Glucose Meters: A Review of Technical Challenges to Obtaining Accurate Results
Open accessThe DOI matches its Crossref record; 44 of the 45 references it cites were found in a registry or an index, and 1 are not checked yet.
Abstract
Glucose meters are universally utilized in the management of hypoglycemic and hyperglycemic disorders in a variety of healthcare settings. Establishing the accuracy of glucose meters, however, is challenging. Glucose meters can only analyze whole blood, and glucose is unstable in whole blood. Technical accuracy is defined as the closeness of agreement between a test result and the true value of that analyte. Truth for glucose is analysis by isotope dilution mass spectrometry, and frozen serum standards analyzed by this method are available from the National Institute of Standards and Technology. Truth for whole blood has not been established, and cells must be separated from the whole blood matrix before analysis by a method like isotope dilution mass spectrometry. Serum cannot be analyzed by glucose meters, and isotope dilution mass spectrometry is not commonly available in most hospitals and diabetes clinics to evaluate glucose meter accuracy. Consensus standards recommend comparing whole blood analysis on a glucose meter against plasma/serum centrifuged from a capillary specimen and analyzed by a clinical laboratory comparative method. Yet capillary samples may not provide sufficient volume to test by both methods, and venous samples may be used as an alternative when differences between venous and capillary blood are considered. There are thus multiple complexities involved in defining technical accuracy and no clear consensus among standards agencies and professional societies on accuracy criteria. Clinicians, however, are more concerned with clinical agreement of the glucose meter with a serum/plasma laboratory result. Acceptance criteria for clinical agreement vary across the range of glucose concentrations and depend on how the result will be used in screening or management of the patient. A variety of factors can affect glucose meter results, including operator technique, environmental exposure, and patient factors, such as medication, oxygen therapy, anemia, hypotension, and other disease states. This article reviews the challenges involved in obtaining accurate glucose meter results.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
National Committee for Clinical Laboratory Standards
Bruck, E. (1980) · Pediatrics · doi:10.1542/peds.65.1.187 · Cited by 14,751
Intensive insulin therapy in the medical ICU
Van den Berghe, G., Wilmer, A., Hermans, G., Meersseman, W., Wouters, P. J., Milants, I., Van Wijngaerden, E., Bobbaers, H., & Bouillon, R. (2006) · New England Journal of Medicine · doi:10.1056/nejmoa052521 · Cited by 3,639
Tietz Textbook of Clinical Chemistry
(1999) · Archives of Pathology and Laboratory Medicine · doi:10.5858/123.7.646b · Cited by 2,495
Capes, S. E., Hunt, D., Malmberg, K., Pathak, P., & Gerstein, H. C. (2001) · Stroke · doi:10.1161/hs1001.096194 · Cited by 1,824
Evaluating clinical accuracy of systems for self-monitoring of blood glucose
Clarke, W. L., Cox, D., Gonder-Frederick, L. A., Carter, W., & Pohl, S. L. (1987) · Diabetes Care · doi:10.2337/diacare.10.5.622 · Cited by 1,435
Standards of Medical Care in Diabetes—2008
Dora, J. M., Kramer, C. K., & Canani, L. H. (2008) · Diabetes Care · doi:10.2337/dc08-0109 · Cited by 1,409
Standards of Medical Care in Diabetes—2008
American Diabetes Association (2008) · Diabetes Care · doi:10.2337/dc08-s012 · Cited by 1,377
Parkes, J. L., Slatin, S. L., Pardo, S., & Ginsberg, B. H. (2000) · Diabetes Care · doi:10.2337/diacare.23.8.1143 · Cited by 620
Vlasselaers, D., Milants, I., Desmet, L., Wouters, P. J., Vanhorebeek, I., van den Heuvel, I., Mesotten, D., Casaer, M. P., Meyfroidt, G., Ingels, C., et al. (2009) · The Lancet · doi:10.1016/s0140-6736(09)60044-1 · Cited by 554
Association of Hyperglycemia with Increased Mortality after Severe Burn Injury
Gore, D. C., Chinkes, D., Heggers, J., Herndon, D. N., Wolf, S. E., & Desai, M. (2001) · Journal of Trauma and Acute Care Surgery · doi:10.1097/00005373-200109000-00021 · Cited by 459
Research works printed but not linked: 9
- Mendosa D. Meter memories. How Tom, Dick, and Charlie did it. Diabetes Wellness Letter. 2000;1(6) [Google Scholar]
- Bernstein RK. London: Little Brown; 1997. Dr. Bernstein's diabetes solution. A complete guide to achieving normal blood sugars. [Google Scholar]
- International Organization for Standardization. ISO/IEC Guide 99. Geneva: International Organization for Standardization; 2007. International vocabulary of metrology—Basic and general concepts and associated terms (VIM) [Google Scholar]
- International Organization for Standardization. Determination of performance criteria for in vitro blood glucose monitoring systems for management of human diabetes mellitus. Geneva: International Organization for Standardization. 2002 ISO 15197. [Google Scholar]
- U.S. Food and Drug Administration. Rockville: Food and Drug Administration. 1989. Review criteria for assessment of portable invasive blood glucose monitoringin vitro diagnostic devices which use glucose oxidase, dehydrogenase, or hexokinase methodology. [Google Scholar]
- Canadian Standards Association. CSA Z316.4-94. Mississauga: Canadian Standards Association. 1994. Performance specifications for portable whole blood glucose monitor systems for use in diabetes management. [Google Scholar]
- Instituto Mexicano del Seguro Social. Mexico City: Instituto Mexicano del Seguro Social. 1998. Reactive strip assessment protocol for determination of glucose in blood. [Google Scholar]
- TNO Center for Medical Technology. Netherlands: TNO Center for Medical Technology; 1991. Quality guideline for non-implantable portable blood glucose monitors for self-monitoring. TNO-CMT/90.021E. [Google Scholar]
- Meneghini LF, Arce RL. Blood glucose data from meter memories often not useful for pattern recognition. Abstract presented at American Diabetes Association 65th Annual Meeting. Diabetes. 2005;54(Suppl 1:A95) [Google Scholar]
Cited by
Items 101–200 of 433
Rajendran, J., & Esfandyarpour, R. (2024) · Biomedical Materials & Devices · doi:10.1007/s44174-024-00226-9 · Cited by 37
A new generation of sensors for non-invasive blood glucose monitoring.
Wu, J., Liu, Y., Yin, H., & Guo, M. (2023) · PubMed Central · Cited by 65
Graphene nanocomposites for transdermal biosensing
Tabish, T. A., Abbas, A., & Narayan, R. J. (2021) · Wiley Interdisciplinary Reviews Nanomedicine and Nanobiotechnology · doi:10.1002/wnan.1699 · Cited by 29
Cook, C. B., Wellik, K. E., & Fowke, M. (2011) · Journal of Diabetes Science and Technology · doi:10.1177/193229681100500402 · Cited by 34
Inpatient Glycemic Management in the Setting of Renal Insufficiency/Failure/Dialysis
Iyengar, R., Franzese, J., & Gianchandani, R. Y. (2018) · Current Diabetes Reports · doi:10.1007/s11892-018-1044-y · Cited by 31
Watabe, S., Sakamoto, Y., Morikawa, M., Okada, R., Miura, T., & Ito, E. (2011) · PLoS ONE · doi:10.1371/journal.pone.0022955 · Cited by 31
Retrospective evaluation of xylitol ingestion in dogs: 192 cases (2007–2012)
DuHadway, M. R., Sharp, C. R., Meyers, K. E., & Koenigshof, A. M. (2015) · Journal of Veterinary Emergency and Critical Care · doi:10.1111/vec.12350 · Cited by 28
Megahed, A., Hiew, M. W. H., Townsend, J. R., Messick, J. B., & Constable, P. D. (2015) · Journal of Veterinary Internal Medicine · doi:10.1111/jvim.13608 · Cited by 31
Bennett, K. A., Turner, L. M., Millward, S., Moss, S. E. W., & Hall, A. J. (2017) · Conservation Physiology · doi:10.1093/conphys/cox013 · Cited by 29
Giordanengo, A., Årsand, E., Woldaregay, A. Z., Bradway, M., Grøttland, A., Hartvigsen, G., Granja, C., Torsvik, T., & Hansen, A. H. (2019) · JMIR Diabetes · doi:10.2196/14002 · Cited by 30
Detection of Enzymes, Viruses, and Bacteria Using Glucose Meters
Das, A., Cui, X., Chivukula, V., & Iyer, S. S. (2018) · Analytical Chemistry · doi:10.1021/acs.analchem.8b02960 · Cited by 28
Liu, W., Yuan, Y., Liu, Q., Chen, M., & Chen, X. (2021) · Sensors and Actuators B Chemical · doi:10.1016/j.snb.2021.131296 · Cited by 27
Accuracy and User Performance Evaluation of a Blood Glucose Monitoring System
Caswell, M., Frank, J., Viggiani, M. T., Pardo, S. A., Dunne, N., Warchal-Windham, M. E., & Morin, R. (2015) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2014.0129 · Cited by 28
A sample volume independent paper microfluidic device for quantifying glucose in real human plasma
Choobbari, M. L., Rad, M. B., Jahanshahi, A., & Ghourchian, H. (2020) · Microfluidics and Nanofluidics · doi:10.1007/s10404-020-02382-y · Cited by 26
A Mediated Enzymatic Electrochemical Sensor Using Paper-Based Laser-Induced Graphene
Gao, P., Kasama, T., Shin, J., Huang, Y., & Miyake, R. (2022) · Biosensors · doi:10.3390/bios12110995 · Cited by 25
Hoang, Q., Pisani, M. A., Inzucchi, S. E., Hu, B., & Honiden, S. (2014) · Journal of Critical Care · doi:10.1016/j.jcrc.2014.06.007 · Cited by 28
Hornig, K. J., Byers, S. R., Callan, R. J., Holt, T. N., Field, M., & Han, H. (2013) · American Journal of Veterinary Research · doi:10.2460/ajvr.74.8.1059 · Cited by 29
Pathophysiology and aetiology of hypoglycaemic crises
Morgan, R. K., Cortes, Y., & Murphy, L. A. (2018) · Journal of Small Animal Practice · doi:10.1111/jsap.12911 · Cited by 26
Sengupta, S., Handoo, A., Haq, I., Dahiya, K., Mehta, S., & Kaushik, M. (2021) · Indian Journal of Clinical Biochemistry · doi:10.1007/s12291-021-00971-4 · Cited by 26
Wu, Y. J., Jian-Mei, Z. H. O. U., Jiang, Y., Li, W., Meng-Jie, H. E., Xiao, Y., & Chen, J. (2022) · Chinese Journal of Analytical Chemistry · doi:10.1016/j.cjac.2022.100187 · Cited by 24
Daly, N., Carroll, C., Flynn, I., Harley, R., Maguire, P., & Turner, M. J. (2016) · BJOG An International Journal of Obstetrics & Gynaecology · doi:10.1111/1471-0528.14255 · Cited by 28
Sensors Based on Molecularly Imprinted Polymers
Pohanka, M. (2017) · International Journal of Electrochemical Science · doi:10.20964/2017.09.67 · Cited by 24
Fornerod, M. J. J., Álvarez‐Fernández, A., Michalska, M., Papakonstantinou, I., & Guldin, S. (2023) · Chemistry of Materials · doi:10.1021/acs.chemmater.3c01202 · Cited by 23
Prophylactic Dextrose Gel Does Not Prevent Neonatal Hypoglycemia: A Quasi-Experimental Pilot Study
Coors, S. M., Cousin, J. J., Hagan, J. L., & Kaiser, J. R. (2018) · The Journal of Pediatrics · doi:10.1016/j.jpeds.2018.02.025 · Cited by 23
Attri, B., Goyal, A., Gupta, Y., & Tandon, N. (2020) · Diabetes Therapy · doi:10.1007/s13300-020-00873-3 · Cited by 23
Lam, M., Louie, R. F., Curtis, C. M., Ferguson, W., Vy, J. H., Truong, A.-T., Sumner, S. L., & Kost, G. J. (2014) · Journal of Diabetes Science and Technology · doi:10.1177/1932296813514325 · Cited by 23
Biswas, S. K., Chatterjee, S., Laha, S., Pakira, V., Som, N. K., Saha, S., & Chakraborty, S. K. (2022) · Lab on a Chip · doi:10.1039/d2lc00824f · Cited by 24
Hermayer, K. L., Loftley, A., Reddy, S. B., Narla, S. N., Epps, N. A., & Zhu, Y. (2015) · Current Diabetes Reports · doi:10.1007/s11892-015-0582-9 · Cited by 23
Privacy-enabled remote health monitoring applications for resource constrained wearable devices
Preuveneers, D., & Joosen, W. (2016) · ACM Symposium on Applied Computing · doi:10.1145/2851613.2851683 · Cited by 22
RuO2/rGO heterostructures as mimic peroxidases for colorimetric detection of glucose
Li, H., Sun, H., Ding, J., Wu, T., Cai, S., Wang, C., & Yang, R. (2022) · Microchimica Acta · doi:10.1007/s00604-022-05319-0 · Cited by 22
Detection of H2O2 by Fluorescence Correlation Spectroscopy
Ito, E., Watabe, S., Morikawa, M., Kodama, H., Okada, R., & Miura, T. (2013) · Methods in enzymology on CD-ROM/Methods in enzymology · doi:10.1016/b978-0-12-405883-5.00008-9 · Cited by 22
Sweet Sensor for the Detection of Aflatoxin M1 in Whole Milk
Giovanni, S. D., Zambrini, V., Varriale, A., & D’Auria, S. (2019) · ACS Omega · doi:10.1021/acsomega.9b01300 · Cited by 21
Shettigar, L., Sivaraman, S., Rao, R. R., Arun, S. A., Chopra, A., Kamath, S. U., & Rana, R. (2024) · Acta Odontologica Scandinavica · doi:10.1080/00016357.2023.2267678 · Cited by 23
Glucose control in non‐critically ill inpatients with diabetes: towards closed‐loop
Thabit, H., & Hovorka, R. (2013) · Diabetes Obesity and Metabolism · doi:10.1111/dom.12228 · Cited by 24
Chelonid Alphaherpesvirus 5 DNA in Fibropapillomatosis-Affected Chelonia mydas
Domiciano, I. G., Broadhurst, M. K., Domit, C., Flaiban, K. K. M. D. C., Goldberg, D. W., Fritzen, J. T. T., & Bracarense, A. P. F. R. L. (2019) · EcoHealth · doi:10.1007/s10393-019-01412-8 · Cited by 21
Rapid exclusion of bacterial arthritis using a glucometer
Omar, M., Reichling, M., Liodakis, E., Ettinger, M., Guenther, D., Decker, S., Krettek, C., Suero, E. M., & Mommsen, P. (2016) · Clinical Rheumatology · doi:10.1007/s10067-016-3255-4 · Cited by 21
Chawla, M., Randhawa, J. K., & Siril, P. F. (2017) · New Journal of Chemistry · doi:10.1039/c6nj03920k · Cited by 22
Glucose meters—fit for clinical purpose
Tirimacco, R., Koumantakis, G., Erasmus, R., Mosca, A., Sandberg, S., Watson, I. D., Goldsmith, B., & Gillery, P. (2013) · Clinical Chemistry and Laboratory Medicine · doi:10.1515/cclm-2013-0011 · Cited by 20
Choukem, S. P., Sih, C., Nebongo, D. N., Tientcheu, P., & Kengne, A. P. (2019) · Pan African Medical Journal · doi:10.11604/pamj.2019.32.118.15553 · Cited by 26
Moradi, S., Hosseini, M., Firoozbakhtian, A., Tashnizi, M. B., & Ganjali, M. R. (2024) · Journal of Electroanalytical Chemistry · doi:10.1016/j.jelechem.2024.118675 · Cited by 22
Solnica, B., SKUPIEN, J. K., Kuśnierz‐Cabala, B., Słowińska-Solnica, K., Witek, P., Cempa, A., & Malecki, M. T. (2011) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm.2011.770 · Cited by 21
Glucose Screening Measurements and Noninvasive Glucose Monitor Methods
Zheng, H., He, J., Li, P., Guo, M., Jin, H., Shen, J., Xie, Z., & Chi, C. (2018) · Procedia Computer Science · doi:10.1016/j.procs.2018.10.202 · Cited by 21
A multicenter evaluation of a point-of-care blood glucose meter system in critically ill patients
Nichols, J. H., Brandler, E. S., Fantz, C. R., Fisher, K., Goodman, M. D., Headden, G., Hoppensteadt, D., Matika, R., Peacock, W. F., Rodrigo, J., et al. (2021) · The Journal of Applied Laboratory Medicine · doi:10.1093/jalm/jfab005 · Cited by 19
“Methods: Overall, 476 arterial (376 pediatric/adult, 100 neonate), 375 venous, and 100 neonatal heel-stick whole-blood samples were collected and evaluated from critical care settings at 10 US hospitals, including the emergency department, medical and surgical intensive care units (ICUs), and neonatal and pediatric IC…”
BackgroundInfluentialSelf-citationGhosh, N., & Verma, S. (2024) · Current Medical Research and Opinion · doi:10.1080/03007995.2024.2422005 · Cited by 19
Portable system for the detection of micromolar concentrations of glucose
Kostov, Y. V., Ge, X., Rao, G., & Tolosa, L. M. (2014) · Measurement Science and Technology · doi:10.1088/0957-0233/25/2/025701 · Cited by 19
Schnell, O., Hinzmann, R., Kulzer, B., Freckmann, G., Erbach, M., Lodwig, V., & Heinemann, L. (2013) · Journal of Diabetes Science and Technology · doi:10.1177/193229681300700619 · Cited by 21
Pastakia, S. D., Cheng, S. Y., Kirui, N., & Kamano, J. H. (2015) · Clinical Diabetes · doi:10.2337/diaclin.33.3.136 · Cited by 23
Mackey, P., Thompson, B., Boyle, M. E., Apsey, H. A., Seifert, K. M., Schlinkert, R. T., Stearns, J. D., & Cook, C. B. (2015) · Journal of Diabetes Science and Technology · doi:10.1177/1932296815592027 · Cited by 19
Assessing the Reliability of Commercially Available Point of Care in Various Clinical Fields
Pezzuto, F., Scarano, A., Marini, C., Rossi, G., Stocchi, R., Cerbo, A. D., & Cerbo, A. D. (2019) · The Open Public Health Journal · doi:10.2174/1874944501912010342 · Cited by 21
The Development, Characteristics, and Challenges of Biosensors: The Example of Blood Glucose Meters
Chen, H., & Chen, C. (2025) · Chemosensors · doi:10.3390/chemosensors13080300 · Cited by 18
Hierarchical Nanobiosensors at the End of the SARS-CoV-2 Pandemic
Medrano-Lopez, J. A., Villalpando, I., Salazar, M. I., & Torres-Torres, C. (2024) · Biosensors · doi:10.3390/bios14020108 · Cited by 18
Non‐invasive Measurement of Blood Glucose by Breath Analysis
Mansouri, S., Boulares, S., & Alhadidi, T. (2020) · IEEJ Transactions on Electrical and Electronic Engineering · doi:10.1002/tee.23216 · Cited by 18
Feasibility of Continuous Glucose Monitoring in Critically Ill Emergency Department Patients
Lee, J. H., Kim, K., Jo, Y. H., Rhee, J. E., Lee, J. C., Kim, K. S., Kwon, W. Y., Suh, G. J., Kim, H. C., Yoon, H. I., et al. (2011) · Journal of Emergency Medicine · doi:10.1016/j.jemermed.2011.06.037 · Cited by 19
Diabetes in the Desert: What Do Patients Know about the Heat?
Nassar, A. A., Childs, R. D., Boyle, M. E., Jameson, K. A., Fowke, M., Waters, K. R., Hovan, M. J., & Cook, C. B. (2010) · Journal of Diabetes Science and Technology · doi:10.1177/193229681000400514 · Cited by 21
Point-of-Care Glucose and Ketone Monitoring
Chong, S. K., & Reineke, E. L. (2016) · Topics in companion animal medicine · doi:10.1053/j.tcam.2016.05.005 · Cited by 21
González-Viveros, N., Castro-Ramos, J., Gómez‐Gil, P., Cerecedo-Núñez, H. H., Gutiérrez-Delgado, F., Torres‐Rasgado, E., Pérez-Fuentes, R., & Flores-Guerrero, J. L. (2022) · Lasers in Medical Science · doi:10.1007/s10103-022-03633-w · Cited by 21
Non Inflammatory Boronate Based Glucose-Responsive Insulin Delivery Systems
Dasgupta, I., Tanifum, E. A., Srivastava, M., Phatak, S. S., Cavasotto, C. N., Analoui, M., & Annapragada, A. (2012) · PLoS ONE · doi:10.1371/journal.pone.0029585 · Cited by 17
Fan, L., Jiang, Y., Deng, R., Zhu, H., Dai, X., Liang, H., Li, N., & Qian, Z. (2024) · ACS Sensors · doi:10.1021/acssensors.3c02474 · Cited by 17
Point of care blood glucose devices in the hospital setting
Tran, N. K., LaValley, C., Bagley, B., & Rodrigo, J. (2023) · Critical Reviews in Clinical Laboratory Sciences · doi:10.1080/10408363.2023.2170316 · Cited by 17
Higbie, C. T., Eshar, D., & Bello, N. M. (2015) · American Journal of Veterinary Research · doi:10.2460/ajvr.76.6.532 · Cited by 16
Critical Care Glucose Point-of-Care Testing
Narla, S. N., Jones, M., Hermayer, K. L., & Zhu, Y. (2016) · Advances in clinical chemistry · doi:10.1016/bs.acc.2016.05.002 · Cited by 17
Agarwal, S., Varghese, R. T., SAVIAN, R. E. N. A. T. O. R. E. Z. E. N. D. E., Wang, C. C. L., & Galindo, R. J. (2025) · Endocrine Practice · doi:10.1016/j.eprac.2025.05.749 · Cited by 15
Development of an exogenous coreactant-free electrochemiluminescent sensor for sensing glucose
Zuo, M., Cui, L., Wang, S., Wei, W., Gao, W., & Zhang, C. (2023) · The Analyst · doi:10.1039/d3an00266g · Cited by 15
Shoji, A., Takahashi, Y., Osato, S., & Sugawara, M. (2018) · Journal of Pharmaceutical and Biomedical Analysis · doi:10.1016/j.jpba.2018.09.028 · Cited by 16
Development of DNA Biosensors Based on DNAzymes and Nucleases
Yang, H., Peng, Y., Xu, M., Xu, S., & Zhou, Y. (2021) · Critical Reviews in Analytical Chemistry · doi:10.1080/10408347.2021.1944046 · Cited by 15
Performance of the CONTOUR® TS Blood Glucose Monitoring System
Frank, J., Wallace, J. F., Pardo, S. A., & Parkes, J. L. (2011) · Journal of Diabetes Science and Technology · doi:10.1177/193229681100500128 · Cited by 18
Bailey, T., Wallace, J. F., Greene, C., Pardo, S., Brown, D., Pflug, B., & Klaff, L. J. (2015) · Clinica Chimica Acta · doi:10.1016/j.cca.2015.06.001 · Cited by 14
Advancements in glucose biosensors: Innovations for wearable and real-time monitoring
Singh, R., Nema, P., Purohit, A., Vishwakarma, S., Tantuway, S., Namdeo, P., Kumar, A., Soni, V., & Kashaw, S. K. (2025) · Talanta · doi:10.1016/j.talanta.2025.129045 · Cited by 17
Chan, H., Lunt, H., Thompson, H., Heenan, H. F., FRAMPTON, C. M. A., & Florkowski, C. M. (2014) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm-2013-1059 · Cited by 14
Electrically wired enzyme/TiO2 composite for glucose detection
Lopes, J. H., Francois-Xavier, C., Barralet, J. E., & Merle, G. E. (2017) · Materials Science and Engineering C · doi:10.1016/j.msec.2017.03.113 · Cited by 14
Tungsten disulfide nanosheets-based colorimetric assay for glucose sensing
Jiao, Y., Li, J., Xiang, J., & Chen, Z. (2020) · Spectrochimica Acta Part A Molecular and Biomolecular Spectroscopy · doi:10.1016/j.saa.2020.118706 · Cited by 14
Colorimetric sensor based on bubble wrap and camera phone for glucose determination
Martinková, P., & Pohanka, M. (2016) · Journal of Applied Biomedicine · doi:10.1016/j.jab.2016.05.003 · Cited by 14
Optimization of an In Vitro Starch Digestibility Assay for Rice
Khatun, A., Waters, D. L., & Liu, L. (2018) · Starch - Stärke · doi:10.1002/star.201700340 · Cited by 17
Ogunbosi, B. O., Jarrett, O. O., Orimadegun, A. E., Ayoola, O., & Osinusi, K. (2022) · Nigerian Journal of Paediatrics · doi:10.4314/njp.v49i3.7 · Cited by 14
Diabetes Mellitus Glucose Prediction by Linear and Bayesian Ensemble Modeling
Ståhl, F. L. (2012) · Lund University Publications (Lund University) · Cited by 22
Sowan, A. K., Vera, A., Malshe, A. V., & Reed, C. C. (2019) · JMIR Medical Informatics · doi:10.2196/11873 · Cited by 15
Yun, H., Campbell, E., Colbourne, B., Power, J., & Randell, E. W. (2018) · Clinical Biochemistry · doi:10.1016/j.clinbiochem.2018.11.008 · Cited by 13
Harrison, B., & Brown, D. (2020) · Expert Review of Medical Devices · doi:10.1080/17434440.2020.1704253 · Cited by 12
“que) can affect the accuracy of BGMS measurements [9], and it has been estimated that over 90% of BGMS errors are operator-dependent [10].”
BackgroundPerformance of diabetes screening tests: an evaluation study of Iranian diabetes screening program
Kianpour, F., Fararouei, M., Hassanzadeh, J., Mohammadi, M., & Dianatinasab, M. (2021) · Diabetology & Metabolic Syndrome · doi:10.1186/s13098-021-00632-9 · Cited by 12
Antenna as sensors: a mini review
Subad, R. A. S. I., Kashouty, M. F. E., & Das, S. (2025) · Journal of Materials Chemistry C · doi:10.1039/d5tc02661j · Cited by 13
Model Reference Adaptive Control of Glucose in Type 1 Diabetics: A Simulation Study
Tárník, M., Miklovičová, E., Murgaš, J., Ottinger, I., & Ludwig, T. (2014) · IFAC Proceedings Volumes · doi:10.3182/20140824-6-za-1003.00321 · Cited by 11
Multidimensional Mapping Method Using an Arrayed Sensing System for Cross-Reactivity Screening
Chocron, S. E., Weisberger, B. M., Ben‐Yoav, H., Winkler, T. E., Kim, E., Kelly, D. L., Payne, G. F., & Ghodssi, R. (2015) · PLoS ONE · doi:10.1371/journal.pone.0116310 · Cited by 11
Investigation of ginger (Zingiber officinale) aqueous extract as an anti-diabetic in vitro
Pakan, P., Lidia, K., & Riwu, M. (2021) · IOP Conference Series Earth and Environmental Science · doi:10.1088/1755-1315/913/1/012108 · Cited by 11
Zhang, J., & Lu, Y. (2018) · Angewandte Chemie · doi:10.1002/ange.201804292 · Cited by 11
Point-of-Care Blood Testing: The Technology Behind the Numbers
Siegrist, K. K., & Rice, M. J. (2019) · Anesthesia & Analgesia · doi:10.1213/ane.0000000000004149 · Cited by 11
Opto-electrochemical transducers for molecular sensing utilizing nanozyme-based sensing technologies
Nandi, I., Patel, V., Srivastava, R., & Chandra, P. (2024) · Microchemical Journal · doi:10.1016/j.microc.2024.111360 · Cited by 10
Giordanengo, A., Öztürk, P., Hansen, A. H., Årsand, E., Grøttland, A., & Hartvigsen, G. (2018) · JMIR Diabetes · doi:10.2196/10431 · Cited by 11
Fast and simple glucose assay based on filter paper as enzymes carrier using phone camera detection
Martinková, P., Brtnický, M., Kynický, J., & Pohanka, M. (2018) · Chemical Papers · doi:10.1007/s11696-018-0527-2 · Cited by 10
Shahim, B., Kjellström, B., Gyberg, V., Jennings, C. S., Smetana, S., & Rydén, L. (2018) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2018.0157 · Cited by 10
Sloand, J. A. (2012) · American Journal of Kidney Diseases · doi:10.1053/j.ajkd.2012.06.020 · Cited by 10
Wang, L., Sievenpiper, J. L., de Souza, R. J., Thomaz, M., Blatz, S., Grey, V., Fusch, C., & Balion, C. M. (2013) · Clinical Chemistry and Laboratory Medicine (CCLM) · doi:10.1515/cclm-2012-0436 · Cited by 10
Non-invasive electrochemistry-driven metals tracing in human biofluids
Faheem, A., & Cinti, S. (2021) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2021.113904 · Cited by 9
Accuracy of Point-of-Care Blood Glucometers in Neonates and Critically Ill Adults
Sheen, Y., Wang, J., Tsai, P.-F., Lee, W., Hsu, Y.-C., Wang, C.-Y., & Sheu, W. H. (2023) · Clinical Therapeutics · doi:10.1016/j.clinthera.2023.05.005 · Cited by 9
Voulgari, C., & Tentolouris, N. (2011) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2011.0035 · Cited by 9
Evaluation of Four Point-Of-Care Glucose Meters in Alpacas
Beemer, O., Byers, S. R., & Bohn, A. A. (2013) · Journal of Veterinary Internal Medicine · doi:10.1111/jvim.12115 · Cited by 11
Darwin, D. (2019) · Asian Journal of Chemistry · doi:10.14233/ajchem.2019.21593 · Cited by 9
Low perfusion index affects the difference in glucose level between capillary and venous blood
Acar, N., Özçelik, H., Çevik, A. A., Özakın, E., Yorulmaz, G., Kebapçı, N., Bilge, U., & Bilgin, M. (2014) · Therapeutics and Clinical Risk Management · doi:10.2147/tcrm.s73359 · Cited by 9
Standards of admission capillary blood glucose levels in cesarean born neonates
Smolkin, T., Ulanovsky, I., Carasso, P., & Makhoul, I. R. (2017) · World Journal of Pediatrics · doi:10.1007/s12519-017-0016-7 · Cited by 9
Fan, K. S., Shojaee‐Moradie, F., Manoli, A., Baumann, P. M., Koehler, G., Edwards, V., Lee, V., Mathieu, C., Mader, J. K., & Russell-Jones, D. L. (2024) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2024.0380 · Cited by 8
Immanuel, J., Kongbrailatpam, T., Rajagopal, R., & Simmons, D. S. (2025) · Diabetes Obesity and Metabolism · doi:10.1111/dom.16209 · Cited by 8
Metadata
- Authors
- Tonyushkina, K., & Nichols, J. H.
- Venue
- Journal of Diabetes Science and Technology
- Publisher
- SAGE Publications
- Published
- 2009-07
- Volume
- 3
- Issue
- 4
- Pages
- 971-980
- ISSN
- 1932-2968 (ISSN-L)
- OpenAlex
- W2112149778
- Cited by
- 406
- References
- 44
- FWCI
- 6.30
- Type
- journal article
- Language
- en
- Open access
- green
- Licence
- journals.sagepub.com/page/policies/text-and-data-mining-license (tdm)
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.