Impact of Glucose Meter Error on Glycemic Variability and Time in Target Range During Glycemic Control After Cardiovascular Surgery
Open accessAbstract
Background: We retrospectively studied the impact of glucose meter error on the efficacy of glycemic control after cardiovascular surgery. Method: Adult patients undergoing intravenous insulin glycemic control therapy after cardiovascular surgery, with 12-24 consecutive glucose meter measurements used to make insulin dosing decisions, had glucose values analyzed to determine glycemic variability by both standard deviation (SD) and continuous overall net glycemic action (CONGA), and percentage glucose values in target glucose range (110-150 mg/dL). Information was recorded for 70 patients during each of 2 periods, with different glucose meters used to measure glucose and dose insulin during each period but no other changes to the glycemic control protocol. Accuracy and precision of each meter were also compared using whole blood specimens from ICU patients. Results: Glucose meter 1 (GM1) had median bias of 11 mg/dL compared to a laboratory reference method, while glucose meter 2 (GM2) had a median bias of 1 mg/dL. GM1 and GM2 differed little in precision (CV = 2.0% and 2.7%, respectively). Compared to the period when GM1 was used to make insulin dosing decisions, patients whose insulin dose was managed by GM2 demonstrated reduced glycemic variability as measured by both SD (13.7 vs 21.6 mg/dL, P < .0001) and CONGA (13.5 vs 19.4 mg/dL, P < .0001) and increased percentage glucose values in target range (74.5 vs 66.7%, P = .002). Conclusions: Decreasing glucose meter error (bias) was associated with decreased glycemic variability and increased percentage of values in target glucose range for patients placed on intravenous insulin therapy following cardiovascular surgery.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“Meters that are designed for professional use have been shown in published studies to be accurate on samples of whole blood (129-131).”
From Guidelines and Recommendations for Laboratory Analysis in the Diagnosis and Management of Diabetes Mellitus (2023) · cites this paper as [131]
In the citing paper
“Changing from one meter to a meter with less meter error (bias) was associated with decreased glycemic variability and increased percentage of values in target glucose range in patients following cardiovascular surgery (131).”
From Guidelines and Recommendations for Laboratory Analysis in the Diagnosis and Management of Diabetes Mellitus (2023) · cites this paper as [131]
Citations
Cites
Intensive insulin therapy in critically ill patients.
Van den Berghe, G., Wouters, P., Weekers, F., Verwaest, C., Bruyninckx, F., Schetz, M., Vlasselaers, D., Ferdinande, P., Lauwers, P., & Bouillon, R. (2001) · New England Journal of Medicine · doi:10.1056/nejmoa011300 · Cited by 9,977
Intensive versus Conventional Glucose Control in Critically Ill Patients
(2009) · New England Journal of Medicine · doi:10.1056/nejmoa0810625 · Cited by 5,119
Reliability of point-of-care testing for glucose measurement in critically ill adults*
Kanji, S., Buffie, J., Hutton, B., Bunting, P. S., Singh, A., McDonald, K., Fergusson, D., McIntyre, L. A., & Hebert, P. C. (2005) · Critical Care Medicine · doi:10.1097/01.ccm.0000189939.10881.60 · Cited by 395 · cited 2×
“is observed at glucose concentrations > 150 mg/dL.(6,26) Because simulation models are limited in the ability to model the effects of changing bias or precision,(16) the inability to adequately model the effects of proportional positive bias observed with GM1 may have caused previous investigators to significantly und…”
BackgroundGlucose Measurement: Confounding Issues in Setting Targets for Inpatient Management
Dungan, K., Chapman, J., Braithwaite, S. S., & Buse, J. (2007) · Diabetes Care · doi:10.2337/dc06-1679 · Cited by 292
Tang, Z., Louie, R. F., Lee, J. H., Lee, D. M., Miller, E. E., & Kost, G. J. (2001) · Critical Care Medicine · doi:10.1097/00003246-200105000-00038 · Cited by 188 · cited 1×
“However, patients in the intensive care unit (ICU) are on multiple medications, and often have abnormal hematocrit, which may affect the performance of glucose meters.(4,5) The degree to which glucose meters correlate with laboratory glucose measurement varies between glucose meter technologies,(4) and correlation in…”
BackgroundTight glucose control in the intensive care unit: Are glucose meters up to the task?
Scott, M. G., Bruns, D. E., Boyd, J. C., & Sacks, D. B. (2009) · Clinical Chemistry · doi:10.1373/clinchem.2008.117291 · Cited by 162 · cited 4×
“Though newer meter technologies may provide more accurate glucose measurement,(8-10) there is still substantial concern about the use of glucose meters for management of glycemic control in the ICU.(11,12) 602099 DSTXXX10.”
BackgroundInfluentialKhan, A. I., Vasquez, Y., Gray, J., Wians, F. H., & Kroll, M. H. (2006) · Archives of Pathology & Laboratory Medicine · doi:10.5858/2006-130-1527-tvorbp · Cited by 145
Karon, B. S., Griesmann, L. D., Scott, R. J., Bryant, S. C., DuBois, J. A., Shirey, T., Presti, S., & Santrach, P. J. (2008) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2007.0257 · Cited by 132 · cited 1×
“However, patients in the intensive care unit (ICU) are on multiple medications, and often have abnormal hematocrit, which may affect the performance of glucose meters.(4,5) The degree to which glucose meters correlate with laboratory glucose measurement varies between glucose meter technologies,(4) and correlation in…”
Self-citationBreton, M. D., & Kovatchev, B. P. (2010) · Journal of Diabetes Science and Technology · doi:10.1177/193229681000400309 · Cited by 103 · cited 1×
“These authors predicted that glycemic variability, risk of hypoglycemia, and observed incidence of hypoglycemia would all increase as the allowable error of the meters increased between 5% and 20%, with a threshold effect for risk of hypoglycemia between 10% and 15% meter error.(17) While this model remains one of the…”
BackgroundKaron, B. S., Gandhi, G. Y., Nuttall, G. A., Bryant, S. C., Schaff, H. V., McMahon, M. M., & Santrach, P. J. (2007) · American Journal of Clinical Pathology · doi:10.1309/6rfqckaajgkwb8m4 · Cited by 98 · cited 1×
“is observed at glucose concentrations > 150 mg/dL.(6,26) Because simulation models are limited in the ability to model the effects of changing bias or precision,(16) the inability to adequately model the effects of proportional positive bias observed with GM1 may have caused previous investigators to significantly und…”
BackgroundSelf-citation
Research works printed but not linked: 1
- Clinical and Laboratory Standards Institute, 2013, Point of Care Blood Glucose Testing in Acute and Chronic Care Facilities: Approved Guideline
Cited by
Sacks, D. B., Arnold, M., Bakris, G. L., Bruns, D. E., Horvath, A. R., Lernmark, Å., Metzger, B. E., Nathan, D. M., & Kirkman, M. S. (2023) · Diabetes Care · doi:10.2337/dci23-0036 · Cited by 351
Sacks, D. B., Arnold, M. A., Bakris, G. L., Bruns, D. E., Horvath, A. R., Lernmark, Å., Metzger, B. E., Nathan, D. M., & Kirkman, M. S. (2023) · Clinical Chemistry · doi:10.1093/clinchem/hvad080 · Cited by 200
Accuracy of Blood Glucose Measurement and Blood Glucose Targets
Eerdekens, G.-J., Rex, S., & Mesotten, D. (2020) · Journal of Diabetes Science and Technology · doi:10.1177/1932296820905581 · Cited by 46
Neonatal Hyperglycemia—Causes, Treatments, and Cautions
Hay, W. W., & Rozance, P. J. (2018) · The Journal of Pediatrics · doi:10.1016/j.jpeds.2018.04.046 · Cited by 38
Karon, B. S., Donato, L. J., Larsen, C. M., Siebenaler, L. K., Wells, A. E., Wood‐Wentz, C. M., Marienau, M. E. S., & Curry, T. B. (2017) · Anesthesiology · doi:10.1097/aln.0000000000001708 · Cited by 21
Self-citation
Metadata
- Authors
- Karon, B. S., Meeusen, J. W., & Bryant, S. C.
- Affiliations
- Mayo Clinic (US)
- Corresponding author
- Karon, Brad S.
- Venue
- Journal of Diabetes Science and Technology
- Publisher
- SAGE Publications
- Published
- 2015-08-25
- Volume
- 10
- Issue
- 2
- Pages
- 336-342
- ISSN
- 1932-2968 (ISSN-L)
- PMC
- PMC4773953
- OpenAlex
- W1849025242
- Semantic Scholar
- f0a1073f2ee192bce4ad53188f23d5a5ecdee174
- Cited by
- 16
- References
- 25
- Citations by year
- 2025: 1 · 2023: 2 · 2022: 2 · 2020: 1 · 2018: 4 · 2017: 5 · 2016: 1
- FWCI
- 2.90
- Type
- journal article
- Language
- en
- Open access
- bronze
- doi.org
- Licence
- journals.sagepub.com/page/policies/text-and-data-mining-license (tdm)
- Locations
- Journal of Diabetes Science and Technology (publishedVersion)
- Europe PMC (PubMed Central) (publishedVersion)
- PubMed Central (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Hyperglycemia and glycemic control in critically ill and hospitalized patients
- Health Sciences › Health Professions › Medical Laboratory Technology › Clinical Laboratory Practices and Quality Control
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
- Keywords
- glycemic variability, time in range, intravenous insulin therapy, postoperative glycemic control, cardiovascular surgery, blood glucose monitoring
- MeSH terms
- Adult, Aged, Aged, 80 and over, Blood Glucose, Female, Humans, Hyperglycemia, Hypoglycemia, Hypoglycemic Agents, Insulin, Intensive Care Units, Male, Middle Aged, Retrospective Studies, Cardiovascular Surgical Procedures, Incidence, Administration, Intravenous
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Impact of Glucose Meter Error on Glycemic Variability and Time in Target Range During Glycemic Control After Cardiovascular Surgery. https://app.cochat.ai/papers/10.1177/1932296815602099
- The paper
- Karon, B. S., Meeusen, J. W., & Bryant, S. C. (2016). Impact of Glucose Meter Error on Glycemic Variability and Time in Target Range During Glycemic Control After Cardiovascular Surgery. Journal of Diabetes Science and Technology. https://doi.org/10.1177/1932296815602099
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.