Technical and clinical accuracy of three blood glucose meters: clinical impact assessment using error grid analysis and insulin sliding scales
The DOI matches its Crossref record; 8 of the 8 references it cites were found in a registry or an index.
Abstract
Aims: Manufacturers of point-of-care testing (POCT) glucose systems must adhere to International Organisation for Standardisation (ISO) 15197 performance standards, a tightened version of which becomes mandatory in May 2016. Using methods deviating from the evaluation requirements of ISO 15197, we investigated the accuracy and clinical utility of three glucose testing systems. Methods: Whole blood glucose was measured on 62 EDTA-preserved samples with each system followed by plasma glucose determination with a laboratory hexokinase method. Technical accuracy was assessed using linear regression, Bland-Altman plots and bias. Clinical utility was evaluated using consensus error grid analysis and insulin dosing algorithms. Results: The correlation coefficient of the glucose measuring devices with the comparator method was between 0.979 and 0.993 (n=61). Bias against the reference method was +0.7–+4.4%. The FreeStyle Lite (0.7±4.4%) and ACCU-CHEK Aviva (1.3±7.3%) systems had the least bias to the reference. Error grid analysis depicts all meters to be clinically accurate, with all results falling within Zones A and B. Conclusions: Conformité Européene-marked devices are presumed to have adequate accuracy, but postmarket evaluations are imperative to assess total system accuracy and whether regulatory targets are achieved.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Parkes, J. L., Slatin, S. L., Pardo, S., & Ginsberg, B. H. (2000) · Diabetes Care · doi:10.2337/diacare.23.8.1143 · Cited by 620
Factors affecting blood glucose monitoring: sources of errors in measurement
Ginsberg, B. H. (2009) · Journal of Diabetes Science and Technology · doi:10.1177/193229680900300438 · Cited by 348
Freckmann, G., Schmid, C., Baumstark, A., Pleus, S., Link, M., & Haug, C. (2012) · Journal of Diabetes Science and Technology · doi:10.1177/193229681200600510 · Cited by 237
System Accuracy Evaluation of 27 Blood Glucose Monitoring Systems According to DIN EN ISO 15197
Freckmann, G., Baumstark, A., Jendrike, N., Zschornack, E., Kocher, S., Tshiananga, J., Heister, F., & Haug, C. (2010) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2009.0128 · Cited by 195
Technical Aspects of the Parkes Error Grid
Pfützner, A., Klonoff, D. C., Pardo, S., & Parkes, J. L. (2013) · Journal of Diabetes Science and Technology · doi:10.1177/193229681300700517 · Cited by 122
Overview on self-monitoring of blood glucose
Montagnana, M., Caputo, M., Giavarina, D., & Lippi, G. (2009) · Clinica Chimica Acta · doi:10.1016/j.cca.2009.01.002 · Cited by 116
Value of Self-Monitoring Blood Glucose Pattern Analysis in Improving Diabetes Outcomes
Parkin, C. G., & Davidson, J. A. (2009) · Journal of Diabetes Science and Technology · doi:10.1177/193229680900300314 · Cited by 96
Klaff, L. J., Brazg, R., Hughes, K., Tideman, A. M., Schachner, H. C., Stenger, P., Pardo, S., Dunne, N., & Parkes, J. L. (2015) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2014.0069 · Cited by 46
Research works printed but not linked: 11
- 10.1016/j.clinthera.2005.10.010 · DOI
- Holko, 2011, Cost effectiveness and cost utility of the noncoding blood glucose meter CONTOUR®TS, Diabetes Metab Syndr Obes, 4, 79
- 10.1016/j.electacta.2009.09.019 · doi:10.1016/j.electacta.2009.09.019
- Hönes, 2008, The technology behind glucose meters: test strips, Diabetes Technol Ther, 10, S10, 10.1089/dia.2008.0005 · DOI · Paper page
- 10.2337/diacare.10.5.622 · DOI · Paper page
- 10.17925/USE.2013.09.01.16 · DOI
- 10.1089/152091503322526969 · DOI · Paper page
- 10.1185/03007995.2012.673479 · DOI · Paper page
- 10.1089/dia.2011.0170 · DOI · Paper page
- Boyd, 2001, Quality specifications for glucose meters: assessment by simulation modeling of errors in insulin dose, Clin Chem, 47, 209, 10.1093/clinchem/47.2.209 · DOI · Paper page
- Skeie, 2001, Patient-derived quality specifications for instruments used in self-monitoring of blood glucose, Clin Chem, 47, 67, 10.1093/clinchem/47.1.67 · DOI · Paper page
Cited by
Zhou, B., Lu, Y., Hajifathalian, K., Bentham, J., Di Cesare, M., Danaei, G., Bixby, H., Cowan, M., Ali, M., Taddei, C., et al. (2016) · The Lancet · doi:10.1016/s0140-6736(16)00618-8 · Cited by 4,311
Comparative accuracy of 17 point-of-care glucose meters
Ekhlaspour, L., Mondesir, D., Lautsch, N., Balliro, C., Hillard, M., Magyar, K., Radocchia, L. G., Esmaeili, A., Sinha, M., & Russell, S. J. (2017) · Journal of Diabetes Science and Technology · doi:10.1177/1932296816672237 · Cited by 127
Global variation in diabetes diagnosis and prevalence based on fasting glucose and hemoglobin A1c
NCD Risk Factor Collaboration (NCD-RisC), Zhou, B., Sheffer, K. E., Bennett, J. E., Gregg, E. W., Danaei, G., Singleton, R. K., Shaw, J. E., Mishra, A., Lhoste, V. P. F., et al. (2023) · Nature Medicine · doi:10.1038/s41591-023-02610-2 · Cited by 94
Baumstark, A., Jendrike, N., Pleus, S., Haug, C., & Freckmann, G. (2017) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2016.0408 · Cited by 22
Metadata
- Authors
- Sutheran, H. L., & Reynolds, T.
- Venue
- Journal of Clinical Pathology
- Publisher
- BMJ
- Published
- 2016-02-05
- Volume
- 69
- Issue
- 10
- Pages
- 899-905
- ISSN
- 0021-9746 (ISSN-L)
- 1472-4146
- OpenAlex
- W2169676099
- Cited by
- 10
- References
- 8
- FWCI
- 1.37
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Technical and clinical accuracy of three blood glucose meters: clinical impact assessment using error grid analysis and insulin sliding scales. https://app.cochat.ai/papers/10.1136/jclinpath-2015-203339
- The paper
- Sutheran, H. L., & Reynolds, T. (2016). Technical and clinical accuracy of three blood glucose meters: clinical impact assessment using error grid analysis and insulin sliding scales. Journal of Clinical Pathology. https://doi.org/10.1136/jclinpath-2015-203339
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.