Feasibility of equivalent performance of 3D TOF [18F]-FDG PET/CT with reduced acquisition time using clinical and semiquantitative parameters
Open accessThe DOI matches its Crossref record; 31 of the 31 references it cites were found in a registry or an index.
Abstract
Background: High-performance time-of-flight (TOF) positron emission tomography (PET) systems have the capability for rapid data acquisition while preserving diagnostic image quality. However, determining a reliable and clinically applicable cut-off of the acquisition time plays an important role in routine practice. This study aimed to assess the diagnostic equivalence of short acquisition time of 57 with routine 75 seconds per bed position (s/BP) of [18 F]-fluoro-deoxy-glucose (FDG) PET. Phantom studies applying EARL criteria suggested the feasibility of shortened acquisition time in routine clinical imaging by 3D TOF PET/CT scanners. Ninety-six patients with melanoma, lung or head and neck cancer underwent a standard whole-body, skull base-to-thigh or vertex-to-thigh [18 F]-FDG PET/CT examination using the 3D TOF Ingenuity TF PET/CT system (Philips, Cleveland, OH). The [18 F]-FDG activity applied was equal to 4MBq per kg body weight. Retrospectively, PET list-mode data were used to calculate a second PET study per patient with a reduced acquisition time of 57 s instead of routine 75 s/BP. PET/CT data were reconstructed using a 3D OSEM TOF algorithm. Blinded patient data were analysed by two nuclear medicine physicians. The number of [18 F]-FDG-avid lesions per body region (head&neck, thorax, abdomen, bone, extremity) and image quality (grade 1–5) were evaluated. Semiquantitative analyses were performed by standardized uptake value (SUV) measurements using 3D volume of interests (VOI). The visual and semiquantitative diagnostic equivalence of 214 [18 F]-FDG-avid lesions were analysed in the routine standard (75 s/BP) as well as the calculated PET/CT studies with short acquisition time. Statistical analyses were performed by equivalence testing and Bland–Altman plots. Results: Lesion detection rate per patient’s body region agreed in > 98% comparing 57 s/BP and 75 s/BP datasets. Overall image quality was determined as equal or superior to 75 s in 80% and 69%, respectively. In the semiquantitative lesion-based analyses, a significant equivalence was found between the 75 s/BP and 57 s/BP PET/CT images both for SUV max (p = 0.004) and SUV mean (p = 0.003). Conclusion: The results of this study demonstrate significant clinical and semiquantitative equivalence between short acquisition time of 57 s/BP and standard 75 s/BP 3D TOF [18 F]-FDG PET/CT scanning, which may improve the patient’s workflow in routine practice.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Figures



![Fig. 4 Comparison of SUVmax 75 s and 57 s PET/CT datasets and SUVmean 75 s and 57 s PET/CT images of 214 [18F]-FDG-avid lesions by performing the TOST equivalence test [20]](https://figures.cochat.ai/paper-figures/ffea5432624547b79ab4ef47c2da82d2/feasibility-of-equivalent-performance-of-3d-tof-18f-fdg-pet-ct-with-reduced-figure-4.jpg)



Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“This is especially true for SUVmean and, to a lesser degree, for SUVpeak, which are both relatively stable even after reduction of acquisition time to 50% or less [76,77,78,79,80].”
From Influences on PET Quantification and Interpretation (2022) · cites this paper as [78]
Experimental — accuracy not yet measured
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
Martin Bland, J., & Altman, D. (1986) · The Lancet · doi:10.1016/s0140-6736(86)90837-8 · Cited by 47,965
FDG PET/CT: EANM procedure guidelines for tumour imaging: version 2.0
Boellaard, R., Delgado-Bolton, R., Oyen, W. J. G., Giammarile, F., Tatsch, K., Eschner, W., Verzijlbergen, F. J., Barrington, S. F., Pike, L. C., Weber, W. A., et al. (2015) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-014-2961-x · Cited by 3,303 · cited 9×
“Abbreviations ALARA : As low as reasonably achievable; Bw: Body weight; CI: Confidence interval; COV: Coefficient of variation; EANM: European Association of Nuclear Medicine; EARL: EANM Research Ltd; [18F]‑FDG: Fluor‑18 fluoro‑deoxy‑glucose; NECR: Noise equivalent count rate; OSEM: Ordered‑subset expectation maxi‑ mi…”
BackgroundMethodInfluentialEquivalence Tests: A Practical Primer for t Tests, Correlations, and Meta-Analyses
Lakens, D. (2017) · Social Psychological and Personality Science · doi:10.1177/1948550617697177 · Cited by 3,001 · cited 4×
“From a statistical point of view, both tests are equivalent in our setting anyway, because the superiority case is highly unlikely if not impossible at all, and therefore, the p values of the TOST and the noninferiority tests will be the same.”
MethodInfluentialFDG PET and PET/CT: EANM procedure guidelines for tumour PET imaging: version 1.0
Boellaard, R., O’Doherty, M. J., Weber, W. A., Mottaghy, F. M., Lonsdale, M. N., Stroobants, S. G., Oyen, W. J. G., Kotzerke, J., Hoekstra, O. S., Pruim, J., et al. (2010) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-009-1297-4 · Cited by 1,326 · cited 1×
“The most commonly employed PET radiopharmaceutical is [18F]-fluoro-deoxy-glucose (FDG), a glucose analogue [2, 5, 6].”
MethodStandards for PET Image Acquisition and Quantitative Data Analysis
Boellaard, R. (2009) · Journal of Nuclear Medicine · doi:10.2967/jnumed.108.057182 · Cited by 902
Recent developments in time-of-flight PET
Vandenberghe, S., Mikhaylova, E., D’Hoe, E., Mollet, P., & Karp, J. S. (2016) · EJNMMI Physics · doi:10.1186/s40658-016-0138-3 · Cited by 368
Impact of Time-of-Flight on PET Tumor Detection
Kadrmas, D. J., Casey, M. E., Conti, M., Jakoby, B. W., Lois, C., & Townsend, D. W. (2009) · Journal of Nuclear Medicine · doi:10.2967/jnumed.109.063016 · Cited by 277 · cited 1×
“One of the major technical advancements in PET/CT scanners was the implementation of the time-of-flight approach leading to better performance of this modality and more accurate localization of the annihilation process [14].”
BackgroundQuantification, improvement, and harmonization of small lesion detection with state-of-the-art PET
van der Vos, C. S., Koopman, D., Rijnsdorp, S., Arends, A. J., Boellaard, R., van Dalen, J. A., Lubberink, M., Willemsen, A. T. M., & Visser, E. P. (2017) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-017-3727-z · Cited by 206 · cited 17×
“Abbreviations ALARA : As low as reasonably achievable; Bw: Body weight; CI: Confidence interval; COV: Coefficient of variation; EANM: European Association of Nuclear Medicine; EARL: EANM Research Ltd; [18F]‑FDG: Fluor‑18 fluoro‑deoxy‑glucose; NECR: Noise equivalent count rate; OSEM: Ordered‑subset expectation maxi‑ mi…”
BackgroundMethodInfluentialRecent Advances and Future Progress in PET Instrumentation
Slomka, P. J., Pan, T., & Germano, G. (2016) · Seminars in Nuclear Medicine · doi:10.1053/j.semnuclmed.2015.09.006 · Cited by 177
Innovations in Instrumentation for Positron Emission Tomography
Berg, E., & Cherry, S. R. (2018) · Seminars in Nuclear Medicine · doi:10.1053/j.semnuclmed.2018.02.006 · Cited by 150
(2004) · PubMed · Cited by 141
Watson, C. C., Casey, M., Bendriem, B., Carney, J., Townsend, D. W., Eberl, S., Meikle, S. R., & DiFilippo, F. P. (2005) · PubMed · Cited by 140
Masuda, Y., Kondo, C., Matsuo, Y., Uetani, M., & Kusakabe, K. (2009) · Journal of Nuclear Medicine · doi:10.2967/jnumed.108.060590 · Cited by 104
van Sluis, J., Boellaard, R., Somasundaram, A., van Snick, J. H., Borra, R., Dierckx, R., Stormezand, G. N., Glaudemans, A. W. J. M., & Noordzij, W. (2019) · Journal of Nuclear Medicine · doi:10.2967/jnumed.119.227801 · Cited by 100
Systematic and Distributed Time-of-Flight List Mode PET Reconstruction
Wang, W., Hu, Z., Gualtieri, E. E., Parma, M. J., Walsh, E. S., Sebok, D., Hsieh, Y.-L., Tung, C.-H., Song, X., Griesmer, J. J., et al. (2006) · 2006 IEEE Nuclear Science Symposium Conference Record · doi:10.1109/nssmic.2006.354229 · Cited by 95 · cited 1×
“vendor-recommended blob-based ordered-subset expectation maximization (OSEM) time of flight (TOF) algorithm [18] with the default setting of 3 iterations and 33 subsets and a matrix of 144 × 144 with a voxel size of 4 × 4 × 4 mm3.”
MethodJapanese guideline for the oncology FDG-PET/CT data acquisition protocol: synopsis of Version 1.0
Fukukita, H., Senda, M., Terauchi, T., Suzuki, K., Daisaki, H., Matsumoto, K., Ikari, Y., & Hayashi, M. (2010) · Annals of Nuclear Medicine · doi:10.1007/s12149-010-0377-7 · Cited by 89
Japanese guideline for the oncology FDG-PET/CT data acquisition protocol: synopsis of Version 2.0
Fukukita, H., Suzuki, K., Matsumoto, K., Terauchi, T., Daisaki, H., Ikari, Y., Shimada, N., & Senda, M. (2014) · Annals of Nuclear Medicine · doi:10.1007/s12149-014-0849-2 · Cited by 89
Initial experience with a SiPM-based PET/CT scanner: influence of acquisition time on image quality
Sonni, I., Baratto, L., Park, S., Hatami, N., Srinivas, S., Davidzon, G., Gambhir, S. S., & Iagaru, A. (2018) · EJNMMI Physics · doi:10.1186/s40658-018-0207-x · Cited by 74 · cited 7×
“minimal PET acquisition time of 90s/BP [15].”
BackgroundMethodInfluentialDosage optimization in positron emission tomography: state-of-the-art methods and future prospects.
Karakatsanis, N. A., Fokou, E., & Tsoumpas, C. (2015) · Europe PMC (PubMed Central) · Cited by 70
van Sluis, J., Boellaard, R., Dierckx, R. A. J. O., Stormezand, G. N., Glaudemans, A. W. J. M., & Noordzij, W. (2020) · Journal of Nuclear Medicine · doi:10.2967/jnumed.119.234351 · Cited by 65 · cited 1×
“Further benefits can be experienced in the paediatric field where PET/CT studies with short acquisition times might lead to a decrease in anaesthesia or sedation time [15, 31].”
BackgroundMurray, I., Kalemis, A., Glennon, J., Hasan, S., Quraishi, S., Beyer, T., & Avril, N. E. (2010) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-010-1466-5 · Cited by 60 · cited 4×
“Comparable studies [15, 23, 25] (see Table 6) either included less PET patient data, evaluated different PET/ CT scanners, used other methods for statistical analyses Fig.”
BackgroundMethodInfluentialWillowson, K. P., Bailey, E. A., & Bailey, D. L. (2011) · Australasian Physical & Engineering Sciences in Medicine · doi:10.1007/s13246-011-0119-8 · Cited by 56
Baratto, L., Park, S. Y., Hatami, N., Davidzon, G., Srinivas, S., Gambhir, S. S., & Iagaru, A. (2017) · PLOS ONE · doi:10.1371/journal.pone.0178936 · Cited by 50 · cited 1×
“The most commonly employed PET radiopharmaceutical is [18F]-fluoro-deoxy-glucose (FDG), a glucose analogue [2, 5, 6].”
MethodPrieto, E., García‐Velloso, M. J., Rodríguez‐Fraile, M., Morán, V., Garcia‐Garcia, B., Guillén, F., Morales, M., Sancho, L., Peñuelas, I., Richter, J. Á., et al. (2018) · Physica Medica · doi:10.1016/j.ejmp.2018.01.021 · Cited by 47 · cited 1×
“Accelerated PET imaging can avoid motion artefacts in the latter patient population and increase patient throughput, which meets the ever-increasing demand for PET studies [1] due to an increasing number of well-evidenced clinical indications.”
BackgroundKoopman, D., van Osch, J. A. C., Jager, P. L., Tenbergen, C. J. A., Knollema, S., Slump, C. H., & van Dalen, J. A. (2016) · EJNMMI Physics · doi:10.1186/s40658-016-0158-z · Cited by 40 · cited 14×
“SUVmax and SUVmean were calculated for 214 [18F]-FDG-avid lesions (164 malignant: mean SUVmax 7.9 ± 4.0, mean SUVmean 4.8 ± 1.9, 50 benign: mean SUVmax 6.0 ± 3.3, mean SUVmean 4 ± 1.4) matched per patient between the 75 s and 57 s [18F]-FDG PET/CT studies.”
BackgroundMethodInfluentialMarner, L., Nysom, K., Sehested, A. M., Borgwardt, L., Mathiasen, R., Mathiasen, R., Wehner, P. S., Henriksen, O. M., Lundemann, M., Thomsen, C., et al. (2019) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-019-04486-2 · Cited by 37
McDermott, G. M., Chowdhury, F. U., & Scarsbrook, A. F. (2013) · Annals of Nuclear Medicine · doi:10.1007/s12149-013-0760-2 · Cited by 25
Patient-specific optimisation of administered activity and acquisition times for 18F-FDG PET imaging
Wickham, F., McMeekin, H., Burniston, M., McCool, D., Pencharz, D., Skillen, A., & Wagner, T. (2017) · EJNMMI Research · doi:10.1186/s13550-016-0250-3 · Cited by 25
Digital PET compliance to EARL accreditation specifications
Koopman, D., Koerkamp, M. G., Jager, P. L., Arkies, H., Knollema, S., Slump, C. H., Sanches, P. G., & van Dalen, J. A. (2017) · EJNMMI Physics · doi:10.1186/s40658-017-0176-5 · Cited by 22
Chen, M.-K., Menard, D. H., & Cheng, D. W. (2016) · Journal of Nuclear Medicine Technology · doi:10.2967/jnmt.115.165258 · Cited by 16 · cited 1×
“SUV) of imaging performance are more appropriate guides for the assessment of lesion detectability and PET image quality [30].”
BackgroundZargan, S., Ghafarian, P., Monfared, A. S., Sharafi, A. A., Bakhshayeshkaram, M., & Ay, M. R. (2017) · Europe PMC (PubMed Central) · Cited by 15
Cited by
Influences on PET Quantification and Interpretation
Rogasch, J. M. M., Hofheinz, F., van Heek, L., Voltin, C.-A., Boellaard, R., & Kobe, C. (2022) · Diagnostics · doi:10.3390/diagnostics12020451 · Cited by 57
Prenosil, G., Hentschel, M., Weitzel, T., Sari, H., Shi, K., Afshar‐Oromieh, A., & Rominger, A. (2022) · EJNMMI Physics · doi:10.1186/s40658-022-00455-1 · Cited by 30
“[33] demonstrated a certain tolerance of EARL compliant FDG PET/CT measurements to image noise, reducing”
BackgroundAlves, V. D. P. V., Brady, S. L., Ata, N. K. A., Li, Y., MacLean, J. R., Zhang, B., Sharp, S. E., & Trout, A. T. (2022) · American Journal of Roentgenology · doi:10.2214/ajr.22.27894 · Cited by 20
Prenosil, G. A., Hentschel, M., Weitzel, T., Sari, H., Shi, K., & Rominger, A. (2021) · Research Square · doi:10.21203/rs.3.rs-864693/v1
Ogasawara, R., Irikawa, A., Watanabe, Y., Harada, T., Hosokawa, S., Koyama, K., Tsuda, K., Kimura, T., Okuda, K., & Takahashi, Y. (2025) · Radiation · doi:10.3390/radiation5020018
“Furthermore, the reported studies primarily focused on shortening the acquisition time, with no changes made to the image reconstruction parameters [13–15].”
Metadata
- Authors
- Pilz, J., Hehenwarter, L., Zimmermann, G., Rendl, G., Schweighofer-Zwink, G., Beheshti, M., & Pirich, C.
- Affiliations
- Paracelsus Medical University (AT)
- Corresponding author
- Pirich, Christian
- ORCID
- Venue
- EJNMMI Research
- Published
- 2021-05-01
- Volume
- 11
- Issue
- 1
- Pages
- 44-44
- ISSN
- 2191-219X (ISSN-L)
- PMC
- PMC8088415
- OpenAlex
- W3159563079
- Cited by
- 6
- References
- 31
- Citations by year
- 2025: 2 · 2022: 3 · 2021: 1
- FWCI
- 0.53
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Licence
- cc-by
- creativecommons.org/licenses/by/4.0 (tdm)
- creativecommons.org/licenses/by/4.0 (vor)
- Locations
- EJNMMI Research (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
- Funders
- Paracelsus Medizinische Privatuniversität (D-17/02/007-PIL)
- Topics
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Medical Imaging Techniques and Applications
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Cardiac Imaging and Diagnostics
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Radiation Dose and Imaging
- Keywords
- 18F-FDG PET, acquisition time reduction, diagnostic equivalence, standardized uptake value, lesion detection rate, image quality, melanoma, lung cancer, head and neck cancer, list-mode data, semi-quantitative analysis, whole-body PET/CT
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.