Diagnostic performance of an automatic blood pressure measurement device, Microlife WatchBP Home A, for atrial fibrillation screening in a real-world primary care setting

Open access
Chan, Pak-Hei, Wong, Chun-Ka, Pun, Louise, Wong, Yu-Fai, Wong, Michelle Man-Ying, Chu, Daniel Wai-Sing, Siu, Chung-Wah · 2017 · BMJ Open
Cited by 58

The DOI matches its Crossref record; 10 of the 10 references it cites were found in a registry or an index.

Abstract

Objective: To evaluate the diagnostic performance of a UK National Institute for Health and Care Excellence-recommended automatic oscillometric blood pressure (BP) measurement device incorporated with an atrial fibrillation (AF) detection algorithm (Microlife WatchBP Home A) for real-world AF screening in a primary healthcare setting. Setting: Primary healthcare setting in Hong Kong. Interventions: This was a prospective AF screening study carried out between 1 September 2014 and 14 January 2015. The Microlife device was evaluated for AF detection and compared with a reference standard of lead-I ECG. Primary outcome measures: Diagnostic performance of Microlife for AF detection. Results: 5969 patients (mean age: 67.2±11.0 years; 53.9% female) were recruited. The mean CHA 2 DS 2-VASc ( C : congestive heart failure [1 point]; H : hypertension [1 point]; A 2 : age 65-74 years [1 point] and age ≥75 years [2 points]; D : diabetes mellitus [1 point]; S : prior stroke or transient ischemic attack [2 points]; VA : vascular disease [1 point]; and Sc : sex category [female] [1 point])score was 2.8±1.3. AF was diagnosed in 72 patients (1.21%) and confirmed by a 12-lead ECG. The Microlife device correctly identified AF in 58 patients and produced 79 false-positives. The corresponding sensitivity and specificity for AF detection were 80.6% (95% CI 69.5 to 88.9) and 98.7% (95% CI 98.3 to 98.9), respectively. Among patients with a false-positive by the Microlife device, 30.4% had sinus rhythm, 35.4% had sinus arrhythmia and 29.1% exhibited premature atrial complexes. With the low prevalence of AF in this population, the positive and negative predictive values of Microlife device for AF detection were 42.4% (95% CI 34.0 to 51.2) and 99.8% (95% CI 99.6 to 99.9), respectively. The overall diagnostic performance of Microlife device to detect AF as determined by area under the curves was 0.90 (95% CI 0.89 to 0.90). Conclusions: In the primary care setting, Microlife WatchBP Home was an effective means to screen for AF, with a reasonable sensitivity of 80.6% and a high negative predictive value of 99.8%, in addition to its routine function of BP measurement. In a younger patient population aged <65 years with a lower prevalence of AF, Microlife WatchBP Home A demonstrated a similar diagnostic accuracy.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1136/bmjopen-2016-013685

Cites

10 of 10 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 10 references

Cited by

See all 58

Metadata

Authors
Chan, P.-H., Wong, C.-K., Pun, L., Wong, Y.-F., Wong, M. M.-Y., Chu, D. W.-S., & Siu, C.-W.
Venue
BMJ Open
Publisher
BMJ
Published
2017-06
Volume
7
Issue
6
Pages
e013685
ISSN
2044-6055 (ISSN-L)
OpenAlex
W2625603853
Cited by
58
References
10
FWCI
4.35
Type
journal article
Language
en
Open access
gold

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Diagnostic performance of an automatic blood pressure measurement device, Microlife WatchBP Home A, for atrial fibrillation screening in a real-world primary care setting. https://app.cochat.ai/papers/10.1136/bmjopen-2016-013685
The paper
Chan, P.-H., Wong, C.-K., Pun, L., Wong, Y.-F., Wong, M. M.-Y., Chu, D. W.-S., & Siu, C.-W. (2017). Diagnostic performance of an automatic blood pressure measurement device, Microlife WatchBP Home A, for atrial fibrillation screening in a real-world primary care setting. BMJ Open. https://doi.org/10.1136/bmjopen-2016-013685

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.