Early Complications After Dual Chamber Versus Single Chamber Pacemaker Implantation
The DOI matches its Crossref record; 1 of the 1 references it cites were found in a registry or an index.
Abstract
This study was performed to compare the frequency of early complications after single chamber versus dual chamber permanent pacemaker implantation. Early complication was defined as one occurring in the 6‐week period following implantation. We prospectively analyzed consecutive pacemaker implantation from January 1987 to June 1993 at our regional center. All complications were also analyzed for the relationship to operator experience, the venous access route, and the presence of temporary pacing wire at the time of implantation of the permanent pacing system. A total of 2019 new pacemaker units were implanted during this period. 1733 patients (85.8%) received a VVI pacemaker and 286 (14.2%) a DDD unit. Wound infection occurredin 11 (0.6%) VVI patients and 6 (2.1%) DDD patients. Lead displacement occurred in 18(1 %) VVI patients and 15 (5.2%) DDD patients (11 [3.8%] atrial and 4 [1.4 %] ventricular). There were 10 (0.6%)pneumothoraces, 9 (0.5%) hematomas requiring drainage, 1 (0.06%) chylocele, and 2 (0.1%) deaths in the VVI group. There were 2 (0.7%) pneumothoraces, 2 (0.7%) hematomas, and no deaths in the DDD group. There was no significant increase in complications for experienced infrequent implanters (< 12 systems per year). In both groups the subclavian approach was associated with a risk of pneumothorax when compared to the cephalic approach. The rate of wound infection was higher in patients who had a temporary pacing wire in place. The use of prophylactic antibiotics does not appear to affect the incidence of wound infection. The early complications in the DDD group were higher than in the VVI group (8.7% vs 2.9%, P < 0.05), being mainly due to an increased incidence of wound infection and atrial lead displacement.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Permanent pacemaker infections: Characterization and management
Choo, M. H., Holmes, D. R., Gersh, B. J., Maloney, J. D., Merideth, J., Pluth, J. R., & Trusty, J. (1981) · The American Journal of Cardiology · doi:10.1016/0002-9149(81)90088-6 · Cited by 177
Research works printed but not linked: 7
- 10.1111/j.1540-8159.1982.tb02281.x · doi:10.1111/j.1540-8159.1982.tb02281.x
- Scheibelhofer W, 1981, Complications of physiologic pacing, PACE, 40, A
- 10.1111/j.1540-8159.1990.tb02095.x · doi:10.1111/j.1540-8159.1990.tb02095.x
- 10.1016/S0003-4975(10)62705-1 · doi:10.1016/s0003-4975(10)62705-1
- 10.1111/j.1540-8159.1986.tb06600.x · doi:10.1111/j.1540-8159.1986.tb06600.x
- 10.1056/NEJM198505303122205 · doi:10.1056/nejm198505303122205
- Winner SJ, 1989, Hazzards associated with the use of temporary pacemakers before permanent pacing, Br Heart J, 61, 96P
Cited by
Kusumoto, F. M., Schoenfeld, M. H., Barrett, C., Edgerton, J. R., Ellenbogen, K. A., Gold, M. R., Goldschlager, N. F., Hamilton, R. M., Joglar, J. A., Kim, R. J., et al. (2019) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2018.10.044 · Cited by 1,008
Klug, D., Balde, M., Pavin, D., Hidden-Lucet, F., Clementy, J., Sadoul, N., Rey, J. L., Lande, G., Lazarus, A., Victor, J., et al. (2007) · Circulation · doi:10.1161/circulationaha.106.678664 · Cited by 700
Kusumoto, F. M., Schoenfeld, M. H., Barrett, C., Edgerton, J. R., Ellenbogen, K. A., Gold, M. R., Goldschlager, N. F., Hamilton, R. M., Joglar, J. A., Kim, R. J., et al. (2019) · Circulation · doi:10.1161/cir.0000000000000628 · Cited by 603
Polyzos, K. A., Konstantelias, A. A., & Falagas, M. E. (2015) · EP Europace · doi:10.1093/europace/euv053 · Cited by 493
Sandoe, J. A. T., Barlow, G., Chambers, J. B., Gammage, M., Guleri, A., Howard, P., Olson, E., Perry, J. D., Prendergast, B. D., Spry, M. J., et al. (2015) · Journal of Antimicrobial Chemotherapy · doi:10.1093/jac/dku383 · Cited by 403
Sandau, K. E., Funk, M., Auerbach, A., Barsness, G. W., Blum, K., Cvach, M., Lampert, R., May, J. L., McDaniel, G. M., Perez, M. V., et al. (2017) · Circulation · doi:10.1161/cir.0000000000000527 · Cited by 342
Wilkoff, B. L., Bello, D., Taborsky, M., Vymazal, J., Kanal, E., Heuer, H., Hecking, K., Johnson, W. B., Young, W., Ramza, B., et al. (2011) · Heart Rhythm · doi:10.1016/j.hrthm.2010.10.002 · Cited by 266
Complications Related to Permanent Pacemaker Therapy
KIVINIEMI, M. S., PIRNES, M. A., ERÄNEN, H. J. K., KETTUNEN, R. V. J., & HARTIKAINEN, J. E. K. (1999) · Pacing and Clinical Electrophysiology · doi:10.1111/j.1540-8159.1999.tb00534.x · Cited by 261
Aggarwal, R. K., Connelly, D. T., Ray, S. G., Ball, J., & Charles, R. G. (1995) · Heart · doi:10.1136/hrt.73.6.571 · Cited by 254
Pocket Hematoma After Pacemaker or Implantable Cardioverter Defibrillator Surgery
Wiegand, U. K. H., LeJeune, D., Boguschewski, F., Bonnemeier, H., Eberhardt, F., Schunkert, H., & Bode, F. (2004) · Chest · doi:10.1378/chest.126.4.1177 · Cited by 223
Complications arising after implantation of DDD pacemakers: the MOST experience
Ellenbogen, K. A., Hellkamp, A. S., Wilkoff, B. L., Camunãs, J. L., Love, J. C., Hadjis, T. A., Lee, K. L., & Lamas, G. A. (2003) · The American Journal of Cardiology · doi:10.1016/s0002-9149(03)00844-0 · Cited by 210
NERY, P. B., FERNANDES, R., NAIR, G. M., SUMNER, G. L., RIBAS, C. S., MENON, S. M. D., WANG, X., KRAHN, A. D., MORILLO, C. A., CONNOLLY, S. J., et al. (2010) · Journal of Cardiovascular Electrophysiology · doi:10.1111/j.1540-8167.2009.01690.x · Cited by 188
Late complications following permanent pacemaker implantation or elective unit replacement
Harcombe, A. A., Newell, S. A., Ludman, P. F., Wistow, T. E., Sharples, L. D., Schofield, P. M., Stone, D. L., Shapiro, L. M., Cole, T., & Petch, M. C. (1998) · Heart · doi:10.1136/hrt.80.3.240 · Cited by 163
“Further studies from our unit and others examined early complication rates for single versus dual chamber systems.”
BackgroundSelf-citationKusumoto, F. M., Schoenfeld, M. H., Barrett, C., Edgerton, J. R., Ellenbogen, K. A., Gold, M. R., Goldschlager, N. F., Hamilton, R. M., Joglar, J. A., Kim, R. J., et al. (2019) · Heart Rhythm · doi:10.1016/j.hrthm.2018.10.037 · Cited by 131
Dretzke, J., Toff, W. D., Lip, G. Y., Raftery, J., Fry-Smith, A., & Taylor, R. S. (2004) · Cochrane Database of Systematic Reviews · doi:10.1002/14651858.cd003710.pub2 · Cited by 89
Metadata
- Authors
- CHAUHAN, A., GRACE, A. A., NEWELL, S. A., STONE, D. L., SHAPIRO, L. M., SCHOFIELD, P. M., & PETCH, M. C.
- Venue
- Pacing and Clinical Electrophysiology
- Publisher
- Wiley
- Published
- 1994-11
- Volume
- 17
- Issue
- 11
- Pages
- 2012-2015
- ISSN
- 0147-8389 (ISSN-L)
- 1540-8159
- OpenAlex
- W1989630240
- Cited by
- 217
- References
- 1
- FWCI
- 2.40
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.