Glucose Meters: A Review of Technical Challenges to Obtaining Accurate Results

Open access
Tonyushkina, Ksenia, Nichols, James H. · 2009 · Journal of Diabetes Science and Technology
Cited by 432

The DOI matches its Crossref record; 44 of the 45 references it cites were found in a registry or an index, and 1 are not checked yet.

Abstract

Glucose meters are universally utilized in the management of hypoglycemic and hyperglycemic disorders in a variety of healthcare settings. Establishing the accuracy of glucose meters, however, is challenging. Glucose meters can only analyze whole blood, and glucose is unstable in whole blood. Technical accuracy is defined as the closeness of agreement between a test result and the true value of that analyte. Truth for glucose is analysis by isotope dilution mass spectrometry, and frozen serum standards analyzed by this method are available from the National Institute of Standards and Technology. Truth for whole blood has not been established, and cells must be separated from the whole blood matrix before analysis by a method like isotope dilution mass spectrometry. Serum cannot be analyzed by glucose meters, and isotope dilution mass spectrometry is not commonly available in most hospitals and diabetes clinics to evaluate glucose meter accuracy. Consensus standards recommend comparing whole blood analysis on a glucose meter against plasma/serum centrifuged from a capillary specimen and analyzed by a clinical laboratory comparative method. Yet capillary samples may not provide sufficient volume to test by both methods, and venous samples may be used as an alternative when differences between venous and capillary blood are considered. There are thus multiple complexities involved in defining technical accuracy and no clear consensus among standards agencies and professional societies on accuracy criteria. Clinicians, however, are more concerned with clinical agreement of the glucose meter with a serum/plasma laboratory result. Acceptance criteria for clinical agreement vary across the range of glucose concentrations and depend on how the result will be used in screening or management of the patient. A variety of factors can affect glucose meter results, including operator technique, environmental exposure, and patient factors, such as medication, oxygen therapy, anemia, hypotension, and other disease states. This article reviews the challenges involved in obtaining accurate glucose meter results.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1177/193229680900300446

Cites

44 of 45 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Research works printed but not linked: 9

  1. Mendosa D. Meter memories. How Tom, Dick, and Charlie did it. Diabetes Wellness Letter. 2000;1(6) [Google Scholar]
  2. Bernstein RK. London: Little Brown; 1997. Dr. Bernstein's diabetes solution. A complete guide to achieving normal blood sugars. [Google Scholar]
  3. International Organization for Standardization. ISO/IEC Guide 99. Geneva: International Organization for Standardization; 2007. International vocabulary of metrology—Basic and general concepts and associated terms (VIM) [Google Scholar]
  4. International Organization for Standardization. Determination of performance criteria for in vitro blood glucose monitoring systems for management of human diabetes mellitus. Geneva: International Organization for Standardization. 2002 ISO 15197. [Google Scholar]
  5. U.S. Food and Drug Administration. Rockville: Food and Drug Administration. 1989. Review criteria for assessment of portable invasive blood glucose monitoringin vitro diagnostic devices which use glucose oxidase, dehydrogenase, or hexokinase methodology. [Google Scholar]
  6. Canadian Standards Association. CSA Z316.4-94. Mississauga: Canadian Standards Association. 1994. Performance specifications for portable whole blood glucose monitor systems for use in diabetes management. [Google Scholar]
  7. Instituto Mexicano del Seguro Social. Mexico City: Instituto Mexicano del Seguro Social. 1998. Reactive strip assessment protocol for determination of glucose in blood. [Google Scholar]
  8. TNO Center for Medical Technology. Netherlands: TNO Center for Medical Technology; 1991. Quality guideline for non-implantable portable blood glucose monitors for self-monitoring. TNO-CMT/90.021E. [Google Scholar]
  9. Meneghini LF, Arce RL. Blood glucose data from meter memories often not useful for pattern recognition. Abstract presented at American Diabetes Association 65th Annual Meeting. Diabetes. 2005;54(Suppl 1:A95) [Google Scholar]

Cited by

See all 432

Metadata

Authors
Tonyushkina, K., & Nichols, J. H.
Venue
Journal of Diabetes Science and Technology
Published
2009-07
Volume
3
Issue
4
Pages
971-980
ISSN
1932-2968 (ISSN-L)
OpenAlex
W2112149778
Cited by
406
References
44
FWCI
6.30
Type
journal article
Language
en
Open access
green
Licence
journals.sagepub.com/page/policies/text-and-data-mining-license (tdm)

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.