Electrical field interactions in different cochlear implant systems
Open accessThe DOI matches its Crossref record; 2 of the 2 references it cites were found in a registry or an index.
Abstract
The goal of this study was to evaluate electrical field interactions produced by the stimulation of different types of intracochlear electrodes in 12 adult subjects (three Ineraid™, four Clarion™ S-Series, three S-Series with the electrode positioning system-EPS and two Clarion™ HiFocus-I with the EPS). Psychophysical measurements were conducted with biphasic stimuli (813 pulse per second, 153.8 μs/phase). “Perturbation” signals (300 ms) were applied to one electrode chosen at the middle of the array and their effects on detection thresholds of “probe” signals (30 ms) were measured on the neighbor basal electrode. Perturbation levels were set below the detection threshold of the perturbation electrode (−2 dB re threshold). Measurements were first conducted for simultaneous stimulation of the probe and of the perturbation electrodes, for monopolar for all subjects and for bipolar stimulus configurations for both Clarion™ HiFocus-I subjects. The tested Clarion™ electrodes did not present lower monopolar interactions than the Ineraid™ electrodes. Nevertheless, considering the shorter distance between electrodes for the Clarion™ than for the Ineraid™, the tested Clarion™ electrodes might be more selective than the Ineraid™. We did not find any significant monopolar electrical field-interaction differences between subjects who received the S-Series array with and without the EPS. We did not find lower interactions for both subjects who received the HiFocus-I array than for subjects who received the S-Series. Electrical field interactions were lower for bipolar than for monopolar configurations for both HiFocus-I subjects. A second set of measurements was conducted for nonsimultaneous stimulation similar to the one used in continuous interleaved sampling sound strategy. These measurements showed that interactions evaluated for simultaneous biphasic stimuli were larger than for nonsimultaneous stimulation.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Transformed Up-Down Methods in Psychoacoustics
Levitt, H. (1971) · The Journal of the Acoustical Society of America · doi:10.1121/1.1912375 · Cited by 5,477
Friesen, L. M., Shannon, R. V., Baskent, D., & Wang, X. (2001) · The Journal of the Acoustical Society of America · doi:10.1121/1.1381538 · Cited by 1,049
Cited by
The Impact of Auditory Spectral Resolution on Listening Effort Revealed by Pupil Dilation
Winn, M. B., Edwards, J. R., & Litovsky, R. Y. (2015) · Ear & Hearing · doi:10.1097/aud.0000000000000145 · Cited by 278
How cochlear implants encode speech
Rubinstein, J. T. (2004) · Current Opinion in Otolaryngology & Head and Neck Surgery · doi:10.1097/01.moo.0000134452.24819.c0 · Cited by 209
Winn, M. B. (2016) · Trends in Hearing · doi:10.1177/2331216516669723 · Cited by 174
Bierer, J. A. (2007) · The Journal of the Acoustical Society of America · doi:10.1121/1.2436712 · Cited by 169
Clinical Evaluation of an Image-Guided Cochlear Implant Programming Strategy
Noble, J. H., Gifford, R. H., Hedley-Williams, A. J., Dawant, B. M., & Labadie, R. F. (2014) · Audiology and Neurotology · doi:10.1159/000365273 · Cited by 162
“Programming efficacy is sensitive to suboptimal electrode positioning [Rubenstein, 2004; Wilson and Dorman, 2008], which can lead to excessive spread of intracochlear electric current, and this is more commonly referred to as ‘channel interaction’ [Boex et al., 2003; Fu and Nogaki, 2005].”
BackgroundSpatiotemporal Interactions in Retinal Prosthesis Subjects
Horsager, A., Greenberg, R. J., & Fine, I. (2010) · Investigative Opthalmology & Visual Science · doi:10.1167/iovs.09-3746 · Cited by 95
Temporal interactions during paired-electrode stimulation in two retinal prosthesis subjects
Horsager, A., Boynton, G. M., Greenberg, R. J., & Fine, I. (2011) · Investigative Opthalmology & Visual Science · doi:10.1167/iovs.10-5282 · Cited by 48
Spencer, T. C., Fallon, J. B., Thien, P. C., & Shivdasani, M. N. (2016) · Investigative Opthalmology & Visual Science · doi:10.1167/iovs.16-19325 · Cited by 31
Metadata
- Authors
- Boëx, C., de Balthasar, C., Kós, M.-I., & Pelizzone, M.
- Venue
- The Journal of the Acoustical Society of America
- Publisher
- Acoustical Society of America (ASA)
- Published
- 2003-10-01
- Volume
- 114
- Issue
- 4
- Pages
- 2049-2057
- ISSN
- 0001-4966 (ISSN-L)
- 1520-8524
- OpenAlex
- W2003040117
- Cited by
- 85
- References
- 2
- FWCI
- 1.46
- Type
- journal article
- Language
- en
- Open access
- bronze
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.