Accuracy, Precision, and Reproducibility of Four T1 Mapping Sequences: A Head-to-Head Comparison of MOLLI, ShMOLLI, SASHA, and SAPPHIRE
Open accessAbstract
PURPOSE: To compare accuracy, precision, and reproducibility of four commonly used myocardial T1 mapping sequences: modified Look-Locker inversion recovery (MOLLI), shortened MOLLI (ShMOLLI), saturation recovery single-shot acquisition (SASHA), and saturation pulse prepared heart rate independent inversion recovery (SAPPHIRE). MATERIALS AND METHODS: This HIPAA-compliant study was approved by the institutional review board. All subjects provided written informed consent. Accuracy, precision, and reproducibility of the four T1 mapping sequences were first compared in phantom experiments. In vivo analysis was performed in seven healthy subjects (mean age ± standard deviation, 38 years ± 19; four men, three women) who were imaged twice on two separate days. In vivo reproducibility of native T1 mapping and extracellular volume (ECV) were measured. Differences between the sequences were assessed by using Kruskal-Wallis and Wilcoxon rank sum tests (phantom data) and mixed-effect models (in vivo data). RESULTS: T1 mapping accuracy in phantoms was lower with ShMOLLI (62 msec) and MOLLI (44 msec) than with SASHA (13 msec; P < .05) and SAPPHIRE (12 msec; P < .05). MOLLI had similar precision to ShMOLLI (4.0 msec vs 5.6 msec; P = .07) but higher precision than SAPPHIRE (6.8 msec; P = .002) and SASHA (8.7 msec; P < .001). All sequences had similar reproducibility in phantoms (P = .1). The four sequences had similar in vivo reproducibility for native T1 mapping (∼25-50 msec; P > .05) and ECV quantification (∼0.01-0.02; P > .05). CONCLUSION: SASHA and SAPPHIRE yield higher accuracy, lower precision, and similar reproducibility compared with MOLLI and ShMOLLI for T1 measurement. Different sequences yield different ECV values; however, all sequences have similar reproducibility for ECV quantification.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Modified Look‐Locker inversion recovery (MOLLI) for high‐resolution T1 mapping of the heart
Messroghli, D. R., Radjenovic, A., Kozerke, S., Higgins, D. M., Sivananthan, M. U., & Ridgway, J. P. (2004) · Magnetic Resonance in Medicine · doi:10.1002/mrm.20110 · Cited by 1,349
Moon, J. C., Messroghli, D. R., Kellman, P., Piechnik, S. K., Robson, M. D., Ugander, M., Gatehouse, P. D., Arai, A. E., Friedrich, M. G., Neubauer, S., et al. (2013) · Journal of Cardiovascular Magnetic Resonance · doi:10.1186/1532-429x-15-92 · Cited by 1,136
Self-citationAssessment of Myocardial Fibrosis With Cardiovascular Magnetic Resonance
Mewton, N., Liu, C. Y., Croisille, P., Bluemke, D., & Lima, J. A. C. (2011) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2010.11.013 · Cited by 984
Time Saving in Measurement of NMR and EPR Relaxation Times
Look, D. C., & Locker, D. R. (1970) · Review of Scientific Instruments · doi:10.1063/1.1684482 · Cited by 819
T1-mapping in the heart: accuracy and precision
Kellman, P., & Hansen, M. S. (2014) · Journal of Cardiovascular Magnetic Resonance · doi:10.1186/1532-429x-16-2 · Cited by 764
Self-citationIles, L., Pfluger, H., Phrommintikul, A., Cherayath, J., Aksit, P., Gupta, S. N., Kaye, D. M., & Taylor, A. J. (2008) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2008.06.049 · Cited by 736
Piechnik, S. K., Ferreira, V. M., Dall'Armellina, E., Cochlin, L. E., Greiser, A., Neubauer, S., & Robson, M. D. (2010) · Journal of Cardiovascular Magnetic Resonance · doi:10.1186/1532-429x-12-69 · Cited by 670
Wong, T. C., Piehler, K., Meier, C. G., Testa, S. M., Klock, A. M., Aneizi, A. A., Shakesprere, J., Kellman, P., Shroff, S. G., Schwartzman, D. S., et al. (2012) · Circulation · doi:10.1161/circulationaha.111.089409 · Cited by 493
Self-citationSaturation recovery single‐shot acquisition (SASHA) for myocardial T1 mapping
Chow, K., Flewitt, J. A., Green, J. D., Pagano, J. J., Friedrich, M. G., & Thompson, R. B. (2014) · Magnetic Resonance in Medicine · doi:10.1002/mrm.24878 · Cited by 396
Self-citationWong, T. C., Piehler, K. M., Kang, I. A., Kadakkal, A., Kellman, P., Schwartzman, D. S., Mulukutla, S. R., Simon, M. A., Shroff, S. G., Kuller, L. H., et al. (2014) · European Heart Journal · doi:10.1093/eurheartj/eht193 · Cited by 379
Self-citationArheden, H., Saeed, M., Higgins, C. B., Gao, D.-W., Bremerich, J., Wyttenbach, R., Dae, M. W., & Wendland, M. F. (1999) · Radiology · doi:10.1148/radiology.211.3.r99jn41698 · Cited by 358
Extracellular volume fraction mapping in the myocardium, part 2: initial clinical experience
Kellman, P., Wilson, J. R., Xue, H., Bandettini, W. P., Shanbhag, S. M., Druey, K. M., Ugander, M., & Arai, A. E. (2012) · Journal of Cardiovascular Magnetic Resonance · doi:10.1186/1532-429x-14-64 · Cited by 259
Self-citationFlacke, S. J., Fischer, S. E., & Lorenz, C. H. (2001) · Radiology · doi:10.1148/radiology.218.3.r01fe18703 · Cited by 211
Weingärtner, S., Akçakaya, M., Basha, T., Kissinger, K. V., Goddu, B., Berg, S., Manning, W. J., & Nezafat, R. (2014) · Magnetic Resonance in Medicine · doi:10.1002/mrm.24761 · Cited by 177
Self-citationRobson, M. D., Piechnik, S. K., Tunnicliffe, E. M., & Neubauer, S. (2013) · Magnetic Resonance in Medicine · doi:10.1002/mrm.24867 · Cited by 161
Three‐dimensional T1 mapping of the mouse heart using variable flip angle steady‐state MR imaging
Coolen, B. F., Geelen, T., Paulis, L. E. M., et al. (2010) · NMR in Biomedicine · doi:10.1002/nbm.1566 · Cited by 70
Arrhythmia insensitive rapid cardiac T1 mapping pulse sequence
Fitts, M., Breton, É., Kholmovski, E. G., Dosdall, D. J., Vijayakumar, S., Hong, K. P., Ranjan, R., MARROUCHE, N. F., Axel, L., & Kim, D. C. (2012) · Magnetic Resonance in Medicine · doi:10.1002/mrm.24586 · Cited by 67
3D myocardial T1 mapping at 3T using variable flip angle method: Pilot study
Poinsignon-Clique, H., Cheng, H. M., Marie, P., Felblinger, J., & Beaumont, M. (2013) · Magnetic Resonance in Medicine · doi:10.1002/mrm.24688 · Cited by 36
Cooper, M. A., Nguyen, T. D., Spincemaille, P., Prince, M. R., Weinsaft, J. W., & Wang, Y. (2012) · Magnetic Resonance in Medicine · doi:10.1002/mrm.24151 · Cited by 22
Knesewitsch, T., Meierhofer, C., Rieger, H., Rößler, J., Frank, M., Martinoff, S. A., Hess, J., Stern, H. C., & Fratz, S. (2013) · Journal of Cardiovascular Magnetic Resonance · doi:10.1186/1532-429x-15-3 · Cited by 18
Cited by
Messroghli, D. R., Moon, J. C., Ferreira, V. M., Grosse-Wortmann, L., He, T., Kellman, P., Mascherbauer, J., Nezafat, R., Salerno, M., Schelbert, E. B., et al. (2016) · Journal of Cardiovascular Magnetic Resonance · doi:10.1186/s12968-017-0389-8 · Cited by 1,727
Self-citationTaylor, A. J., Salerno, M., Dharmakumar, R., & Jerosch-Herold, M. (2016) · JACC: Cardiovascular Imaging · doi:10.1016/j.jcmg.2015.11.005 · Cited by 563
T1 Mapping in Characterizing Myocardial Disease: A Comprehensive Review
Puntmann, V. O., Peker, E., Chandrashekhar, Y., & Nagel, E. (2016) · Circulation Research · doi:10.1161/circresaha.116.307974 · Cited by 337
Magnetic resonance multitasking for motion-resolved quantitative cardiovascular imaging
Christodoulou, A. G., Shaw, J. L., Nguyen, C., Yang, Q., Xie, Y., Wang, N., & Li, D. (2018) · Nature Biomedical Engineering · doi:10.1038/s41551-018-0217-y · Cited by 323
Bane, O., Hectors, S. J., Wagner, M., Arlinghaus, L. L., Aryal, M. P., Cao, Y., Chenevert, T. L., Fennessy, F., Huang, W., Hylton, N. M., et al. (2018) · Magnetic Resonance in Medicine · doi:10.1002/mrm.26903 · Cited by 109
Metadata
- Authors
- Roujol, S., Weingärtner, S., Foppa, M., Chow, K., Kawaji, K., Ngo, L. H., Kellman, P., Manning, W. J., Thompson, R. B., & Nezafat, R.
- Venue
- Radiology
- Published
- 2014-09
- Volume
- 272
- Issue
- 3
- Pages
- 683-689
- ISSN
- 0033-8419 (ISSN-L)
- 1527-1315
- OpenAlex
- W2144511843
- Cited by
- 317
- References
- 20
- FWCI
- 23.57
- Type
- journal article
- Language
- en
- Open access
- green
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.