Cited by 27Non-Invasive Vital Sign MonitoringAdvanced Sensor and Energy Harvesting MaterialsHeart Rate Variability and Autonomic Control

A contact-force regulated photoplethysmography (PPG) platform

Open access
Sim, Jai Kyoung, Ahn, Bongyoung, Doh, Il·2018·AIP Advances

Abstract

A photoplethysmography (PPG) platform integrated with a miniaturized force-regulator is proposed in this study. Because a thermo-pneumatic type regulator maintains a consistent contact-force between the PPG probe and the measuring site, a consistent and stable PPG signal can be obtained. We designed and fabricated a watch-type PPG platform with an overall size of 35 mm × 19 mm. In the PPG measurement on the radial artery wrist while posture of the wrist is changed to extension, neutral, or flexion, regulation of the contact-force provides consistent PPG measurements for which the variations in the PPG amplitude (PPGA) was 7.2 %. The proposed PPG platform can be applied to biosignal measurements in various fields such as PPG-based ANS monitoring to estimate nociception, sleep apnea syndrome, and psychological stress.

Integrity signals

Registry

Crossref · 10.1063/1.5020914

Cites

28 of 28 references verified0 citation errors · 0 not checkable

Statistics

Recomputing p-values needs the paper’s full text.

Claims the paper makes

We couldn’t retrieve this paper’s full text, so its claims and self-check weren’t run. The registry and citation checks still apply.

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

  • Without force regulation, the coefficient of variation between five posture stages was 50.9%; the addition of the force regulator dropped the coefficient of variation to 1.8%.

    “Without force regulation, the coefficient of variation between five posture stages was calculated at 50.9%. The addition of the force regulator dropped the coefficient of variation to 1.8%.”

Cites

Flagged first, then by citations

Cited by

Sorted by citations

Authors
Sim, J. K., Ahn, B., & Doh, I.
Affiliations
Korea Research Institute of Standards and Science (KR)
Venue
AIP Advances
Published
2018-04-01
Volume
8
Issue
4
ISSN
2158-3226 (ISSN-L)
OpenAlex
W2795906808
Cited by
27
References
28
Citations by year
2026: 4 · 2025: 5 · 2024: 4 · 2023: 2 · 2022: 5 · 2021: 3 · 2020: 2 · 2019: 2
FWCI
0.70
Type
journal article
Language
en
Open access
Locations
  • AIP Advances (publishedVersion, cc-by)
  • DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
Funders
  • Korea Research Institute of Standards and Science (18011083)
  • National Research Foundation of Korea (17442002)
Topics
  • Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
  • Physical Sciences › Engineering › Biomedical Engineering › Advanced Sensor and Energy Harvesting Materials
  • Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Heart Rate Variability and Autonomic Control
Keywords
photoplethysmography, contact force control, wrist posture, nociception, sleep apnea, psychological stress

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review