Delays in Minimally Invasive Continuous Glucose Monitoring Devices: A Review of Current Technology
Open accessAbstract
Through the use of enzymatic sensors—inserted subcutaneously in the abdomen or ex vivo by means of microdialysis fluid extraction—real-time minimally invasive continuous glucose monitoring (CGM) devices estimate blood glucose by measuring a patient's interstitial fluid (ISF) glucose concentration. Signals acquired from the interstitial space are subsequently calibrated with capillary blood glucose samples, a method that has raised certain questions regarding the effects of physiological time lags and of the duration of processing delays built into these devices. The time delay between a blood glucose reading and the value displayed by a continuous glucose monitor consists of the sum of the time lag between ISF and plasma glucose, in addition to the inherent electrochemical sensor delay due to the reaction process and any front-end signal-processing delays required to produce smooth traces. Presented is a review of commercially available, minimally invasive continuous glucose monitors with manufacturer-reported device delays. The data acquisition process for the Medtronic MiniMed (Northridge, CA) continuous glucose monitoring system—CGMS® Gold—and the Guardian® RT monitor is described with associated delays incurred for each processing step. Filter responses for each algorithm are examined using in vitro hypoglycemic and hyperglycemic clamps, as well as with an analysis of fast glucose excursions from a typical meal response. Results demonstrate that the digital filters used by each algorithm do not cause adverse effects to fast physiologic glucose excursions, although nonphysiologic signal characteristics can produce greater delays.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The check wasn’t confident either way, so it gives no verdict — never disputed.
In the citing paper
“The Medtronic Guardian® REAL-Time has been validated against plasma glucose [49] and is widely used in clinical settings [50], [51].”
From Impact of Five Nights of Sleep Restriction on Glucose Metabolism, Leptin and Testosterone in Young Adult Men (2012) · cites this paper as [49]
Experimental — accuracy not yet measuredIn the citing paper
“The lag time between blood and subcutaneous tissue glucose concentrations cause further inaccuracies for CGM devices.79–80”
From Biocompatible Materials for Continuous Glucose Monitoring Devices (2013) · cites this paper as [79]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
(1993) · New England Journal of Medicine · doi:10.1056/nejm199309303291401 · Cited by 24,667
The scientist and engineer's guide to digital signal processing
Smith, S. W. (1997) · Cited by 2,805
Deiss, D., Bolinder, J., Riveline, J.-P., Battelino, T., Bosi, E., Tubiana-Rufi, N., Kerr, D., & Phillip, M. (2006) · Diabetes Care · doi:10.2337/dc06-1134 · Cited by 539
Improvement in Glycemic Excursions With a Transcutaneous, Real-Time Continuous Glucose Sensor
Garg, S. K., Zisser, H., Schwartz, S. L., Bailey, T. S., Kaplan, R. A., Ellis, S. L., & Jovanovič, L. (2006) · Diabetes Care · doi:10.2337/diacare.29.01.06.dc05-1686 · Cited by 429
The MiniMed Continuous Glucose Monitoring System
Mastrototaro, J. J. (2000) · Diabetes Technology & Therapeutics · doi:10.1089/15209150050214078 · Cited by 394
Self-citationPerformance Evaluation of the MiniMed® Continuous Glucose Monitoring System During Patient Home Use
Gross, T. M., Bode, B. W., EINHORN, D. F., Kayne, D. M., Reed, J. H. C., White, N. H., & Mastrototaro, J. J. (2000) · Diabetes Technology & Therapeutics · doi:10.1089/152091500316737 · Cited by 304
Self-citationNoninvasive Glucose Monitoring
Tamada, J. A., Garg, S. K., Jovanovič, L., Pitzer, K. R., Fermi, S. J., & Potts, R. O. (1999) · JAMA · doi:10.1001/jama.282.19.1839 · Cited by 281
Bode, B. W., Gross, T. M., Thornton, K. R., & Mastrototaro, J. J. (1999) · Diabetes Research and Clinical Practice · doi:10.1016/s0168-8227(99)00113-8 · Cited by 232
Self-citationAccuracy of the 5-Day FreeStyle Navigator Continuous Glucose Monitoring System
Weinstein, R. L., Schwartz, S. L., Brazg, R., Bugler, J., Peyser, T. A., & McGarraugh, G. V. (2007) · Diabetes Care · doi:10.2337/dc06-1602 · Cited by 228
Can Interstitial Glucose Assessment Replace Blood Glucose Measurements?
Rebrin, K., & Steil, G. M. (2000) · Diabetes Technology & Therapeutics · doi:10.1089/15209150050194332 · Cited by 227
Self-citation
Research works printed but not linked: 2
- Mastrototaro JJ, 1999, J Pediatr Endoncrinol Metab., 12, 751
- Bode B, 2008, Managed Care., 17, 40
Cited by
Sorted by citations
Managing diabetes with nanomedicine: challenges and opportunities
Omid Veiseh, A., Tang, B. C., Whitehead, K. A., Anderson, D. G., & Langer, R. (2014) · Nature Reviews Drug Discovery · doi:10.1038/nrd4477 · Cited by 623
Biocompatible Materials for Continuous Glucose Monitoring Devices
Nichols, S. P., Koh, A., Storm, W. L., Shin, J. H., & Schoenfisch, M. H. (2013) · Chemical Reviews · doi:10.1021/cr300387j · Cited by 347
Closed-loop insulin delivery from bench to clinical practice
Hovorka, R. (2011) · Nature Reviews Endocrinology · doi:10.1038/nrendo.2011.32 · Cited by 342
Continuous glucose monitoring: A review of the technology and clinical use
Klonoff, D. C., Ahn, D., & Drincic, A. (2017) · Diabetes Research and Clinical Practice · doi:10.1016/j.diabres.2017.08.005 · Cited by 320
Advances in wearable chemical sensor design for monitoring biological fluids
Matzeu, G., Florea, L., & Diamond, D. (2015) · Sensors and Actuators B Chemical · doi:10.1016/j.snb.2015.01.077 · Cited by 307
Metadata
- Authors
- Keenan, D. B., Mastrototaro, J. J., Voskanyan, G., & Steil, G. M.
- Affiliations
- Medtronic (United States) (US)
- Boston Children's Hospital (US)
- Corresponding author
- Keenan, D. Barry
- Venue
- Journal of Diabetes Science and Technology
- Publisher
- SAGE Publications
- Published
- 2009-09
- Volume
- 3
- Issue
- 5
- Pages
- 1207-1214
- ISSN
- 1932-2968 (ISSN-L)
- PMC
- PMC2769894
- OpenAlex
- W2127835324
- Cited by
- 198
- References
- 27
- Citations by year
- 2026: 9 · 2025: 12 · 2024: 2 · 2023: 8 · 2022: 11 · 2021: 13 · 2020: 15 · 2019: 11 · 2018: 17 · 2017: 18 · 2016: 10 · 2015: 9 · 2014: 13 · 2013: 22 · 2012: 12
- FWCI
- 6.80
- Type
- journal article
- Language
- en
- Open access
- green
- www.ncbi.nlm.nih.gov
- Locations
- Journal of Diabetes Science and Technology (publishedVersion)
- PubMed Central (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
- Physical Sciences › Chemistry › Analytical Chemistry › Electrochemical sensors and biosensors
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Pancreatic function and diabetes
- Keywords
- continuous glucose monitoring, interstitial glucose, sensor calibration, digital filters, hypoglycemic clamp, hyperglycemic clamp, meal response, diabetes management
- MeSH terms
- Algorithms, Blood Glucose, Diagnosis, Computer-Assisted, Dietary Carbohydrates, Equipment Design, Humans, Predictive Value of Tests, Signal Processing, Computer-Assisted, Time Factors, Blood Glucose Self-Monitoring, Reproducibility of Results, Artifacts, Diagnostic Equipment, Extracellular Fluid, Electrochemical Techniques
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review