A Viscoelastic Biomechanical Model of the Cornea Describing the Effect of Viscosity and Elasticity on Hysteresis
The DOI matches its Crossref record; 21 of the 21 references it cites were found in a registry or an index.
Abstract
PURPOSE: To develop a method for evaluating viscosity and elasticity of the cornea and to examine the effect that both properties have on hysteresis. METHODS: A three-component spring and dashpot model was created in Simulink in Matlab to represent the purely elastic and viscoelastic behavior of the cornea during a measurement using device called an ocular response analyzer (ORA). Values for elasticity and viscosity were varied while sinusoidal stress was applied to the model. The simulated stresses were used to determine how hysteresis is affected by the individual components of elasticity, viscosity, and maximum stress. To validate the model, high-speed photography was used to measure induced strain in a corneal phantom during ORA measurement. This measured strain was compared with the strains simulated by the model. RESULTS: When the spring in the viscoelastic portion of the model was stiffened, hysteresis decreased. When the spring in the purely elastic element was stiffened, hysteresis increased. If both springs were stiffened together, hysteresis peaked strongly as a function of the viscosity of the viscoelastic element. Below the peak value, lower elasticity was associated with higher hysteresis. Above the peak value, higher elasticity was associated with higher hysteresis. In addition, hysteresis increased as the air maximum pressure was increased. Measurements from phantom corresponded to predictions from the model. CONCLUSIONS: A viscoelastic model is presented to illustrate how changing viscosity and elasticity may affect hysteresis. Low hysteresis can be associated with either high elasticity or low elasticity, depending on the viscosity, a finding consistent with clinical reports.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Kass, M. A. (2002) · Archives of Ophthalmology · doi:10.1001/archopht.120.6.701 · Cited by 3,705
Roark's Formulas for Stress and Strain
Roark, R. J., & Young, W. C. (1938) · Cited by 3,020
Determining in vivo biomechanical properties of the cornea with an ocular response analyzer
Luce, D. A. (2005) · Journal of Cataract and Refractive Surgery · doi:10.1016/j.jcrs.2004.10.044 · Cited by 1,311
Wollensak, G., Spoerl, E., & Seiler, T. (2003) · Journal of Cataract and Refractive Surgery · doi:10.1016/s0886-3350(03)00407-3 · Cited by 1,024
Liu, J., & Roberts, C. J. (2005) · Journal of Cataract and Refractive Surgery · doi:10.1016/j.jcrs.2004.09.031 · Cited by 753
Self-citationCouncil for International Organizations of Medical Sciences
Bankowski, Z. (1991) · Health Systems Research · doi:10.1007/978-3-642-61250-3_3 · Cited by 575
Biomechanical properties of keratoconus and normal corneas
Andreassen, T. T., Hjorth Simonsen, A., & Oxlund, H. (1980) · Experimental Eye Research · doi:10.1016/s0014-4835(80)80027-3 · Cited by 540 · cited 1×
“0 Larc L0/L0 ( 10 ) Using the values for stress experienced by the corneal surface and the experimental strain on the corneal surface, E1 ,E 2, and 1 were determined using equations 1 and 2 and implemented in a computer simulation created with Simulink in Matlab R2006b.”
MethodCentral Corneal Thickness and Corneal Hysteresis Associated With Glaucoma Damage
Congdon, N. G., Broman, A. T., Bandeen-Roche, K., Grover, D., & Quigley, H. A. (2006) · American Journal of Ophthalmology · doi:10.1016/j.ajo.2005.12.007 · Cited by 507 · cited 3×
“E2(0 1) E11 1˙1 ( 1 ) E2(0 1 )( 2)”
BackgroundInfluentialKotecha, A., Elsheikh, A., Roberts, C. R., Zhu, H., & Garway-Heath, D. F. (2006) · Investigative Opthalmology & Visual Science · doi:10.1167/iovs.06-0557 · Cited by 423
Self-citationAssessment of Corneal Biomechanical Properties and Their Variation with Age
Elsheikh, A., Wang, D., Brown, M., Rama, P., Campanelli, M., & Pye, D. (2007) · Current Eye Research · doi:10.1080/02713680601077145 · Cited by 394
Nonlinear material properties of intact cornea and sclera
Woo, S. L., KOBAYASHI, A. S., Schlegel, W. A., & Lawrence, C. (1972) · Experimental Eye Research · doi:10.1016/0014-4835(72)90139-x · Cited by 345
A comparison of biomechanical properties between human and porcine cornea
Zeng, Y., Yang, J., Huang, K., Lee, Z., & Lee, X. (2001) · Journal of Biomechanics · doi:10.1016/s0021-9290(00)00219-0 · Cited by 316
Determination of the True Intraocular Pressure and Modulus of Elasticity of the Human Cornea in vivo
Orssengo, G. (1999) · Bulletin of Mathematical Biology · doi:10.1006/bulm.1999.0102 · Cited by 304
Strip Extensiometry for Comparison of the Mechanical Response of Bovine, Rabbit, and Human Corneas
Hoeltzel, D. A., Altman, P., Buzard, K., & Choe, K.-I. (1992) · Journal of Biomechanical Engineering · doi:10.1115/1.2891373 · Cited by 263
Downs, J. C., Suh, J.-K. F., Thomas, K. A., Bellezza, A. J., Hart, R. T., & Burgoyne, C. F. (2005) · Investigative Ophthalmology & Visual Science · doi:10.1167/iovs.04-0114 · Cited by 252
Regional elastic performance of the human cornea
Hjortdal, J. Ø. (1996) · Journal of Biomechanics · doi:10.1016/0021-9290(95)00152-2 · Cited by 231
The mechanical properties of the rabbit and human cornea
Jue, B., & Maurice, D. M. (1986) · Journal of Biomechanics · doi:10.1016/0021-9290(86)90135-1 · Cited by 222
Comparison of mechanical properties of keratoconus and normal corneas
Nash, I. S., Greene, P. S., & Foster, C. S. (1982) · Experimental Eye Research · doi:10.1016/0014-4835(82)90040-9 · Cited by 204 · cited 1×
“h h0 hdepressed ( 6 ) Using h0 and R0, the original arc length of the cornea (L0 )i s calculated with the use of equation 7. As the air pressure applied to the cornea increases, the arc length of the cornea decreases, and the changing arc length is found using either equation 8 or 9 as applicable.”
MethodA microstructurally-based finite element model of the incised human cornea
Pinsky, P. M., & Datye, D. V. (1991) · Journal of Biomechanics · doi:10.1016/0021-9290(91)90169-n · Cited by 159
Influence of Corneal Variables on Accuracy of Intraocular Pressure Measurement
Damji, K. F., Muni, R. H., & Munger, R. (2003) · Journal of Glaucoma · doi:10.1097/00061198-200302000-00015 · Cited by 104
Viscoelastic properties of human cornea
KOBAYASHI, A. S., Staberg, L. G., & Schlegel, W. A. (1973) · Experimental Mechanics · doi:10.1007/bf02322337 · Cited by 44
Cited by
Corneal Biomechanics and Biomaterials
Ruberti, J. W., Sinha Roy, A., & Roberts, C. J. (2011) · Annual Review of Biomedical Engineering · doi:10.1146/annurev-bioeng-070909-105243 · Cited by 241
Self-citationCorneal Deformation Measurement Using Scheimpflug Noncontact Tonometry
Hon, Y., & Lam, A. K. C. (2013) · Optometry and Vision Science · doi:10.1097/opx.0b013e318279eb87 · Cited by 235
Designing Scaffolds for Corneal Regeneration
Ahearne, M., Fernández‐Pérez, J., Masterton, S., Madden, P. W., & Bhattacharjee, P. (2020) · Advanced Functional Materials · doi:10.1002/adfm.201908996 · Cited by 174
Blackburn, B. J., Jenkins, M. W., Rollins, A. M., & Dupps, W. J. (2019) · Frontiers in Bioengineering and Biotechnology · doi:10.3389/fbioe.2019.00066 · Cited by 167
“Both viscous and elastic properties resist deformation, but viscous material resists deformation from applied force over time, while elastic materials more immediately deform with applied force and return to the original state when that force is removed (Glass et al., 2008).”
BackgroundGuo, H., Hosseini-Moghaddam, S. M., & Hodge, W. (2019) · BMC Ophthalmology · doi:10.1186/s12886-019-1165-3 · Cited by 159
Metadata
- Authors
- Glass, D. H., Roberts, C. J., Litsky, A. S., & Weber, P. A.
- Venue
- Investigative Opthalmology & Visual Science
- Published
- 2008-09-01
- Volume
- 49
- Issue
- 9
- Pages
- 3919
- ISSN
- 0146-0404 (ISSN-L)
- 1552-5783
- OpenAlex
- W2069067786
- Cited by
- 201
- References
- 21
- FWCI
- 7.88
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.