The 2020 “WHO Technical Specifications for Automated Non-Invasive Blood Pressure Measuring Devices With Cuff”

John, Oommen, Campbell, Norm R.C., Brady, Tammy M., Farrell, Margret, Varghese, Cherian, Velazquez Berumen, Adriana, Velez Ruiz Gaitan, Laura A., Toffelmire, Nicola, Ameel, Mohammad, Mideksa, Mulugeta, Jaffe, Marc G., Schutte, Aletta E., Khan, Taskeen, Lopez Meneses, Laura Patricia · 2021 · Hypertension
Cited by 53

The DOI matches its Crossref record; 16 of the 16 references it cites were found in a registry or an index.

Abstract

High systolic blood pressure (BP) is the single leading modifiable risk factor for death worldwide. Accurate BP measurement is the cornerstone for screening, diagnosis, and management of hypertension. Inaccurate BP measurement is a leading patient safety challenge. A recent World Health Organization report has outlined the technical specifications for automated noninvasive clinical BP measurement with cuff. The report is applicable to ambulatory, home, and office devices used for clinical purposes. The report recommends that for routine clinical purposes, (1) automated devices be used, (2) an upper arm cuff be used, and (3) that only automated devices that have passed accepted international accuracy standards (eg, the International Organization for Standardization 81060-2; 2018 protocol) be used. Accurate measurement also depends on standardized patient preparation and measurement technique and a quiet, comfortable setting. The World Health Organization report provides steps for governments, manufacturers, health care providers, and their organizations that need to be taken to implement the report recommendations and to ensure accurate BP measurement for clinical purposes. Although, health and scientific organizations have had similar recommendations for many years, the World Health Organization as the leading governmental health organization globally provides a potentially synergistic nongovernment government opportunity to enhance the accuracy of clinical BP assessment.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1161/hypertensionaha.120.16625

Cites

16 of 16 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 16 references

Research works printed but not linked: 14

  1. 10.1001/jama.2013.184182 · DOI · Paper page
  2. 10.1016/S0140-6736(16)31134-5 · DOI · Paper page
  3. World Health Organization. WHO technical specifications for automated non-invasive blood pressure measuring devices with cuff. World Health Organization; 2020:1–68.
  4. 10.1111/jch.12336 · DOI · Paper page
  5. 10.1111/jch.13735 · DOI
  6. 10.1111/jch.13655 · DOI
  7. Campbell, NR, McKay, DW. Accurate blood pressure measurement: why does it matter? CMAJ. 1999;161:277–278.
  8. 10.1111/jch.13915 · DOI
  9. 10.1001/jama.289.8.1027 · DOI · Paper page
  10. World Health Assembly. Health technologies. Sixtieth World Health Assembly Resolution WHA60.29. World Health Organization; 2007.
  11. World Health Organization. Affordable Technology: Blood Pressure Measuring Devices for Low Resource Settings. World Health Organization; 2005:92 4 159264 8.
  12. 10.1161/CIRCULATIONAHA.120.046205 · DOI
  13. World Health Organization. Fourth WHO Global Forum on Medical Devices Report. World Health Organization; 2019:1–174.
  14. 10.1111/jch.14017 · DOI · Paper page

Cited by

See all 53

Metadata

Authors
John, O., Campbell, N. R. C., Brady, T. M., Farrell, M., Varghese, C., Velazquez Berumen, A., Velez Ruiz Gaitan, L. A., Toffelmire, N., Ameel, M., Mideksa, M., Jaffe, M. G., Schutte, A. E., Khan, T., & Lopez Meneses, L. P.
Affiliations
  • From the George Institute for Global Health, University of New South Wales, New Delhi, India (O.J.)
  • Department of Medicine, Physiology and Pharmacology and Community Health Sciences, Libin Cardiovascular Institute of Alberta, University of Calgary, Canada (N.R.C.C.)
  • Department of Pediatrics, Division of Nephrology, Johns Hopkins University School of Medicine, Baltimore, MD (T.M.B.)
  • Resolve to Save Lives, an initiative of Vital Strategies, New York, NY (M.F.)
  • Cross Cutting Lead, Non-Communicable Diseases and Special Initiatives (C.V.), World Health Organization, Geneva, Switzerland
  • Team Lead Medical Devices and In Vitro Diagnostics (A.V.B.), World Health Organization, Geneva, Switzerland
  • Biomedical Engineer, Operations Support and Logistic, WHO Health Emergency Programme (L.A.V.R.G.)
  • Department of Non-Communicable Diseases (N.T.), World Health Organization, Geneva, Switzerland
  • Healthcare Technology Division, National Health Systems Resource Centre, Baba Gangnath Marg, Munirka, New Delhi, India (M.A.)
  • Biomedical Engineer, Medical Service Directorat, Federal Ministry of Health, Ethiopia (M.M.)
  • Department of Endocrinology, Kaiser Permanente San Francisco Medical Center, CA (M.G.J.)
  • School of Population Health, University of New South Wales (A.E.S.)
  • Department of Non-Communicable Diseases (T.K.), World Health Organization, Geneva, Switzerland
  • Biomedical Engineer, Independent Senior Consultant, Veracrus, México (L.P.L.M.).
Venue
Hypertension
Published
2021-03-03
Volume
77
Issue
3
Pages
806-812
ISSN
  • 0194-911X
  • 1524-4563
Cited by
53
References
16
Language
en

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: The 2020 “WHO Technical Specifications for Automated Non-Invasive Blood Pressure Measuring Devices With Cuff”. https://app.cochat.ai/papers/10.1161/hypertensionaha.120.16625
The paper
John, O., Campbell, N. R. C., Brady, T. M., Farrell, M., Varghese, C., Velazquez Berumen, A., Velez Ruiz Gaitan, L. A., Toffelmire, N., Ameel, M., Mideksa, M., Jaffe, M. G., Schutte, A. E., Khan, T., & Lopez Meneses, L. P. (2021). The 2020 “WHO Technical Specifications for Automated Non-Invasive Blood Pressure Measuring Devices With Cuff”. Hypertension. https://doi.org/10.1161/hypertensionaha.120.16625

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.