Effects of different contacting pressure on the transfer function between finger photoplethysmographic and radial blood pressure waveforms

Hsiu, H, Hsu, C-L, Wu, T-L·2011·Proceedings of the Institution of Mechanical Engineers Part H Journal of Engineering in Medicine

Abstract

BACKGROUND: The blood pressure (BP) waveform is suggested to reflect the whole-body blood supply distribution, but its non-invasive assessment is not sufficiently user friendly for practical applications. The present authors studied the correlation between BP and photoplethysmography (PPG) waveforms, with the aim of determining the optimal range for contact pressure stimulation (PS) to produce a reliable transfer function in their harmonic parameters. METHODS: Finger PPG and radial-artery BP signals were measured simultaneously and noninvasively on healthy volunteers (n = 45). PS of 0-200 mmHg was applied to the finger, and 1 min data sequences were recorded. In frequency-domain analysis, linear regression was applied to the calculated amplitude ratios or the first five harmonics between BP and PPG waveforms. RESULTS: In the 60 mmHg-PS group, the BP-PPG regression of amplitude ratios was highest, and the agreements between them were also the best verified by Bland-Altman analysis. CONCLUSION: In the present study, frequency-domain analysis was performed to study the correlation between BP and PPG waveforms. Differences in pressure-induced-vasodilation responses underlie the different BP-PPG waveform correlations obtained by applying different PSs. The non-invasively derived PPG parameters might help to provide an easier method to acquire the radial-artery BP waveform, and hence broaden the application of BP waveform analysis.

Integrity signals

Registry

Crossref · 10.1177/0954411910396288

Cites

19 of 19 references verified0 citation errors · 0 not checkable

Statistics

Recomputing p-values needs the paper’s full text.

Claims the paper makes

This paper’s full text isn’t openly available, so its claims and self-check weren’t run. The registry and citation checks still apply.

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

  • The change in these fundamental features can affect other measurements such as the b/a ratio, the relationship between the frequency response of a PPG signal and a blood pressure signal, or the pulse transit time (PTT).

    “The change in these fundamental features can affect other measurements such as the b/a ratio, derived from the a and b peaks of the second derivative of a PPG signal, used to characterize arterial stiffness [132], the relationship between the frequency response of a PPG signal and a blood pressure signal [207], or the pulse transit time (PTT) which is used as an indicator for fluctuating stiffness”
  • At 60 mmHg, there were the greatest R2 values for four out of five harmonics, and with different applied pressures, the R2 values ranged from 0.13 to 0.77 for the first harmonic.

    “At 60 mmHg, there were the greatest R2 values for four out of five harmonics, and with the different applied pressures, the R2 values ranged from 0.13 to 0.77 for the first harmonic.”

Cites

Flagged first, then by citations

Cited by

Sorted by citations

Authors
Hsiu, H., Hsu, C.-L., & Wu, T.-L.
Affiliations
  • National Taiwan University of Science and Technology (TW)
  • Yuan Ze University (TW)
Corresponding author
Hsiu, H
ORCID
Hsiu, H
Venue
Proceedings of the Institution of Mechanical Engineers Part H Journal of Engineering in Medicine
Published
2011-06
Volume
225
Issue
6
Pages
575-583
ISSN
  • 0954-4119 (ISSN-L)
  • 2041-3033
OpenAlex
W2323988838
Cited by
46
References
19
Citations by year
2026: 1 · 2025: 2 · 2024: 1 · 2023: 9 · 2022: 3 · 2021: 4 · 2020: 5 · 2019: 5 · 2018: 3 · 2017: 3 · 2016: 1 · 2015: 2 · 2014: 3 · 2013: 2 · 2012: 1
FWCI
1.32
Type
journal article
Language
en
Open access
closed
Locations
  • Proceedings of the Institution of Mechanical Engineers Part H Journal of Engineering in Medicine (publishedVersion)
  • PubMed (publishedVersion)
Funders
National Science Council
Topics
  • Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
  • Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Cardiovascular Health and Disease Prevention
  • Health Sciences › Medicine › Surgery › Hemodynamic Monitoring and Therapy
Keywords
finger photoplethysmography, amplitude ratio, harmonic analysis, non-invasive blood pressure measurement, frequency domain analysis
MeSH terms
Adult, Algorithms, Blood Pressure, Blood Pressure Determination, Fingers, Humans, Male, Linear Models, Photoplethysmography, Radial Artery

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review