The DOI matches its Crossref record; 1 of the 1 references it cites were found in a registry or an index.
The ingestible telemetric body core temperature sensor: a review of validity and exercise applications
Open accessAbstract
An ingestible telemetric temperature sensor for measuring body core temperature (T c) was first described 45 years ago, although the method has only recently gained widespread use for exercise applications. This review aims to (1) use Bland and Altman’s limits of agreement (LoA) method as a basis for quantitatively reviewing the agreement between intestinal sensor temperature (T intestinal), oesophageal temperature (T oesophageal) and rectal temperature (T rectal) across numerous previously published validation studies; (2) review factors that may affect agreement; and (3) review the application of this technology in field-based exercise studies. The agreement between T intestinal and T oesophageal is suggested to meet our delimitation for an acceptable level of agreement (ie, systematic bias <0.1°C and 95% LoA within ±0.4°C). The agreement between T intestinal and T rectal shows a significant systematic bias >0.1°C, although the 95% LoA is acceptable. T intestinal responds less rapidly than T oesophageal at the start or cessation of exercise or to a change in exercise intensity, but more rapidly than T rectal. When using this technology, care should be taken to ensure adequate control over sensor calibration and data correction, timing of ingestion and electromagnetic interference. The ingestible sensor has been applied successfully in numerous sport and occupational applications such as the continuous measurement of T c in deep sea saturation divers, distance runners and soldiers undertaking sustained military training exercises. It is concluded that the ingestible telemetric temperature sensor represents a valid index of T c and shows excellent utility for ambulatory field-based applications.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
Martin Bland, J., & Altman, D. (1986) · The Lancet · doi:10.1016/s0140-6736(86)90837-8 · Cited by 47,965
Research works printed but not linked: 37
- 1996, Med Sci Sports Exerc, 28, 1
- 10.1136/bjsm.37.6.548 · DOI
- 10.2165/00007256-200636010-00004 · DOI
- 1999, J Appl Physiol, 86, 1032, 10.1152/jappl.1999.86.3.1032 · DOI · Paper page
- Blatteis CM. Methods of body temperature measurement. In: Blatteis CM, ed. Physiology and pathophysiology of temperature regulation. Singapore: World Scientific, 1998:273–9. · DOI
- 1961, New Sci, 16, 419
- 10.1088/0143-0815/2/4/001 · DOI
- 1988, Johns Hopkins APL Tech Dig, 9, 16
- 2004, J Therm Biol, 29, 605, 10.1016/j.jtherbio.2004.08.020 · DOI
- 1983, J Appl Physiol, 54, 1025, 10.1152/jappl.1983.54.4.1025 · DOI
- 1993, J Therm Biol, 18, 203, 10.1016/0306-4565(93)90004-D · DOI
- 1993, Aviat Space Environ Med, 64, 760
- 1997, J Therm Biol, 22, 343, 10.1016/S0306-4565(97)00032-6 · DOI
- 1998, Med Sci Sports Exerc, 30, 468, 10.1097/00005768-199803000-00020 · DOI
- 10.1081/CBI-120004227 · DOI
- 2006, Med Sci Sports Exerc, 38, 1926, 10.1249/01.mss.0000233800.69776.ef · DOI
- 1962, J Physiol, 160, 22, 10.1113/jphysiol.1962.sp006830 · DOI
- 2001, Pro Aust Physiol Pharmacol Soc, 32, 28
- 10.2165/00007256-199826040-00002 · DOI · Paper page
- 1998, Prehosp Emerg Care, 2, 280, 10.1080/10903129808958880 · DOI
- 2000, J Swimming Res, 14, 26
- 2001, Undersea Hyperb Med, 28, 149
- 2004, J Athl Train, 39, 235
- 2006, Med Sci Sports Exerc, 38, 803, 10.1249/01.mss.0000218134.74238.6a · DOI
- 2006, Med Sci Sports Exerc, 38, 894, 10.1249/01.mss.0000218122.59968.eb · DOI
- 10.1136/bjsm.2005.021980 · DOI
- 2006, J Athl Train, 41, 8
- 10.1136/bjsm.2005.022426 · DOI
- 1966, J Physiol, 184, 581, 10.1113/jphysiol.1966.sp007932 · DOI
- 1971, Int J Biometeorol, 15, 176, 10.1007/BF01803894 · DOI
- 1991, Int J Radiat Oncol Biol Phys, 21, 1353, 10.1016/0360-3016(91)90297-H · DOI
- 1975, J Appl Physiol, 38, 1030, 10.1152/jappl.1975.38.6.1030 · DOI
- 1989, J Physiol, 412, 42
- 1990, J Physiol, 430, 93
- 1977, J Appl Physiol, 42, 909, 10.1152/jappl.1977.42.6.909 · DOI
- 1999, Eur J Appl Physiol, 79, 341, 10.1007/s004210050518 · DOI
- 1997, Mil Med, 162, 623, 10.1093/milmed/162.9.623 · DOI
Cited by
Considerations for the measurement of core, skin and mean body temperatures
Taylor, N. A. S., Tipton, M. J., & Kenny, G. P. (2014) · Journal of Thermal Biology · doi:10.1016/j.jtherbio.2014.10.006 · Cited by 420
Circadian rhythm of wrist temperature in normal-living subjects
Sarabia, J. A., Rol, M. A., Mendiola, P., & Madrid, J. A. (2008) · Physiology & Behavior · doi:10.1016/j.physbeh.2008.08.005 · Cited by 222
The effect of stress on core and peripheral body temperature in humans
Vinkers, C. H., Penning, R., Hellhammer, J., Verster, J. C., Klaessens, J. H. G. M., Olivier, B., & Kalkman, C. J. (2013) · Stress · doi:10.3109/10253890.2013.807243 · Cited by 220
Circadian rhythmicity of body temperature and metabolism
Refinetti, R. (2020) · Temperature · doi:10.1080/23328940.2020.1743605 · Cited by 155
Thermal pleasure in built environments: alliesthesia in different thermoregulatory zones
Parkinson, T., de Dear, R., & Candido, C. (2016) · Building Research & Information · doi:10.1080/09613218.2015.1059653 · Cited by 126
Metadata
- Authors
- Byrne, C., & Lim, C. L.
- Venue
- British Journal of Sports Medicine
- Publisher
- BMJ
- Published
- 2006-12-18
- Volume
- 41
- Issue
- 3
- Pages
- 126-133
- ISSN
- 0306-3674 (ISSN-L)
- 1473-0480
- OpenAlex
- W2141006808
- Cited by
- 456
- References
- 1
- FWCI
- 3.72
- Type
- journal article
- Language
- en
- Open access
- bronze
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: The ingestible telemetric body core temperature sensor: a review of validity and exercise applications. https://app.cochat.ai/papers/10.1136/bjsm.2006.026344
- The paper
- Byrne, C., & Lim, C. L. (2007). The ingestible telemetric body core temperature sensor: a review of validity and exercise applications. British Journal of Sports Medicine. https://doi.org/10.1136/bjsm.2006.026344
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.