Transpulmonary thermodilution: advantages and limits
Open accessThe DOI matches its Crossref record; 63 of the 63 references it cites were found in a registry or an index.
Abstract
BACKGROUND: For complex patients in the intensive care unit or in the operating room, many questions regarding their haemodynamic management cannot be answered with simple clinical examination. In particular, arterial pressure allows only a rough estimation of cardiac output. Transpulmonary thermodilution is a technique that provides a full haemodynamic assessment through cardiac output and other indices. MAIN BODY: Through the analysis of the thermodilution curve recorded at the tip of an arterial catheter after the injection of a cold bolus in the venous circulation, transpulmonary thermodilution intermittently measures cardiac output. This measure allows the calibration of pulse contour analysis. This provides continuous and real time monitoring of cardiac output, which is not possible with the pulmonary artery catheter. Transpulmonary thermodilution provides several variables beyond cardiac output. It estimates the end-diastolic volume of the four cardiac cavities, which is a marker of cardiac preload. It provides an estimation of the systolic function of the combined ventricles. It is more direct than the pulmonary artery catheter, but does not allow the distinct estimation of right and left cardiac function. It is easier and faster to perform than echocardiography, but does not provide a full evaluation of the cardiac structure and function. Transpulmonary thermodilution has the unique advantage of being able to estimate at the bedside extravascular lung water, which quantifies the volume of pulmonary oedema, and pulmonary vascular permeability, which quantifies the degree of a pulmonary capillary leak. Both indices are helpful for guiding fluid strategy, especially in case of acute respiratory distress syndrome. CONCLUSIONS: Transpulmonary thermodilution provides a full cardiovascular evaluation that allows one to answer many questions regarding haemodynamic management. It belongs to the category of "advanced" devices that are indicated for the most critically ill and/or complex patients.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Chen, Y., Wang, L., Kang, Q., Zhang, X., Yu, G., Wan, X., Wang, J., & Zhu, K. (2017) · Cellular Physiology and Biochemistry · doi:10.1159/000477308 · Cited by 10,679
Acute Respiratory Distress Syndrome: The Berlin Definition
(2012) · JAMA · doi:10.1001/jama.2012.5669 · Cited by 8,731
Cecconi, M., De Backer, D., Antonelli, M., Beale, R., Bakker, J., Hofer, C., Jaeschke, R., Mebazaa, A., Pinsky, M. R., Teboul, J. L., et al. (2014) · Intensive Care Medicine · doi:10.1007/s00134-014-3525-z · Cited by 1,767
Self-citationAssessment of cardiac preload and extravascular lung water by single transpulmonary thermodilution
Sakka, S. G., Rühl, C. C., Pfeiffer, U. J., Beale, R., McLuckie, A., Reinhart, K., & Meier-Hellmann, A. (2000) · Intensive Care Medicine · doi:10.1007/s001340050043 · Cited by 920
Kumar, A., Anel, R., Bunnell, E., Habet, K., Zanotti, S., Marshall, S., Neumann, A., Ali, A., Cheang, M., Kavinsky, C., et al. (2004) · Critical Care Medicine · doi:10.1097/01.ccm.0000114996.68110.c9 · Cited by 800
Prognostic Value of Extravascular Lung Water in Critically Ill Patients
Sakka, S. G., Klein, M., Reinhart, K., & Meier‐Hellmann, A. (2002) · CHEST Journal · doi:10.1378/chest.122.6.2080 · Cited by 672
Prediction of fluid responsiveness: an update
Monnet, X., Marik, P. E., & Teboul, J.-L. (2016) · Annals of Intensive Care · doi:10.1186/s13613-016-0216-7 · Cited by 600
Self-citationMitchell, J. P., Schuller, D., Calandrino, F. S., & Schuster, D. P. (1992) · American Review of Respiratory Disease · doi:10.1164/ajrccm/145.5.990 · Cited by 587
Fluid challenges in intensive care: the FENICE study: A global inception cohort study
on behalf of the FENICE Investigators and the ESICM Trial Group, Cecconi, M., Hofer, C., Teboul, J.-L., Pettila, V., Wilkman, E., Molnar, Z., Della Rocca, G., Aldecoa, C., Artigas, A., et al. (2015) · Intensive Care Medicine · doi:10.1007/s00134-015-3850-x · Cited by 573
Self-citationStetz, C. W., Miller, R. G., Kelly, G. E., & Raffin, T. A. (1982) · PubMed · doi:10.1164/arrd.1982.126.6.1001 · Cited by 447
Passive leg raising: five rules, not a drop of fluid!
Monnet, X., & Teboul, J.-L. (2015) · Critical Care · doi:10.1186/s13054-014-0708-5 · Cited by 425
Self-citationClinical review: Update on hemodynamic monitoring - a consensus of 16
Vincent, J.-L., Rhodes, A., Perel, A., Martin, G. S., Rocca, G., Vallet, B., Pinsky, M. R., Hofer, C. K., Teboul, J.-L., de Boode, W.-P., et al. (2011) · Critical Care · doi:10.1186/cc10291 · Cited by 419
Self-citationReuter, D. A., Felbinger, T. W., Schmidt, C., et al. (2002) · Intensive Care Medicine · doi:10.1007/s00134-002-1211-z · Cited by 375
Global End-Diastolic Volume as an Indicator of Cardiac Preload in Patients With Septic Shock *
Michard, F., Alaya, S., Zarka, V., Bahloul, M., Richard, C., & Teboul, J.-L. (2003) · Chest · doi:10.1378/chest.124.5.1900 · Cited by 352
Self-citationLess invasive hemodynamic monitoring in critically ill patients
Teboul, J.-L., Saugel, B., Cecconi, M., De Backer, D., Hofer, C. K., Monnet, X., Perel, A., Pinsky, M. R., Reuter, D. A., Rhodes, A., et al. (2016) · Intensive Care Medicine · doi:10.1007/s00134-016-4375-7 · Cited by 328
Self-citationMonnet, X., Osman, D., Ridel, C., Lamia, B., Richard, C., & Teboul, J.-L. (2009) · Critical Care Medicine · doi:10.1097/ccm.0b013e3181968fe1 · Cited by 301
Self-citationRobotham, J. L., Takata, M., Berman, M., & Harasawa, Y. (1991) · Anesthesiology · doi:10.1097/00000542-199101000-00026 · Cited by 272
Cordemans, C., De laet, I., Van Regenmortel, N., Schoonheydt, K., Dits, H., Huber, W., & Malbrain, M. L. (2012) · Annals of Intensive Care · doi:10.1186/2110-5820-2-s1-s1 · Cited by 271
Jozwiak, M., Silva, S., Persichini, R., Anguel, N., Osman, D., Richard, C., Teboul, J.-L., & Monnet, X. (2013) · Critical Care Medicine · doi:10.1097/ccm.0b013e31826ab377 · Cited by 259
Self-citationBenes, J., Giglio, M., Brienza, N., & Michard, F. (2014) · Critical Care · doi:10.1186/s13054-014-0584-z · Cited by 257
Monnet, X., Anguel, N., Osman, D., Hamzaoui, O., Richard, C., & Teboul, J.-L. (2007) · Intensive Care Medicine · doi:10.1007/s00134-006-0498-6 · Cited by 227
Self-citationExtravascular lung water in critical care: recent advances and clinical applications
Jozwiak, M., Teboul, J.-L., & Monnet, X. (2015) · Annals of Intensive Care · doi:10.1186/s13613-015-0081-9 · Cited by 203
Self-citationExtravascular lung water in patients with severe sepsis: a prospective cohort study
Martin, G. S., Eaton, S., Mealer, M., & Moss, M. (2005) · Critical Care · doi:10.1186/cc3025 · Cited by 201
A Prospective Study of Lung Water Measurements during Patient Management in an Intensive Care Unit
Eisenberg, P. R., Hansbrough, J. R., Anderson, D., & Schuster, D. P. (1987) · American Review of Respiratory Disease · doi:10.1164/ajrccm/136.3.662 · Cited by 197
Precision of the transpulmonary thermodilution measurements
Monnet, X., Persichini, R., Ktari, M., Jozwiak, M., Richard, C., & Teboul, J.-L. (2011) · Critical Care · doi:10.1186/cc10421 · Cited by 194
Self-citationMatthieu, B., Karine, N.-G., Vincent, C., Alain, V., François, C. J., Philippe, R., & François, S. (2008) · Anesthesia & Analgesia · doi:10.1213/ane.0b013e318168b309 · Cited by 187
Hamzaoui, O., Monnet, X., Richard, C., Osman, D., Chemla, D., & Teboul, J.-L. (2008) · Critical Care Medicine · doi:10.1097/01.ccm.ob013e318161fec4 · Cited by 179
Self-citationPhillips, C. R., Chesnutt, M. S., & Smith, S. M. (2008) · Critical Care Medicine · doi:10.1097/01.ccm.0000295314.01232.be · Cited by 178
Experts’ recommendations for the management of adult patients with cardiogenic shock
Levy, B., Bastien, O., Bendjelid, K., Cariou, A., Chouihed, T., Combes, A., Mebazaa, A., Megarbane, B., Plaisance, P., Ouattara, A., et al. (2015) · Annals of Intensive Care · doi:10.1186/s13613-015-0052-1 · Cited by 172
Self-citationPierrakos, C., Velissaris, D., Scolletta, S., Heenen, S., De Backer, D., & Vincent, J.-L. (2012) · Intensive Care Medicine · doi:10.1007/s00134-011-2457-0 · Cited by 171
Kushimoto, S., Taira, Y., Kitazawa, Y., Okuchi, K., Sakamoto, T., Ishikura, H., Endo, T., Yamanouchi, S., Tagami, T., Yamaguchi, J., et al. (2012) · Critical Care · doi:10.1186/cc11898 · Cited by 159
Tagami, T., Kushimoto, S., Yamamoto, Y., Atsumi, T., Tosa, R., Matsuda, K., Oyama, R., Kawaguchi, T., Masuno, T., Hirama, H., et al. (2010) · Critical Care · doi:10.1186/cc9250 · Cited by 154
Estimation of left ventricular systolic function by single transpulmonary thermodilution
Combes, A., Berneau, J.-B., Luyt, C. E., & Trouillet, J. (2004) · Intensive Care Medicine · doi:10.1007/s00134-004-2289-2 · Cited by 149
Hofer, C. K., Furrer, L., Matter-Ensner, S., Maloigne, M., Klaghofer, R., Genoni, M., & Zollinger, A. (2005) · British Journal of Anaesthesia · doi:10.1093/bja/aei123 · Cited by 146
Monnet, X., Anguel, N., Naudin, B., Jabot, J., Richard, C., & Teboul, J. (2010) · Critical Care · doi:10.1186/cc9058 · Cited by 140
Self-citationMonnet, X., Letierce, A., Hamzaoui, O., Chemla, D., Anguel, N., Osman, D., Richard, C., & Teboul, J. (2011) · Critical Care Medicine · doi:10.1097/ccm.0b013e31820edcf0 · Cited by 133
Self-citationAccurate characterization of extravascular lung water in acute respiratory distress syndrome*
Berkowitz, D. M., Danai, P. A., Eaton, S., Moss, M. M., & Martin, G. S. (2008) · Critical Care Medicine · doi:10.1097/ccm.0b013e3181743eeb · Cited by 129
Staub, N. C. (1978) · CHEST Journal · doi:10.1378/chest.74.5.559 · Cited by 118
Huber, W., Umgelter, A., Reindl, W., Franzen, M., Schmidt, C., von Delius, S., Geisler, F., Eckel, F., Fritsch, R., Siveke, J. T., et al. (2008) · Critical Care Medicine · doi:10.1097/ccm.0b013e3181809928 · Cited by 114
Pulmonary vascular dysfunction in ARDS
Ryan, D., Fröhlich, S., & McLoughlin, P. (2014) · Annals of Intensive Care · doi:10.1186/s13613-014-0028-6 · Cited by 107
Zhang, Z., Ni, H., & Qian, Z. (2015) · Intensive Care Medicine · doi:10.1007/s00134-014-3638-4 · Cited by 104
Effect of the degree of tricuspid regurgitation on cardiac output measurements by thermodilution
Martin, B., Pachl, J., & Jan, H. (2002) · Intensive Care Medicine · doi:10.1007/s00134-002-1352-0 · Cited by 97
Predictive value of EEG-awakening for behavioral awakening from coma
Kang, X., Yang, F., Li, W., Ma, C., Li, L., & Jiang, W. (2015) · Annals of Intensive Care · doi:10.1186/s13613-015-0094-4 · Cited by 87
Kiefer, N., Hofer, C. K., Marx, G., Geisen, M., Giraud, R., Siegenthaler, N., Hoeft, A., Bendjelid, K., & Rex, S. (2012) · Critical Care · doi:10.1186/cc11366 · Cited by 86
Complications related to less-invasive haemodynamic monitoring
Belda, F. J., Aguilar, G., Teboul, J., Pestaña, D., Calvo, F. J. R., Malbrain, M. L. N. G., Luís, J., Ramasco, F., Umgelter, A., Wendon, J. A., et al. (2010) · British Journal of Anaesthesia · doi:10.1093/bja/aeq377 · Cited by 84
Self-citationChew, M. S., Ihrman, L., Düring, J., Bergenzaun, L., Ersson, A., Undén, J., Rydén, J., Åkerman, E., & Larsson, M. (2012) · Critical Care · doi:10.1186/cc10599 · Cited by 80
Roch, A., Michelet, P., Lambert, D., Delliaux, S., Saby, C., Perrin, G., Ghez, O. Y., Brégeon, F., Thomas, P. A., de Carpentier, J. P., et al. (2004) · Critical Care Medicine · doi:10.1097/01.ccm.0000114831.59185.02 · Cited by 73
Bendjelid, K., Giraud, R., Siegenthaler, N., et al. (2010) · Critical Care · doi:10.1186/cc9332 · Cited by 70
Volume-limited versus pressure-limited hemodynamic management in septic and nonseptic shock*
Trof, R. J., Beishuizen, A., Cornet, A. D., de Wit, R. J., Girbes, A. R. J., & Groeneveld, A. B. J. (2012) · Critical Care Medicine · doi:10.1097/ccm.0b013e31823bc5f9 · Cited by 62
Bendjelid, K., Marx, G., Kiefer, N., Simon, T. P., Geisen, M., Hoeft, A., Siegenthaler, N., & Hofer, C. K. (2013) · British Journal of Anaesthesia · doi:10.1093/bja/aet116 · Cited by 60
Hemodynamic monitoring in the era of digital health
Michard, F. (2016) · Annals of Intensive Care · doi:10.1186/s13613-016-0119-7 · Cited by 57
The precision of PiCCO ® measurements in hypothermic post‐cardiac arrest patients
Tagami, T., Kushimoto, S., Tosa, R., Omura, M., Hagiwara, J., Hirama, H., & Yokota, H. (2012) · Anaesthesia · doi:10.1111/j.1365-2044.2011.06981.x · Cited by 53
Perel, A., Saugel, B., Teboul, J., Malbrain, M. L. N. G., Belda, F. J., Fernández‐Mondejar, E., Кirov, М. Y., Wendon, J. A., Lussmann, R., & Maggiorini, M. (2015) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-015-9811-7 · Cited by 53
Self-citationExtravascular lung water after pneumonectomy and one-lung ventilation in sheep
Владимирович, К. В., Suborov, E. V., Кirov, М. Y., Kuklin, V. N., Sobhkhez, M., Johnsen, S., Wærhaug, K., & Bjertnæs, L. J. (2007) · Critical Care Medicine · doi:10.1097/01.ccm.0000265739.51887.2b · Cited by 46
Sakka, S. G., Hanusch, T., Thuemer, O., & Wegscheider, K. (2007) · Anesthesia & Analgesia · doi:10.1213/01.ane.0000280440.08530.fb · Cited by 40
Roch, A., Michelet, P., D’Journo, X. B., Brousse, D., Blayac, D., Lambert, D., & Auffray, J. (2005) · CHEST Journal · doi:10.1378/chest.128.2.927 · Cited by 37
Identifying patients with ARDS: time for a different approach
Schuster, D. (1997) · Intensive Care Medicine · doi:10.1007/s001340050486 · Cited by 31
Room-temperature vs iced saline indicator injection for transpulmonary thermodilution
Huber, W., Kraski, T., Haller, B., Mair, S., Saugel, B., Beitz, A., Schmid, R. M., & Malbrain, M. L. N. G. (2014) · Journal of Critical Care · doi:10.1016/j.jcrc.2014.08.005 · Cited by 31
Faybik, P., Hetz, H., Baker, A., Iankovskaya, E., Krenn, C. G., & Steltzer, H. (2004) · Journal of Critical Care · doi:10.1016/j.jcrc.2004.04.002 · Cited by 31
Dres, M., Teboul, J., Guerin, L., et al. (2014) · Critical Care Medicine · doi:10.1097/ccm.0000000000000341 · Cited by 23
Self-citationResults of questionable management protocols are inherently questionable
Teboul, J., Monnet, X., & Perel, A. (2012) · Critical Care Medicine · doi:10.1097/ccm.0b013e3182545759 · Cited by 10
Self-citationHaas, S., Trepte, C. J. C., Nitzschke, R., Jürgens, T. P., Goepfert, M. S., Goetz, A. E., & Reuter, D. A. (2013) · Anesthesia & Analgesia · doi:10.1213/ane.0b013e31828f2c39 · Cited by 8
Suria, S., Wyniecki, A., Eghiaian, A., et al. (2014) · PLoS ONE · doi:10.1371/journal.pone.0104369 · Cited by 2
Cited by
Malbrain, M. L. N. G., Van Regenmortel, N., Saugel, B., De Tavernier, B., Van Gaal, P.-J., Joannes-Boyau, O., Teboul, J.-L., Rice, T. W., Mythen, M., & Monnet, X. (2018) · Annals of Intensive Care · doi:10.1186/s13613-018-0402-x · Cited by 632
“Extravascular lung water measured by transpulmonary thermodilution, as well as the pulmonary vascular permeability index which is inferred from it, might reflect the pulmonary risk of fluid infusion more directly [40, 48, 74].”
BackgroundSelf-citationArterial Pulse Pressure Variation with Mechanical Ventilation
Teboul, J.-L., Monnet, X., Chemla, D., & Michard, F. (2019) · American Journal of Respiratory and Critical Care Medicine · doi:10.1164/rccm.201801-0088ci · Cited by 200
“In cases where the risks of lung edema are not negligible, additional variables, such as extravascular lung water (74) or the presence of B-lines on lung ultrasound examination (75), are useful in making the appropriate decision.”
BackgroundSelf-citationHow can assessing hemodynamics help to assess volume status?
De Backer, D., Aissaoui, N., Cecconi, M., Chew, M. S., Denault, A., Hajjar, L., Hernandez, G., Messina, A., Myatra, S. N., Ostermann, M., et al. (2022) · Intensive Care Medicine · doi:10.1007/s00134-022-06808-9 · Cited by 199
“Transpulmonary thermodilution Transpulmonary thermodilution allows measurement of intrathoracic blood volumes, EVLW and CO [26].”
BackgroundSelf-citationA plea for personalization of the hemodynamic management of septic shock
De Backer, D., Cecconi, M., Chew, M. S., Hajjar, L., Monnet, X., Ospina-Tascón, G. A., Ostermann, M., Pinsky, M. R., & Vincent, J.-L. (2022) · Critical Care · doi:10.1186/s13054-022-04255-y · Cited by 177
“In a patient without comorbidities and with minimal organ dysfunction, non-calibrated or internally calibrated CO-monitoring devices [48] may be used but more complex patients (based on comorbidities, associated organ dysfunction or poor evolution) would benefit from the use of transpulmonary thermodilution [49] or ev…”
MethodSelf-citationManagement of sepsis and septic shock in the emergency department
Gavelli, F., Castello, L. M., & Avanzi, G. C. (2021) · Internal and Emergency Medicine · doi:10.1007/s11739-021-02735-7 · Cited by 161
Metadata
- Authors
- Monnet, X., & Teboul, J.-L.
- Venue
- Critical Care
- Published
- 2017-06-19
- Volume
- 21
- Issue
- 1
- Pages
- 147-147
- ISSN
- 1364-8535 (ISSN-L)
- OpenAlex
- W2625069600
- Cited by
- 285
- References
- 63
- FWCI
- 21.73
- Type
- journal article
- Language
- en
- Open access
- gold
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.