Incidence and timing of potentially high-risk arrhythmias detected through long term continuous ambulatory electrocardiographic monitoring
Open accessAbstract
BACKGROUND: Ambulatory electrocardiographic (ECG) monitoring is the standard to screen for high-risk arrhythmias. We evaluated the clinical utility of a novel, leadless electrode, single-patient-use ECG monitor that stores up to 14 days of a continuous recording to measure the burden and timing of potentially high-risk arrhythmias. METHODS: We examined data from 122,815 long term continuous ambulatory monitors (iRhythm ZIO® Service, San Francisco) prescribed from 2011 to 2013 and categorized potentially high-risk arrhythmias into two types: (1) ventricular arrhythmias including non-sustained and sustained ventricular tachycardia and (2) bradyarrhythmias including sinus pauses >3 s, atrial fibrillation pauses >5 s, and high-grade heart block (Mobitz Type II or third-degree heart block). RESULTS: Of 122,815 ZIO® recordings, median wear time was 9.9 (IQR 6.8-13.8) days and median analyzable time was 9.1 (IQR 6.4-13.1) days. There were 22,443 (18.3%) with at least one episode of non-sustained ventricular tachycardia (NSVT), 238 (0.2%) with sustained VT, 1766 (1.4%) with a sinus pause >3 s (SP), 520 (0.4%) with a pause during atrial fibrillation >5 s (AFP), and 1486 (1.2%) with high-grade heart block (HGHB). Median time to first arrhythmia was 74 h (IQR 26-149 h) for NSVT, 22 h (IQR 5-73 h) for sustained VT, 22 h (IQR 7-64 h) for SP, 31 h (IQR 11-82 h) for AFP, and 40 h (SD 10-118 h) for HGHB. CONCLUSIONS: A significant percentage of potentially high-risk arrhythmias are not identified within 48-h of ambulatory ECG monitoring. Longer-term continuous ambulatory ECG monitoring provides incremental detection of these potentially clinically relevant arrhythmic events.
AI snapshot
Written from the paper’s full text.
TL;DR
In 122,815 long-term continuous ambulatory ECG records, potentially high-risk ventricular and bradyarrhythmias were common, and many were first detected after 48 hours of monitoring. Extending monitoring toward 14 days increased detection of these events beyond traditional 24- or 48-hour Holter strategies.
Key findings
- Among 122,815 eligible records, median wear time was 9.9 days (IQR 6.8–13.8) and median analyzable time was 9.1 days (IQR 6.4–13.1). Detected arrhythmias included non-sustained VT in 22,443 (18%), sustained VT in 238 (0.2%), sinus pause >3 s in 1,766 (1.4%), AF pause >5 s in 521 (0.4%), and high-grade heart block in 1,468 (0.4%).
- Median time to first arrhythmia was 74 h (IQR 26–149) for non-sustained VT, 22 h (IQR 5–73) for sustained VT, 22 h (IQR 7–64) for sinus pause, 31 h (IQR 11–82) for AF pause, and 40 h (SD 10–118) for high-grade heart block.
- Cumulative detection increased with longer monitoring: sustained VT was 52.5% detected at 24 h, 65.5% at 48 h, and 92.9% by 7 days, with an additional 7.1% detected between 7 and 14 days. Non-sustained VT was 23.4%, 38.0%, and 79.4% detected by 1, 2, and 7 days, respectively.
- For sinus pauses >3 s, 31.7% were detected within 1 day, 46.6% within 2 days, and 83.1% within 3 days. Diagnostic yields at 2 days versus 7 days for ventricular arrhythmias and bradyarrhythmias were significantly different in the overall populations (P < 0.01).
Methods
- Retrospective analysis of all ZIO® Service long-term continuous ambulatory ECG monitors prescribed from November 2011 to December 2013 (N = 128,401), cleaned to 122,815 eligible records from 122,454 unique patients; exclusions included age <18 years, wear or analyzable time <24 h, heart rates >300 or <20 beats/min, and start times outside study dates.
- The ZIO® Patch is a leadless, single-patient-use monitor that records up to 14 days of continuous single-channel ECG. Outcomes were categorized as ventricular arrhythmias (non-sustained and sustained VT >30 s) and bradyarrhythmias (sinus pauses >3 s, AF pauses >5 s, and high-grade heart block including Mobitz Type II or third-degree heart block); symptomatic pauses occurred within 45 s of a patient trigger.
- Analyses used SAS 9.3 and calculated proportions with 95% confidence intervals, cumulative yield per additional day of monitoring, and chi-squared tests comparing high-risk arrhythmia detection at 1, 2, and 7 days versus 14 days.
Limitations
- The paper states: no data on changes in clinical management or patient outcomes after monitoring; incomplete information on all symptomatic triggers; limited demographic data and no comorbidity data; some bradyarrhythmias such as asymptomatic sinus and AF pauses may occur nocturnally in normal subjects; clinical indications for monitor ordering were not validated; and at least 25% of patients wore the device <7 days (25th percentile 6.8 days), artificially reducing observed yield from days 7 to 14, so actual yield may be higher.
- Evident from the text: this is an observational analysis without a control or comparison monitoring strategy, and arrhythmia identification relied on manufacturer algorithms and Certified Cardiac Technician review.
- Evident from the text: there are minor numerical inconsistencies between the abstract and results/table for AF pause >5 s (520 vs 521) and high-grade heart block (1,486 vs 1,468).
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
0 of 17 cited sources checked · 0 supported · 0 flagged
- Findingown data
The diagnostic yields at 2 days versus 7 days for ventricular arrhythmias and bradyarrhythmias were significantly different from each other in the overall populations (P < 0.01).
“The diagnostic yields at 2 days versus 7 days for ventricular arrhythmias and bradyarrhythmias were significantly different from each other in the overall populations (P < 0.01).”
Quote in the paperSubstantial claim - Findingown data
For sustained VT, 65.5% were identified by 48 h.
“For sustained VT, only 52.5 % of the total identified arrhythmias were identified at 24 h, and approximately two-thirds (65.5 %) were identified by 48 h.”
Quote in the paperSubstantial claim - Findingown data
Additional 7 days of monitoring between 7 and 14 days yielded an additional 7.1% of sustained VT arrhythmias.
“Most arrhythmias were identified by 7 days (92.9 %), but the additional 7 days of monitoring between 7 and 14 days yielded an additional 7.1 % of these potentially lethal arrhythmias.”
Quote in the paperSubstantial claim - Findingown data
There were 22,443 (18%) recordings with nonsustained VT.
“Of the 122,815 eligible records, there were 22,443 (18 %) with nonsustained VT, 238 (0.2 %) with sustained VT, 1766 (1.4 %) with sinus pauses >3 s, 521 (0.4 %) with AF pauses >5 s, 249 (0.2 %) with symptomatic pauses and 1468 (0.4 %) with high-grade heart block (Table 1).”
Quote in the paperNot a substantial claim - Findingown data
For non-sustained VT, 38.0% identified by 2 days.
“A similar trend was seen for the more common non-sustained VT, with 23.4, 38.0, and 79.4 % of all non-sustained VT being identified by 1, 2, and 7 days respectively.”
Quote in the paperNot a substantial claim
- Findingown data
Median time to first arrhythmia was 22 h (IQR 5-73 h) for sustained VT.
“Median time to first arrhythmia was 74 h (IQR 26–149 h) for NSVT, 22 h (IQR 5–73 h) for sustained VT, 22 h (IQR 7–64 h) for SP, 31 h (IQR 11–82 h) for AFP, and 40 h (SD 10–118 h) for HGHB.”
Quote in the paperNot a substantial claim - Findingown data
For sinus pauses >3 s, 31.7% of total detected arrhythmias were found within 1 day.
“For the most common bradyarrhythmia, sinus pauses >3 s (N = 1766), 31.7 % of the total detected arrhythmias were found within 1 day, 46.6 % within 2 days, and 83.1 % within 3 days of monitoring.”
Quote in the paperSubstantial claim - Findingown data
For sustained VT, 92.9% were identified by 7 days.
“Most arrhythmias were identified by 7 days (92.9 %), but the additional 7 days of monitoring between 7 and 14 days yielded an additional 7.1 % of these potentially lethal arrhythmias.”
Quote in the paperSubstantial claim - Findingown data
For non-sustained VT, 79.4% identified by 7 days.
“A similar trend was seen for the more common non-sustained VT, with 23.4, 38.0, and 79.4 % of all non-sustained VT being identified by 1, 2, and 7 days respectively.”
Quote in the paperSubstantial claim - Findingown data
There were 1766 (1.4%) recordings with sinus pauses >3 s.
“Of the 122,815 eligible records, there were 22,443 (18 %) with nonsustained VT, 238 (0.2 %) with sustained VT, 1766 (1.4 %) with sinus pauses >3 s, 521 (0.4 %) with AF pauses >5 s, 249 (0.2 %) with symptomatic pauses and 1468 (0.4 %) with high-grade heart block (Table 1).”
Quote in the paperNot a substantial claim - Findingown data
Median time to first arrhythmia was 22 h (IQR 7-64 h) for SP.
“Median time to first arrhythmia was 74 h (IQR 26–149 h) for NSVT, 22 h (IQR 5–73 h) for sustained VT, 22 h (IQR 7–64 h) for SP, 31 h (IQR 11–82 h) for AFP, and 40 h (SD 10–118 h) for HGHB.”
Quote in the paperNot a substantial claim - Findingown data
The overall mean wear time was 9.6 ± 4.0 days.
“The overall mean wear time was 9.6 ± 4.0 days, and more than 25 % of the recorders were worn for at least 13.8 days.”
Quote in the paperNot a substantial claim - Findingown data
For sinus pauses >3 s, 83.1% within 3 days.
“For the most common bradyarrhythmia, sinus pauses >3 s (N = 1766), 31.7 % of the total detected arrhythmias were found within 1 day, 46.6 % within 2 days, and 83.1 % within 3 days of monitoring.”
Quote in the paperSubstantial claim - Findingown data
There were 238 (0.2%) recordings with sustained VT.
“Of the 122,815 eligible records, there were 22,443 (18 %) with nonsustained VT, 238 (0.2 %) with sustained VT, 1766 (1.4 %) with sinus pauses >3 s, 521 (0.4 %) with AF pauses >5 s, 249 (0.2 %) with symptomatic pauses and 1468 (0.4 %) with high-grade heart block (Table 1).”
Quote in the paperNot a substantial claim - Findingown data
For sinus pauses >3 s, 46.6% within 2 days.
“For the most common bradyarrhythmia, sinus pauses >3 s (N = 1766), 31.7 % of the total detected arrhythmias were found within 1 day, 46.6 % within 2 days, and 83.1 % within 3 days of monitoring.”
Quote in the paperSubstantial claim - Findingown data
For non-sustained VT, 23.4% identified by 1 day.
“A similar trend was seen for the more common non-sustained VT, with 23.4, 38.0, and 79.4 % of all non-sustained VT being identified by 1, 2, and 7 days respectively.”
Quote in the paperNot a substantial claim - Findingown data
More than 25% of the recorders were worn for at least 13.8 days.
“The overall mean wear time was 9.6 ± 4.0 days, and more than 25 % of the recorders were worn for at least 13.8 days.”
Quote in the paperNot a substantial claim - Findingown data
Median time to first arrhythmia was 74 h (IQR 26-149 h) for NSVT.
“Median time to first arrhythmia was 74 h (IQR 26–149 h) for NSVT, 22 h (IQR 5–73 h) for sustained VT, 22 h (IQR 7–64 h) for SP, 31 h (IQR 11–82 h) for AFP, and 40 h (SD 10–118 h) for HGHB.”
Quote in the paperNot a substantial claim - Findingown data
Of 122,815 ZIO recordings, median analyzable time was 9.1 (IQR 6.4-13.1) days.
“Of 122,815 ZIO® recordings, median wear time was 9.9 (IQR 6.8–13.8) days and median analyzable time was 9.1 (IQR 6.4–13.1) days.”
Quote in the paperNot a substantial claim - Findingown data
Median time to first arrhythmia was 40 h (SD 10-118 h) for HGHB.
“Median time to first arrhythmia was 74 h (IQR 26–149 h) for NSVT, 22 h (IQR 5–73 h) for sustained VT, 22 h (IQR 7–64 h) for SP, 31 h (IQR 11–82 h) for AFP, and 40 h (SD 10–118 h) for HGHB.”
Quote in the paperNot a substantial claim - Findingown data
For nearly all arrhythmias except symptomatic pauses, there were more detected arrhythmias among men than women.
“More than half (53 %) of the recorders were worn by women; but for nearly all arrhythmias except symptomatic pauses, there were more detected arrhythmias among men than women.”
Quote in the paperNot a substantial claim - Findingown data
Of 122,815 ZIO recordings, median wear time was 9.9 (IQR 6.8-13.8) days.
“Of 122,815 ZIO® recordings, median wear time was 9.9 (IQR 6.8–13.8) days and median analyzable time was 9.1 (IQR 6.4–13.1) days.”
Quote in the paperNot a substantial claim - Findingown data
More than 20% of non-sustained VT was identified in the 7-14 day window.
“This was more pronounced for non-sustained VT than sustained VT, with more than 20 % of non-sustained VT being identified in the 7–14 day window.”
Quote in the paperSubstantial claim - Findingown data
For sustained VT, 52.5% of total identified arrhythmias were identified at 24 h.
“For sustained VT, only 52.5 % of the total identified arrhythmias were identified at 24 h, and approximately two-thirds (65.5 %) were identified by 48 h.”
Quote in the paperSubstantial claim - Findingown data
For all arrhythmias examined, longer monitoring times significantly increased the yield of detected arrhythmias.
“For all arrhythmias examined, longer monitoring times significantly increased the yield of detected arrhythmias.”
Quote in the paperSubstantial claim - Findingown data
There were 249 (0.2%) recordings with symptomatic pauses.
“Of the 122,815 eligible records, there were 22,443 (18 %) with nonsustained VT, 238 (0.2 %) with sustained VT, 1766 (1.4 %) with sinus pauses >3 s, 521 (0.4 %) with AF pauses >5 s, 249 (0.2 %) with symptomatic pauses and 1468 (0.4 %) with high-grade heart block (Table 1).”
Quote in the paperNot a substantial claim - Findingown data
There were 1468 (0.4%) recordings with high-grade heart block.
“Of the 122,815 eligible records, there were 22,443 (18 %) with nonsustained VT, 238 (0.2 %) with sustained VT, 1766 (1.4 %) with sinus pauses >3 s, 521 (0.4 %) with AF pauses >5 s, 249 (0.2 %) with symptomatic pauses and 1468 (0.4 %) with high-grade heart block (Table 1).”
Quote in the paperNot a substantial claim - Findingown data
There were 521 (0.4%) recordings with AF pauses >5 s.
“Of the 122,815 eligible records, there were 22,443 (18 %) with nonsustained VT, 238 (0.2 %) with sustained VT, 1766 (1.4 %) with sinus pauses >3 s, 521 (0.4 %) with AF pauses >5 s, 249 (0.2 %) with symptomatic pauses and 1468 (0.4 %) with high-grade heart block (Table 1).”
Quote in the paperNot a substantial claim - Findingown data
Median time to first arrhythmia was 31 h (IQR 11-82 h) for AFP.
“Median time to first arrhythmia was 74 h (IQR 26–149 h) for NSVT, 22 h (IQR 5–73 h) for sustained VT, 22 h (IQR 7–64 h) for SP, 31 h (IQR 11–82 h) for AFP, and 40 h (SD 10–118 h) for HGHB.”
Quote in the paperNot a substantial claim - Findingnone given
We observed similar findings across gender and age subgroups.
“We observed similar findings across gender and age subgroups.”
Quote in the paperNot a substantial claim - Premisecitation
Traditional 24-h monitoring devices (Holter monitors) often do not detect symptomatic or clinically meaningful arrhythmias.
“Traditional 24-h monitoring devices (i.e., Holter monitors) often do not detect symptomatic or clinically meaningful arrhythmias [6, 7].”
Quote in the paperSubstantial claim0 of 2 sources checked- cites [6] Healey JS, Connolly SJ, Gold MR, Israel CW, Van Gelder IC, Capucci A, et al. Subclinical atrial fibrillation and the risk of stroke. N Engl J Med. 2012;366(2):120–9. source
Not checked in the free run.
- cites [7] Reiffel JA, Schwarzberg R, Murry M. Comparison of autotriggered memory loop recorders versus standard loop recorders versus 24-hour Holter monitors for arrhythmia detection. Am J Cardiol. 2005;95(9):1055–9. source
Not checked in the free run.
- Premisecitation
Emerging evidence suggests that longer wear times yield greater arrhythmia detection in selected at-risk patients that could impact clinical decision-making and outcomes.
“Emerging evidence suggests that longer wear times yield greater arrhythmia detection in selected at-risk patients that could impact clinical decision-making and outcomes [8].”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [8] Gladstone DJ, Spring M, Dorian P, Panzov V, Thorpe KE, Hall J, et al. Atrial fibrillation in patients with cryptogenic stroke. N Engl J Med. 2014;370(26):2467–77. source
Not checked in the free run.
- Premisecitation
Ambulatory ECG monitoring is the standard of care to screen symptomatic outpatient adults for high-risk ventricular and atrial arrhythmias.
“Ambulatory electrocardiographic (ECG) monitoring is the standard of care to screen symptomatic outpatient adults for high-risk ventricular and atrial arrhythmias [1–3].”
Quote in the paperSubstantial claim0 of 3 sources checked- cites [1] Crawford MH, Bernstein SJ, Deedwania PC, DiMarco JP, Ferrick KJ, Garson Jr A, et al. ACC/AHA guidelines for ambulatory electrocardiography: executive summary and recommendations. A report of the American College of Cardiology/American Heart Association task force on practice guidelines (committee to revise the guidelines for ambulatory electrocardiography). Circulation. 1999;100(8):886–93. source
Not checked in the free run.
- cites [2] Katritsis DG, Siontis GC, Camm AJ. Prognostic significance of ambulatory ECG monitoring for ventricular arrhythmias. Prog Cardiovasc Dis. 2013;56(2):133–42. source
Not checked in the free run.
- cites [3] Liao J, Khalid Z, Scallan C, Morillo C, O’Donnell M. Noninvasive cardiac monitoring for detecting paroxysmal atrial fibrillation or flutter after acute ischemic stroke: a systematic review. Stroke. 2007;38(11):2935–40. source
Not checked in the free run.
- Premisecitation
In a registry of 239 patients who wore 30-day loop recorders after discharge for cryptogenic ischemic stroke, 24% of all detected cases of occult atrial fibrillation were found in the final 10 days of 30-day monitoring.
“in a registry of 239 patients who wore 30-day loop recorders after discharge for cryptogenic ischemic stroke, researchers found that 24 % of all detected cases of occult atrial fibrillation were found in the final 10 days of 30-day monitoring (i.e., between days 20 and 30) [14].”
Quote in the paperSubstantial claim0 of 1 sources checked- cites [14] Flint AC, Banki NM, Ren X, Rao VA, Go AS. Detection of Paroxysmal Atrial Fibrillation by 30-Day Event Monitoring in Cryptogenic Ischemic Stroke: The Stroke and Monitoring for PAF in Real Time (SMART) Registry. Stroke. 2012. source
Not checked in the free run.
- Interpretationown data
A significant percentage of potentially high-risk arrhythmias are not identified within 48-h of ambulatory ECG monitoring.
“A significant percentage of potentially high-risk arrhythmias are not identified within 48-h of ambulatory ECG monitoring.”
Quote in the paperSubstantial claim - Interpretationown data
Longer-term continuous ambulatory ECG monitoring provides incremental detection of these potentially clinically relevant arrhythmic events.
“Longer-term continuous ambulatory ECG monitoring provides incremental detection of these potentially clinically relevant arrhythmic events.”
Quote in the paperSubstantial claim - Premisecitation
Recent evidence has demonstrated that longer monitoring may be useful to detect arrhythmias in high-risk patient populations, such as those with a recent history of cryptogenic ischemic stroke.
“recent evidence has demonstrated that longer monitoring may be useful to detect arrhythmias in high-risk patient populations, such as those with a recent history of cryptogenic ischemic stroke [8, 14]”
Quote in the paperSubstantial claim0 of 2 sources checked- cites [8] Gladstone DJ, Spring M, Dorian P, Panzov V, Thorpe KE, Hall J, et al. Atrial fibrillation in patients with cryptogenic stroke. N Engl J Med. 2014;370(26):2467–77. source
Not checked in the free run.
- cites [14] Flint AC, Banki NM, Ren X, Rao VA, Go AS. Detection of Paroxysmal Atrial Fibrillation by 30-Day Event Monitoring in Cryptogenic Ischemic Stroke: The Stroke and Monitoring for PAF in Real Time (SMART) Registry. Stroke. 2012. source
Not checked in the free run.
- Premisecitation
In a larger controlled trial of a similar patient population where 24-h Holter monitoring was compared to 30-day monitoring, 17% of all cases of atrial fibrillation were detected in the final week of monitoring.
“Similarly, in a larger controlled trial of a similar patient population where 24-h Holter monitoring was compared to 30-day monitoring, 17 % of all cases of atrial fibrillation were detected in the final week of monitoring [8].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [8] Gladstone DJ, Spring M, Dorian P, Panzov V, Thorpe KE, Hall J, et al. Atrial fibrillation in patients with cryptogenic stroke. N Engl J Med. 2014;370(26):2467–77. source
Not checked in the free run.
- Interpretationown data
The actual yield of detected arrhythmias from days 7 to 14 of monitoring may be even higher than observed.
“Thus, the actual yield of detected arrhythmias from days 7 to 14 or monitoring may be even higher than we observed.”
Quote in the paperNot a substantial claim
- Classified as probably not a substantial claim. 9
- May combine two claims. 17
Probably not a substantial claim: may not pass S3 (load-bearing).
Findingown dataMore than half (53%) of the recorders were worn by women.
“More than half (53 %) of the recorders were worn by women; but for nearly all arrhythmias except symptomatic pauses, there were more detected arrhythmias among men than women.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing).
Findingown data25% of recorders contained greater than 13 days of analyzable time.
“Analyzable time was similar, with 25 % of recorders containing greater than 13 days of analyzable time.”
Quote in the paperNot a substantial claimMay combine two claims.
Findingown dataAlthough nearly half (49.8%) of analyzed patients were aged less than 65 years old, the majority of all detected arrhythmias were among patients aged 65 years or older.
“Although nearly half (49.8 %) of the analyzed patients were for patients aged less than 65 years old, the majority of all detected arrhythmias were among patients aged 65 years or older.”
Quote in the paperNot a substantial claimMay combine two claims.
Findingown dataOverall, ventricular arrhythmias were more prevalent than bradyarrhythmias, driven by the large burden of episodes of nonsustained VT.
“Overall, ventricular arrhythmias were more prevalent than bradyarrhythmias, although this was driven by the large burden of episodes of nonsustained VT.”
Quote in the paperNot a substantial claimMay combine two claims.
Findingown dataFor ventricular arrhythmias, although the majority are identified in the first 7 days, a significant proportion were still detected in the 7 to 14-day monitoring window.
“For ventricular arrhythmias, we found that although the majority of arrhythmias are identified in the first 7 days, a significant proportion of arrhythmias were still detected in the 7 to 14-day monitoring window.”
Quote in the paperSubstantial claimMay combine two claims.
Findingown dataSimilar trends were observed for the other bradyarrhythmias, and these trends were consistent for both genders and across age categories.
“Similar trends were observed for the other bradyarrhythmias, and these trends were consistent for both genders and across age categories.”
Quote in the paperNot a substantial claimMay combine two claims.
Findingown dataDuring the study period, 122,815 eligible ZIO Patch records were identified, contributed by 122,454 unique patients.
“During the study period, we identified 122,815 eligible ZIO® Patch records contributed by 122,454 unique patients (Fig. 1).”
Quote in the paperNot a substantial claimMay combine two claims.
Findingown dataThese trends were similar for both men and women and across all age ranges, although sustained VT was detected earlier among very elderly patients (aged 80 years and older).
“These trends were similar for both men and women and across all age ranges, although sustained VT was detected earlier among very elderly patients (aged 80 years and older).”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Study design and conductnone givenWe did not validate the data on the clinical indication for the ordered monitors, and differences among providers' thresholds for ordering the monitors could have an impact on patient selection and arrhythmia yield.
“Further, we did not validate the data on the clinical indication for the ordered monitors, and differences among providers’ thresholds for ordering the monitors could have an impact on patient selection and arrhythmia yield.”
Quote in the paperNot a substantial claimMay combine two claims.
Interpretationown dataPatient compliance with extended monitoring was high, with at least 25% of patients achieving greater than 13 days of continuous monitoring.
“Patient compliance with extended monitoring was high, with at least 25 % of patients achieving greater than 13 days of continuous monitoring.”
Quote in the paperSubstantial claimClassified as probably not a substantial claim.
PremisecitationPrior research suggests that traditional 24-h Holter monitoring is not sufficiently long enough to detect many types of arrhythmias.
“Prior research suggests that traditional 24-h Holter monitoring is not sufficiently long enough to detect many types of arrhythmias [10–13]”
Quote in the paperSubstantial claim0 of 4 sources checked- cites [10] Seet RC, Friedman PA, Rabinstein AA. Prolonged rhythm monitoring for the detection of occult paroxysmal atrial fibrillation in ischemic stroke of unknown cause. Circulation. 2011;124(4):477–86. source
Not checked in the free run.
- cites [11] Higgins P, MacFarlane PW, Dawson J, McInnes GT, Langhorne P, Lees KR. Noninvasive cardiac event monitoring to detect atrial fibrillation after ischemic stroke: a randomized, controlled trial. Stroke. 2013;44(9):2525–31. source
Not checked in the free run.
- cites [12] Schreiber D, Sattar A, Drigalla D, Higgins S. Ambulatory cardiac monitoring for discharged emergency department patients with possible cardiac arrhythmias. West J Emerg Med. 2014;15(2):194–8. source
Not checked in the free run.
- cites [13] Rothman SA, Laughlin JC, Seltzer J, Walia JS, Baman RI, Siouffi SY, et al. The diagnosis of cardiac arrhythmias: a prospective multi-center randomized study comparing mobile cardiac outpatient telemetry versus standard loop event monitoring. J Cardiovasc Electrophysiol. 2007;18(3):241–7. source
Not checked in the free run.
Probably not a substantial claim: may not pass S3 (load-bearing).
PremisecitationThe consideration of therapeutic interventions such as permanent pacemaker implantation is recommended by the joint ACC/AHA/HRS guidelines for the high-risk bradyarrhythmias evaluated in our study.
“The consideration of therapeutic interventions such as permanent pacemaker implantation is recommended by the joint American College of Cardiology, American Heart Association, and Heart Rhythm Society guidelines for the high-risk bradyarrhythmias evaluated in our study [15].”
Quote in the paperNot a substantial claim0 of 1 sources checked- cites [15] Epstein AE, DiMarco JP, Ellenbogen KA, Estes III NA, Freedman RA, Gettes LS, et al. ACC/AHA/HRS 2008 Guidelines for device-based therapy of cardiac rhythm abnormalities: a report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (Writing Committee to Revise the ACC/AHA/NASPE 2002 guideline update for implantation of cardiac pacemakers and antiarrhythmia devices): developed in collaboration with the American Association for Thoracic Surgery and Society of Thoracic Surgeons. Circulation. 2008;117(21):e350–408. source
Not checked in the free run.
May combine two claims.
Interpretationown dataThe relatively high wear and analyzable time for the longer term continuous monitors suggests that outpatient ECG monitoring using this approach is feasible and can have significant yield of clinically important arrhythmias beyond atrial fibrillation.
“The relatively high wear and analyzable time for the longer term continuous monitors suggests that outpatient ECG monitoring using this approach is feasible and can have significant yield of clinically important arrhythmias beyond atrial fibrillation.”
Quote in the paperSubstantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Study design and conductown dataWe did not have data on any changes in clinical management or patient outcomes following monitoring, so we were unable to delineate the direct clinical impact from the detection of arrhythmias found from the monitors in our study.
“We did not have data on any changes in clinical management or patient outcomes following monitoring, so we were unable to delineate the direct clinical impact from the detection of arrhythmias found from the monitors in our study.”
Quote in the paperNot a substantial claimMay combine two claims.
Study design and conductnone givenPatient information was limited to demographic characteristics, and data were unavailable on patients' comorbidities.
“Patient information was limited to demographic characteristics, and data were unavailable on patients’ comorbidities, which could potentially help further risk stratify patients and allow for predictive modeling to help identify those most at-risk for high-risk arrhythmias.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
InterpretationreasoningOur study examined a more recent time period than prior investigations of long-term continuous monitors and suggests that changes in the device technology and accumulated operator experience may have resulted in improved patient compliance.
“Our study examined a more recent time period than prior investigations of long-term continuous monitors [5, 9] and suggests that changes in the device technology and accumulated operator experience may have resulted in improved patient compliance.”
Quote in the paperNot a substantial claim0 of 2 sources checked- cites [5] Barrett PM, Komatireddy R, Haaser S, Topol S, Sheard J, Encinas J, et al. Comparison of 24-hour Holter monitoring with 14-day novel adhesive patch electrocardiographic monitoring. Am J Med. 2014;127(1):95. e11-97. source
Not checked in the free run.
- cites [9] Turakhia MP, Hoang DD, Zimetbaum P, Miller JD, Froelicher VF, Kumar UN, et al. Diagnostic utility of a novel leadless arrhythmia monitoring device. Am J Cardiol. 2013;112(4):520–4. source
Not checked in the free run.
May combine two claims.
Study design and conductown dataAt least a quarter of patients wore the device for less than 7 days (25th percentile of 6.8 days), thus artificially reducing the yield of detected arrhythmias from 7 to 14 days.
“Finally, while average wear time was high, at least a quarter of patients wore the device for less than 7 days (25th percentile of 6.8 days), thus artificially reducing the yield of detected arrhythmias from 7 to 14 days.”
Quote in the paperNot a substantial claimProbably not a substantial claim: may not pass S3 (load-bearing). May combine two claims.
Study design and conductown dataWe did not have information on all symptomatic triggers, and thus did not analyze the proportion of all symptomatic triggers that correlate to true arrhythmias.
“We did not have information on all symptomatic triggers, and thus did not analyze the proportion of all symptomatic triggers that correlate to true arrhythmias.”
Quote in the paperNot a substantial claimMay combine two claims.
Interpretationown dataLonger term monitoring up to 14 days resulted in high patient compliance, and greater detection of high-risk arrhythmias than 24- or 48-h monitoring strategies.
“In sum, our study suggests that longer term monitoring up to 14 days resulted in high patient compliance, and greater detection of high-risk arrhythmias than 24- or 48-h monitoring strategies.”
Quote in the paperSubstantial claimMay combine two claims.
Interpretationown dataAlthough the bulk of arrhythmias were detected within the first 7 days, longer-term monitoring between 7 and 14 days yielded a significant number of likely clinically meaningful, potentially high-risk arrhythmias.
“Although the bulk of arrhythmias were detected within the first 7 days, longer-term monitoring between 7 and 14 days yielded a significant number of likely clinically meaningful, potentially high-risk arrhythmias.”
Quote in the paperSubstantial claimClassified as probably not a substantial claim. May combine two claims.
Study design and conductnone givenMobitz II heart block and third degree heart block were identified by the manufacturer according to their FDA-cleared algorithms and 100% data curation and quality review by CCTs trained in advanced arrhythmia detection.
“Mobitz II heart block and third degree heart block were identified by the manufacturer according to their FDA-cleared algorithms and 100 % data curation and quality review by CCT’s trained in advanced arrhythmia detection.”
Quote in the paperSubstantial claim
The paper relies on this without support.
There is limited empirical evidence on the impact and diagnostic yield of longer continuous monitoring for clinically meaningful arrhythmias outside of atrial fibrillation.
“Although there has been very limited evaluation of this approach outside of detecting the presence of atrial fibrillation, longer monitoring periods are emerging as a new standard of care for selected patients.”
“Although conventional wisdom suggests that longer monitoring may be useful for detecting rarer, potentially high-risk arrhythmias, such as ventricular arrhythmias and bradyarrhythmias, there is little empirical evidence on the impact and diagnostic yield of longer continuous monitoring for other clinically meaningful arrhythmias outside of atrial fibrillation.”
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Some large-scale population-based studies (11/179, 6.1%) relied on secondary data collected with mobile apps or web-based wearable platforms or provided through a wearable company.
“Some large-scale population-based studies (11/179, 6.1%) relied on secondary data collected with mobile apps [87] or web-based wearable platforms [153] or provided through a wearable company [189].”
From The impact of wearable technologies in health research: scoping review (2022) · cites this paper as [189]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
Each contradiction between two passages listed here was confirmed by an independent check.
High-grade heart-block percentage is understated
MajorThe reported 1,468 high-grade heart-block records are approximately 1.2% of 122,815 eligible records, not 0.4%.
Reported 0.4 · recomputed 1.195294
“1468 (0.4 %) with high-grade heart block (Table 1)”
Sinus-pause median conflicts with the one-day detection rate
MajorA median time to first sinus pause of 22 hours means at least half of first sinus pauses occurred by 22 hours. That conflicts with the statement that only 31.7% were found within one day.
“22 h (IQR 7–64 h) for SP”
“31.7 % of the total detected arrhythmias were found within 1 day”
Atrial-fibrillation pause counts differ
MinorThe abstract reports 520 recordings with an atrial-fibrillation pause, while the Results section reports 521.
“520 (0.4 %) with a pause during atrial fibrillation >5 s (AFP)”
“521 (0.4 %) with AF pauses >5 s”
High-grade heart-block counts differ between abstract and table
MinorThe abstract reports 1,486 recordings with high-grade heart block, whereas Table 1 reports 1,468.
“1486 (1.2 %) with high-grade heart block (HGHB)”
“High-grade heart blocka N = 1,468”
Cites
ACC/AHA/HRS 2008 guidelines for device-based therapy of cardiac rhythm abnormalities
Epstein, A. E., DiMarco, J. P., Ellenbogen, K. A., Estes, N. A. M., Freedman, R. A., Gettes, L. S., Gillinov, A. M., Gregoratos, G., Hammill, S. C., Hayes, D. L., et al. (2008) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2008.02.032 · Cited by 2,546 · see citing works
Subclinical atrial fibrillation and the risk of stroke
Healey, J. S., Connolly, S. J., Gold, M. R., Israel, C. W., Van Gelder, I. C., Capucci, A., Lau, C. P., Fain, E., Yang, S., Bailleul, C., et al. (2012) · New England Journal of Medicine · doi:10.1056/nejmoa1105575 · Cited by 2,100 · see citing works
ACC/AHA/HRS 2008 Guidelines for Device-Based Therapy of Cardiac Rhythm Abnormalities
Epstein, A. E., Dimarco, J., Ellenbogen, K. A., Estes, N. A. M., Freedman, R. A., Gettes, L. S., Gillinov, A. M., Gregoratos, G., Hammill, S. C., Hayes, D. L., et al. (2008) · Circulation · doi:10.1161/circualtionaha.108.189742 · Cited by 1,768 · see citing works
Atrial fibrillation in patients with cryptogenic stroke
Gladstone, D. J., Spring, M., Dorian, P., Panzov, V., Thorpe, K. E., Hall, J., Vaid, H., O'Donnell, M., Laupacis, A., Côté, R., et al. (2014) · New England Journal of Medicine · doi:10.1056/nejmoa1311376 · Cited by 1,276 · see citing works
Barrett, P. M., Komatireddy, R., Haaser, S., Topol, S., Sheard, J., Encinas, J., Fought, A. J., & Topol, E. J. (2014) · The American Journal of Medicine · doi:10.1016/j.amjmed.2013.10.003 · Cited by 498 · see citing works
Liao, J., Khalid, Z., Scallan, C., Morillo, C., & O’Donnell, M. (2007) · Stroke · doi:10.1161/strokeaha.106.478685 · Cited by 331 · see citing works
Seet, R. C. S., Friedman, P. A., & Rabinstein, A. A. (2011) · Circulation · doi:10.1161/circulationaha.111.029801 · Cited by 274 · see citing works
Diagnostic utility of a novel leadless arrhythmia monitoring device.
Turakhia, M. P., Hoang, D. D., Zimetbaum, P., Miller, J. D., Froelicher, V. F., Kumar, U. N., Xu, X., Yang, F., & Heidenreich, P. A. (2013) · The American Journal of Cardiology · doi:10.1016/j.amjcard.2013.04.017 · Cited by 258 · cited 2× · see citing works
“Further details on the ZIO® Service and its analytic algorithms have been described previously [5, 9].”
MethodResultCrawford, M. H., Bernstein, S. J., Deedwania, P. C., DiMarco, J. P., Ferrick, K. J., Garson, A., Green, L. A., Greene, H. L., Silka, M. J., Stone, P. H., et al. (1999) · Circulation · doi:10.1161/01.cir.100.8.886 · Cited by 215 · see citing works
ROTHMAN, S. A., LAUGHLIN, J. C., SELTZER, J., WALIA, J. S., BAMAN, R. I., SIOUFFI, S. Y., SANGRIGOLI, R. M., & KOWEY, P. R. (2007) · Journal of Cardiovascular Electrophysiology · doi:10.1111/j.1540-8167.2006.00729.x · Cited by 194 · see citing works
Flint, A. C., Banki, N. M., Ren, X., Rao, V. A., & Go, A. S. (2012) · Stroke · doi:10.1161/strokeaha.112.665844 · Cited by 187 · see citing works
Self-citationReiffel, J. A., Schwarzberg, R., & Murry, M. (2005) · The American Journal of Cardiology · doi:10.1016/j.amjcard.2005.01.025 · Cited by 131 · see citing works
Noninvasive Cardiac Event Monitoring to Detect Atrial Fibrillation After Ischemic Stroke
Higgins, P., Macfarlane, P. W., Dawson, J., McInnes, G. T., Langhorne, P., & Lees, K. R. (2013) · Stroke · doi:10.1161/strokeaha.113.001927 · Cited by 127 · see citing works
Schreiber, D., Sattar, A., Drigalla, D., & Higgins, S. (2014) · Western Journal of Emergency Medicine · doi:10.5811/westjem.2013.11.18973 · Cited by 65 · see citing works
Long-term continuous external electrocardiographic recording: a review
Enseleit, F., & Duru, F. (2006) · EP Europace · doi:10.1093/europace/euj054 · Cited by 56 · see citing works
Prognostic significance of ambulatory ECG monitoring for ventricular arrhythmias
Katritsis, D. G., Siontis, G. C. M., & Camm, A. J. (2013) · Progress in Cardiovascular Diseases · doi:10.1016/j.pcad.2013.07.005 · Cited by 44 · see citing works
Flagged first, then by citations
Cited by
Sorted by citations
Heart Disease and Stroke Statistics—2020 Update: A Report From the American Heart Association
Virani, S. S., Alonso, A., Benjamin, E. J., Bittencourt, M. S., Callaway, C. W., Carson, A. P., Chamberlain, A. M., Chang, A. R., Cheng, S., Delling, F. N., et al. (2020) · Circulation · doi:10.1161/cir.0000000000000757 · Cited by 9,175 · see citing works
Heart Disease and Stroke Statistics—2019 Update: A Report From the American Heart Association
Benjamin, E. J., Muntner, P., Alonso, A., Bittencourt, M. S., Callaway, C. W., Carson, A. P., Chamberlain, A. M., Chang, A. R., Cheng, S., Das, S. R., et al. (2019) · Circulation · doi:10.1161/cir.0000000000000659 · Cited by 9,036 · see citing works
Heart disease and stroke statistics-2018 update: a report from the american heart association
Benjamin, E. J., Virani, S. S., Callaway, C. W., Chamberlain, A. M., Chang, A. R., Cheng, S., Chiuve, S. E., Cushman, M., Delling, F. N., Deo, R., et al. (2018) · Circulation · doi:10.1161/cir.0000000000000558 · Cited by 7,146 · see citing works
Heart Disease and Stroke Statistics—2023 Update: A Report From the American Heart Association
Tsao, C. W., Aday, A. W., Almarzooq, Z. I., Anderson, C. A. M., Arora, P., Avery, C. L., Baker-Smith, C. M., Beaton, A. Z., Boehme, A. K., Buxton, A. E., et al. (2023) · Circulation · doi:10.1161/cir.0000000000001123 · Cited by 8,300 · see citing works
Heart Disease and Stroke Statistics—2022 Update: A Report From the American Heart Association
Tsao, C. W., Aday, A. W., Almarzooq, Z. I., Alonso, A., Beaton, A. Z., Bittencourt, M. S., Boehme, A. K., Buxton, A. E., Carson, A. P., Commodore-Mensah, Y., et al. (2022) · Circulation · doi:10.1161/cir.0000000000001052 · Cited by 5,987 · see citing works
Heart Disease and Stroke Statistics—2021 Update: A Report From the American Heart Association
Virani, S. S., Alonso, A., Aparicio, H. J., Benjamin, E. J., Bittencourt, M. S., Callaway, C. W., Carson, A. P., Chamberlain, A. M., Cheng, S., Delling, F. N., et al. (2021) · Circulation · doi:10.1161/cir.0000000000000950 · Cited by 5,784 · see citing works
Martin, S. S., Aday, A. W., Almarzooq, Z. I., Anderson, C. A. M., Arora, P., Avery, C. L., Baker-Smith, C. M., Barone Gibbs, B., Beaton, A. Z., Boehme, A. K., et al. (2024) · Circulation · doi:10.1161/cir.0000000000001209 · Cited by 2,800 · see citing works
Wearables and the Medical Revolution
Dunn, J., Runge, R., & Snyder, M. (2018) · Personalized Medicine · doi:10.2217/pme-2018-0044 · Cited by 654 · see citing works
Mobile Photoplethysmographic Technology to Detect Atrial Fibrillation
Guo, Y., Wang, H., Zhang, H., Liu, T., Liang, Z., Xia, Y., Yan, L., Xing, Y., Shi, H., Li, S., et al. (2019) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2019.08.019 · Cited by 489 · see citing works
The impact of wearable technologies in health research: scoping review
Huhn, S., Axt, M., Gunga, H.-C., Maggioni, M. A., Munga, S., Obor, D., Sié, A., Boudo, V., Bunker, A., Sauerborn, R., et al. (2022) · JMIR mHealth and uHealth · doi:10.2196/34384 · Cited by 379 · see citing works
Self-powered cardiovascular electronic devices and systems
Zheng, Q., Tang, Q., Wang, Z. L., & Li, Z. (2021) · Nature Reviews Cardiology · doi:10.1038/s41569-020-0426-4 · Cited by 348 · see citing works
Review—Energy Autonomous Wearable Sensors for Smart Healthcare: A Review
Dahiya, A. S., Thireau, J., Boudaden, J., Lal, S., Gulzar, U., Zhang, Y., Gil, T., Azémard, N., Ramm, P., Kiessling, T., et al. (2019) · Journal of The Electrochemical Society · doi:10.1149/2.0162003jes · Cited by 106 · see citing works
Accuracy of mHealth Devices for Atrial Fibrillation Screening: Systematic Review
Giebel, G. D., & Gissel, C. (2019) · JMIR mHealth and uHealth · doi:10.2196/13641 · Cited by 99 · see citing works
Diagnostic Accuracy of [11C]PIB Positron Emission Tomography for Detection of Cardiac Amyloidosis
Rosengren, S., Clemmensen, T. S., Tolbod, L. P., Granstam, S.-O., Eiskjær, H., Wikström, G., Vedin, O., Kero, T., Lubberink, M., Harms, H. J., et al. (2020) · JACC. Cardiovascular imaging · doi:10.1016/j.jcmg.2020.02.023 · Cited by 95 · see citing works
Kazanskiy, N. L., Butt, M. A., & Khonina, S. N. (2022) · Nanomaterials · doi:10.3390/nano12030334 · Cited by 87 · see citing works
Kołtowski, Ł., Balsam, P., Główczyńska, R., Rokicki, J. K., Peller, M., Maksym, J., Blicharz, L., Maciejewski, K., Niedziela, M., Opolski, G., et al. (2019) · Cardiology Journal · doi:10.5603/cj.a2019.0001 · Cited by 76 · see citing works
Rooney, M. R., Soliman, E. Z., Lutsey, P. L., Norby, F. L., Loehr, L. R., Mosley, T. H., Zhang, M. J., Gottesman, R. F., Coresh, J., Folsom, A. R., et al. (2019) · Circulation Arrhythmia and Electrophysiology · doi:10.1161/circep.119.007390 · Cited by 67 · see citing works
Yenikomshian, M. A., Jarvis, J. L., Patton, C., Yee, C. W., Mortimer, R., Birnbaum, H. G., & Topash, M. (2019) · Current Medical Research and Opinion · doi:10.1080/03007995.2019.1610370 · Cited by 61 · see citing works
Genetic and environmental determinants of diastolic heart function
Thanaj, M., Mielke, J., McGurk, K. A., Bai, W., Savioli, N., de Marvao, A., Meyer, H. V., Zeng, L., Sohler, F., Lumbers, R. T., et al. (2022) · Nature Cardiovascular Research · doi:10.1038/s44161-022-00048-2 · Cited by 52 · see citing works
Park, K. J., Sharma, G., Kennedy, J. D., & Seyal, M. (2017) · Epilepsia · doi:10.1111/epi.13934 · Cited by 56 · see citing works
Hennig, A., Salel, M., Sacher, F., Camaioni, C., Sridi, S., Denis, A., Montaudon, M., Laurent, F., Jaı̈s, P., & Cochet, H. (2017) · EP Europace · doi:10.1093/europace/eux278 · Cited by 53 · see citing works
Michler, F., Shi, K., Schellenberger, S., Steigleder, T., Malessa, A., Hameyer, L., Neumann, N., Lurz, F., Ostgathe, C., Weigel, R., et al. (2019) · Sensors · doi:10.3390/s19112492 · Cited by 54 · see citing works
Enhanced detection of cardiac arrhythmias utilizing 14-day continuous ECG patch monitoring
Liu, C., Chang, S., Yeh, Y., Chung, F., Hu, Y., Chou, C.-C., Hung, K., Chang, P., Liao, J., Chan, Y., et al. (2021) · International Journal of Cardiology · doi:10.1016/j.ijcard.2021.03.015 · Cited by 49 · see citing works
Santala, O. E., Halonen, J., Martikainen, S., Jäntti, H., Rissanen, T. T., Tarvainen, M. P., Laitinen, T. P., Laitinen, T. M., Väliaho, E., Hartikainen, J. E. K., et al. (2021) · JMIR mhealth and uhealth · doi:10.2196/29933 · Cited by 40 · see citing works
Diagnostic and therapeutic value of implantable loop recorder: A tertiary care center experience
Padmanabhan, D., Kancharla, K., El‐Harasis, M. A., Isath, A. M., Makkar, N., Noseworthy, P. A., Friedman, P. A., Cha, Y., & Kapa, S. (2018) · Pacing and Clinical Electrophysiology · doi:10.1111/pace.13533 · Cited by 32 · see citing works
Janus electrode with stable asymmetric wettability for robust biosignal monitoring on sweaty skin
Harimurti, S., Wang, W., Sasaki, K., Okuda, C., Wijaya, T. J., Nayeem, M. O. G., Lee, S., Yokota, T., & Someya, T. (2024) · Materials Today · doi:10.1016/j.mattod.2024.03.009 · Cited by 31 · see citing works
Nonsustained ventricular tachycardia in heart failure with preserved ejection fraction
Gutiérrez, A., Ash, J., Akdemir, B., Alexy, T., Cogswell, R. J., Chen, J., & Adabağ, S. (2020) · Pacing and Clinical Electrophysiology · doi:10.1111/pace.14043 · Cited by 30 · see citing works
2020 Clinical practice guidelines for Supraventricular tachycardia in adults
Bokeria, L. A., Golukhova, E. Z., Popov, S. V., Artyukhina, Е. А., Базаев, В. А., Batalov, R. E., Bokeria, O. L., Vygovsky, A. B., Gilyarov, M. Y., Sp, G., et al. (2021) · Russian Journal of Cardiology · doi:10.15829/1560-4071-2021-4484 · Cited by 23 · see citing works
Santala, O. E., Lipponen, J. A., Jäntti, H., Rissanen, T. T., Tarvainen, M. P., Laitinen, T. P., Laitinen, T. M., Castrén, M. K., Väliaho, E., Rantula, O., et al. (2022) · JMIR Cardio · doi:10.2196/31230 · Cited by 15 · see citing works
Arrhythmia Patterns in Patients on Ibrutinib
Fazal, M., Kapoor, R., Cheng, P. P., Rogers, A. J., Narayan, S. M., Wang, P. J., Witteles, R. M., Perino, A. C., Baykaner, T., & Rhee, J. (2022) · Frontiers in Cardiovascular Medicine · doi:10.3389/fcvm.2021.792310 · Cited by 15 · see citing works
Willcox, M. E., Compton, S. J., & Bardy, G. H. (2021) · Heart Rhythm O2 · doi:10.1016/j.hroo.2021.09.008 · Cited by 14 · see citing works
Gupta, N., Yang, J., Reynolds, K., Lenane, J., Garcia, E., Sung, S. H., Harrison, T. N., Solomon, M. D., & Go, A. S. (2022) · The American Journal of Cardiology · doi:10.1016/j.amjcard.2021.11.027 · Cited by 14 · see citing works
Self-citationHaimovich, J. S., Achille, P. D., Nauffal, V. D., Singh, P., Reeder, C., Wang, X., Sarma, G., Kornej, J., Benjamin, E. J., Philippakis, A., et al. (2024) · JACC Advances · doi:10.1016/j.jacadv.2024.101004 · Cited by 13 · see citing works
Bechtlufft, B. M., Nascimento, B. R., Sable, C., Fraga, C. L., Barbosa, M. M., Reis, S. D. P., Diamantino, A. C., Meira, Z. M. A., Castilho, S. R. T., Gomes, N. F. A., et al. (2020) · BMJ Open · doi:10.1136/bmjopen-2020-036827 · Cited by 13 · see citing works
2025 Clinical practice guidelines for Supraventricular tachycardias
Popov, S. V., Davtyan, K. V., Shubik, Y. V., Artyukhina, Е. А., Базаев, В. А., Batalov, R. E., Bokeria, L. A., Bubnova, M. G., Vygovsky, A. B., Sp, G., et al. (2025) · Russian Journal of Cardiology · doi:10.15829/1560-4071-2025-6448 · Cited by 12 · see citing works
Piccini, J. P., Clark, R. L., Kowey, P. R., Mittal, S., Dunnmon, P. M., Stockbridge, N., Reiffel, J. A., Turakhia, M. P., Ziegler, P., Kleiman, R., et al. (2017) · American Heart Journal · doi:10.1016/j.ahj.2017.01.012 · Cited by 12 · see citing works
Safety and outcome of nurse-led syncope clinics and implantable loop recorder implants
Eftekhari, H., He, H., Lee, J. D., Paul, G., Zhupaj, A., Lachlan, T., Kuehl, M. W., Dhanjal, T., Panikker, S., Yusuf, S., et al. (2021) · Heart Rhythm · doi:10.1016/j.hrthm.2021.11.006 · Cited by 12 · see citing works
Rowin, E. J., Das, G., Madias, C. A., Hsu, M., Crosson, L. A., Turakhia, M. P., Maron, B. J., & Maron, M. S. (2024) · Heart Rhythm · doi:10.1016/j.hrthm.2024.09.040 · Cited by 11 · see citing works
A review of patient-led data acquisition for atrial fibrillation detection to prevent stroke
Kareem, M., Lei, N., Ali, A., Ciaccio, E. J., Acharya, U. R., & Faust, O. (2021) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2021.102818 · Cited by 10 · see citing works
Clinical impact of a novel ambulatory rhythm monitor in children
May, J. W., Carter, E. L., Hitt, J. R., & Burklow, T. R. (2018) · Cardiology in the Young · doi:10.1017/s1047951118001142 · Cited by 9 · see citing works
Kolkenbeck‐Ruh, A., Soepnel, L. M., Crouch, S. H., Naidoo, S., Smith, W. T., Norris, S. A., Davies, J. I., & Ware, L. J. (2022) · BMC Cardiovascular Disorders · doi:10.1186/s12872-022-02837-w · Cited by 8 · see citing works
Use of Patient Health Records to Quantify Drug-Related Pro-arrhythmic Risk
Davies, M. R., Martinec, M., Walls, R., Schwarz, R., Mirams, G. R., Wang, K., Steiner, G., Surinach, A., Flores, C., Lavé, T., et al. (2020) · Cell Reports Medicine · doi:10.1016/j.xcrm.2020.100076 · Cited by 7 · see citing works
Kasenda, S., Crampin, A. C., Davies, J. I., Malava, J., Manganizithe, S., Kumambala, A., & Sandford, B. (2022) · BMJ Open · doi:10.1136/bmjopen-2021-055501 · Cited by 7 · see citing works
Santala, O. E., Lipponen, J. A., Jäntti, H., Rissanen, T. T., Tarvainen, M. P., Väliaho, E., Rantula, O., Naukkarinen, N. S., Hartikainen, J. E. K., Martikainen, T. J., et al. (2023) · Cardiology in Review · doi:10.1097/crd.0000000000000526 · Cited by 8 · see citing works
Patch versus traditional ambulatory ECG monitoring in children
Hitt, J. R., Carter, E., & May, J. W. (2021) · Progress in Pediatric Cardiology · doi:10.1016/j.ppedcard.2021.101408 · Cited by 5 · see citing works
Bokhari, S. F. H., Waseem, A. B., Raza, H., Iqbal, A., Javaid, S., Idrees, B., Saad, K. A. D., Bakht, D., & Dost, W. (2025) · World Journal of Cardiology · doi:10.4330/wjc.v17.i7.106841 · Cited by 5 · see citing works
Safabakhsh, S., Du, D., Liew, J., Parker, J., McIlroy, C., Khasanova, E., Indraratna, P., Blanke, P., Leipsic, J. A., Andrade, J. G., et al. (2021) · CJC Open · doi:10.1016/j.cjco.2021.10.010 · Cited by 5 · see citing works
Wearable nanosensor-based hardware and software complex for dynamic cardiac monitoring
Ivanov, M. L., Enshin, S. I., Аvdeeva, D. K., Yuzhakov, M. M., Kodermyatov, R. E., & Цой, Е. И. (2019) · Journal of Physics Conference Series · doi:10.1088/1742-6596/1327/1/012029 · Cited by 5 · see citing works
Battisti, A. J., Pinkerton, R., Fokin, V., Wright, B., Alexander, J. H., Berger, R. D., & Turakhia, M. P. (2025) · Heart Rhythm · doi:10.1016/j.hrthm.2025.11.007 · Cited by 4 · see citing works
Bestetti, R. B., Bocchi, E. A., Bestetti, R., Issa, V. S., Furlan-Daniel, R., & Nakazone, M. A. (2022) · Frontiers in Medicine · doi:10.3389/fmed.2022.910388 · Cited by 4 · see citing works
Premature Beats Rejection Strategy on Paroxysmal Atrial Fibrillation Detection
Zhang, X., Li, J., Cai, Z., Zhao, L., & Liu, C. (2022) · Frontiers in Physiology · doi:10.3389/fphys.2022.890139 · Cited by 3 · see citing works
Rwebembera, J., Beaton, A., Okello, E., Engelman, D., Fall, N., Mirabel, M., Nakitto, M., Nunes, M. C. P., Pulle, J., Sarnacki, R., et al. (2022) · Open Heart · doi:10.1136/openhrt-2022-002160 · Cited by 3 · see citing works
Evaluating the benefits of home-based management of atrial fibrillation: current perspectives
Sheikh, A. B., Felzer, J., Munir, A., Morin, D. P., & Lavie, C. J. (2016) · Pragmatic and Observational Research · doi:10.2147/por.s96670 · Cited by 2 · see citing works
Long-Term Follow-Up of Second-Degree Heart Block in Children
Epstein, R., Liberman, L., & Silver, E. S. (2023) · Pediatric Cardiology · doi:10.1007/s00246-023-03195-w · Cited by 2 · see citing works
Wijesurendra, R. S., Pessoa-Amorim, G., Buck, G. A. N., Harper, C., Bulbulia, R. A., Jones, N. R., A’Court, C., Kurien, R., Taylor, K., Casadei, B., et al. (2025) · American Heart Journal · doi:10.1016/j.ahj.2025.07.004 · Cited by 2 · see citing works
Turushev, N. V., Аvdeeva, D. K., Enshin, S. I., & Yevtushenko, G. S. (2023) · Biomedical Engineering · doi:10.1007/s10527-023-10262-2 · Cited by 1 · see citing works
Mao, L., Wang, L., & Lv, D.-M. (2023) · Frontiers in Pediatrics · doi:10.3389/fped.2023.1167278 · Cited by 1 · see citing works
Evaluating the Process of Ambulatory Electrocardiographic Monitor Application and Completion Rates
Bhatnagar, A. S. (2019) · KU ScholarWorks (University of Kansas)
The Substrate of Sudden Death in Long-QT Syndrome is localized in the EpicardiumPreprint
Pappone, C., Ciconte, G., Anastasia, L., Borrelli, V., Grant, E., Vicedomini, G., & Santinelli, V. (2021) · medRxiv · doi:10.1101/2021.11.22.21266568
Health and biosensor technology—A revolution underway for the well-being of the population
Bastidas, G. A., & Iglesias, R. (2024) · Wearable Technology · doi:10.54517/wt.v4i1.2618 · Cited by 1 · see citing works
An ambulatory electrocardiograph with a high-resolution signal recording function
Turushev, N. V., Аvdeeva, D. K., Enshin, S. I., Ivanov, M. L., Максимов, И. В., & Evtushenko, G. S. (2024) · Biomedical Engineering · doi:10.1007/s10527-024-10370-7
Homme, J. L., Kranz, L. A., House, M., & Greenmyer, J. R. (2025) · The American Journal of Emergency Medicine · doi:10.1016/j.ajem.2025.04.012
Fischer, Q., Nault, I., Guillén, P. L. C., Avvedimento, M., Benavent-Garcia, C., Pelletier-Beaumont, É., Philippon, F., & Rodés‐Cabau, J. (2026) · Circulation Cardiovascular Interventions · doi:10.1161/circinterventions.125.016288
Russo, P., Nathan, R., Shokoohi, M., Hendrickson, E., Boyle, K., Turakhia, M. P., & Wright, B. (2026) · Journal of Medical Economics · doi:10.1080/13696998.2026.2707021
Ambulatory Electrocardiography Modalities in Pediatric Arrhythmia Detection and Management
Hsu, Y. T., Tseng, W. C., Hsu, Y. L., Wu, M. H., Lu, C. W., Wu, K. L., Peng, W., & Chiu, S. N. (2026) · Archives of Pediatrics and Adolescent Medicine · doi:10.1001/jamapediatrics.2026.4910
- Authors
- Solomon, M. D., Yang, J., Sung, S. H., Livingston, M. L., Sarlas, G., Lenane, J. C., & Go, A. S.
- Affiliations
- Kaiser Permanente (US)
- iRhythm (United States) (US)
- Corresponding author
- Solomon, Matthew D.
- Venue
- BMC Cardiovascular Disorders
- Published
- 2016-02-17
- Volume
- 16
- Issue
- 1
- Pages
- 35-35
- ISSN
- 1471-2261 (ISSN-L)
- PMC
- PMC4756401
- OpenAlex
- W2276194925
- Semantic Scholar
- 2645e0816df31a03d380c9819341fce7ade0b286
- Cited by
- 65
- References
- 16
- Citations by year
- 2026: 3 · 2025: 5 · 2024: 6 · 2023: 5 · 2022: 11 · 2021: 11 · 2020: 6 · 2019: 10 · 2018: 4 · 2017: 3 · 2016: 1
- FWCI
- 4.95
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Licence
- cc-by
- Locations
- BMC Cardiovascular Disorders (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- Funders
- iRhythm, Inc
- Topics
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › ECG Monitoring and Analysis
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Cardiac electrophysiology and arrhythmias
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Cardiac pacing and defibrillation studies
- Keywords
- non-sustained ventricular tachycardia, sustained ventricular tachycardia, sinus pause, arrhythmia incidence
- MeSH terms
- Aged, Aged, 80 and over, Arrhythmias, Cardiac, Bradycardia, Female, Heart Block, Humans, Male, Middle Aged, Retrospective Studies, Time Factors, Cohort Studies, Electrocardiography, Ambulatory, Incidence, Tachycardia, Ventricular, Sinus Arrest, Cardiac, Atrioventricular Block
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Claims read by deepseek-v4.1-flash under definition version 3 (extract-v3) on 2026-10-09.
Classified by jev-1.13-20260917 (gold-guideline-v0.2) on 2026-10-09.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review