Helium-4 magnetometers for room-temperature biomedical imaging: toward collective operation and photon-noise limited sensitivity
Open accessThe DOI matches its Crossref record; 8 of the 8 references it cites were found in a registry or an index.
Abstract
Optically-pumped magnetometers constitute a valuable tool for imaging biological magnetic signals without cryogenic cooling. Nowadays, numerous developments are being pursued using alkali-based magnetometers, which have demonstrated excellent sensitivities in the spin-exchange relaxation free (SERF) regime that requires heating to >100 °C. In contrast, metastable helium-4 based magnetometers work at any temperature, which allows a direct contact with the scalp, yielding larger signals and a better patient comfort. However former 4 He magnetometers displayed large noises of >200 fT/Hz 1/2 with 300-Hz bandwidth. We describe here an improved magnetometer reaching a sensitivity better than 50 fT/Hz 1/2, nearly the photon shot noise limit, with a bandwidth of 2 kHz. Like other zero-field atomic magnetometers, these magnetometers can be operated in closed-loop architecture reaching several hundredths nT of dynamic range. A small array of 4 magnetometers operating in a closed loop has been tested with a successful correction of the cross-talks.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
A subfemtotesla multichannel atomic magnetometer
Kominis, I. K., Kornack, T. W., Allred, J. C., & Romalis, M. V. (2003) · Nature · doi:10.1038/nature01484 · Cited by 1,631
Moving magnetoencephalography towards real-world applications with a wearable system
Boto, E., Holmes, N., Leggett, J., Roberts, G., Shah, V., Meyer, S. S., Muñoz, L. D., Mullinger, K. J., Tierney, T. M., Bestmann, S., et al. (2018) · Nature · doi:10.1038/nature26147 · Cited by 1,377
High-Sensitivity Atomic Magnetometer Unaffected by Spin-Exchange Relaxation
Allred, J. C., Lyman, R. N., Kornack, T. W., & Romalis, M. V. (2002) · Physical Review Letters · doi:10.1103/physrevlett.89.130801 · Cited by 1,049
A microfabricated optically-pumped magnetic gradiometer
Sheng, D., Perry, A. R., Krzyzewski, S. P., Geller, S., Kitching, J., & Knappe, S. (2017) · Applied Physics Letters · doi:10.1063/1.4974349 · Cited by 179
Magnetocardiography with a 16-channel fiber-coupled single-cell Rb optically pumped magnetometer
Kim, Y. J., Savukov, I., & Newman, S. (2019) · Applied Physics Letters · doi:10.1063/1.5094339 · Cited by 133
Characterizing atomic magnetic gradiometers for fetal magnetocardiography
Sulai, I. A., DeLand, Z. J., Bulatowicz, M. D., Wahl, C. P., Wakai, R. T., & Walker, T. G. (2019) · Review of Scientific Instruments · doi:10.1063/1.5091007 · Cited by 79
Jensen, K., Skarsfeldt, M. A., Stærkind, H., Arnbak, J., Balabas, M. V., Olesen, S.-P., Bentzen, B. H., & Polzik, E. S. (2018) · Scientific Reports · doi:10.1038/s41598-018-34535-z · Cited by 79
Clinical utility of magnetocardiography in cardiology for the detection of myocardial ischemia
Camm, A. J., Henderson, R., Brisinda, D., Body, R., Charles, R. G., Varcoe, B., & Fenici, R. (2019) · Journal of Electrocardiology · doi:10.1016/j.jelectrocard.2019.07.009 · Cited by 42
Cited by
Brookes, M. J., Leggett, J., Rea, M., Hill, R. M., Holmes, N., Boto, E., & Bowtell, R. (2022) · Trends in Neurosciences · doi:10.1016/j.tins.2022.05.008 · Cited by 412
Using OPMs to measure neural activity in standing, mobile participants
Seymour, R. A., Alexander, N., Mellor, S., O'Neill, G. C., Tierney, T. M., Barnes, G. R., & Maguire, E. A. (2021) · NeuroImage · doi:10.1016/j.neuroimage.2021.118604 · Cited by 102
Wearable OPM-MEG: a changing landscape for epilepsy
Pedersen, M., Abbott, D. F., & Jackson, G. D. (2022) · Epilepsia · doi:10.1111/epi.17368 · Cited by 68
Four-channel optically pumped magnetometer for a magnetoencephalography sensor array
Iivanainen, J., Carter, T. R., Dhombridge, J. E., Read, T. S., Campbell, K., Abate, Q., Ridley, D. M., Borna, A., & Schwindt, P. D. D. (2024) · Optics Express · doi:10.1364/oe.517961 · Cited by 18
“The majority of the OPM sensors are based on alkali metal atoms such as rubidium-87 ( 87 Rb; [25–27]), potassium [7,28] and cesium [29,30] while some sensors use helium-4 [31].”
Metadata
- Authors
- Fourcault, W., Romain, R., Le Gal, G., Bertrand, F., Josselin, V., Le Prado, M., Labyt, E., & Palacios-Laloy, A.
- Venue
- Optics Express
- Publisher
- Optica Publishing Group
- Published
- 2021-04-28
- Volume
- 29
- Issue
- 10
- Pages
- 14467
- ISSN
- 1094-4087 (ISSN-L)
- OpenAlex
- W3136666376
- Cited by
- 65
- References
- 8
- FWCI
- 5.04
- Type
- journal article
- Language
- en
- Open access
- gold
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Helium-4 magnetometers for room-temperature biomedical imaging: toward collective operation and photon-noise limited sensitivity. https://app.cochat.ai/papers/10.1364/oe.420031
- The paper
- Fourcault, W., Romain, R., Le Gal, G., Bertrand, F., Josselin, V., Le Prado, M., Labyt, E., & Palacios-Laloy, A. (2021). Helium-4 magnetometers for room-temperature biomedical imaging: toward collective operation and photon-noise limited sensitivity. Optics Express. https://doi.org/10.1364/oe.420031
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.