Clinical assessment of the accuracy of blood glucose measurement devices

Pfützner, Andreas, Mitri, Michael, Musholt, Petra B., Sachsenheimer, Daniela, Borchert, Marcus, Yap, Andrew, Forst, Thomas · 2012 · Current Medical Research and Opinion
Cited by 64

The DOI matches its Crossref record; 5 of the 5 references it cites were found in a registry or an index.

Abstract

OBJECTIVES: Blood glucose meters for patient self-measurement need to comply with the accuracy standards of the ISO 15197 guideline. We investigated the accuracy of the two new blood glucose meters BG*Star and iBG*Star (Sanofi-Aventis) in comparison to four other competitive devices (Accu-Chek Aviva, Roche Diagnostics; FreeStyle Freedom Lite, Abbott Medisense; Contour, Bayer; OneTouch Ultra 2, Lifescan) at different blood glucose ranges in a clinical setting with healthy subjects and patients with type 1 and type 2 diabetes. BGStar and iBGStar are employ dynamic electrochemistry, which is supposed to result in highly accurate results. METHODS: The study was performed on 106 participants (53 female, 53 male, age (mean ± SD): 46 ± 16 years, type 1: 32 patients, type 2: 34 patients, and 40 healthy subjects). Two devices from each type and strips from two different production lots were used for glucose assessment (∼200 readings/meter). Spontaneous glucose assessments and glucose or insulin interventions under medical supervision were applied to perform measurements in the different glucose ranges in accordance with the ISO 15197 requirements. Sample values <50 mg/dL and >400 mg/dL were prepared by laboratory manipulations. The YSI glucose analyzer (glucose oxidase method) served as the standard reference method which may be considered to be a limitation in light of glucose hexokinase-based meters. RESULTS: For all devices, there was a very close correlation between the glucose results compared to the YSI reference method results. The correlation coefficients were r = 0.995 for BGStar and r = 0.992 for iBGStar (Aviva: 0.995, Freedom Lite: 0.990, Contour: 0.993, Ultra 2: 0.990). Error-grid analysis according to Parkes and Clarke revealed both 100% of the readings to be within the clinically acceptable areas (Clarke: A + B with BG*Star (100 + 0), Aviva (97 + 3), and Contour (97 + 3); and 99.5% with iBG*Star (97.5 + 2), Freedom Lite (98 + 1.5), and Ultra 2 (97.5 + 2)). CONCLUSIONS: This study demonstrated the very high accuracy of BG*Star, iBG*Star, and the competitive blood glucose meters in a clinical setting.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1185/03007995.2012.673479

Cites

5 of 5 references verified0 citation errors · 0 not checkable

Statistics

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 5 references

Research works printed but not linked: 18

  1. 10.4158/EP.13.S1.1 · DOI · Paper page
  2. 10.1016/j.amjmed.2005.07.055 · DOI
  3. 10.1007/s00125-008-0946-7 · DOI
  4. 10.1185/030079906X148274 · DOI · Paper page
  5. Lunt H, 2010, NZ Med J, 123, 74
  6. 10.1177/193229681000400111 · DOI
  7. Passing H, 1983, J Clin Chem Clin Biochem, 21, 709
  8. 10.2337/diacare.10.5.622 · DOI · Paper page
  9. 10.1039/b005967f · DOI
  10. Iyengar S, 2007, Diabetes, Stoffw Herz,, 16, 15
  11. 10.1053/j.ajkd.2005.09.007 · DOI
  12. 10.1089/dia.2007.0257 · DOI · Paper page
  13. Pfützner A, 2009, Diabetes Stoffw Herz, 18, 387
  14. 10.1177/193229681100500520 · DOI · Paper page
  15. 10.1055/s-2006-924011 · DOI
  16. 10.1177/193229680900300438 · DOI · Paper page
  17. 10.1177/193229681100500128 · DOI
  18. 10.1089/dia.2010.0116 · DOI

Cited by

See all 64

Metadata

Authors
Pfützner, A., Mitri, M., Musholt, P. B., Sachsenheimer, D., Borchert, M., Yap, A., & Forst, T.
Venue
Current Medical Research and Opinion
Published
2012-03-21
Volume
28
Issue
4
Pages
525-531
ISSN
  • 0300-7995 (ISSN-L)
  • 1473-4877
OpenAlex
W2017141557
Cited by
64
References
5
FWCI
8.51
Type
journal article
Language
en
Open access
closed

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: Clinical assessment of the accuracy of blood glucose measurement devices. https://app.cochat.ai/papers/10.1185/03007995.2012.673479
The paper
Pfützner, A., Mitri, M., Musholt, P. B., Sachsenheimer, D., Borchert, M., Yap, A., & Forst, T. (2012). Clinical assessment of the accuracy of blood glucose measurement devices. Current Medical Research and Opinion. https://doi.org/10.1185/03007995.2012.673479

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.