Evaluating the accuracy of heart rate sensors based on photoplethysmography for in-the-wild analysis
Abstract
Continuous measurement of physiological functions, like heart rate (HR) and heart rate variability (HRV), using commercially available wearable sensors provides the prospects of improving the healthcare of individuals with a positive impact on society, bringing pervasiveness, lower cost, and broader access. However, common wearable devices use photoplethysmography (PPG) to derive data on HR and HRV, and it is yet unclear to which extent PPG signals can be used as a proxy for data collected using medical-grade devices. To address this challenge, we consider five consumer devices to assess the signal quality of HR and two devices measuring HRV and compare them with a standard electrocardiography (ECG) Holter monitor. We collect data from fourteen participants who followed a 55 minutes protocol for at least two sessions. Using this data set, which we make publicly available to the research community, we show that PPG is a valid proxy for both HR and standard time- and frequency-domain measurements of HRV. Further, we demonstrate that wearable devices are suitable for monitoring both HR and HRV in daily life but might be limited during strenuous exercise. The study indicates that armband-based devices are more reliable than wrist-based wearables for HRV assessment.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The Empatica E4 may be unsuited for monitoring everyday activities.
“The device has also been evaluated in general purpose real-world settings, with results indicating that the Empatica E4 may be unsuited for monitoring everyday activities70,71.”
From Data quality evaluation in wearable monitoring (2022) · cites this paper as [71]
Cites
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
Martin Bland, J., & Altman, D. (1986) · The Lancet · doi:10.1016/s0140-6736(86)90837-8 · Cited by 47,965 · see citing works
A guideline of selecting and reporting intraclass correlation coefficients for reliability research
Koo, T. K., & Li, M. Y. (2016) · Journal of Chiropractic Medicine · doi:10.1016/j.jcm.2016.02.012 · Cited by 29,639 · see citing works
An overview of heart rate variability metrics and norms
Shaffer, F., & Ginsberg, J. P. (2017) · Frontiers in Public Health · doi:10.3389/fpubh.2017.00258 · Cited by 7,185 · see citing works
The Internet of Things for Health Care: A Comprehensive Survey
Islam, S. M. R., Kwak, D., Kabir, M. H., Hossain, M., & Kwak, K.-S. (2015) · IEEE Access · doi:10.1109/access.2015.2437951 · Cited by 3,199 · see citing works
Lee, H., Choi, T. K., Lee, Y. B., Cho, H. R., Ghaffari, R., Wang, L., Choi, H. J., Chung, T. D., Lu, N., Hyeon, T., et al. (2016) · Nature Nanotechnology · doi:10.1038/nnano.2016.38 · Cited by 1,802 · see citing works
Kemp, A. H., Quintana, D. S., Gray, M. A., Felmingham, K. L., Brown, K., & Gatt, J. M. (2010) · Biological Psychiatry · doi:10.1016/j.biopsych.2009.12.012 · Cited by 1,286 · see citing works
A dynamical model for generating synthetic electrocardiogram signals
McSharry, P., Clifford, G. D., Tarassenko, L., & Smith, L. A. (2003) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2003.808805 · Cited by 1,232 · see citing works
The role of vagal function in the risk for cardiovascular disease and mortality
Thayer, J. F., & Lane, R. D. (2007) · Biological Psychology · doi:10.1016/j.biopsycho.2005.11.013 · Cited by 1,033 · see citing works
Recommendations for the Standardization and Interpretation of the Electrocardiogram
Kligfield, P., Gettes, L. S., Bailey, J. J., Childers, R., Deal, B. J., Hancock, E. W., van Herpen, G., Kors, J. A., Macfarlane, P. W., Mirvis, D. M., et al. (2007) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2007.01.024 · Cited by 815 · see citing works
Kligfield (2007) · doi:10.1161/circulationaha.106.180200 · Cited by 802 · see citing works
Work stress and coronary heart disease: what are the mechanisms?
Chandola, T., Britton, A., Brunner, E., Hemingway, H., Malik, M., Kumari, M., Badrick, E., Kivimaki, M., & Marmot, M. (2008) · European Heart Journal · doi:10.1093/eurheartj/ehm584 · Cited by 673 · see citing works
How accurate is pulse rate variability as an estimate of heart rate variability?
Schäfer, A., & Vagedes, J. (2013) · International Journal of Cardiology · doi:10.1016/j.ijcard.2012.03.119 · Cited by 661 · see citing works
Michael, S., Graham, K. S., & Davis, G. M. (2017) · Frontiers in Physiology · doi:10.3389/fphys.2017.00301 · Cited by 626 · see citing works
Applying neural network analysis on heart rate variability data to assess driver fatigue
Patel, M., Lal, S., Kavanagh, D., & Rossiter, P. (2010) · Expert Systems with Applications · doi:10.1016/j.eswa.2010.12.028 · Cited by 418 · see citing works
Validity of the Polar V800 heart rate monitor to measure RR intervals at rest
Giles, D., Draper, N., & Neil, W. (2015) · European Journal of Applied Physiology · doi:10.1007/s00421-015-3303-9 · Cited by 376 · see citing works
Accuracy of Heart Rate Watches: Implications for Weight Management
Wallen, M. P., Gomersall, S. R., Keating, S. E., Wisløff, U., & Coombes, J. S. (2016) · PLOS ONE · doi:10.1371/journal.pone.0154420 · Cited by 336 · see citing works
Can Wearable Devices Accurately Measure Heart Rate Variability? A Systematic Review
Georgiou, K., Larentzakis, A. V., Khamis, N. N., Alsuhaibani, G. I., Alaska, Y. A., & Giallafos, E. J. (2018) · Folia Medica · doi:10.2478/folmed-2018-0012 · Cited by 244 · see citing works
Validation of the Empatica E4 wristband
McCarthy, C., Pradhan, N., Redpath, C., & Adler, A. (2016) · 2016 IEEE EMBS International Student Conference (ISC) · doi:10.1109/embsisc.2016.7508621 · Cited by 235 · see citing works
Validation of Biofeedback Wearables for Photoplethysmographic Heart Rate Tracking.
(2016) · PubMed Central · Cited by 150 · see citing works
Evaluation of wearable consumer heart rate monitors based on photopletysmography
Parak, J., & Korhonen, I. (2014) · 2014 36th Annual International Conference of the IEEE Engineering in Medicine and Biology Society · doi:10.1109/embc.2014.6944419 · Cited by 148 · see citing works
A photoelectric plethysmograph for the measurement of cutaneous blood flow
CHALLONER, A. V. J., & Ramsay, C. A. (1974) · Physics in Medicine and Biology · doi:10.1088/0031-9155/19/3/003 · Cited by 145 · see citing works
Validation of Heart Rate Monitor Polar RS800 for Heart Rate Variability Analysis During Exercise
Hernando, D., Garatachea, N., Almeida, R., Casajús, J. A., & Bailón, R. (2016) · The Journal of Strength and Conditioning Research · doi:10.1519/jsc.0000000000001662 · Cited by 140 · see citing works
Nunan, D., Jakovljević, D. G., Donovan, G., Hodges, L. D., SANDERCOCK, G. R. H., & Brodie, D. A. (2008) · European Journal of Applied Physiology · doi:10.1007/s00421-008-0742-6 · Cited by 123 · see citing works
Comparison of finger plethysmograph to ECG in the measurement of heart rate variability
Giardino, N. D., Lehrer, P. M., & Edelberg, R. (2002) · Psychophysiology · doi:10.1111/1469-8986.3920246 · Cited by 119 · see citing works
Bonaduce, D., Petretta, M., Marcianò, F., Vicario, M. L. E., Apicella, C., Rao, M. A. E., Nicolai, E., & Volpe, M. (1999) · American Heart Journal · doi:10.1016/s0002-8703(99)70112-2 · Cited by 110 · see citing works
Mazilu, S., Blanke, U., Hardegger, M., Tröster, G., Gazit, E., & Hausdorff, J. M. (2014) · CHI Conference on Human Factors in Computing Systems · doi:10.1145/2556288.2557278 · Cited by 84 · see citing works
Vescio, B., Salsone, M., Gambardella, A., & Quattrone, A. (2018) · Sensors · doi:10.3390/s18030844 · Cited by 73 · see citing works
Validity of wrist-worn consumer products to measure heart rate and energy expenditure
Thiebaud, R. S., Funk, M. D., Patton, J. C., Massey, B. L., Shay, T. E., Schmidt, M. G., & Giovannitti, N. (2018) · Digital Health · doi:10.1177/2055207618770322 · Cited by 56 · see citing works
Bailón, R., Garatachea, N., de la Iglesia, I., et al. (2013) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2013.2242328 · Cited by 37 · see citing works
Recognizing Digital Biomarkers for Fatigue Assessment in Patients with Multiple Sclerosis
Barrios, L., Oldrati, P., Santini, S., et al. (2018) · doi:10.4108/eai.20-4-2018.2276340 · Cited by 16 · see citing works
Self-citation
Flagged first, then by citations
Cited by
Sorted by citations
Li, K., Cardoso, C., Moctezuma-Ramirez, A., Elgalad, A., & Perin, E. (2023) · International Journal of Environmental Research and Public Health · doi:10.3390/ijerph20247146 · Cited by 160 · see citing works
Zhang, Y., Weaver, R. G., Armstrong, B., Burkart, S., Zhang, S., & Beets, M. W. (2020) · Journal of Sports Sciences · doi:10.1080/02640414.2020.1767348 · Cited by 88 · see citing works
Data quality evaluation in wearable monitoring
Böttcher, S., Vieluf, S., Bruno, E., Joseph, B., Epitashvili, N., Biondi, A., Zabler, N., Glasstetter, M., Dümpelmann, M., Van Laerhoven, K., et al. (2022) · Scientific Reports · doi:10.1038/s41598-022-25949-x · Cited by 91 · see citing works
- Authors
- Barrios, L., Oldrati, P., Santini, S., & Lutterotti, A.
- Affiliations
- ETH Zurich (CH)
- Università della Svizzera Italiana (USI), Lugano, Switzerland
- University of Zurich (CH)
- Venue
- Proceedings of the 13th EAI International Conference on Pervasive Computing Technologies for Healthcare
- Publisher
- ACM
- Published
- 2019-05-20
- Pages
- 251-261
- ISSN
- 2153-1633 (ISSN-L)
- OpenAlex
- W2953978148
- Cited by
- 43
- References
- 31
- Citations by year
- 2026: 1 · 2025: 3 · 2024: 9 · 2023: 5 · 2022: 10 · 2021: 7 · 2020: 7 · 2019: 1
- FWCI
- 12.97
- Type
- conference paper
- Language
- en
- Open access
- closed
- Locations
- International Conference on Pervasive Computing Technologies for Healthcare (publishedVersion)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Heart Rate Variability and Autonomic Control
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › ECG Monitoring and Analysis
- Keywords
- photoplethysmography, heart rate monitoring, heart rate variability, wearable sensors, consumer health devices, sensor accuracy, physical activity
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review