Micron: An Actively Stabilized Handheld Tool for Microsurgery

Open access
MacLachlan, Robert A., Becker, Brian C., Tabares, Jaime Cuevas, Podnar, Gregg W., Lobes, Louis A., Riviere, Cameron N.·2012·IEEE Transactions on Robotics

Abstract

We describe the design and performance of a hand-held actively stabilized tool to increase accuracy in micro-surgery or other precision manipulation. It removes involuntary motion such as tremor by actuating the tip to counteract the effect of the undesired handle motion. The key components are a three-degree-of-freedom piezoelectric manipulator that has 400 μm range of motion, 1 N force capability, and bandwidth over 100 Hz, and an optical position measurement subsystem that acquires the tool pose with 4 μm resolution at 2000 samples/s. A control system using these components attenuates hand motion by at least 15 dB (a fivefold reduction). By considering the effect of the frequency response of Micron on the human visual feedback loop, we have developed a filter that reduces unintentional motion, yet preserves intuitive eye-hand coordination. We evaluated the effectiveness of Micron by measuring the accuracy of the human/machine system in three simple manipulation tasks. Handheld testing by three eye surgeons and three non-surgeons showed a reduction in position error of between 32% and 52%, depending on the error metric.

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1109/tro.2011.2169634

Cites

60 of 60 references verified0 citation errors · 0 not checkable

Statistics

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

  • Unintentional tool motion and excessive forces during membrane peeling can cause vision loss or irreversible retinal damage.

    “unintentional tool motion and excessive forces can easily give rise to vision loss or irreversible damage to the retina”

    From Preliminary evaluation of a micro-force sensing handheld robot for vitreoretinal surgery (2012) · cites this paper as [5]

    The Introduction’s discussion of microsurgical manipulation aids says that forces too small for a surgeon to feel can still create damaging pressure, and the Introduction describes involuntary tool motion as a source of positioning error. The cited passages do not establish that these factors during membrane peeling can easily cause vision loss or irreversible retinal damage. (LLM, conf=0.84)

    Experimental — accuracy not yet measured

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Cites

Flagged first, then by citations

Cited by

Sorted by citations

Authors
MacLachlan, R. A., Becker, B. C., Tabares, J. C., Podnar, G. W., Lobes, L. A., & Riviere, C. N.
Affiliations
  • Carnegie Mellon University (US)
  • Universidad de Valladolid (ES)
  • University of Pittsburgh (US)
Venue
IEEE Transactions on Robotics
Published
2012-02
Volume
28
Issue
1
Pages
195-212
ISSN
  • 1552-3098 (ISSN-L)
  • 1941-0468
OpenAlex
W2087665419
Cited by
270
References
60
Citations by year
2026: 16 · 2025: 14 · 2024: 12 · 2023: 12 · 2022: 16 · 2021: 22 · 2020: 22 · 2019: 16 · 2018: 23 · 2017: 22 · 2016: 16 · 2015: 26 · 2014: 17 · 2013: 16 · 2012: 16
FWCI
28.89
Type
journal article
Language
en
Open access
Locations
  • IEEE Transactions on Robotics (publishedVersion)
  • PubMed (publishedVersion)
  • Europe PMC (PubMed Central) (acceptedVersion)
  • PubMed Central (acceptedVersion)
Funders
  • National Eye Institute (R21EY016359)
  • National Institute of Biomedical Imaging and Bioengineering (R01EB000526, R01EB007969)
Topics
  • Physical Sciences › Engineering › Mechanical Engineering › Teleoperation and Haptic Systems
  • Physical Sciences › Engineering › Control and Systems Engineering › Piezoelectric Actuators and Control
  • Life Sciences › Neuroscience › Cognitive Neuroscience › Motor Control and Adaptation
Keywords
microsurgery, motion stabilization, tremor compensation, eye-hand coordination, visual feedback, precision manipulation, surgical accuracy

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review