Delays in Minimally Invasive Continuous Glucose Monitoring Devices: A Review of Current Technology
Open accessAbstract
Through the use of enzymatic sensors—inserted subcutaneously in the abdomen or ex vivo by means of microdialysis fluid extraction—real-time minimally invasive continuous glucose monitoring (CGM) devices estimate blood glucose by measuring a patient's interstitial fluid (ISF) glucose concentration. Signals acquired from the interstitial space are subsequently calibrated with capillary blood glucose samples, a method that has raised certain questions regarding the effects of physiological time lags and of the duration of processing delays built into these devices. The time delay between a blood glucose reading and the value displayed by a continuous glucose monitor consists of the sum of the time lag between ISF and plasma glucose, in addition to the inherent electrochemical sensor delay due to the reaction process and any front-end signal-processing delays required to produce smooth traces. Presented is a review of commercially available, minimally invasive continuous glucose monitors with manufacturer-reported device delays. The data acquisition process for the Medtronic MiniMed (Northridge, CA) continuous glucose monitoring system—CGMS® Gold—and the Guardian® RT monitor is described with associated delays incurred for each processing step. Filter responses for each algorithm are examined using in vitro hypoglycemic and hyperglycemic clamps, as well as with an analysis of fast glucose excursions from a typical meal response. Results demonstrate that the digital filters used by each algorithm do not cause adverse effects to fast physiologic glucose excursions, although nonphysiologic signal characteristics can produce greater delays.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The check wasn’t confident either way, so it gives no verdict — never disputed.
In the citing paper
“The Medtronic Guardian® REAL-Time has been validated against plasma glucose [49] and is widely used in clinical settings [50], [51].”
From Impact of Five Nights of Sleep Restriction on Glucose Metabolism, Leptin and Testosterone in Young Adult Men (2012) · cites this paper as [49]
Experimental — accuracy not yet measuredIn the citing paper
“The lag time between blood and subcutaneous tissue glucose concentrations cause further inaccuracies for CGM devices.79–80”
From Biocompatible Materials for Continuous Glucose Monitoring Devices (2013) · cites this paper as [79]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
(1993) · New England Journal of Medicine · doi:10.1056/nejm199309303291401 · Cited by 24,667
The scientist and engineer's guide to digital signal processing
Smith, S. W. (1997) · Cited by 2,805
Deiss, D., Bolinder, J., Riveline, J.-P., Battelino, T., Bosi, E., Tubiana-Rufi, N., Kerr, D., & Phillip, M. (2006) · Diabetes Care · doi:10.2337/dc06-1134 · Cited by 539
Improvement in Glycemic Excursions With a Transcutaneous, Real-Time Continuous Glucose Sensor
Garg, S. K., Zisser, H., Schwartz, S. L., Bailey, T. S., Kaplan, R. A., Ellis, S. L., & Jovanovič, L. (2006) · Diabetes Care · doi:10.2337/diacare.29.01.06.dc05-1686 · Cited by 429
The MiniMed Continuous Glucose Monitoring System
Mastrototaro, J. J. (2000) · Diabetes Technology & Therapeutics · doi:10.1089/15209150050214078 · Cited by 394
Self-citationPerformance Evaluation of the MiniMed® Continuous Glucose Monitoring System During Patient Home Use
Gross, T. M., Bode, B. W., EINHORN, D. F., Kayne, D. M., Reed, J. H. C., White, N. H., & Mastrototaro, J. J. (2000) · Diabetes Technology & Therapeutics · doi:10.1089/152091500316737 · Cited by 304
Self-citationNoninvasive Glucose Monitoring
Tamada, J. A., Garg, S. K., Jovanovič, L., Pitzer, K. R., Fermi, S. J., & Potts, R. O. (1999) · JAMA · doi:10.1001/jama.282.19.1839 · Cited by 281
Bode, B. W., Gross, T. M., Thornton, K. R., & Mastrototaro, J. J. (1999) · Diabetes Research and Clinical Practice · doi:10.1016/s0168-8227(99)00113-8 · Cited by 232
Self-citationAccuracy of the 5-Day FreeStyle Navigator Continuous Glucose Monitoring System
Weinstein, R. L., Schwartz, S. L., Brazg, R., Bugler, J., Peyser, T. A., & McGarraugh, G. V. (2007) · Diabetes Care · doi:10.2337/dc06-1602 · Cited by 228
Can Interstitial Glucose Assessment Replace Blood Glucose Measurements?
Rebrin, K., & Steil, G. M. (2000) · Diabetes Technology & Therapeutics · doi:10.1089/15209150050194332 · Cited by 227
Self-citation
Research works printed but not linked: 2
- Mastrototaro JJ, 1999, J Pediatr Endoncrinol Metab., 12, 751
- Bode B, 2008, Managed Care., 17, 40
Cited by
Sorted by citations
Managing diabetes with nanomedicine: challenges and opportunities
Omid Veiseh, A., Tang, B. C., Whitehead, K. A., Anderson, D. G., & Langer, R. (2014) · Nature Reviews Drug Discovery · doi:10.1038/nrd4477 · Cited by 623
Biocompatible Materials for Continuous Glucose Monitoring Devices
Nichols, S. P., Koh, A., Storm, W. L., Shin, J. H., & Schoenfisch, M. H. (2013) · Chemical Reviews · doi:10.1021/cr300387j · Cited by 347
Closed-loop insulin delivery from bench to clinical practice
Hovorka, R. (2011) · Nature Reviews Endocrinology · doi:10.1038/nrendo.2011.32 · Cited by 342
Continuous glucose monitoring: A review of the technology and clinical use
Klonoff, D. C., Ahn, D., & Drincic, A. (2017) · Diabetes Research and Clinical Practice · doi:10.1016/j.diabres.2017.08.005 · Cited by 320
Advances in wearable chemical sensor design for monitoring biological fluids
Matzeu, G., Florea, L., & Diamond, D. (2015) · Sensors and Actuators B Chemical · doi:10.1016/j.snb.2015.01.077 · Cited by 307
Lee, I., Probst, D., Klonoff, D. C., & Sode, K. (2021) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2021.113054 · Cited by 305
Reynolds, A. C., Dorrian, J., Liu, P. Y., Van Dongen, H. P. A., Wittert, G. A., Harmer, L. J., & Banks, S. (2012) · PLoS ONE · doi:10.1371/journal.pone.0041218 · Cited by 231
“The Medtronic GuardianH REAL-Time has been validated against plasma glucose [49] and is widely used in clinical settings [50,51].”
MethodFokkert, M. J., van Dijk, P. R., Edens, M. A., Abbes, S., de Jong, D., Slingerland, R. J., & Bilo, H. J. G. (2017) · BMJ Open Diabetes Research & Care · doi:10.1136/bmjdrc-2016-000320 · Cited by 209
Time Lag of Glucose From Intravascular to Interstitial Compartment in Humans
Basu, A., Dube, S., Slama, M. Q., Errazuriz, I., Amezcua, J. C., Kudva, Y. C., Peyser, T. A., Carter, R. E., Cobelli, C., & Basu, R. (2013) · Diabetes · doi:10.2337/db13-1132 · Cited by 223
Continuous Glucose Monitoring: Real-Time Algorithms for Calibration, Filtering, and Alarms
Bequette, B. W. (2010) · Journal of Diabetes Science and Technology · doi:10.1177/193229681000400222 · Cited by 193
Electrogenetic cellular insulin release for real-time glycemic control in type 1 diabetic mice
Krawczyk, K., Xue, S., Buchmann, P., Hamri, G. C., Saxena, P., Hussherr, M., Shao, J., Ye, H., Xie, M., & Fussenegger, M. (2020) · Science · doi:10.1126/science.aau7187 · Cited by 196
Schmelzeisen‐Redeker, G., Schoemaker, M., Kirchsteiger, H., Freckmann, G., Heinemann, L., & del Re, L. (2015) · Journal of Diabetes Science and Technology · doi:10.1177/1932296815590154 · Cited by 162
Zaharieva, D. P., Turksoy, K., MCGAUGH, S. M., Pooni, R., Vienneau, T. E., Ly, T. T., & Riddell, M. C. (2019) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2018.0364 · Cited by 156
Non-Invasive Glucose Monitoring: A Review of Challenges and Recent Advances
Lin, T. (2017) · Current Trends in Biomedical Engineering & Biosciences · doi:10.19080/ctbeb.2017.06.555696 · Cited by 144
Recent advances in optical sensors for continuous glucose monitoring
Ahmed, I., Jiang, N., Shao, X., Elsherif, M., Alam, F., Salih, A. E., Butt, H., & Yetisen, A. K. (2022) · Sensors & Diagnostics · doi:10.1039/d1sd00030f · Cited by 144
A review of optical methods for continuous glucose monitoring
Jernelv, I. L., Mileńko, K., Fuglerud, S. S., Hjelme, D. R., Ellingsen, R., & Aksnes, A. (2018) · Applied Spectroscopy Reviews · doi:10.1080/05704928.2018.1486324 · Cited by 137
Multivariate calibration of NIR spectroscopic sensors for continuous glucose monitoring
Goodarzi, M., Sharma, S., Ramón, H., & Saeys, W. (2015) · TrAC Trends in Analytical Chemistry · doi:10.1016/j.trac.2014.12.005 · Cited by 130
The use of reinforcement learning algorithms to meet the challenges of an artificial pancreas
Bothe, M. K., Dickens, L., Reichel, K., Tellmann, A., Ellger, B., Westphal, M., & Faisal, A. A. (2013) · Expert Review of Medical Devices · doi:10.1586/17434440.2013.827515 · Cited by 126
Acciaroli, G., Vettoretti, M., Facchinetti, A., & Sparacino, G. (2018) · Biosensors · doi:10.3390/bios8010024 · Cited by 130
Siegmund, T., Heinemann, L., Kolassa, R., & Thomas, A. (2017) · Journal of Diabetes Science and Technology · doi:10.1177/1932296817699637 · Cited by 108
A metabolic function of the hippocampal sharp wave-ripple
Tingley, D., McClain, K., Kaya, E., Carpenter, J., & Buzsáki, G. (2021) · Nature · doi:10.1038/s41586-021-03811-w · Cited by 108
ISPAD Clinical Practice Consensus Guidelines 2022: Diabetes technologies: Glucose monitoring
Tauschmann, M., Forlenza, G. P., Hood, K. K., Cardona‐Hernandez, R., Giani, E., Hendrieckx, C., DeSalvo, D. J., Laffel, L. M., Saboo, B., Wheeler, B. J., et al. (2022) · Pediatric Diabetes · doi:10.1111/pedi.13451 · Cited by 123
Accuracy of the Enlite 6-Day Glucose Sensor with Guardian and Veo Calibration Algorithms
Keenan, D. B., Mastrototaro, J. J., Zisser, H., Cooper, K. A., Gautham, R., Lee, S. W., Yusi, J., Bailey, T. S., Brazg, R., & Shah, R. V. (2011) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2011.0199 · Cited by 101
Self-citationWearable physiological systems and technologies for metabolic monitoring
Gao, W., Brooks, G. A., & Klonoff, D. C. (2018) · Journal of Applied Physiology · doi:10.1152/japplphysiol.00407.2017 · Cited by 91
Design and Evaluation of a Robust PID Controller for a Fully Implantable Artificial Pancreas
Huyett, L. M., Dassau, E., Zisser, H., & Doyle, F. J. (2015) · Industrial & Engineering Chemistry Research · doi:10.1021/acs.iecr.5b01237 · Cited by 90
Application of Microfluidics in Wearable Devices
Chen, G., Zheng, J., Liu, L., & Xu, L. (2019) · Small Methods · doi:10.1002/smtd.201900688 · Cited by 90
Implantable Nanosensors: Toward Continuous Physiologic Monitoring
Ruckh, T. T., & Clark, H. A. (2013) · Analytical Chemistry · doi:10.1021/ac402688k · Cited by 84
Hanna, J., Tawk, Y., Azar, S. T., Ramadan, A., Dia, B., Shamieh, E., Zoghbi, S., Kanj, R. N., Costantine, J., & Eid, A. A. (2022) · Scientific Reports · doi:10.1038/s41598-022-19251-z · Cited by 79
Accuracy of a New Real-Time Continuous Glucose Monitoring Algorithm
Keenan, D. B., Cartaya, R., & Mastrototaro, J. J. (2010) · Journal of Diabetes Science and Technology · doi:10.1177/193229681000400114 · Cited by 79
Self-citationWolkowicz, K. L., Aiello, E. M., Vargas, E., Teymourian, H., Tehrani, F., Wang, J., Pinsker, J. E., Doyle, F. J., Patti, M., Laffel, L. M., et al. (2020) · Bioengineering & Translational Medicine · doi:10.1002/btm2.10201 · Cited by 84
Closed-Loop Insulin Delivery in Type 1 Diabetes
Thabit, H., & Hovorka, R. (2012) · Endocrinology and Metabolism Clinics of North America · doi:10.1016/j.ecl.2011.12.003 · Cited by 77
Massa, G., Gys, I., Eyndt, A. O., Bevilacqua, E., Wijnands, A., Declercq, P. E., & Zeevaert, R. (2018) · Hormone Research in Paediatrics · doi:10.1159/000487361 · Cited by 74
Thermoresponsive biodegradable PEG‐PCL‐PEG based injectable hydrogel for pulsatile insulin delivery
Payyappilly, S. S., Dhara, S., & Chattopadhyay, S. (2013) · Journal of Biomedical Materials Research Part A · doi:10.1002/jbm.a.34800 · Cited by 70
Adaptive Control of Artificial Pancreas Systems ‐ A Review
Turksoy, K., & Çınar, A. (2014) · Journal of Healthcare Engineering · doi:10.1260/2040-2295.5.1.1 · Cited by 65
Huang, C., Hao, Z., Qi, T., Pan, Y., & Zhao, X. (2020) · Journal of Materiomics · doi:10.1016/j.jmat.2020.02.002 · Cited by 62
Moser, O., Mader, J. K., Tschakert, G., Mueller, A., Groeschl, W., Pieber, T. R., Koehler, G., Messerschmidt, J., & Hofmann, P. (2016) · Nutrients · doi:10.3390/nu8080489 · Cited by 63
McGarraugh, G. V., Brazg, R., & Weinstein, R. L. (2011) · Journal of Diabetes Science and Technology · doi:10.1177/193229681100500114 · Cited by 57
Microfluidic chip connected to porous microneedle array for continuous ISF sampling
Takeuchi, K., Takama, N., Sharma, K., Paul, O., Ruther, P., Suga, T., & Kim, B. (2021) · Drug Delivery and Translational Research · doi:10.1007/s13346-021-01050-0 · Cited by 56
Nonadjunctive Use of Continuous Glucose Monitoring for Diabetes Treatment Decisions
Castle, J. R., & Jacobs, P. G. (2016) · Journal of Diabetes Science and Technology · doi:10.1177/1932296816631569 · Cited by 58
Hypoglycemia Reduction and Changes in Hemoglobin A1c in the ASPIRE In-Home Study
Weiss, R., Garg, S. K., Bode, B. W., Bailey, T. S., Ahmann, A., Schultz, K. A., Welsh, J. B., Shin, J., & for the ASPIRE In-Home Study Group (2015) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2014.0306 · Cited by 54
Device and method for noninvasive glucose assessment
Segman, Y. (. (2018) · Journal of Diabetes Science and Technology · doi:10.1177/1932296818763457 · Cited by 55
Tauschmann, M., Cardona‐Hernandez, R., DeSalvo, D. J., Hood, K. K., Лаптев, Д. Н., Olinder, A. L., Wheeler, B. J., & Smart, C. E. (2025) · Hormone Research in Paediatrics · doi:10.1159/000543156 · Cited by 69
Evaluation of the Performance of a Novel System for Continuous Glucose Monitoring
Zschornack, E., Schmid, C., Pleus, S., Link, M., Klötzer, H.-M., Obermaier, K., Schoemaker, M., Strasser, M., Frisch, G., Schmelzeisen‐Redeker, G., et al. (2013) · Journal of Diabetes Science and Technology · doi:10.1177/193229681300700403 · Cited by 54
Burnett, D. R., Huyett, L. M., Zisser, H. C., Doyle, F. J., & Mensh, B. D. (2014) · Diabetes · doi:10.2337/db13-1649 · Cited by 52
Monitoring of Pediatric Type 1 Diabetes
Marks, B. E., & Wolfsdorf, J. I. (2020) · Frontiers in Endocrinology · doi:10.3389/fendo.2020.00128 · Cited by 55
A Comparison of Time Delay in Three Continuous Glucose Monitors for Adolescents and Adults
Sinha, M., McKeon, K., Parker, S., Goergen, L. G., Zheng, H., El-Khatib, F. H., & Russell, S. J. (2017) · Journal of Diabetes Science and Technology · doi:10.1177/1932296817704443 · Cited by 50
Zecchin, C., Facchinetti, A., Sparacino, G., & Cobelli, C. (2013) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2012.0208 · Cited by 47
A Bio-Inspired Glucose Controller Based on Pancreatic β-Cell Physiology
Herrero, P., Georgiou, P., Oliver, N. S., Johnston, D. G., & Toumazou, C. (2012) · Journal of Diabetes Science and Technology · doi:10.1177/193229681200600316 · Cited by 46
Huyett, L. M., Dassau, E., Zisser, H., & Doyle, F. J. (2018) · IEEE Control Systems · doi:10.1109/mcs.2017.2766322 · Cited by 47
Comparative assessment of blood glucose monitoring techniques: a review
Ahmadian, N., Manickavasagan, A., & Ali, A. (2022) · Journal of Medical Engineering & Technology · doi:10.1080/03091902.2022.2100496 · Cited by 46
Leelarathna, L., English, S. W., Thabit, H., Caldwell, K., Allen, J. M., Kumareswaran, K., Wilinska, M. E., Nodale, M., Haidar, A., Evans, M. L., et al. (2013) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2013.0221 · Cited by 44
Baldo, F. D., Canton, C., Testa, S., Swales, H., Drudi, I., Golinelli, S., & Fracassi, F. (2020) · Journal of Veterinary Internal Medicine · doi:10.1111/jvim.15930 · Cited by 51
Glucose Level Sensing Using Single Asymmetric Split Ring Resonator
Saleh, G., Ateeq, I. S., & Al-Naib, I. A. I. (2021) · Sensors · doi:10.3390/s21092945 · Cited by 44
Interstitial fluid glucose time-lag correction for real-time continuous glucose monitoring
Keenan, D. B., Mastrototaro, J. J., WEINZIMER, S. A., & Steil, G. M. (2012) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2012.05.007 · Cited by 44
Self-citationVincent, G. E., Jay, S. M., Sargent, C., Kovac, K., Vandelanotte, C., Ridgers, N. D., & Ferguson, S. A. (2017) · Neurobiology of Sleep and Circadian Rhythms · doi:10.1016/j.nbscr.2017.09.001 · Cited by 42
van Beers, C. A. J., & DeVries, J. H. (2016) · Journal of Diabetes Science and Technology · doi:10.1177/1932296816653411 · Cited by 41
Long short-term memory neural network for glucose prediction
Carrillo-Moreno, J., Gandía, C. P., Sendra-Arranz, R., García-Sáez, G., Hernando, M. E., & Gutiérrez, Á. (2020) · Neural Computing and Applications · doi:10.1007/s00521-020-05248-0 · Cited by 44
Seo, W., Park, S. W., Kim, N., Jin, S., & Park, S. (2021) · Computer Methods and Programs in Biomedicine · doi:10.1016/j.cmpb.2021.106424 · Cited by 43
Toward Closing the Loop: An Update on Insulin Pumps and Continuous Glucose Monitoring Systems
Aye, T., Block, J. M., & Buckingham, B. A. (2010) · Endocrinology and Metabolism Clinics of North America · doi:10.1016/j.ecl.2010.05.005 · Cited by 40
The Evolution of Non-invasive Blood Glucose Monitoring System for Personal Application
Salam, N. A. A., Saad, W. H. M., Manap, Z., & Salehuddin, F. (2016) · Journal of Telecommunication Electronic and Computer Engineering (JTEC) · Cited by 40
Fibrotic Encapsulation Is the Dominant Source of Continuous Glucose Monitor Delays
McClatchey, P. M., McClain, E. S., Williams, I. M., Malabanan, C. M., James, F. D., Lord, P., Gregory, J. M., Cliffel, D. E., & Wasserman, D. H. (2019) · Diabetes · doi:10.2337/db19-0229 · Cited by 36
Insight into continuous glucose monitoring: from medical basics to commercialized devices
Chmayssem, A., Nadolska, M., Tubbs, E., Sadowska, K., Vadgma, P., Shitanda, I., Tsujimura, S., Lattach, Y., Peacock, M., Tingry, S., et al. (2023) · Microchimica Acta · doi:10.1007/s00604-023-05743-w · Cited by 37
Sato, H., Hosojima, M., Ishikawa, T., Aoki, K., Okamoto, T., Saito, A., & Tsuchida, M. (2017) · Annals of Thoracic and Cardiovascular Surgery · doi:10.5761/atcs.oa.17-00045 · Cited by 31
Devices for continuous monitoring of glucose: update in technology
Gómez, A. M., Carrillo, D. C. H., & Múñoz, Ó. (2017) · Medical Devices Evidence and Research · doi:10.2147/mder.s110121 · Cited by 30
Marks, B. E., Williams, K. M., Sherwood, J. S., & Putman, M. S. (2021) · Journal of Clinical & Translational Endocrinology · doi:10.1016/j.jcte.2021.100282 · Cited by 30
Mahmoudi, Z., Johansen, M. D., Christiansen, J. S., & Hejlesen, O. K. (2013) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2013.0041 · Cited by 29
Mahmoudi, Z., Johansen, M. D., Christiansen, J. S., & Hejlesen, O. K. (2014) · Journal of Diabetes Science and Technology · doi:10.1177/1932296814531356 · Cited by 28
Althobaiti, M., & Al-Naib, I. A. I. (2021) · IEEE photonics journal · doi:10.1109/jphot.2021.3079408 · Cited by 27
Sampled-data Observer-based Glucose Control for the Artificial Pancreas
Borri, A., Palumbo, P. J., Manes, C., Panunzi, S., & Gaetano, A. D. (2017) · Acta Polytechnica Hungarica · doi:10.12700/aph.14.1.2017.1.6 · Cited by 27
Predicting Glucose Levels in Patients with Type1 Diabetes Based on Physiological and Activity Data
Vahedi, M. R., MacBride, K. B., Wunsik, W., Kim, Y., Fong, C., Padilla, A. J., Pourhomayoun, M., Zhong, A., Kulkarni, S., Arunachalam, S., et al. (2018) · doi:10.1145/3220127.3220133 · Cited by 26
From islet of Langerhans transplantation to the bioartificial pancreas
Berney, T., Wassmer, C., Lebreton, F., Bellofatto, K., Fonseca, L. M., Bignard, J., Hanna, R., Peloso, A., & Berishvili, E. (2022) · La Presse Médicale · doi:10.1016/j.lpm.2022.104139 · Cited by 26
A differential dielectric affinity glucose sensor
Huang, X. L., LeDuc, C. A., Ravussin, Y., Li, S., Davis, E. N., Song, B., Li, D., Xu, K., Accili, D., Wang, Q., et al. (2013) · Lab on a Chip · doi:10.1039/c3lc51026c · Cited by 24
Symbolic Control Design of an Artificial Pancreas for Type-2 Diabetes
Borri, A., Pola, G., Pepe, P., Benedetto, M. D. D., & Palumbo, P. J. (2021) · IEEE Transactions on Control Systems Technology · doi:10.1109/tcst.2021.3135320 · Cited by 23
Glucose control in non‐critically ill inpatients with diabetes: towards closed‐loop
Thabit, H., & Hovorka, R. (2013) · Diabetes Obesity and Metabolism · doi:10.1111/dom.12228 · Cited by 21
Kriventsov, S., Lindsey, A., & Hayeri, A. (2020) · JMIR Diabetes · doi:10.2196/18660 · Cited by 22
Hug, S. A., Riond, B., & Schwarzwald, C. C. (2013) · Journal of Veterinary Emergency and Critical Care · doi:10.1111/vec.12072 · Cited by 21
Design of non-invasive glucose meter using near-infrared technique
Saleh, G., Alkaabi, F., Al-Hajhouj, N., Al-Towailib, F., & Al-Hamza, S. (2018) · Journal of Medical Engineering & Technology · doi:10.1080/03091902.2018.1439114 · Cited by 21
Ward, W. K., Engle, J. M., Branigan, D. L., Youssef, J. E., Massoud, R. G., & Castle, J. R. (2011) · Diabetic Medicine · doi:10.1111/j.1464-5491.2011.03545.x · Cited by 20
Effects of External Factors in CGM Sensor Glucose Concentration Prediction
Ahmed, H. B., & Serener, A. (2016) · Procedia Computer Science · doi:10.1016/j.procs.2016.09.452 · Cited by 20
Sekido, K., Sekido, T., Kaneko, A., Hosokawa, M., Sato, A., Sato, Y., Yamazaki, M., & Komatsu, M. (2017) · Endocrine Journal · doi:10.1507/endocrj.ej17-0193 · Cited by 19
Urbano, F., Farella, I., Brunetti, G., & Faienza, M. F. (2023) · International Journal of Molecular Sciences · doi:10.3390/ijms241511980 · Cited by 22
Scuffi, C. (2010) · touchREVIEWS in Endocrinology · doi:10.17925/ee.2014.10.01.36 · Cited by 20
Thabit, H., & Hovorka, R. (2014) · Current Opinion in Endocrinology Diabetes and Obesity · doi:10.1097/med.0000000000000048 · Cited by 18
Kalita, D., Sharma, H., & Mirza, K. B. (2024) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2024.3396880 · Cited by 18
Mader, J. K., Feichtner, F., Bock, G., Köhler, G., Schaller, R., Plank, J., Pieber, T. R., & Ellmerer, M. (2012) · Diabetes Research and Clinical Practice · doi:10.1016/j.diabres.2012.02.010 · Cited by 19
Carrara, M. A., Batista, M. R., Saruhashi, T. R., Felisberto‐Junior, A. M., Guilhermetti, M., & Bazotte, R. B. (2012) · Life Sciences · doi:10.1016/j.lfs.2012.03.037 · Cited by 18
Quantization improves stabilization of dynamical systems with delayed feedback
Stépàn, G., MILTON, J. G., & Insperger, T. (2017) · Chaos An Interdisciplinary Journal of Nonlinear Science · doi:10.1063/1.5006777 · Cited by 18
Marics, G., Koncz, L., Eitler, K., Vatai, B., Szénási, B., Zakariás, D., Mikos, B., Körner, A., & Tóth‐Heyn, P. (2015) · The Italian Journal of Pediatrics/Italian journal of pediatrics · doi:10.1186/s13052-015-0122-x · Cited by 18
Keenan, D. B., Grosman, B., Clark, H. W., Roy, A., WEINZIMER, S. A., Shah, R. V., & Mastrototaro, J. J. (2011) · Journal of Diabetes Science and Technology · doi:10.1177/193229681100500603 · Cited by 18
Self-citationExploring Minimally Invasive Approach to Define Stages of Type 1 Diabetes Remotely
Kontola, H., Alanko, I., Koskenniemi, J. J., Löyttyniemi, E., Itoshima, S., Knip, M., Veijola, R., Toppari, J., & Kero, J. T. (2022) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2021.0554 · Cited by 18
Althobaiti, M. (2022) · Biosensors · doi:10.3390/bios12100805 · Cited by 17
Youssef, J. E., Castle, J. R., Engle, J. M., Massoud, R. G., & Ward, W. K. (2010) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2010.0020 · Cited by 17
Raza, A., Khan, A. H., Sharma, P., & Ahmad, K. (2020) · Mathematical sciences · doi:10.1007/s40096-020-00355-4 · Cited by 17
Linear Modeling and Prediction in Diabetes Physiology
Cescon, M., & Johansson, R. (2014) · Lecture notes in bioengineering · doi:10.1007/978-3-642-54464-4_9 · Cited by 23
Finger-Actuated Micropump of Constant Flow Rate without Backflow
Ahmad, N. N., Ghazali, N. N. N., Rani, A. T. A., Othman, M. H., Kee, C. C., Jiwanti, P. K., Rodríguez‐Gómez, A., & Wong, Y. H. (2023) · Micromachines · doi:10.3390/mi14040881 · Cited by 15
Dermal-fluid-enabled detection platforms for non-invasive ambulatory monitoring
Veronica, A., Li, Y., Li, Y., Hsing, I., & Nyein, H. Y. Y. (2023) · Sensors & Diagnostics · doi:10.1039/d3sd00165b · Cited by 15
Processing Diabetes Mellitus Composite Events in MAGPIE
Brugués, A., Bromuri, S., Barry, M., del Toro, Ó. J., Mazurkiewicz, M. R., Kardas, P., Pegueroles, J., & Schumacher, M. (2015) · Journal of Medical Systems · doi:10.1007/s10916-015-0377-1 · Cited by 16
(Hybrid) Closed-Loop Systems: From Announced to Unannounced Exercise
Zimmer, R. T., Auth, A., Schierbauer, J., Haupt, S., Wachsmuth, N. B., Zimmermann, P., Voit, T., Battelino, T., Sourij, H., & Moser, O. (2023) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2023.0293 · Cited by 15
Reporting of hypoglycaemia in clinical trials of basal insulins: A need for consensus
Frier, B. M., Ratzki‐Leewing, A. A., & Harris, S. B. (2019) · Diabetes Obesity and Metabolism · doi:10.1111/dom.13732 · Cited by 12
Langeland, L. B. L., Salvesen, Ø. O., Selle, H., Carlsen, S. M., & Fougner, K. J. (2012) · International Journal of Clinical Practice · doi:10.1111/j.1742-1241.2012.02947.x · Cited by 13
Metadata
- Authors
- Keenan, D. B., Mastrototaro, J. J., Voskanyan, G., & Steil, G. M.
- Affiliations
- Medtronic (United States) (US)
- Boston Children's Hospital (US)
- Corresponding author
- Keenan, D. Barry
- Venue
- Journal of Diabetes Science and Technology
- Publisher
- SAGE Publications
- Published
- 2009-09
- Volume
- 3
- Issue
- 5
- Pages
- 1207-1214
- ISSN
- 1932-2968 (ISSN-L)
- PMC
- PMC2769894
- OpenAlex
- W2127835324
- Cited by
- 198
- References
- 27
- Citations by year
- 2026: 9 · 2025: 12 · 2024: 2 · 2023: 8 · 2022: 11 · 2021: 13 · 2020: 15 · 2019: 11 · 2018: 17 · 2017: 18 · 2016: 10 · 2015: 9 · 2014: 13 · 2013: 22 · 2012: 12
- FWCI
- 6.80
- Type
- journal article
- Language
- en
- Open access
- green
- www.ncbi.nlm.nih.gov
- Locations
- Journal of Diabetes Science and Technology (publishedVersion)
- PubMed Central (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
- Physical Sciences › Chemistry › Analytical Chemistry › Electrochemical sensors and biosensors
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Pancreatic function and diabetes
- Keywords
- continuous glucose monitoring, interstitial glucose, sensor calibration, digital filters, hypoglycemic clamp, hyperglycemic clamp, meal response, diabetes management
- MeSH terms
- Algorithms, Blood Glucose, Diagnosis, Computer-Assisted, Dietary Carbohydrates, Equipment Design, Humans, Predictive Value of Tests, Signal Processing, Computer-Assisted, Time Factors, Blood Glucose Self-Monitoring, Reproducibility of Results, Artifacts, Diagnostic Equipment, Extracellular Fluid, Electrochemical Techniques
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review