CoChat

Limitations and Late Complications of Third-Generation Automatic Cardioverter-Defibrillators

Nunain, S. O., Roelke, M., Trouton, T., Osswald, S., Kim, Y. H., Sosa-Suarez, G., Brooks, D. R., McGovern, B., Guy, M., Torchiana, D. F., Vlahakes, G. J., Garan, H., Ruskin, J. N. · 1995 · Circulation
Cited by 158

The DOI matches its Crossref record; 14 of the 14 references it cites were found in a registry or an index.

Abstract

Background This study examines the limitations and complex management problems associated with the use of tiered-therapy, implantable cardioverter-defibrillators (ICDs). Methods and Results The study group comprises the first 154 patients undergoing implantation of tiered-therapy ICDs at our institution. Pulse generators from three different manufacturers were used. In 39 patients, a complete nonthoracotomy lead system was used. The perioperative mortality was 1.3%. Of these 154 patients, 37% experienced late postoperative problems. Twenty-one patients required system revision within 36.5 months (mean, 8.57±11.3) of surgery. Reasons for revision were spurious shocks due to electrode fractures (3) or electrode adapter malfunction (2), inadequate signal from endocardial rate-sensing electrodes (3), superior vena cava or right ventricular coil migration (5), failure to correct tachyarrhythmias due to a postimplant rise in defibrillation threshold (5), or pulse generator failure (3). One of these patients required system removal for infection after revision of an endocardial lead. A further 32 patients received inappropriate shocks for atrial fibrillation with a rapid ventricular response or sinus tachycardia. Two of these patients also received shocks for ventricular tachycardia initiated by antitachycardia pacing triggered by atrial fibrillation. Ventricular pacing for bradycardia was associated with inappropriate shocks due to excessive autogain in 2 patients. Conclusions Despite the major diagnostic and therapeutic advantages of tiered-therapy ICDs, a significant proportion of patients continue to experience hardware-related complications or receive inappropriate shocks.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1161/01.cir.91.8.2204

Cites

14 of 14 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 14 references

Cited by

See all 158

Metadata

Authors
Nunain, S. O., Roelke, M., Trouton, T., Osswald, S., Kim, Y. H., Sosa-Suarez, G., Brooks, D. R., McGovern, B., Guy, M., Torchiana, D. F., Vlahakes, G. J., Garan, H., & Ruskin, J. N.
Venue
Circulation
Published
1995-04-15
Volume
91
Issue
8
Pages
2204-2213
ISSN
  • 0009-7322 (ISSN-L)
  • 1524-4539
OpenAlex
W1984232457
Cited by
158
References
14
FWCI
13.88
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Limitations and Late Complications of Third-Generation Automatic Cardioverter-Defibrillators. https://app.cochat.ai/papers/10.1161/01.cir.91.8.2204
The paper
Nunain, S. O., Roelke, M., Trouton, T., Osswald, S., Kim, Y. H., Sosa-Suarez, G., Brooks, D. R., McGovern, B., Guy, M., Torchiana, D. F., Vlahakes, G. J., Garan, H., & Ruskin, J. N. (1995). Limitations and Late Complications of Third-Generation Automatic Cardioverter-Defibrillators. Circulation. https://doi.org/10.1161/01.cir.91.8.2204

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.