Glucose Meters: A Review of Technical Challenges to Obtaining Accurate Results
Open accessThe DOI matches its Crossref record; 44 of the 45 references it cites were found in a registry or an index, and 1 are not checked yet.
Abstract
Glucose meters are universally utilized in the management of hypoglycemic and hyperglycemic disorders in a variety of healthcare settings. Establishing the accuracy of glucose meters, however, is challenging. Glucose meters can only analyze whole blood, and glucose is unstable in whole blood. Technical accuracy is defined as the closeness of agreement between a test result and the true value of that analyte. Truth for glucose is analysis by isotope dilution mass spectrometry, and frozen serum standards analyzed by this method are available from the National Institute of Standards and Technology. Truth for whole blood has not been established, and cells must be separated from the whole blood matrix before analysis by a method like isotope dilution mass spectrometry. Serum cannot be analyzed by glucose meters, and isotope dilution mass spectrometry is not commonly available in most hospitals and diabetes clinics to evaluate glucose meter accuracy. Consensus standards recommend comparing whole blood analysis on a glucose meter against plasma/serum centrifuged from a capillary specimen and analyzed by a clinical laboratory comparative method. Yet capillary samples may not provide sufficient volume to test by both methods, and venous samples may be used as an alternative when differences between venous and capillary blood are considered. There are thus multiple complexities involved in defining technical accuracy and no clear consensus among standards agencies and professional societies on accuracy criteria. Clinicians, however, are more concerned with clinical agreement of the glucose meter with a serum/plasma laboratory result. Acceptance criteria for clinical agreement vary across the range of glucose concentrations and depend on how the result will be used in screening or management of the patient. A variety of factors can affect glucose meter results, including operator technique, environmental exposure, and patient factors, such as medication, oxygen therapy, anemia, hypotension, and other disease states. This article reviews the challenges involved in obtaining accurate glucose meter results.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
National Committee for Clinical Laboratory Standards
Bruck, E. (1980) · Pediatrics · doi:10.1542/peds.65.1.187 · Cited by 14,751
Intensive insulin therapy in the medical ICU
Van den Berghe, G., Wilmer, A., Hermans, G., Meersseman, W., Wouters, P. J., Milants, I., Van Wijngaerden, E., Bobbaers, H., & Bouillon, R. (2006) · New England Journal of Medicine · doi:10.1056/nejmoa052521 · Cited by 3,639
Tietz Textbook of Clinical Chemistry
(1999) · Archives of Pathology and Laboratory Medicine · doi:10.5858/123.7.646b · Cited by 2,495
Capes, S. E., Hunt, D., Malmberg, K., Pathak, P., & Gerstein, H. C. (2001) · Stroke · doi:10.1161/hs1001.096194 · Cited by 1,824
Evaluating clinical accuracy of systems for self-monitoring of blood glucose
Clarke, W. L., Cox, D., Gonder-Frederick, L. A., Carter, W., & Pohl, S. L. (1987) · Diabetes Care · doi:10.2337/diacare.10.5.622 · Cited by 1,435
Standards of Medical Care in Diabetes—2008
Dora, J. M., Kramer, C. K., & Canani, L. H. (2008) · Diabetes Care · doi:10.2337/dc08-0109 · Cited by 1,409
Standards of Medical Care in Diabetes—2008
American Diabetes Association (2008) · Diabetes Care · doi:10.2337/dc08-s012 · Cited by 1,377
Parkes, J. L., Slatin, S. L., Pardo, S., & Ginsberg, B. H. (2000) · Diabetes Care · doi:10.2337/diacare.23.8.1143 · Cited by 620
Vlasselaers, D., Milants, I., Desmet, L., Wouters, P. J., Vanhorebeek, I., van den Heuvel, I., Mesotten, D., Casaer, M. P., Meyfroidt, G., Ingels, C., et al. (2009) · The Lancet · doi:10.1016/s0140-6736(09)60044-1 · Cited by 554
Association of Hyperglycemia with Increased Mortality after Severe Burn Injury
Gore, D. C., Chinkes, D., Heggers, J., Herndon, D. N., Wolf, S. E., & Desai, M. (2001) · Journal of Trauma and Acute Care Surgery · doi:10.1097/00005373-200109000-00021 · Cited by 459
Research works printed but not linked: 9
- Mendosa D. Meter memories. How Tom, Dick, and Charlie did it. Diabetes Wellness Letter. 2000;1(6) [Google Scholar]
- Bernstein RK. London: Little Brown; 1997. Dr. Bernstein's diabetes solution. A complete guide to achieving normal blood sugars. [Google Scholar]
- International Organization for Standardization. ISO/IEC Guide 99. Geneva: International Organization for Standardization; 2007. International vocabulary of metrology—Basic and general concepts and associated terms (VIM) [Google Scholar]
- International Organization for Standardization. Determination of performance criteria for in vitro blood glucose monitoring systems for management of human diabetes mellitus. Geneva: International Organization for Standardization. 2002 ISO 15197. [Google Scholar]
- U.S. Food and Drug Administration. Rockville: Food and Drug Administration. 1989. Review criteria for assessment of portable invasive blood glucose monitoringin vitro diagnostic devices which use glucose oxidase, dehydrogenase, or hexokinase methodology. [Google Scholar]
- Canadian Standards Association. CSA Z316.4-94. Mississauga: Canadian Standards Association. 1994. Performance specifications for portable whole blood glucose monitor systems for use in diabetes management. [Google Scholar]
- Instituto Mexicano del Seguro Social. Mexico City: Instituto Mexicano del Seguro Social. 1998. Reactive strip assessment protocol for determination of glucose in blood. [Google Scholar]
- TNO Center for Medical Technology. Netherlands: TNO Center for Medical Technology; 1991. Quality guideline for non-implantable portable blood glucose monitors for self-monitoring. TNO-CMT/90.021E. [Google Scholar]
- Meneghini LF, Arce RL. Blood glucose data from meter memories often not useful for pattern recognition. Abstract presented at American Diabetes Association 65th Annual Meeting. Diabetes. 2005;54(Suppl 1:A95) [Google Scholar]
Cited by
Items 301–400 of 433
Chesa, J. L. B., Nieto, M. L. H., Hernández, F. J. L., Oliver, P., & de la Presa, B. G. (2017) · Revista del Laboratorio Clínico · doi:10.1016/j.labcli.2017.01.003 · Cited by 2
Slugocki, M., Bialonczyk, D., & Özdener-Poyraz, A. E. (2018) · Journal of Pharmacy Technology · doi:10.1177/8755122518813889 · Cited by 2
Nayeem, J., Kamaluddin, S. M., Chowdhury, H. A., & Ali, L. (2019) · International Journal of Diabetes in Developing Countries · doi:10.1007/s13410-019-00712-7 · Cited by 2
Utility of point‐of‐care vs reference laboratory testing for the evaluation of glucose levels
Andriankaja, O. M., Muñoz‐Torres, F. J., Vergara, J. L., Pérez, C. M., & JOSHIPURA, K. J. (2019) · Diabetic Medicine · doi:10.1111/dme.13922 · Cited by 2
Lim, J., & Kwon, G. C. (2021) · Imsang byeong'ri wa jeongdo gwanri/Journal of laboratory medicine and quality assurance · doi:10.15263/jlmqa.2021.43.2.43 · Cited by 2
Naugler, C. T., Zhang, Z., & Redman, L. (2014) · Canadian Journal of Diabetes · doi:10.1016/j.jcjd.2014.04.004 · Cited by 2
Design and Development of an Internet of things Based Glucometer with Wireless Transmission
Okafor, S. A., Arukalam, F. M., Ekuma, I. C., Eziefuna, E. O., Ihetu, C. A., Okey-Mbata, C. C., Ezeamaku, U. L., Iheaturu, N. C., & Okafor, A. L. (2022) · Journal of Engineering Research and Reports · doi:10.9734/jerr/2022/v22i1217582 · Cited by 2
Fitriyani, N. L., Syafrudin, M., Ulyah, S. M., Alfian, G., Qolbiyani, S. L., Yang, C., Rhee, J., & Anshari, M. (2023) · Mathematics · doi:10.3390/math11102266 · Cited by 2
Tessou, K. Z., Lawson-Evi, P., Metowogo, K., Diallo, A., Eklu-Gadegkeku, K., Aklikokou, K. A., & Gbéassor, M. (2013) · International Journal of Biomedical Science · doi:10.59566/ijbs.2013.9255 · Cited by 2
Shukla, G., Vishvakarma, S., Sharma, N., Singh, S., Agarwal, R., Jha, P., Srivastava, S., Fatima, G., Tiwari, A. M., Gupta, P. K., et al. (2023) · Journal of basic and applied research international · doi:10.56557/jobari/2023/v29i58472 · Cited by 2
Wang, T., Zhang, L., Gong, L., Xin, Y., Han, L., Li, N., Peng, P., Zhao, X., & Li, R. (2024) · iLABMED · doi:10.1002/ila2.37 · Cited by 2
Naugler, C. T., Zhang, Z., & Redman, L. (2016) · Canadian Journal of Diabetes · doi:10.1016/j.jcjd.2016.04.004 · Cited by 2
RetractedPrerna, P., & Tripathy, S. S. (2020) · Lecture notes in electrical engineering · doi:10.1007/978-981-15-2854-5_40 · Cited by 1
Farfán-Albarracín, J. D., Ramírez-Sierra, C. L., Segura, O. M. E., Cárdenas, S. H. P., Prada, H. A. T., Rueda, M. C., Espitia, I. L., & Morales, A. M. B. (2020) · Brain and Development · doi:10.1016/j.braindev.2020.07.013 · Cited by 1
Salarizadeh, N., Shojai, S., Pebdeni, A. B., Nojoki, F., Zargar, S. J., & Habibi-Rezaei, M. (2023) · Elsevier eBooks · doi:10.1016/b978-0-323-90222-9.00020-0 · Cited by 2
Fukuoka, M., Yasuhi, I., Yamashita, H., Ashimoto, K., Kurata, N., Yamaguchi, J., Koga, M., Sugimi, S., Suga, S., & Fukuda, M. (2023) · Journal of Diabetes Research · doi:10.1155/2023/9574563 · Cited by 1
Estimation of Blood Glucose Level of Human by Measuring Key Parameters in Electrocardiogram
Lin, Y., Tsai, H., Li, C., Chen, C., Liu, C., & Lee, W. (2023) · IEEE International Instrumentation and Measurement Technology Conference · doi:10.1109/i2mtc53148.2023.10175995 · Cited by 2
Implementation of a Novel Personalized Non-Invasive Glucometer Using Machine Learning
Rajasingam, N., Padmanaban, S., Bhuvaneswari, B., Madhan Kumar, U., Imrankhan, R., & Kesavarani, G. (2024) · International Conference on Bio Signals, Images, and Instrumentation (ICBSII) · doi:10.1109/icbsii61384.2024.10564074 · Cited by 1
Patient and glucometer: how can glucose self-monitoring be improved?
Kononova, Y. A., Babenko, A. Y., & Bregovskiy, V. B. (2024) · Meditsinskiy sovet = Medical Council · doi:10.21518/ms2024-259 · Cited by 1
Jensen, O. C., Mujica, N. D., Gyawali, B., & Gyntelberg, F. (2025) · Primary care diabetes · doi:10.1016/j.pcd.2025.05.008 · Cited by 1
Garcia, M. R., Mann, N., Perelló, M., Silic, V., Sparks, T. H., & Titmarsh, H. F. (2025) · Frontiers in Veterinary Science · doi:10.3389/fvets.2025.1694081 · Cited by 1
Assessment of analytical performance of a glucometer in Indian adults at a tertiary care hospital
Khan, N., Khan, H., Baviskar, P. A., Mehta, S. P., Patro, H., Nair, J., Kotkar, S., & Rojekar, M. V. (2025) · BMC Endocrine Disorders · doi:10.1186/s12902-025-02118-5 · Cited by 1
Varalakshmi, S. D., Choudhari, K. S., Lewis, L. E. S., Koushik, H., Kartha, V. B., & Unnikrishnan, V. K. (2025) · Microchemical Journal · doi:10.1016/j.microc.2025.114595 · Cited by 1
Sánchez, C. C. G., Salazar, L. A. C., Delgadillo, D. S., Paredes‐Madrid, L., & Vargas, F. M. L. (2025) · IEEE Colombian Caribbean Conference (C3) · doi:10.1109/c366505.2025.11340406 · Cited by 1
Luka, S., Ishaku, A., Stephen, L. W., & Christian, I. O. (2026) · American Journal of Laboratory Medicine · doi:10.11648/j.ajlm.20261101.14 · Cited by 1
Patra, S., Sahu, K. M., & Swain, S. K. (2026) · International Journal of Biological Macromolecules · doi:10.1016/j.ijbiomac.2026.151286 · Cited by 1
Paul, T. K., Rakib, M. A., & Islam, M. S. (2026) · IEEE International Conference on Quantum Photonics, Artificial Intelligence & Networking (QPAIN) · doi:10.1109/qpain69676.2026.11546602 · Cited by 1
Gama, M. P. R., Cruzeta, Ossowski, Bay, & Michaelis (2012) · Research and Reports in Endocrine Disorders · doi:10.2147/rred.s24631 · Cited by 4
Intérêt du sang veineux de cordon ombilical dans le diagnostic de l’hypoglycémie néonatale
Djohan, Y. F., Cisse-Camara, M., Niamke, A. G., Koffi, G., Dere, L., Monde, A., Tiahou, G. G., Djessou, P., & Sess, D. (2016) · International Journal of Biological and Chemical Sciences · doi:10.4314/ijbcs.v9i5.26 · Cited by 1
Mdlalose, T. I., Christison, K. W., & Vosloo, A. (2017) · Aquaculture · doi:10.1016/j.aquaculture.2016.12.024 · Cited by 2
Glucose Meter Performance in a Tertiary Care Institution
Mogarekar, M. R., Phad, M. B., & Dhabe, M. G. (2016) · Point of Care The Journal of Near-Patient Testing & Technology · doi:10.1097/poc.0000000000000085 · Cited by 1
Avaliação da acurácia de dois glicosímetros veterinários para uso em cães
Correa, M. L., Leme, F. D. O. P., & da Costa‐Val, A. P. (2018) · Semina Ciências Biológicas e da Saúde · doi:10.5433/1679-0367.2018v39n1p19 · Cited by 1
Determination of Glucose and Continuous Glucose Monitoring
Wang, Y. F., & Jia, W. (2018) · Springer eBooks · doi:10.1007/978-981-10-7074-7_1 · Cited by 1
Megerssa, Y. C., & Jima, N. T. (2020) · doi:10.35248/2161-1009.19.8.387 · Cited by 1
Mondal, H., & Mondal, S. (2020) · Indian Journal of Medical Sciences · doi:10.25259/ijms_150_2020 · Cited by 1
Blutzuckerselbstmessung – wie kann ein korrekter Wert erzielt werden?
Baumstark, A., & Jendrike, N. R. M. (2020) · Diabetes aktuell · doi:10.1055/a-1296-1337 · Cited by 1
Evaluation of the clinical accuracy of six portable blood glucose meters in dogs
Shin, M. H., Kim, H., Yun, T., Kang, J., & Kang, B. (2020) · Korean Journal of Veterinary Research · doi:10.14405/kjvr.2020.60.3.123 · Cited by 1
Nanomaterials as glucose sensors for diabetes monitoring
Danquah, M. K., & Jeevanandam, J. (2022) · Elsevier eBooks · doi:10.1016/b978-0-323-85396-5.00004-x · Cited by 1
(2016) · Cambridge University Press eBooks · doi:10.1017/cbo9781139519557.004 · Cited by 1
Determining Zero Point of Blood Glucose Detection Through Potentiostat Circuit Configuration
Sharifdini, A., Sanaei, H. S., Ghazzaghi, H., & Ghobadi, M. Z. (2022) · IEEE Sensors Journal · doi:10.1109/jsen.2022.3206502 · Cited by 1
Glucose Monitoring Techniques and Their Calibration
Yadav, D., Singh, S. P., & Dubey, P. K. (2023) · Springer eBooks · doi:10.1007/978-981-99-2074-7_99 · Cited by 1
Commercialized Enzymatic Biosensors in Healthcare Against the Conventional Methods
Satyaprasad, A. U. (2023) · Springer eBooks · doi:10.1007/978-981-15-6982-1_14 · Cited by 2
Blood Glucose Measurement in Rabbits - Performance Evaluation of Portable Glucometers
Fuchs, T., Saldanha, A., Dusi, G. A., Medeiros, N. D. C., de Sales, B. R., Muehlbauer, E., Wrobel, L. A., Cavalcante, C. Z., & Pimpão, C. T. (2024) · ACTA SCIENTIAE VETERINARIAE · doi:10.22456/1679-9216.140309 · Cited by 1
UV/Ozone-Assisted Covalent Bioconjugation on Graphene Tapes
Jiang, F., Lin, Q., Li, S. C., Song, G., Ge, B., & Yu, H. (2025) · Langmuir · doi:10.1021/acs.langmuir.4c03901 · Cited by 1
Which sample matrix is suitable for glucometer external quality control?
Springer, D., Omastová, K., Budina, M., & Zima, T. (2012) · Klinická biochemie a metabolismus · doi:10.61568/kbm.2012.027 · Cited by 1
КОМБІНОВАНА ОБРОБКА ДАНИХ САМОСТІЙНОГО МОНІТОРИНГУ РІВНЯ ГЛЮКОЗИ В КРОВІ
Chuiko, G. P., Dvornik, O. V., & Darnapuk, Y. S. (2019) · Medical Informatics and Engineering · doi:10.11603/mie.1996-1960.2019.3.10433 · Cited by 1
Baines, C. R., Cooper, P. D., O’Rourke, G. A., & Miller, C. (2020) · Diving and Hyperbaric Medicine Journal · doi:10.28920/dhm50.2.144-151 · Cited by 3
The effects of herbal extracts and compounds on the glucose metabolism in Hep G2 cells
Kandunuri, K. K. (2023) · London Met Repository (London Metropolitan University) · Cited by 8
Glucometrics and Measuring Blood Glucose in Critically Ill Patients
Evans, G. E., Crabtree, D., & Weavind, L. (2013) · Springer eBooks · doi:10.1007/978-1-4614-8557-5_35
Utilisation des lecteurs de glycémie portables en réanimation
Roman, A., & Preiser, J. (2011) · Réanimation · doi:10.1007/s13546-011-0272-9
Crèmes éclaircissantes et mesures glycémiques
Bihan, H., Hoareau, F., Guyot, A., Fysekidis, M., Cohen, R., & Reach, G. (2012) · Médecine des Maladies Métaboliques · doi:10.1016/s1957-2557(12)70422-4
Hypertriglyceridemia Masking Hyperglycemia
Kunjal, R. R. R. (2015) · Journal of Metabolic Syndrome · doi:10.4172/2167-0943.1000i101
Interfering Factors in Quality of Glucose Measurement
Hellman, R. (2014) · Frontiers in diabetes · doi:10.1159/000363475 · Cited by 1
Korpi-Steiner, N. L. (2014) · Point of Care The Journal of Near-Patient Testing & Technology · doi:10.1097/poc.0000000000000029
Scott, M. G. (2014) · Point of Care The Journal of Near-Patient Testing & Technology · doi:10.1097/poc.0000000000000025
Khan, A. I. (2014) · Point of Care The Journal of Near-Patient Testing & Technology · doi:10.1097/poc.0000000000000031
Clemmons, E. A. (2015) · Journal of the American Veterinary Medical Association · doi:10.2460/javma.247.4.343
Endocrine and Metabolic Crises in the Pediatric Intensive Care Unit
Argent, A. C. (2011) · Elsevier eBooks · doi:10.1016/b978-1-4377-1367-1.00168-3
Capillary Blood Glucose Meters in the Home, Clinic, and Hospital
Lewandrowski, K. B. (2017) · Point of Care The Journal of Near-Patient Testing & Technology · doi:10.1097/poc.0000000000000119
Das, S. (2016) · Nursing & Care Open Access Journal · doi:10.15406/ncoaj.2016.01.00011 · Cited by 1
Sajjadi, M., Bijari, H., Soltani, M., & Kianmehr, M. (2017) · Quarterly of Horizon of Medical Sciences · doi:10.18869/acadpub.hms.23.1.1
The reliability of blood glucose point-of-care testing (POCT) system in China teaching hospital
Ma, J. (2017) · International Journal of Hematology and Therapy · doi:10.15436/2381-1404.17.1546
A Novel Approach to Assess Performance of Glucometer: Error Grid Analysis
Kumavat, V., Rojekar, M. V., Panot, J., & Adhe-Rojekar, A. (2017) · Point of Care The Journal of Near-Patient Testing & Technology · doi:10.1097/poc.0000000000000150
Use of a Simple Fabrication Process to Produce a Biosensor: The 3-Hydroxybutyrate Dehydrogenase Case
Ribau, I., & Fortunato, E. M. C. (2018) · Journal of Pharmacy and Pharmacology · doi:10.17265/2328-2150/2018.02.005
Development of glucose biosensors using nanostructured composites
Al-Sagur, H. (2018) · Sheffield Hallam University · doi:10.7190/shu-thesis-00120
Screening strategies for dysglycaemia in relation to cardiovascular risk
Shahim, B. (2019) · Figshare · Cited by 1
Ans, A. Q., Jawad, A. M., Abdul-Razak, M. Y., & Abdul-Razak, N.-A. Y. (2018) · The Medical Journal of Basrah University · doi:10.33762/mjbu.2018.145030
Response to Flash Glucose Monitoring: A Patient's and Clinician's Caveats and Concerns
Virdi, N. S., Budiman, E. S., & Karinka, S. A. (2019) · Endocrine Practice · doi:10.4158/1934-2403-25.5.506
Needle and Potentiostat Study for Blood Glucose Monitoring: Third-Generation E-Mosquito
Rojas, J. T. D. C. (2020) · doi:10.11575/prism/37628
Giordanengo, A. (2020) · NVA (Nasjonalt vitenarkiv) · Cited by 1
Photoacoustic technique for noninvasive glucose sensing : principles and implementations
Zhang, R. (2020) · DR-NTU (Nanyang Technological University) · doi:10.32657/10356/137014
DiAbVi: IoT-based Recommender System for Diabetic Patients
Lopez, A. L. (2018)
A comparative study of blood glucose level measurement between glucometer and semi autoanalyser
Lakshmi, N. V., & Kiranmai, P. (2018) · International Journal of Clinical Biochemistry and Research · doi:10.18231/2394-6377.2018.0089
Self-monitoring of blood glucose as a basis for diabetes management
Mayorov, A. Y., Mel'nikova, O. G., & Filippov, Y. I. (2013) · Cardiosomatics · doi:10.26442/cs45036
Morais, A. D., & Pozzatti, P. N. (2021) · PubVet · doi:10.31533/pubvet.v15n08a897.1-5
Lee, D., Ahn, H., Seo, B.-G., & Lee, S. (2021) · Journal of Sensor Science and Technology · doi:10.46670/jsst.2021.30.4.229
Enhanced Accuracy for Glucose Prediction Using Neural Network
Sequeira, M. D., Parab, J. S., Pinto, C. F., & Naik, G. M. (2021) · Lecture notes in networks and systems · doi:10.1007/978-3-030-84760-9_33
A semiparametric Gumbel regression model for analyzing longitudinal data with non‐normal tails
Hyun, N., Couper, D. J., & Zeng, D. (2021) · Statistics in Medicine · doi:10.1002/sim.9248
The development of molecularly imprinted polymers for sensor and colorimetric assay applications
Lowdon, J. (2021) · Research Publications (Maastricht University) · doi:10.26481/dis.20211130jl
Aboona, F., Zahedi, S., & Reddy, S. S. K. (2019) · Springer eBooks · doi:10.1007/978-3-030-22720-3_9 · Cited by 2
Hipszer, B. R., Heitz, J. W., & Joseph, J. I. (2016) · Cambridge University Press eBooks · doi:10.1017/cbo9781139519557.017
Medical Monitoring of Preexisting DM and GDM
Kintiraki, E., & Goulis, D. G. (2022) · Springer eBooks · doi:10.1007/978-3-030-89243-2_7
Apaza, P. F. V., & de la Torre, C. A. L. (2022) · Revista Brasileira de Ciência Veterinária · doi:10.4322/rbcv.2022.023
Implementation of Six Sigma in Glucose POCT Quality Control at Dr. Soetomo General Academic Hospital
Oktaviani, S., & Fuadi, M. R. (2022) · INDONESIAN JOURNAL OF CLINICAL PATHOLOGY AND MEDICAL LABORATORY · doi:10.24293/ijcpml.v28i2.1848
Applications of IC Products in Automotive, Industrial, and Medical Equipment
Yan, Q., Zhang, Q., Wang, X., & Liu, H. (2023) · Springer eBooks · doi:10.1007/978-981-99-2836-1_20
Correlation of Glucometer and Laboratory Blood Glucose Values in a Hospital Set-Up
Dhananjayan, R. (2023) · International Journal of Science and Research (IJSR) · doi:10.21275/mr231209172202
Non-invasive Glucometry: A New Frontier in Blood Sugar Monitoring
Lohia, H., Singhal, R., Bhardwaj, S., Agarwal, D., & Virmani, D. (2024) · Lecture notes in networks and systems · doi:10.1007/978-981-97-4152-6_40
ҮЗДІКСІЗ ГЛИКЕМИЯЛЫҚ МОНИТОРИНГ – ДИАБЕТОЛОГИЯДАҒЫ ЖАҢА ТЕХНОЛОГИЯЛАР (әдеби шолу)
Kaibagarova, I., Мавлюдова, Н. М., Ж.Ж., Н., Овсянникова, А. К., & Kalzhanova, M. (2024) · Farmaciâ Kazahstana · doi:10.53511/pharmkaz.2024.62.25.006
Micro/millifluidic platforms for bioanalytical applications
Henry, C. S., Ataide, V. N., Cherwin, A. E., Baldo, T. A., Russo, M. J., Meloni, G. N., & Paixão, T. R. L. C. (2025) · Elsevier eBooks · doi:10.1016/b978-0-443-15675-5.00017-3 · Cited by 1
Bonetti, G., Garzaniti, A., VUVAN, G., SALVADORI, S., ZALTIERI, G., & Carta, M. (2025) · La Rivista Italiana della Medicina di Laboratorio - Italian Journal of Laboratory Medicine · doi:10.23736/s1825-859x.25.00277-4
Yılmaz, E., Arpa, M., Şen, B., & Çingirt, M. (2025) · Ahi Evran Medical Journal · doi:10.46332/aemj.1456343
Determination of the Precision of Glucometers Used in Saudi Arabia
Al-Othman, S. A., Al-Zaidany, Z. H., Al-Ghannam, S. H., Al-Turki, A. M., Al-Abdulazeem, A. A., Vatte, C., Habara, A. H., Al‐Ali, A. K., & Alawami, M. F. (2025) · Sensors · doi:10.3390/s25113561
In Vivo Electrochemical Biosensors: Technology and Personalized Medicine Go Hand in Hand
Lou, Y., Chen, S., Peng, S., Lan, F., 高卓维, Zhang, Y., & Luo, M. (2025) · Health and Metabolism · doi:10.53941/hm.2025.100025 · Cited by 1
Cosmas, F. N., Urioa, F. H., & Mbugi, E. V. (2025) · East Africa Science · doi:10.24248/easci.v7i1.122
Stavrova, V., Crcarevska, M. S., Naumovska, Z., Bozhinovska, B., Andonoska, J., Stavric, K., Maksimova, V., Meštrović, A., & Rouse, M. J. (2025) · Healthcare · doi:10.3390/healthcare13172107
Brouwer, K. A. (2025) · Digital Commons - Michigan Tech (Michigan Technological University) · doi:10.37099/mtu.dc.etdr/1977
Establishing indicators to monitor the utilization of POC glucose meters in glycemic control
Yun, H., & Ly, B. (2025) · Clinical Biochemistry · doi:10.1016/j.clinbiochem.2025.111021
Metadata
- Authors
- Tonyushkina, K., & Nichols, J. H.
- Venue
- Journal of Diabetes Science and Technology
- Publisher
- SAGE Publications
- Published
- 2009-07
- Volume
- 3
- Issue
- 4
- Pages
- 971-980
- ISSN
- 1932-2968 (ISSN-L)
- OpenAlex
- W2112149778
- Cited by
- 406
- References
- 44
- FWCI
- 6.30
- Type
- journal article
- Language
- en
- Open access
- green
- Licence
- journals.sagepub.com/page/policies/text-and-data-mining-license (tdm)
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.