System Accuracy Evaluation of 43 Blood Glucose Monitoring Systems for Self-Monitoring of Blood Glucose according to DIN EN ISO 15197
Open accessThe DOI matches its Crossref record; 36 of the 36 references it cites were found in a registry or an index.
Abstract
Background: The accuracy of systems for self-monitoring of blood glucose is important, as reliable measurement results are a prerequisite for therapeutic decisions. Methods: This system accuracy evaluation study was performed according to DIN EN ISO 15197:2003 for 43 Conformité Européenne (CE)-labeled blood glucose (BG) monitoring systems. Measurement results of each system were compared with results of the designated comparison method (manufacturer's measurement procedure): glucose oxidase method (YSI 2300 glucose analyzer) or hexokinase method (Hitachi 917/ cobas 501). Results: Complete assessment according to the International Organization for Standardization (ISO) standard was performed for 34 out of 43 systems, and 27 (79.4%) meet the requirements of the standard, i.e., ≥95% of their results showed at least the minimum acceptable accuracy. For 9 of the 43 systems, complete accuracy assessment was not performed due to an oxygen sensitivity (manufacturer's labeling). The bias (according to Bland and Altman) of all 43 evaluated systems ranged from −14.1% to +12.4%. Conclusions: From the 34 systems completely assessed, 7 systems did not fulfill the minimal accuracy requirements of the ISO standard. The CE mark apparently does not guarantee that all BG systems provide accuracy according to the standard. Because inaccurate systems bear the risk of false therapeutic decisions, regular and standardized evaluation of BG meters and test strips should be requested in order to ensure adherence to quality standards.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“The current requirement of traceability according to ISO 17511 as stated by ISO 15197, without specifying further requirements or performance goals, might be insufficient in light of the considerable systematic differences between measurement procedures observed in this and other studies.21,33”
From Impact of Two Different Reference Measurement Procedures on Apparent System Accuracy of 18 CE-Marked Current-Generation Blood Glucose Monitoring Systems (2022) · cites this paper as [33]
Experimental — accuracy not yet measuredThe source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“Yet, only a few studies investigated possible systematic differences between different comparison methods (GOD or HK) or measurement procedures up to now, whereas a bias of up to 8% between the different comparison measurement procedures has been reported.21,31-35”
From Impact of Two Different Reference Measurement Procedures on Apparent System Accuracy of 18 CE-Marked Current-Generation Blood Glucose Monitoring Systems (2022) · cites this paper as [33]
Experimental — accuracy not yet measuredThe cited source says what the claim attributes to it.
In the citing paper
“Nevertheless, several evaluation or surveillance studies within recent years repeatedly revealed that a non-negligible number of CE-marked BGMS, which do not fulfill the minimum accuracy criteria of ISO 15197, had been or currently still are on sale on the European market.12–14”
From System accuracy evaluation of 18 CE-marked current-generation blood glucose monitoring systems based on EN ISO 15197:2015 (2020) · cites this paper as [13]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
Martin Bland, J., & Altman, D. (1986) · The Lancet · doi:10.1016/s0140-6736(86)90837-8 · Cited by 47,965
(1993) · New England Journal of Medicine · doi:10.1056/nejm199309303291401 · Cited by 24,667
International Organization for Standardization
(2011) · SpringerReference · doi:10.1007/springerreference_16847 · Cited by 12,534
International Organization for Standardization
(2015) · Encyclopedia of Astrobiology · doi:10.1007/978-3-662-44185-5_100574 · Cited by 6,716
(2007) · Endocrine Practice · doi:10.4158/ep.13.s1.1 · Cited by 699
Global guideline for type 2 diabetes: recommendations for standard, comprehensive, and minimal care
IDF Clinical Guidelines Task Force (2006) · Diabetic Medicine · doi:10.1111/j.1464-5491.2006.01918.x · Cited by 433
Polonsky, W. H., Fisher, L., Schikman, C. H., Hinnen, D. A., Parkin, C. G., Jelsovsky, Z., Petersen, B., Schweitzer, M., & Wagner, R. S. (2011) · Diabetes Care · doi:10.2337/dc10-1732 · Cited by 429
Factors affecting blood glucose monitoring: sources of errors in measurement
Ginsberg, B. H. (2009) · Journal of Diabetes Science and Technology · doi:10.1177/193229680900300438 · Cited by 348
Martin, S., Schneider, B., Heinemann, L., Lodwig, V., Kurth, H.-J., Kolb, H., Scherbaum, W. A., & for the ROSSO Study Group (2006) · Diabetologia · doi:10.1007/s00125-005-0083-5 · Cited by 284
Approved IFCC recommendation on reporting results for blood glucose (abbreviated)
D'Orazio, P. (2005) · Clinical Chemistry · doi:10.1373/clinchem.2005.051979 · Cited by 208
Research works printed but not linked: 2
- 17. Bundesärztekammer. Richtlinie der Bundesärztekammer zur Qualitätssicherung laboratoriumsmedizinischer Untersuchungen. Dtsch Ärztebl. 2008;105(7):341-55.
- 23. Freckmann G, Haug C, Heinemann L. Blutzuckerselbstmessung heute: Sind alle Geräte gleich? Diabetes Stoffwechsel Herz. 2011;20(4):235-41.
Cited by
Items 101–200 of 244
Susairaj, P., Snehalatha, C., Raghavan, A., Nanditha, A., Vinitha, R., Satheesh, K., Johnston, D. G., Wareham, N. J., & Ramachandran, A. (2019) · Journal of Diabetes and Clinical Research · doi:10.33696/diabetes.1.009 · Cited by 17
“Concurrently, capillary random plasma glucose was also measured using a glucometer (Accu-Check Performa, Roche Diagnostics, Mannheim Germany), calibrated to infer venous plasma glucose values [12].”
MethodLarge-scale performance evaluation of Accu-Chek inform II point-of-care glucose meters
Jeong, T., Cho, E., Ko, D., Lee, W., Chun, S., Hong, K.-S., & Min, W. (2016) · Scandinavian Journal of Clinical and Laboratory Investigation · doi:10.1080/00365513.2016.1233575 · Cited by 16
“Freckmann and colleagues have reported that the accuracy results of 34 types of POC glucose meters ranged from 60–100% and that 18 (53%) satisfied the accuracy requirements for ISO 15197:2013 [10].”
BackgroundUsing uncertain data from body-worn sensors to gain insight into type 1 diabetes
Heintzman, N. D., & Kleinberg, S. (2016) · Journal of Biomedical Informatics · doi:10.1016/j.jbi.2016.08.022 · Cited by 16
“So discretizing 68 mg/dL as “low glucose” and 70 mg/dL as “normal” overstates the confidence in these measurements [49].”
BackgroundLower treatment targets for gestational diabetes: is lower really better?
Kevat, D., Sinha, A. K., & McLean, A. (2014) · The Medical Journal of Australia · doi:10.5694/mja14.00099 · Cited by 13
Electrochemical Glucose Biosensors for Diabetes Care
Ocvirk, G., Buck, H., & DuVall, S. H. (2016) · Bioanalytical reviews · doi:10.1007/11663_2016_3 · Cited by 16
“Publications where SMBG systems are evaluated according to ISO15197:2003, ISO15197:2013, FDA draft guidance, and so-called “10/10” and “5/5” criteria are becoming more common [195, 201–204].”
BackgroundMethodInfluentialFreckmann, G., Baumstark, A., Jendrike, N., Mende, J., Schauer, S., Link, M., Pleus, S., & Haug, C. (2022) · Journal of Diabetes Science and Technology · doi:10.1177/1932296820948873 · Cited by 12
Self-citationZakian, A., Tehrani-Sharif, M., Mokhber‐Dezfouli, M. R., Nouri, M., & Constable, P. (2017) · Australian Veterinary Journal · doi:10.1111/avj.12568 · Cited by 12
Henao-Pabon, G., Gao, N., Prasad, K. S., & Li, X. (2023) · Biosensors · doi:10.3390/bios13050566 · Cited by 12
“Only two of them meet the ISO (International Organization for Standardization) 2013 criteria for the international standards for blood glucose monitoring systems [14,15].”
BackgroundHeinemann, L., Freckmann, G., & Koschinsky, T. (2013) · Journal of Diabetes Science and Technology · doi:10.1177/193229681300700230 · Cited by 14
Self-citationMoreira, F. D., Reis, C. E. G., Gallassi, A. D., Moreira, D. C., & Welker, A. F. (2024) · PLoS ONE · doi:10.1371/journal.pone.0311501 · Cited by 12
Bahramzadeh, A., Tehrani, S. S., Goodarzi, G., Seyyedebrahimi, S., & Meshkani, R. (2023) · Phytotherapy Research · doi:10.1002/ptr.8086 · Cited by 11
“Immediately, 2 g/kg of glucose was injected intraperitoneally, and then blood was collected five times (at 15, 30, 60, 90, and 120 min), and blood glucose was measured using the Roche Diagnostics Accu-check Aviva system, (Burgess Hill, UK) (Freckmann et al., 2012).”
MethodPfützner, A., Demircik, F., Kirsch, V., Pfützner, J. W., Strobl, S., Hanna, M., Spatz, J., & Pfützner, A. H. (2019) · Journal of Diabetes Science and Technology · doi:10.1177/1932296819841353 · Cited by 11
Moreira, F. D., Reis, C. E. G., Welker, A. F., & Gallassi, A. D. (2022) · Nutrients · doi:10.3390/nu14183736 · Cited by 17
“998 with the hexokinase method) [25], at fasting and at 15, 30, 45, 60, 90, and 120 min postprandially.”
MethodSutheran, H. L., & Reynolds, T. (2016) · Journal of Clinical Pathology · doi:10.1136/jclinpath-2015-203339 · Cited by 10
Pandolfo, A., Messina, B., & Russo, G. M. (2021) · Foods · doi:10.3390/foods10092221 · Cited by 11
“2012 [22] that argued the best type of systems available for guaranteeing accuracy evaluation according to DIN EN ISO 15.”
MethodChlup, R., Doubravová, B., Bartek, J., Zapletalová, J., Krystyník, O., & Procházka, V. (2013) · Disease Markers · doi:10.1155/2013/702427 · Cited by 11
Joseph, J. I. (2013) · Journal of Diabetes Science and Technology · doi:10.1177/193229681300700519 · Cited by 11
If SMBG Accuracy Is Critical to Patient Safety, Why Are Inaccurate Meters Still on the Market?
Parkin, C. G. (2017) · Journal of Diabetes Science and Technology · doi:10.1177/1932296817693256 · Cited by 11
Glycemic Index (GI) Values for Major Sources of Dietary Carbohydrates in Iran
Kazemi, F., Danaei, G., Farzadfar, F., Malik, V., Parsaeian, M., Pouraram, H., Zamaninour, N., Rahmani, J., & Motlagh, A. D. (2020) · Intern J Endocrinol Metab · doi:10.5812/ijem.99793 · Cited by 9
Tai, J. C. J., Wong, L. Z., & Richardson, A. (2021) · Cureus · doi:10.7759/cureus.15597 · Cited by 10
“highlighted some variation in the quality and accuracy of various monitoring systems [24], and while all SMBG kits prescribed in this general practice were compliant with DVLA requirements and ISO standards, it is clear that not all SMBG kits are equal.”
MethodZayar, N. N., Sangthong, R., Saw, S., Aung, S. T., & Chongsuvivatwong, V. (2020) · Tropical Medicine and Infectious Disease · doi:10.3390/tropicalmed5030107 · Cited by 10
“The prevalence of DM may be different had the standard glucose concentrations in plasma been used [31,44].”
MethodCorrelation between capillary and venous blood glucose levels in diabetic patients
Adnan, M., Imam, S. F., Shabbir, I., Ali, Z., & Rahat, T. (2015) · Asian Biomedicine · doi:10.5372/1905-7415.0901.368 · Cited by 9
“ADA guidelines recommend an analytical error of ≤5% for all values [2, 3].”
BackgroundDiabetes Technology and Therapy in the Pediatric Age Group
Maahs, D. M., & Shalitin, S. (2016) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2016.2509 · Cited by 8
“Over the last years several publications have pointed out that not all previously approved glucose-meter devices conform to the current and/or planned ISO standards (12,13), which may have an impact on clinical decision making and glucose control.”
Assessing System Accuracy of Blood Glucose Monitoring Systems Using Rectangle Target Plots
Müller, P., Hattemer, A., & Stephan, P. (2015) · Journal of Diabetes Science and Technology · doi:10.1177/1932296815612496 · Cited by 10
“Nearly 80% of the system accuracy results have been published.(12-17) The objective of the analysis was to evaluate performance characteristics of SMBG systems per manufacturer.”
BackgroundAccuracy of Sensocard Glucose Meter: Comparing with Reference Glucose Oxidase Method
Kassahun, M., & Melak, T. (2014) · Journal of Medical Diagnostic Methods · doi:10.4172/2168-9784.1000162 · Cited by 8
Demircik, F., Kirsch, V., Ramljak, S., Vogg, M., Pfützner, A. H., & Pfützner, A. (2019) · Journal of Diabetes Science and Technology · doi:10.1177/1932296819841357 · Cited by 9
Shin, J., Park, H., Cho, S., Nam, H., & Lee, K.-J. (2014) · Computers in Biology and Medicine · doi:10.1016/j.compbiomed.2014.06.005 · Cited by 10
Possible Impacts of New Accuracy Standards of Self-Monitoring of Blood Glucose
Garg, S. K., Reed, K., Grey, J. M. S., Westerman, A., & Pediatrics, P. O. M. A. (2013) · US Endocrinology · doi:10.17925/use.2013.09.01.16 · Cited by 8
Best, T., Miller, J., & Teo, W. (2024) · Nutritional Neuroscience · doi:10.1080/1028415x.2024.2325227 · Cited by 8
“On testing days, blood glucose tests were undertaken via AccuChek Performa glucometer, test strips and single-use Lancet Safe-T-Pro Plus lancets (Roche, Castle Hill, NSW) as an accurate and reliable standard for glucose testing [15].”
Noninvasive evaluation of glucose concentration in the human blood based on electrocardiograms
Pulavskyi, A. A., Krivenko, S., & Krivenko, S. A. (2015) · IEEE International Conference on Electronics and Nanotechnology (ELNANO) · doi:10.1109/elnano.2015.7146890 · Cited by 7
Leigh, S., IDRIS, I. R., Collins, B., Granby, P., Noble, M., & Parker, M. (2015) · Diabetic Medicine · doi:10.1111/dme.12977 · Cited by 7
Glycemic Index and Amylose Content of 25 Japonica Rice Italian Cultivar
Rondanelli, M., Haxhari, F., Gasparri, C., Barrile, G. C., Cavioni, A., Guido, D., Mansueto, F., Zese, M., Mazzola, G., Moroni, A., et al. (2023) · Starch - Stärke · doi:10.1002/star.202300031 · Cited by 7
Wolde, M., Tarekegn, G., & Kebede, T. (2018) · Journal of Diabetes Science and Technology · doi:10.1177/1932296817751747 · Cited by 7
‘Knowing where I am’: self-monitoring of blood glucose in diabetes
Meetoo, D. D., Wong, L., & Fatani, T. (2018) · British Journal of Nursing · doi:10.12968/bjon.2018.27.10.537 · Cited by 8
Strilka, R., Sjulin, T., Huprikar, N., Cameron, L., Woody, P. W., & Armen, S. B. (2020) · International Journal of Critical Illness and Injury Science · doi:10.4103/ijciis.ijciis_112_19 · Cited by 7
“According to the International Organization for Standardization, measurements shall fall within ± 15 (mg/dl) of the results of the manufacturer’s measurement procedure at glucose concentrations <75 (mg/dl) and within ± 20% at glucose concentrations ≥(75 mg/dl).[31] A larger clinical study would be required to mitigate…”
George, K., Joy, S., Palliyalil, H. K., Shaji, S. M., Vishakan, A. S., & Sreedevi, A. (2024) · Cureus · doi:10.7759/cureus.70981 · Cited by 8
“A study conducted in Germany showed that out of 34 blood glucose systems assessed, only half fulfilled the minimal accuracy requirements of the ISO 15197 standards [11].”
Zijlstra, E., Heinemann, L., Fischer, A., & Kapitza, C. (2016) · Journal of Diabetes Science and Technology · doi:10.1177/1932296816668373 · Cited by 6
Jenie, R. P., Damayanthi, E., Irzaman, I., Rimbawan, R., Sukandar, D., & Alatas, H. (2018) · IOP Conference Series Earth and Environmental Science · doi:10.1088/1755-1315/187/1/012012 · Cited by 6
Analysis of five glucometers available in India
Kumar, V., Indukuri, D., & Bhat, N. (2014) · International Journal of Diabetes in Developing Countries · doi:10.1007/s13410-014-0216-9 · Cited by 6
The application of glucose point of care testing in three metropolitan hospitals
Sharp, L. K., Farrance, I., & Greaves, R. F. (2015) · Pathology · doi:10.1016/j.pathol.2015.11.014 · Cited by 6
Clemmons, E. A., Stovall, M. I., Owens, D. C., Scott, J. A., Wilkes, A. C., Kempf, D., & Ethun, K. F. (2016) · PubMed · Cited by 6
The comparison of two glucose measurement systems: POCT devices versus central laboratory
Baygutalp, N. K., Bakan, E., Bayraktutan, Z., & Umudum, F. Z. (2018) · Turkish Journal of Biochemistry · doi:10.1515/tjb-2017-0196 · Cited by 7
“To date, several studies have been conducted on the performances of GMSs, and different aspects of the topic have been evaluated [2, 31, 32].”
BackgroundHeald, A. H., Livingston, M. A., Fryer, A. A., Cortés, G., Anderson, S. G., Gadsby, R., Laing, I., Lunt, M., Young, R. J., & Stedman, M. (2018) · International Journal of Clinical Practice · doi:10.1111/ijcp.13252 · Cited by 6
Li, F.-F., Xie, Y., Shi, B., Niu, M. Y., Guo, H., Cao, Y., Liu, B., Yan, R., Su, X., Wu, J.-D., et al. (2018) · Expert Review of Medical Devices · doi:10.1080/17434440.2018.1473031 · Cited by 6
Jenie, R. P., Damayanthi, E., Irzaman, I., Rimbawan, R., Sukandar, D., Nurdin, N. M., & Alatas, H. (2019) · Journal of Medical Devices · doi:10.1115/1.4044336 · Cited by 8
Cutler, D. C., Koenig, A., Girolamo, N. D., & Mayer, J. (2020) · American Journal of Veterinary Research · doi:10.2460/ajvr.81.8.642 · Cited by 7
Blood Glucose Meter Market: This World is Undergoing Drastic Changes
Heinemann, L., & Klonoff, D. C. (2013) · Journal of Diabetes Science and Technology · doi:10.1177/193229681300700301 · Cited by 6
Effects of Different Methods Used to Take Blood Samples on Blood Glucose Measurements
Ergın, E., & Zaybak, A. (2021) · Clinical Nursing Research · doi:10.1177/10547738211024782 · Cited by 5
“It provides 100% reliable results in compliance with the international accuracy standards (Freckmann et al., 2012).”
MethodTopuz, İ., & Gözüm, S. (2022) · Clinical and Experimental Health Sciences · doi:10.33808/clinexphealthsci.984039 · Cited by 5
Rondanelli, M., Ferrario, R. A., Barrile, G. C., Guido, D., Gasparri, C., Ferraris, C., Cavioni, A., Mansueto, F., Mazzola, G., Patelli, Z., et al. (2023) · Journal of Medicinal Food · doi:10.1089/jmf.2022.0120 · Cited by 6
Exercise for People with Type 1 Diabetes
GALLEN, I. W. (2014) · Medicine and sport science/Medicine and sport · doi:10.1159/000357344 · Cited by 7
Clinical Evaluation of a Novel On-Strip Calibration Method for Blood Glucose Measurement
Noble, M., Rippeth, J. J., Edington, D., Rayman, G. A., Brandon-Jones, S., Hollowood, Z., & Kew, S. J. (2014) · Journal of Diabetes Science and Technology · doi:10.1177/1932296814532575 · Cited by 5
Jeanny, G., & Hope, P. (2016) · Journal of Diabetes Science and Technology · doi:10.1177/1932296815608871 · Cited by 5
“This is a 1-time process, and postmarket release evaluations or independent evaluations are not mandatory, despite studies repeatedly showing test strip lots of available systems not adhering with ISO 15197 accuracy criteria.(3-6) In 2 studies, performed in April and May 2013 and in October 2014, the SMBG systems Gluc…”
MethodHin, B. F. Y. Y., Bueno, I., Lowe, C. R., & Jones, S. (2016) · Journal of Diabetes Science and Technology · doi:10.1177/1932296816664362 · Cited by 5
“The performance requirements for blood glucose monitoring systems (BGMSs) set out in the International Organization for Standardization (ISO) standard 15197:20031 was revised with tighter standards in 2013 (ISO: 15197:2013),2 following concerns about the accuracy of many BGMSs.3 Klonoff and Prahalad4 in a recent revie…”
MethodImpact of the 2013 National Rollout of CMS Competitive Bidding Program: The Disruption Continues
Puckrein, G. A., Hirsch, I. B., Parkin, C. G., Taylor, B. T., Xu, L., & Marrero, D. G. (2017) · Diabetes Care · doi:10.2337/dc17-0960 · Cited by 5
Östman, E. M., Samigullin, A., Heyman‐Lindén, L., Andersson, K. E., Björck, I., Öste, R., & Humpert, P. M. (2020) · PLoS ONE · doi:10.1371/journal.pone.0234237 · Cited by 7
“This can be explained by the use of patient glucometers, which are known for their imprecision [66, 67] and cannot be expected to produce reliable data for the calculation of differences in glucose iAUC.”
BackgroundAccuracy of devices for self-monitoring of blood glucose: A stochastic error model
Vettoretti, M., Facchinetti, A., Sparacino, G., & Cobelli, C. (2015) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2015.7318867 · Cited by 6
“Indeed, in the literature several evidences suggest that the single-zone Gaussian model is inappropriate to describe SMBG error distribution [7-9].”
BackgroundNon-enzymatic sweat-based glucose sensors: progress, difficulties, and prospects
Perumal, K., Sreeja, B. S., & Sulthana, S. F. (2026) · Sensors & Diagnostics · doi:10.1039/d6sd00042h · Cited by 5
System Accuracy Evaluation of the GlucoRx Nexus Voice TD-4280 Blood Glucose Monitoring System
Khan, M. A., Broadbent, K., Morris, M., Ewins, D. L., & Joseph, F. (2014) · Disease Markers · doi:10.1155/2014/602586 · Cited by 4
“Although we are the first to document such a finding for the GlucoRx Nexus TD-4280 BG monitoring system, previous studies have noted a similar lack of accuracy compliance to the latest guidelines, albeit using a draft version of the ISO 15197:2013 guidelines and not the full protocol released in June 2013 [23].”
BackgroundMethodResultInfluentialRetractedAlzaid, A., Schlaeger, C., & Hinzmann, R. (2013) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2013.0260 · Cited by 4
Angelides, K., Matsunami, R. K., & Engler, D. A. (2015) · Journal of Diabetes Science and Technology · doi:10.1177/1932296815586016 · Cited by 4
Pleus, S., Flacke, F., Sieber, J., Haug, C., & Freckmann, G. (2017) · Journal of Diabetes Science and Technology · doi:10.1177/1932296817707292 · Cited by 4
Self-citationLuppa, P. B., Zeller, M., Pieper, M., Kaiser, P., Weiss, N., Vierbaum, L., & Freckmann, G. (2024) · Frontiers in Molecular Biosciences · doi:10.3389/fmolb.2024.1371426 · Cited by 4
Self-citationNew Technologies for Glucose Monitoring and Insulin Administration
Horwitz, D. L., & Klonoff, D. C. (2016) · Wiley eBooks · doi:10.1002/9781118924853.ch30 · Cited by 4
Accuracy of Glucose Meter Among Adults in a Semi-urban Area in Kathmandu, Nepal
Pokhrel, A., Silvanus, V., Pokhrel, B. R., Baral, B. K., Khanal, M. P., Gyawali, P., Pokhrel, L., & Regmi, D. (2019) · Journal of Nepal Medical Association · doi:10.31729/jnma.4247 · Cited by 6
Heart Rate Variability as a Potential Non-invasive Marker of Blood Glucose Level
Jarman, L. R., Elliott, J. L., Lees, T., Clifton-Bligh, R. J., Simpson, A. M., Nassif, N. T., & Lal, S. (2021) · Human Physiology · doi:10.1134/s0362119721020031 · Cited by 5
Grammatical Evolution-Based Approach for Extracting Interpretable Glucose-Dynamics Models
Falco, I. D., Cioppa, A. D., Koutný, T., Scafuri, U., Tarantino, E., & Úbl, M. (2021) · IEEE Symposium on Computers and Communications (ISCC) · doi:10.1109/iscc53001.2021.9631483 · Cited by 4
Javeriana, P. U., & Medina, A. M. G. (2017) · Revista Colombiana de Endocrinología Diabetes & Metabolismo · doi:10.53853/encr.3.2.29 · Cited by 7
Krishnan, G., & Raghuram, J. N. V. (2024) · International Journal of Pharmaceutical and Healthcare Marketing · doi:10.1108/ijphm-06-2023-0044 · Cited by 4
Recommendations for the use, selection and kontrol of glucometers
Friedecky, B., Springer, D., Kratochvíla, J., Škrha, J., & Zima, T. (2014) · Klinická biochemie a metabolismus · doi:10.61568/kbm.2014.026 · Cited by 4
Cai, W., Li, Z., Huang, H., Li, X., Zhao, J. K., Liu, P., Wu, Y., Liu, M., & Wang, J. (2023) · IEEE Sensors Journal · doi:10.1109/jsen.2023.3257834 · Cited by 3
Aghakachoei, S., Zargarzadeh, A. H., & Amini, M. (2015) · Journal of Pharmaceutical Care · Cited by 3
Exactitud y concordancia entre glucómetros: un estudio en condiciones habituales de práctica clínica
Rodrigo, E., Deib-Morgan, K., de Diego, O. G., García-Velasco, P., Sgaramella, G. A., & González, I. G. (2016) · Medicina de Familia SEMERGEN · doi:10.1016/j.semerg.2016.01.011 · Cited by 6
“…de pacientes diabéticos utilizan actualmente estos dispositivos de forma rutinaria Los glucómetros han demostrado un alto grado de exactitud 2,3 , y en la práctica clínica, los valores basales de glucemia en sangre capilar (GC) y de glucemia plasmática en sangre venosa (GP) son a menudo…”
Implementing formulary change in diabetes
Swift, J., Pearson, K., Hill, J., Cheung, J., Richardson, D., Myers, A. J., & Barker, C. (2017) · British Journal of Healthcare Management · doi:10.12968/bjhc.2017.23.4.167 · Cited by 3
Do We Need the Replacement of YSI 2300? A View from the Clinical Laboratory
Spanou, L., & Makris, K. (2021) · Journal of Diabetes Science and Technology · doi:10.1177/19322968211007170 · Cited by 4
Teodorczyk, M., Nandagopalan, S., Maguire, P., & Stegmann, J. (2013) · Journal of Diabetes Science and Technology · doi:10.1177/193229681300700325 · Cited by 3
Impact of Meal Constituents on Artificial Pancreas Algorithms
Campos-Náñez, E., & Kovatchev, B. P. (2016) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2016.0335 · Cited by 3
“Each bar presents A1c levels obtained using 43 realistic (commercially available) meters.(16) 608 CAMPOS-NÁÑEZ AND KOVATCHEV”
Method7 th Annual Symposium on Self-Monitoring of Blood Glucose (SMBG), May 8–10, 2014, Helsinki, Finland
Parkin, C. G., Mlinac, A., & Hinzmann, R. (2014) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2014.0239 · Cited by 2
“, miscoding meters to test strips) can exacerbate errors and increase the risk for hypoglycemia.(55,56) Because clinical outcome trials are inherently difficult to execute and likely unethical because of the potential danger to study participants, use of computer simulations to assess the influence of blood glucose me…”
BackgroundExcellence in acute cardiac care: an overview of the seminar proceedings
Thygesen, K. A., Jaffe, A. S., Apple, F. S., Möckel, M., Venge, P., Plebani, M., Defillipi, C., Frye, B., Wallseth, A., & Stokfisch, V. L. (2014) · European Heart Journal Acute Cardiovascular Care · doi:10.1177/2048872614541905 · Cited by 2
Alabeidi, F., Alanazi, A. R., Alsaeed, S. A., & Alanzi, H. O. (2021) · International Journal of Health Sciences · doi:10.53730/ijhs.v1ns1.15256 · Cited by 2
Glukosemessung in der Diabetestherapie
Pleus, S., Buck, S., Heinemann, L., & Freckmann, G. (2025) · Die Diabetologie · doi:10.1007/s11428-025-01338-6 · Cited by 2
Self-citationThe potential for POCT in the Internet of Things (IoT)
Rode-Schubert, C., Norgall, T., & Bietenbeck, A. (2018) · Springer eBooks · doi:10.1007/978-3-662-54497-6_42 · Cited by 2
Dianov, O. A., Мальцев, В. В., Мурина, Е. С., & Анфилов, И. С. (2020) · Problems of Endocrinology · doi:10.14341/probl12311 · Cited by 3
Moriuchi, T., Otaki, Y., Satou, H., Chai, F., Hayashida, Y., Aikawa, R., Sugiyama, T., & Sode, K. (2021) · Journal of Diabetes Science and Technology · doi:10.1177/19322968211060865 · Cited by 2
Zayar, N. N., Chotipanvithayakul, R., Bjertness, E., Htet, A. S., Geater, A. F., & Chongsuvivatwong, V. (2023) · International Journal of General Medicine · doi:10.2147/ijgm.s439141 · Cited by 2
Smith, R., Borg, R., Wong, V., Russell, H., & Mak, M. (2025) · Nutrients · doi:10.3390/nu17030400 · Cited by 2
“Participants were provided with either the Freestyle Lite ® (Abbott Diabetes Care) or Accu-Chek Nano ® (Roche Diagnostics) blood glucose meters, both of which met ISO accuracy standards [25].”
Estimation and personalization of clinical insulin therapy parameters
Palma, R. C. (2013) · Texas ScholarWorks (Texas Digital Library) · Cited by 6
Chesa, J. L. B., Nieto, M. L. H., Hernández, F. J. L., Oliver, P., & de la Presa, B. G. (2017) · Revista del Laboratorio Clínico · doi:10.1016/j.labcli.2017.01.003 · Cited by 2
Surapaneni, K. M., Kandikattu, R., Singh, A., & Joshi, A. (2017) · Annals of Tropical Medicine and Public Health · doi:10.4103/1755-6783.205553 · Cited by 2
“Previous studies have advocated home blood pressure monitoring (HBPM) and home blood sugar monitoring (HBSM) as adjuncts to prevent adverse effects of hypertension and diabetes.[14-18] The third Indian guidelines on hypertension have encouraged HBPM due to its usefulness at follow-ups.”
BackgroundDelevatti, R. S., Kanitz, A. C., Alberton, C. L., Marson, E. C., Pantoja, P. D., Pinho, C. D. F., Lisboa, S. C., & Kruel, L. F. M. (2018) · Frontiers in Physiology · doi:10.3389/fphys.2018.01609 · Cited by 3
Software-based Monitoring for Calibration of Measurement Units in Real-time Systems
Maruf, M. A., & Azim, A. (2018) · Annual Conference of the IEEE Industrial Electronics Society (IECON) · doi:10.1109/iecon.2018.8591626 · Cited by 2
Real-Time Continuous Glucose Monitoring in Adult Outpatients
Bonomo, M. A., Grassi, G., Bartolo, P. D., & Maran, A. (2014) · Frontiers in diabetes · doi:10.1159/000363482 · Cited by 1
Pfützner, A. (2013) · Journal of Diabetes Science and Technology · doi:10.1177/193229681300700118 · Cited by 1
Fournier, P. A. (2013) · Journal of Diabetes Science and Technology · doi:10.1177/193229681300700521 · Cited by 1
Rott, H., & Koschinsky, T. (2018) · Springer eBooks · doi:10.1007/978-3-662-54497-6_33 · Cited by 1
Baumstark, A., Jendrike, N. R. M., Pleus, S., Waldenmaier, D., Zschornack, E., Link, M., Kamecke, U., Mende, J., & Freckmann, G. (2019) · Diabetologie und Stoffwechsel · doi:10.1055/a-0987-0015 · Cited by 1
“[12] Freckmann G, Schmid C, Baumstark A et al.”
BackgroundSelf-citationSilvanus, V., Kafle, P. P., Pokhrel, A., Baral, B. K., & Pokhrel, B. R. (2019) · Nepal Medical College Journal · doi:10.3126/nmcj.v21i4.27615 · Cited by 1
Modeling the SMBG measurement error
Vettoretti, M. (2020) · Elsevier eBooks · doi:10.1016/b978-0-12-816714-4.00005-3 · Cited by 1
Biosensor-Based Point-of-Care Devices: Metabolites and Pulse Oximetry
Hwang, I. M., Lou, X. A., Toubian, A. A., & Kamei, D. T. (2021) · Springer eBooks · doi:10.1007/978-981-16-2782-8_4 · Cited by 1
Renugopalakrishnan, V., Koester, S. J., McDevitt, M. R., Ajayan, P. M., Liepmann, D., Demirci, U., Paulmurugan, R., & Viswanathan, S. (2015) · MRS Proceedings · doi:10.1557/opl.2015.344 · Cited by 1
Metadata
- Authors
- Freckmann, G., Schmid, C., Baumstark, A., Pleus, S., Link, M., & Haug, C.
- Affiliations
- Universität Ulm (DE)
- Corresponding author
- Freckmann, Guido
- Venue
- Journal of Diabetes Science and Technology
- Publisher
- SAGE Publications
- Published
- 2012-09
- Volume
- 6
- Issue
- 5
- Pages
- 1060-1075
- ISSN
- 1932-2968 (ISSN-L)
- PMC
- PMC3570840
- OpenAlex
- W2153556745
- Semantic Scholar
- f51002e49bd0efca6a14d70d53eb7f2e22fb03d4
- Cited by
- 237
- References
- 36
- Citations by year
- 2026: 1 · 2025: 4 · 2024: 9 · 2023: 13 · 2022: 8 · 2021: 11 · 2020: 16 · 2019: 19 · 2018: 30 · 2017: 28 · 2016: 23 · 2015: 22 · 2014: 31 · 2013: 20 · 2012: 1
- FWCI
- 18.99
- Type
- journal article
- Language
- en
- Open access
- bronze
- doi.org
- Licence
- journals.sagepub.com/page/policies/text-and-data-mining-license (tdm)
- Locations
- Journal of Diabetes Science and Technology (publishedVersion)
- Europe PMC (PubMed Central) (publishedVersion)
- PubMed Central (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
- Health Sciences › Health Professions › Medical Laboratory Technology › Clinical Laboratory Practices and Quality Control
- Physical Sciences › Physics and Astronomy › Instrumentation › Scientific Measurement and Uncertainty Evaluation
- Keywords
- self-monitoring of blood glucose, blood glucose meters, test strips, measurement accuracy, CE marking, glucose oxidase method, hexokinase method, bias, diabetes management
- MeSH terms
- Adolescent, Adult, Aged, Aged, 80 and over, Blood Glucose, Diabetes Mellitus, Equipment and Supplies, European Union, Female, Humans, Insulin Infusion Systems, Male, Middle Aged, Oxygen, Reference Standards, Blood Glucose Self-Monitoring, Reproducibility of Results, Device Approval, Young Adult
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.