuous glucose monitor) in parallel, we obtained encouraging performance for our sensor comparable with those of the commercial sensors. The effects of endogenous interferents common to all subjects, i.e., common sugars, vitamins (ascorbic acid), and metabolites (uric acid) have also been investigated by using a large Franz cell assembly. From the interference test, it is shown that the change in sensor response is dominated by changes in glucose level for concentrations relevant to blood, and the effects of interferents are negligible in comparison.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Wild, S., Roglic, G., Green, A., Sicree, R., & King, H. (2004) · Diabetes Care · doi:10.2337/diacare.27.5.1047 · Cited by 15,297
Gabriel, S., Lau, R. W., & Gabriel, C. (1996) · Physics in Medicine and Biology · doi:10.1088/0031-9155/41/11/002 · Cited by 4,270
Percutaneous Absorption. On the Relevance of in Vitro Data
Franz, T. J. (1975) · Journal of Investigative Dermatology · doi:10.1111/1523-1747.ep12533356 · Cited by 1,132
A 3-$\mu\hbox{W}$ CMOS Glucose Sensor for Wireless Contact-Lens Tear Glucose Monitoring
Liao, Y.-T., Yao, H., Lingley, A., Parviz, B., & Otis, B. P. (2012) · IEEE Journal of Solid-State Circuits · doi:10.1109/jssc.2011.2170633 · Cited by 372
Factors affecting blood glucose monitoring: sources of errors in measurement
Ginsberg, B. H. (2009) · Journal of Diabetes Science and Technology · doi:10.1177/193229680900300438 · Cited by 348
Spectroscopic and Clinical Aspects of Noninvasive Glucose Measurements
Khalil, O. S. (1999) · Clinical Chemistry · doi:10.1093/clinchem/45.2.165 · Cited by 331
Microwave-Based Noninvasive Concentration Measurements for Biomedical Applications
Hofmann, M., Fischer, G., Weigel, R., & Kissinger, D. (2013) · IEEE Transactions on Microwave Theory and Techniques · doi:10.1109/tmtt.2013.2250516 · Cited by 288
Integrated Broadband Microwave and Microfluidic Sensor Dedicated to Bioengineering
Grenier, K., Dubuc, D., Poleni, P.-E., Kumemura, M., Toshiyoshi, H., Fujii, T., & Fujita, H. (2009) · IEEE Transactions on Microwave Theory and Techniques · doi:10.1109/tmtt.2009.2034226 · Cited by 278
Continuous Glucose Monitoring Systems: A Review
Vashist, S. (2013) · Diagnostics · doi:10.3390/diagnostics3040385 · Cited by 261
Transdermal monitoring of glucose and other analytes using ultrasound
Kost, J., Mitragotri, S., Gabbay, R. A., Pishko, M., & Langer, R. (2000) · Nature Medicine · doi:10.1038/73213 · Cited by 253
First human experiments with a novel non-invasive, non-optical continuous glucose monitoring system
Caduff, A., Hirt, E., Feldman, Y., Ali, Z., & Heinemann, L. (2003) · Biosensors and Bioelectronics · doi:10.1016/s0956-5663(03)00196-9 · Cited by 247
Noninvasive Blood Glucose Monitoring
Klonoff, D. C. (1997) · Diabetes Care · doi:10.2337/diacare.20.3.433 · Cited by 208
Microwave Dielectric Behavior of Wet Soils
Behari, J. (2005) · Remote sensing and digital image processing · doi:10.1007/1-4020-3288-9 · Cited by 175
Modulated insulin delivery from glucose-sensitive hydrogel dosage forms
Kim, J. J., & Park, K. (2001) · Journal of Controlled Release · doi:10.1016/s0168-3659(01)00447-3 · Cited by 174
Yilmaz, T., Foster, R., & Hao, Y. (2014) · IEEE Transactions on Antennas and Propagation · doi:10.1109/tap.2014.2313139 · Cited by 168
A microwave frequency sensor for non-invasive blood-glucose measurement
Jean, B. R., Green, E. C., & McClung, M. J. (2008) · IEEE Sensors Applications Symposium (SAS) · doi:10.1109/sas.2008.4472932 · Cited by 133
Improved Split-Ring Resonator for Microfluidic Sensing
Rowe, D. J., al-Malki, S., Abduljabar, A. A., Porch, A., Barrow, D. A., & Allender, C. J. (2014) · IEEE Transactions on Microwave Theory and Techniques · doi:10.1109/tmtt.2014.2299514 · Cited by 127
Self-citationContinuous Noninvasive Glucose Monitoring Technology Based on “Occlusion Spectroscopy”
Amir, O., Weinstein, D., Zilberman, S., Less, M., Perl‐Treves, D., Primack, H., Weinstein, A., Gabis, E., Fikhte, B. A., & Karasik, A. (2007) · Journal of Diabetes Science and Technology · doi:10.1177/193229680700100403 · Cited by 125
The correlation of the complex dielectric constant and blood glucose at low frequency
Park, J.-H., Kim, C.-S., Choi, B. C., & Ham, K.-K. (2003) · Biosensors and Bioelectronics · doi:10.1016/s0956-5663(03)00188-x · Cited by 94
Millimeter-wave non-invasive monitoring of glucose in anesthetized rats
Siegel, P. H., Lee, Y., & Pikov, V. (2014) · International Conference on Infrared, Millimeter, and Terahertz Waves (IRMMW-THz) · doi:10.1109/irmmw-thz.2014.6956294 · Cited by 62
Choi, H., Nylon, J., Luzio, S. D., Beutler, J., & Porch, A. (2014) · IEEE MTT-S International Microwave Workshop Series on RF and Wireless Technologies for Biomedical and Healthcare Applications · doi:10.1109/imws-bio.2014.7032398 · Cited by 43
Self-citation
Research works printed but not linked: 12
- klein, 1983, Light Scattering in Solids I
- schoemaker, 2014, CGM sensor design principles for reliable and accurate glucose monitoring in the subcutaneous tissue, 7th Int Adv Technol Treatments for Diabetes Conf
- 10.1109/MeMeA.2011.5966704 · DOI
- chowdhury, 2014, Comparative performance between the former Glucowatch® and an improved non-invasive continuous glucose monitoring watch (CGM Watch™), 7th Int Adv Technol Treatments for Diabetes Conf
- dehennis, 2014, An integrated wireless fluorimeter for a long term implantable, continuous glucose monitoring system, 7th Int Adv Technol Treatments for Diabetes Conf
- gal, 2014, Calibration schemes of a truly non-invasive glucose monitor for variety of diabetics, 7th Int Ad Technol Treatments for Diabetes Conf
- 2015, 2014 annual report
- 10.1109/TMTT.1984.1132619 · doi:10.1109/tmtt.1984.1132619
- 10.1109/TMTT.1983.1131661 · doi:10.1109/tmtt.1983.1131661
- 10.1109/TMTT.2014.2300066 · doi:10.1109/tmtt.2014.2300066
- 10.1007/0-306-48533-8_2 · doi:10.1007/0-306-48533-8_2
- 10.1063/1.4809823 · doi:10.1063/1.4809823
Cited by
Villena Gonzales, W., Mobashsher, A., & Abbosh, A. (2019) · Sensors · doi:10.3390/s19040800 · Cited by 772
Non-Invasive Blood Glucose Monitoring Technology: A Review
Tang, L., Chang, S. J., Chen, C.-J., & Liu, J.-T. (2020) · Sensors · doi:10.3390/s20236925 · Cited by 481
Flexible Wearable Sensors for Cardiovascular Health Monitoring
Chen, S., Qi, J., Fan, S., Qiao, Z., Yeo, J. C., & Lim, C. T. (2021) · Advanced Healthcare Materials · doi:10.1002/adhm.202100116 · Cited by 453
Strongly Enhanced Sensitivity in Planar Microwave Sensors Based on Metamaterial Coupling
Abdolrazzaghi, M., Daneshmand, M., & Iyer, A. K. (2018) · IEEE Transactions on Microwave Theory and Techniques · doi:10.1109/tmtt.2018.2791942 · Cited by 353
Sacks, D. B., Arnold, M., Bakris, G. L., Bruns, D. E., Horvath, A. R., Lernmark, Å., Metzger, B. E., Nathan, D. M., & Kirkman, M. S. (2023) · Diabetes Care · doi:10.2337/dci23-0036 · Cited by 351
Metadata
- Authors
- Choi, H., Naylon, J., Luzio, S., Beutler, J., Birchall, J., Martin, C., & Porch, A.
- Venue
- IEEE Transactions on Microwave Theory and Techniques
- Published
- 2015-10
- Volume
- 63
- Issue
- 10
- Pages
- 3016-3025
- ISSN
- 0018-9480 (ISSN-L)
- 1557-9670
- OpenAlex
- W1654139916
- Cited by
- 312
- References
- 21
- FWCI
- 8.66
- Type
- journal article
- Language
- en
- Open access
- hybrid
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.