NEMA NU 2-2012 performance studies for the SiPM-based ToF-PET component of the GE SIGNA PET/MR system: PET performance measurements of the GE SIGNA PET/MR
Open accessThe DOI matches its Crossref record; 7 of the 7 references it cites were found in a registry or an index.
Abstract
PURPOSE: The GE SIGNA PET/MR is a new whole body integrated time-of-flight (ToF)-PET/MR scanner from GE Healthcare. The system is capable of simultaneous PET and MR image acquisition with sub-400 ps coincidence time resolution. Simultaneous PET/MR holds great potential as a method of interrogating molecular, functional, and anatomical parameters in clinical disease in one study. Despite the complementary imaging capabilities of PET and MRI, their respective hardware tends to be incompatible due to mutual interference. In this work, the GE SIGNA PET/MR is evaluated in terms of PET performance and the potential effects of interference from MRI operation. METHODS: The NEMA NU 2-2012 protocol was followed to measure PET performance parameters including spatial resolution, noise equivalent count rate, sensitivity, accuracy, and image quality. Each of these tests was performed both with the MR subsystem idle and with continuous MR pulsing for the duration of the PET data acquisition. Most measurements were repeated at three separate test sites where the system is installed. RESULTS: The scanner has achieved an average of 4.4, 4.1, and 5.3 mm full width at half maximum radial, tangential, and axial spatial resolutions, respectively, at 1 cm from the transaxial FOV center. The peak noise equivalent count rate (NECR) of 218 kcps and a scatter fraction of 43.6% are reached at an activity concentration of 17.8 kBq/ml. Sensitivity at the center position is 23.3 cps/kBq. The maximum relative slice count rate error below peak NECR was 3.3%, and the residual error from attenuation and scatter corrections was 3.6%. Continuous MR pulsing had either no effect or a minor effect on each measurement. CONCLUSIONS: Performance measurements of the ToF-PET whole body GE SIGNA PET/MR system indicate that it is a promising new simultaneous imaging platform.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Diffusion tensor imaging: Concepts and applications
Le Bihan, D., Mangin, J., Poupon, C., Clark, C. A., Pappata, S., Molko, N., & Chabriat, H. (2001) · Journal of Magnetic Resonance Imaging · doi:10.1002/jmri.1076 · Cited by 3,547
Focus on time-of-flight PET: the benefits of improved time resolution
Conti, M. (2011) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-010-1711-y · Cited by 335
Update on time-of-flight PET imaging
Surti, S. (2015) · Journal of Nuclear Medicine · doi:10.2967/jnumed.114.145029 · Cited by 303
Rausch, I., Cal-González, J., Dapra, D., Gallowitsch, H. J., Lind, P., Beyer, T., & Minear, G. (2015) · EJNMMI Physics · doi:10.1186/s40658-015-0132-1 · Cited by 133
Boellaard, R., Rausch, I., Beyer, T., Delso, G., Yaqub, M., Quick, H. H., & Sattler, B. (2015) · Medical Physics · doi:10.1118/1.4930962 · Cited by 62
Self-citationSystematic Evaluation of Phantom Fluids for Simultaneous PET/MR Hybrid Imaging
Ziegler, S., Braun, H., Ritt, P., Hocke, C., Kuwert, T., & Quick, H. H. (2013) · Journal of Nuclear Medicine · doi:10.2967/jnumed.112.116376 · Cited by 50
Improvements in PET Image Quality in Time of Flight (TOF) Simultaneous PET/MRI
Minamimoto, R., Levin, C., Jamali, M., Holley, D., Barkhodari, A., Zaharchuk, G., & Iagaru, A. (2016) · Molecular Imaging and Biology · doi:10.1007/s11307-016-0939-8 · Cited by 32
Self-citation
Cited by
Spencer, B. A., Berg, E., Schmall, J. P., Omidvari, N., Leung, E. K., Abdelhafez, Y. G., Tang, S., Deng, Z., Dong, Y., Lv, Y., et al. (2021) · Journal of Nuclear Medicine · doi:10.2967/jnumed.120.250597 · Cited by 403
Aide, N., Lasnon, C., Veit-Haibach, P., Sera, T., Sattler, B., & Boellaard, R. (2017) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-017-3740-2 · Cited by 344
Studies of a Next-Generation Silicon-Photomultiplier–Based Time-of-Flight PET/CT System
Hsu, D. F. C., Ilan, E., Peterson, W. T., Uribe, J., Lubberink, M., & Levin, C. S. (2017) · Journal of Nuclear Medicine · doi:10.2967/jnumed.117.189514 · Cited by 241
“A similar detector system design was used in the SIGNA PET/MR system (GE Healthcare) and achieved a coincidence time resolution of less than 400 ps FWHM, making it capable of advanced time-of-flight (TOF) performance (17).”
MethodSelf-citationThe silicon photomultiplier: fundamentals and applications of a modern solid-state photon detector
Gundacker, S., & Heering, A. (2020) · Physics in Medicine & Biology · doi:10.1088/1361-6560/ab7b2d · Cited by 237
“Nevertheless, SiPMs have paved their way into TOF-PET due to the promise of excellent performances proven to deliver 100 ps CTR (in laboratory) and with the prospect of breaking this value even further towards 10 ps. Prominent examples of commercial systems are the Vereos Digital TOF-PET/CT developed by Philips using…”
MethodQuantification, improvement, and harmonization of small lesion detection with state-of-the-art PET
van der Vos, C. S., Koopman, D., Rijnsdorp, S., Arends, A. J., Boellaard, R., van Dalen, J. A., Lubberink, M., Willemsen, A. T. M., & Visser, E. P. (2017) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-017-3727-z · Cited by 206
Performance Evaluation of the Vereos PET/CT System According to the NEMA NU2-2012 Standard
Rausch, I., Ruiz, A., Valverde-Pascual, I., Cal-González, J., Beyer, T., & Carrio, I. (2019) · Journal of Nuclear Medicine · doi:10.2967/jnumed.118.215541 · Cited by 168
“Therefore, the Vereos PET/CT system is able to handle approximately twice the activity concentration within the FOV than other available PET/CT systems, with peak NEC rates occurring at activity concentrations of between 20 and 30 kBq/mL (8,9,20,24,30,31).”
MethodInfluentialPan, T., Einstein, S. A., Kappadath, S. C., Grogg, K. S., Lois Gomez, C., Alessio, A. M., Hunter, W. C., El Fakhri, G., Kinahan, P. E., & Mawlawi, O. R. (2019) · Medical Physics · doi:10.1002/mp.13576 · Cited by 122
Practical issues and limitations of brain attenuation correction on a simultaneous PET-MR scanner
Mackewn, J. E., Stirling, J., Jeljeli, S., Gould, S.-M., Johnstone, R. I., Merida, I., Pike, L. C., McGinnity, C. J., Beck, K., Howes, O., et al. (2020) · EJNMMI Physics · doi:10.1186/s40658-020-00295-x · Cited by 20
Metadata
- Authors
- Grant, A. M., Deller, T. W., Khalighi, M. M., Maramraju, S. H., Delso, G., & Levin, C. S.
- Venue
- Medical Physics
- Publisher
- Wiley
- Published
- 2016-04-13
- Volume
- 43
- Issue
- 5
- Pages
- 2334-2343
- ISSN
- 0094-2405 (ISSN-L)
- OpenAlex
- W2337912037
- Cited by
- 247
- References
- 7
- FWCI
- 25.23
- Type
- journal article
- Language
- en
- Open access
- hybrid
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.