Effects of different contacting pressure on the transfer function between finger photoplethysmographic and radial blood pressure waveforms
Abstract
BACKGROUND: The blood pressure (BP) waveform is suggested to reflect the whole-body blood supply distribution, but its non-invasive assessment is not sufficiently user friendly for practical applications. The present authors studied the correlation between BP and photoplethysmography (PPG) waveforms, with the aim of determining the optimal range for contact pressure stimulation (PS) to produce a reliable transfer function in their harmonic parameters. METHODS: Finger PPG and radial-artery BP signals were measured simultaneously and noninvasively on healthy volunteers (n = 45). PS of 0-200 mmHg was applied to the finger, and 1 min data sequences were recorded. In frequency-domain analysis, linear regression was applied to the calculated amplitude ratios or the first five harmonics between BP and PPG waveforms. RESULTS: In the 60 mmHg-PS group, the BP-PPG regression of amplitude ratios was highest, and the agreements between them were also the best verified by Bland-Altman analysis. CONCLUSION: In the present study, frequency-domain analysis was performed to study the correlation between BP and PPG waveforms. Differences in pressure-induced-vasodilation responses underlie the different BP-PPG waveform correlations obtained by applying different PSs. The non-invasively derived PPG parameters might help to provide an easier method to acquire the radial-artery BP waveform, and hence broaden the application of BP waveform analysis.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The change in these fundamental features can affect other measurements such as the b/a ratio, the relationship between the frequency response of a PPG signal and a blood pressure signal, or the pulse transit time (PTT).
“The change in these fundamental features can affect other measurements such as the b/a ratio, derived from the a and b peaks of the second derivative of a PPG signal, used to characterize arterial stiffness [132], the relationship between the frequency response of a PPG signal and a blood pressure signal [207], or the pulse transit time (PTT) which is used as an indicator for fluctuating stiffness”
From Sources of inaccuracy in photoplethysmography for continuous cardiovascular monitoring (2021) · cites this paper as [207]
At 60 mmHg, there were the greatest R2 values for four out of five harmonics, and with different applied pressures, the R2 values ranged from 0.13 to 0.77 for the first harmonic.
“At 60 mmHg, there were the greatest R2 values for four out of five harmonics, and with the different applied pressures, the R2 values ranged from 0.13 to 0.77 for the first harmonic.”
From Sources of inaccuracy in photoplethysmography for continuous cardiovascular monitoring (2021) · cites this paper as [207]
Cites
Clinical measurement of arterial stiffness obtained from noninvasive pressure waveforms
NICHOLS, W. (2005) · American Journal of Hypertension · doi:10.1016/j.amjhyper.2004.10.009 · Cited by 1,002 · see citing works
The effect of contacting force on photoplethysmographic signals
Teng, X. F., & Zhang, Y. T. (2004) · Physiological Measurement · doi:10.1088/0967-3334/25/5/020 · Cited by 171 · see citing works
The relationship between the photoplethysmographic waveform and systemic vascular resistance
Alian, A. A., Haddadin, A. S., Tantawy, H., Badr, T., Stout, R. G., Silverman, D. G., & Shelley, K. H. (2007) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-007-9097-5 · Cited by 136 · see citing works
Tyan, C.-C., Liu, S.-H., Chen, J., et al. (2007) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2007.910681 · Cited by 92 · see citing works
Allen, J., & Murray, A. (1999) · Physiological Measurement · doi:10.1088/0967-3334/20/3/306 · Cited by 71 · see citing works
TSAI, W., CHEN, J., Wang, M., WU, H., CHI, C., Chen, Y., CHEN, J., & Lin, L.-J. (2005) · American Journal of Hypertension · doi:10.1016/j.amjhyper.2005.03.739 · Cited by 70 · see citing works
Dynamics of Local Pressure-Induced Cutaneous Vasodilation in the Human Hand
Abraham, P., Fromy, B., Merzeau, S., et al. (2001) · Microvascular Research · doi:10.1006/mvre.2000.2290 · Cited by 57 · see citing works
Verylow frequency variability in arterial blood pressure and blood volume pulse
Nitzan, M., Babchenko, A., & Khanokh, B. (1999) · Medical & Biological Engineering & Computing · doi:10.1007/bf02513266 · Cited by 56 · see citing works
Awad, A. A., Ghobashy, M. A. M., Stout, R. G., Silverman, D. G., & Shelley, K. H. (2001) · Anesthesia & Analgesia · doi:10.1097/00000539-200112000-00022 · Cited by 50 · see citing works
Wang, W. K., Hsu, T. L., Chang, H. C., et al. (2000) · The American Journal of Chinese Medicine · doi:10.1142/s0192415x00000076 · Cited by 44 · see citing works
Middleton, P. M., Chan, G. S. H., O’Lone, E., Steel, E., Carroll, R., Celler, B. G., & Lovell, N. H. (2008) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-008-9140-1 · Cited by 41 · see citing works
Hsiu, H., Hsu, W.-C., Chen, B.-H., et al. (2010) · Physiological Measurement · doi:10.1088/0967-3334/31/6/007 · Cited by 32 · see citing works
Self-citationMicrocirculatory effect of different skin contacting pressures around the blood pressure
Hsiu, H., Hsu, W.-C., Chang, S.-L., et al. (2008) · Physiological Measurement · doi:10.1088/0967-3334/29/12/006 · Cited by 30 · see citing works
Self-citationAlteration of Pulse in Human Subjects by Three Chinese Herbs
Wang, W. K., Chen, H. L., Hsu, T. L., et al. (1994) · The American Journal of Chinese Medicine · doi:10.1142/s0192415x94000243 · Cited by 27 · see citing works
Hsiu, H., Hsu, W.-C., Hsu, C.-L., Jan, M.-Y., & Wang, Y. L. (2008) · Lasers in Medical Science · doi:10.1007/s10103-008-0610-y · Cited by 21 · see citing works
Self-citationHsiu, H., Hsu, W.-C., Hsu, C.-L., Huang, S.-M., & Lin, Y.-Y. W. (2009) · Photomedicine and Laser Surgery · doi:10.1089/pho.2008.2390 · Cited by 17 · see citing works
Self-citationConnection between RR-interval length and the pulsatile microcirculatory flow
Hsiu, H., Hsu, C.-L., Chiang, W.-R., et al. (2008) · Physiological Measurement · doi:10.1088/0967-3334/29/2/007 · Cited by 17 · see citing works
Self-citationThe possible role of arterial radial vibration in heart rate and blood pressure matching
Hsiu, H., Chao, P.-T., Hsu, W.-C., Jan, M. Y., Wang, Y.-Y. L., & Wang, W. K. (2008) · Proceedings of the Institution of Mechanical Engineers Part H Journal of Engineering in Medicine · doi:10.1243/09544119jeim349 · Cited by 15 · see citing works
Self-citationHsiu, H., Jan, M. Y., Wang, W. K., et al. (2006) · Physiological Measurement · doi:10.1088/0967-3334/27/2/004 · Cited by 4 · see citing works
Self-citation
Flagged first, then by citations
Cited by
Sorted by citations
Wearable photoplethysmographic sensors—past and present
Tamura, T., Maeda, Y., Sekine, M., & Yoshida, M. (2014) · Electronics · doi:10.3390/electronics3020282 · Cited by 912 · see citing works
“In one particular study, blood pressure and finger PPG waveforms were simultaneously measured in healthy volunteers under different contact pressures ranging from 0 to 200 mmHg [42].”
MethodSources of inaccuracy in photoplethysmography for continuous cardiovascular monitoring
Fine, J., Branan, K. L., Rodriguez, A. J., Boonya-ananta, T., Ajmal, Ramella-Roman, J. C., McShane, M. J., & Coté, G. L. (2021) · Biosensors · doi:10.3390/bios11040126 · Cited by 348 · see citing works
A new, short-recorded photoplethysmogram dataset for blood pressure monitoring in China
Liang, Y., Chen, Z., Liu, G., & Elgendi, M. (2018) · Scientific Data · doi:10.1038/sdata.2018.20 · Cited by 215 · see citing works
Scardulla, F., D’Acquisto, L., Colombarini, R., Hu, S., Pasta, S., & Bellavia, D. (2020) · Sensors · doi:10.3390/s20185052 · Cited by 72 · see citing works
A contact-force regulated photoplethysmography (PPG) platform
Sim, J. K., Ahn, B., & Doh, I. (2018) · AIP Advances · doi:10.1063/1.5020914 · Cited by 27 · see citing works
- Authors
- Hsiu, H., Hsu, C.-L., & Wu, T.-L.
- Affiliations
- National Taiwan University of Science and Technology (TW)
- Yuan Ze University (TW)
- Corresponding author
- Hsiu, H
- ORCID
- Hsiu, H
- Venue
- Proceedings of the Institution of Mechanical Engineers Part H Journal of Engineering in Medicine
- Publisher
- SAGE Publications
- Published
- 2011-06
- Volume
- 225
- Issue
- 6
- Pages
- 575-583
- ISSN
- 0954-4119 (ISSN-L)
- 2041-3033
- OpenAlex
- W2323988838
- Cited by
- 46
- References
- 19
- Citations by year
- 2026: 1 · 2025: 2 · 2024: 1 · 2023: 9 · 2022: 3 · 2021: 4 · 2020: 5 · 2019: 5 · 2018: 3 · 2017: 3 · 2016: 1 · 2015: 2 · 2014: 3 · 2013: 2 · 2012: 1
- FWCI
- 1.32
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- Proceedings of the Institution of Mechanical Engineers Part H Journal of Engineering in Medicine (publishedVersion)
- PubMed (publishedVersion)
- Funders
- National Science Council
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Cardiovascular Health and Disease Prevention
- Health Sciences › Medicine › Surgery › Hemodynamic Monitoring and Therapy
- Keywords
- finger photoplethysmography, amplitude ratio, harmonic analysis, non-invasive blood pressure measurement, frequency domain analysis
- MeSH terms
- Adult, Algorithms, Blood Pressure, Blood Pressure Determination, Fingers, Humans, Male, Linear Models, Photoplethysmography, Radial Artery
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review