Cone-beam computed tomography with a flat-panel imager: Magnitude and effects of x-ray scatter
Open accessAbstract
A system for cone-beam computed tomography (CBCT) based on a flat-panel imager (FPI) is used to examine the magnitude and effects of x-ray scatter in FPI-CBCT volume reconstructions. The system is being developed for application in image-guided therapies and has previously demonstrated spatial resolution and soft-tissue visibility comparable or superior to a conventional CT scanner under conditions of low x-ray scatter. For larger objects consistent with imaging of human anatomy (e.g., the pelvis) and for increased cone angle (i.e., larger volumetric reconstructions), however, the effects of x-ray scatter become significant. The magnitude of x-ray scatter with which the FPI-CBCT system must contend is quantified in terms of the scatter-to-primary energy fluence ratio (SPR) and scatter intensity profiles in the detector plane, each measured as a function of object size and cone angle. For large objects and cone angles (e.g., a pelvis imaged with a cone angle of 6 degrees), SPR in excess of 100% is observed. Associated with such levels of x-ray scatter are cup and streak artifacts as well as reduced accuracy in reconstruction values, quantified herein across a range of SPR consistent with the clinical setting. The effect of x-ray scatter on the contrast, noise, and contrast-to-noise ratio (CNR) in FPI-CBCT reconstructions was measured as a function of SPR and compared to predictions of a simple analytical model. The results quantify the degree to which elevated SPR degrades the CNR. For example, FPI-CBCT images of a breast-equivalent insert in water were degraded in CNR by nearly a factor of 2 for SPR ranging from approximately 2% to 120%. The analytical model for CNR provides a quantitative understanding of the relationship between CNR, dose, and spatial resolution and allows knowledgeable selection of the acquisition and reconstruction parameters that, for a given SPR, are required to restore the CNR to values achieved under conditions of low x-ray scatter. For example, for SPR = 100%, the CNR in FPI-CBCT images can be fully restored by: (1) increasing the dose by a factor of 4 (at full spatial resolution); (2) increasing dose and slice thickness by a factor of 2; or (3) increasing slice thickness by a factor of 4 (with no increase in dose). Other reconstruction parameters, such as transaxial resolution length and reconstruction filter, can be similarly adjusted to achieve CNR equal to that obtained in the scatter-free case.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Feldkamp, L. A., Davis, L. C., & Kress, J. W. (1984) · Journal of the Optical Society of America A · doi:10.1364/josaa.1.000612 · Cited by 6,319
Photon cross sections from 1 keV to 100 MeV for elements Z=1 to Z=100
Storm, L., & Israel, H. I. (1970) · Atomic Data and Nuclear Data Tables · doi:10.1016/s0092-640x(70)80017-1 · Cited by 1,539
Siewerdsen, J. H., & Jaffray, D. A. (2000) · Medical Physics · doi:10.1118/1.1286590 · Cited by 216
Self-citationThe effects of scatter in x‐ray computed tomography
Joseph, P. M., & Spital, R. D. (1982) · Medical Physics · doi:10.1118/1.595111 · Cited by 211
Grids or air gaps for scatter reduction in digital radiography: A model calculation
Neitzel, U. (1992) · Medical Physics · doi:10.1118/1.596836 · Cited by 206
Compton scatter effects in CT reconstructions
Glover, G. H. (1982) · Medical Physics · doi:10.1118/1.595197 · Cited by 167
Cone-beam tomography: recent advances and a tutorial review
Smith, B. D. (1990) · Optical Engineering · doi:10.1117/12.55621 · Cited by 157
Efficient object scatter correction algorithm for third and fourth generation CT scanners
Ohnesorge, B. M., Flohr, T. G., & Klingenbeck-Regn, K. W. (1999) · European Radiology · doi:10.1007/s003300050710 · Cited by 148
Scattered radiation in fan beam imaging systems
Johns, P. C., & Yaffe, M. J. (1982) · Medical Physics · doi:10.1118/1.595076 · Cited by 121
Statistical limitations in transaxial tomography
Barrett, H. H., Gordon, S. K., & Hershel, R. S. (1976) · Computers in Biology and Medicine · doi:10.1016/0010-4825(76)90068-8 · Cited by 110
Feasible cone beam scanning methods for exact reconstruction in three-dimensional tomography
Kudo, H., & Saito, T. (1990) · Journal of the Optical Society of America A · doi:10.1364/josaa.7.002169 · Cited by 50
Overview of reconstruction algorithms for exact cone-beam tomography
Clackdoyle, R., & Defrise, M. (1994) · Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE · doi:10.1117/12.179253 · Cited by 35
Cited by
Flat-panel cone-beam computed tomography for image-guided radiation therapy
Jaffray, D. A., Siewerdsen, J. H., Wong, J. W., & Martinez, A. A. (2002) · International Journal of Radiation Oncology*Biology*Physics · doi:10.1016/s0360-3016(02)02884-5 · Cited by 1,330
Self-citationImage guidance for precise conformal radiotherapy
Mackie, T. R., Kapatoes, J., Ruchala, K., Lu, W., Wu, C., Olivera, G., Forrest, L., Tome, W., Welsh, J., Jeraj, R., et al. (2003) · International Journal of Radiation Oncology*Biology*Physics · doi:10.1016/s0360-3016(03)00090-7 · Cited by 484
Radiation dose reduction in computed tomography: techniques and future perspective
Yu, L., Liu, X., Leng, S., Kofler, J. M., Ramirez-Giraldo, J. C., Qu, M., Christner, J., Fletcher, J. G., & McCollough, C. H. (2009) · Imaging in Medicine · doi:10.2217/iim.09.5 · Cited by 395
Conebeam CT of the Head and Neck, Part 1: Physical Principles
Miracle, A. C., & Mukherji, S. K. (2009) · American Journal of Neuroradiology · doi:10.3174/ajnr.a1653 · Cited by 387
Bissonnette, J., Balter, P. A., Dong, L., Langen, K. M., Lovelock, D. M., Miften, M., Moseley, D. J., Pouliot, J., Sonke, J., & Yoo, S. (2012) · Medical Physics · doi:10.1118/1.3690466 · Cited by 359
Metadata
- Authors
- Siewerdsen, J. H., & Jaffray, D. A.
- Venue
- Medical Physics
- Publisher
- Wiley
- Published
- 2001-02
- Volume
- 28
- Issue
- 2
- Pages
- 220-231
- ISSN
- 0094-2405 (ISSN-L)
- OpenAlex
- W2043193153
- Cited by
- 612
- References
- 12
- FWCI
- 7.09
- Type
- journal article
- Language
- en
- Open access
- bronze
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.