Cardiovascular Flow Measurement with Phase-Contrast MR Imaging: Basic Facts and Implementation
Abstract
Phase-contrast magnetic resonance (MR) imaging is a well-known but undervalued method of obtaining quantitative information on blood flow. Applications of this technique in cardiovascular MR imaging are expanding. According to the sequences available, phase-contrast measurement can be performed in a breath hold or during normal respiration. Prospective as well as retrospective gating techniques can be used. Common errors in phase-contrast imaging include mismatched encoding velocity, deviation of the imaging plane, inadequate temporal resolution, inadequate spatial resolution, accelerated flow and spatial misregistration, and phase offset errors. Flow measurements are most precise if the imaging plane is perpendicular to the vessel of interest and flow encoding is set to through-plane flow. The sequence should be repeated at least once, with a high encoding velocity used initially. If peak velocity has to be estimated, flow measurement is repeated with an adapted encoding velocity. The overall error of a phase-contrast flow measurement comprises errors during prescription as well as errors that occur during image analysis of the flow data. With phase-contrast imaging, the overall error in flow measurement can be reduced to less than 10%, an acceptable level of error for routine clinical use.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
A flow velocity zeugmatographic interlace for NMR imaging in humans
Moran, P. R. (1982) · Magnetic Resonance Imaging · doi:10.1016/0730-725x(82)90170-9 · Cited by 745
Natural history of coarctation of the aorta
Campbell, M. (1970) · Heart · doi:10.1136/hrt.32.5.633 · Cited by 614
Kondo, C., Caputo, G. R., Semelka, R., Foster, E., Shimakawa, A., & Higgins, C. B. (1991) · American Journal of Roentgenology · doi:10.2214/ajr.157.1.2048544 · Cited by 260
Blood Flow Rates by Nuclear Magnetic Resonance Measurements
Singer, J. R. (1959) · Science · doi:10.1126/science.130.3389.1652 · Cited by 248
Hundley, W. G., Li, H. F., Willard, J. E., Landau, C., Lange, R. A., Meshack, B. M., Hillis, L. D., & Peshock, R. M. (1995) · Circulation · doi:10.1161/01.cir.92.5.1151 · Cited by 229
Schwitter, J., DeMarco, T., Kneifel, S., von Schulthess, G. K., Jörg-Ciopor, M., Arheden, H., Rühm, S., Stumpe, K. O., Buck, A., Parmley, W. W., et al. (2000) · Circulation · doi:10.1161/01.cir.101.23.2696 · Cited by 159
NMR blood flow imaging using multiecho, phase contrast sequences
O'Donnell, M. (1985) · Medical Physics · doi:10.1118/1.595736 · Cited by 158
Magnetic Resonance Imaging of Blood Flow with a Phase Subtraction Technique
EVANS, A. J., IWAI, F., Grist, T. M., Sostman, H. D., Hedlund, L. W., Spritzer, C. E., Negro‐Vilar, R., Beam, C. A., & Pelc, N. J. (1993) · Investigative Radiology · doi:10.1097/00004424-199302000-00004 · Cited by 153
Hofman, M. B. M., Visser, F. C., van Rossum, A. C., Vink, G. Q. M., Sprenger, M., & Westerhof, N. (1995) · Magnetic Resonance in Medicine · doi:10.1002/mrm.1910330606 · Cited by 152
Maier, S. E., Meier, D., Boesiger, P., Moser, U., & Vieli, A. (1989) · Radiology · doi:10.1148/radiology.171.2.2649924 · Cited by 140
Cited by
Schulz-Menger, J., Bluemke, D. A., Bremerich, J., Flamm, S. D., Fogel, M. A., Friedrich, M. G., Kim, R. J., von Knobelsdorff-Brenkenhoff, F., Kramer, C. M., Pennell, D. J., et al. (2013) · Journal of Cardiovascular Magnetic Resonance · doi:10.1186/1532-429x-15-35 · Cited by 2,097
Normal values for cardiovascular magnetic resonance in adults and children
Kawel-Boehm, N., Maceira, A., Valsangiacomo-Buechel, E. R., Vogel-Claussen, J., Turkbey, E. B., Williams, R., Plein, S., Tee, M., Eng, J., & Bluemke, D. A. (2015) · Journal of Cardiovascular Magnetic Resonance · doi:10.1186/s12968-015-0111-7 · Cited by 746
Sun, L., Macgowan, C. K., Sled, J. G., Yoo, S.-J., Manlhiot, C., Porayette, P., Grosse-Wortmann, L., Jaeggi, E., McCrindle, B. W., Kingdom, J., et al. (2015) · Circulation · doi:10.1161/circulationaha.114.013051 · Cited by 533
Fratz, S., Chung, T., Greil, G. F., Samyn, M. M., Taylor, A. M., Valsangiacomo Buechel, E. R., Yoo, S.-J., & Powell, A. J. (2013) · Journal of Cardiovascular Magnetic Resonance · doi:10.1186/1532-429x-15-51 · Cited by 507
E XPERIMENTAL AND C OMPUTATIONAL M ETHODS IN C ARDIOVASCULAR F LUID M ECHANICS
Taylor, C. A., & Draney, M. T. (2004) · Annual Review of Fluid Mechanics · doi:10.1146/annurev.fluid.36.050802.121944 · Cited by 285
Metadata
- Authors
- Lotz, J., Meier, C., Leppert, A., & Galanski, M.
- Venue
- RadioGraphics
- Published
- 2002-05
- Volume
- 22
- Issue
- 3
- Pages
- 651-671
- ISSN
- 0271-5333 (ISSN-L)
- 1527-1323
- OpenAlex
- W2100082339
- Cited by
- 673
- References
- 20
- FWCI
- 7.47
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Cardiovascular Flow Measurement with Phase-Contrast MR Imaging: Basic Facts and Implementation. https://app.cochat.ai/papers/10.1148/radiographics.22.3.g02ma11651
- The paper
- Lotz, J., Meier, C., Leppert, A., & Galanski, M. (2002). Cardiovascular Flow Measurement with Phase-Contrast MR Imaging: Basic Facts and Implementation. RadioGraphics. https://doi.org/10.1148/radiographics.22.3.g02ma11651
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.