Insights into the Problem of Alarm Fatigue with Physiologic Monitor Devices: A Comprehensive Observational Study of Consecutive Intensive Care Unit Patients
Open accessThe DOI matches its Crossref record; 23 of the 23 references it cites were found in a registry or an index.
Abstract
PURPOSE: Physiologic monitors are plagued with alarms that create a cacophony of sounds and visual alerts causing "alarm fatigue" which creates an unsafe patient environment because a life-threatening event may be missed in this milieu of sensory overload. Using a state-of-the-art technology acquisition infrastructure, all monitor data including 7 ECG leads, all pressure, SpO(2), and respiration waveforms as well as user settings and alarms were stored on 461 adults treated in intensive care units. Using a well-defined alarm annotation protocol, nurse scientists with 95% inter-rater reliability annotated 12,671 arrhythmia alarms. RESULTS: A total of 2,558,760 unique alarms occurred in the 31-day study period: arrhythmia, 1,154,201; parameter, 612,927; technical, 791,632. There were 381,560 audible alarms for an audible alarm burden of 187/bed/day. 88.8% of the 12,671 annotated arrhythmia alarms were false positives. Conditions causing excessive alarms included inappropriate alarm settings, persistent atrial fibrillation, and non-actionable events such as PVC's and brief spikes in ST segments. Low amplitude QRS complexes in some, but not all available ECG leads caused undercounting and false arrhythmia alarms. Wide QRS complexes due to bundle branch block or ventricular pacemaker rhythm caused false alarms. 93% of the 168 true ventricular tachycardia alarms were not sustained long enough to warrant treatment. DISCUSSION: The excessive number of physiologic monitor alarms is a complex interplay of inappropriate user settings, patient conditions, and algorithm deficiencies. Device solutions should focus on use of all available ECG leads to identify non-artifact leads and leads with adequate QRS amplitude. Devices should provide prompts to aide in more appropriate tailoring of alarm settings to individual patients. Atrial fibrillation alarms should be limited to new onset and termination of the arrhythmia and delays for ST-segment and other parameter alarms should be configurable. Because computer devices are more reliable than humans, an opportunity exists to improve physiologic monitoring and reduce alarm fatigue.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Figures











![Figure 12 True accelerated ventricular rhythm alarm showing why this arrhythmia is not considered an “actionable” alarm condition. Accelerated ventricular rhythm at a rate of 56 for the first 5 beats followed by 2 fusion beats; the last 2 beats are normal sinus rhythm. The invasive arterial pressure waveform (bottom tracing) shows no change between normal rhythm and accelerated ventricular rhythm which confirms the rationale for the published guidelines stating no treatment is indicated for this arrhythmia in hospital settings [5].](https://figures.cochat.ai/paper-figures/440dfa9937e3409b986af268dec77e1a/insights-into-the-problem-of-alarm-fatigue-with-physiologic-monitor-devices-a-figure-12.jpg)




Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“One study conducted a comprehensive observational analysis over 31 days across five ICUs, involving 461 adults and monitoring over 2.5 million alarm instances. Of 1,154,201 unique alarms, 88.8% were for arrhythmias with a majority being false positive98.”
From Bias recognition and mitigation strategies in artificial intelligence healthcare applications (2025) · cites this paper as [98]
Experimental — accuracy not yet measured
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
The Cardiac Arrhythmia Suppression Trial (CAST) Investigators (1989) · New England Journal of Medicine · doi:10.1056/nejm198908103210629 · Cited by 3,300
Third Universal Definition of Myocardial Infarction
Thygesen, K., Alpert, J. S., Jaffe, A. S., Simoons, M. L., Chaitman, B. R., & White, H. D. (2012) · Circulation · doi:10.1161/cir.0b013e31826e1058 · Cited by 2,985
Third universal definition of myocardial infarction
the Writing Group on behalf of the Joint ESC/ACCF/AHA/WHF Task Force for the Universal Definition of Myocardial Infarction, Thygesen, K., Alpert, J. S., Jaffe, A. S., Simoons, M. L., Chaitman, B. R., & White, H. D. (2012) · Nature Reviews Cardiology · doi:10.1038/nrcardio.2012.122 · Cited by 2,053
Zipes, D. P., Camm, A. J., Borggrefe, M., Buxton, A. E., Chaitman, B., Fromer, M., Gregoratos, G., Klein, G., Myerburg, R. J., Quinones, M. A., et al. (2006) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2006.07.010 · Cited by 1,973
Weir, M. (1990) · Cardiopulmonary Physical Therapy Journal · doi:10.1097/01823246-199001020-00009 · Cited by 1,058
The association for the advancement of medical instrumentation
(1974) · Ultrasound in Medicine & Biology · doi:10.1016/0301-5629(74)90035-0 · Cited by 1,031
Drew, B. J., Ackerman, M. J., Funk, M., Gibler, W. B., Kligfield, P., Menon, V., Philippides, G. J., Roden, D. M., & Zareba, W. (2010) · Circulation · doi:10.1161/circulationaha.109.192704 · Cited by 821
Self-citationAuthors/Task Force Members (2006) · European Heart Journal · doi:10.1093/eurheartj/ehl199 · Cited by 523
ACC/AHA guidelines for ambulatory electrocardiography
Crawford, M. H., Bernstein, S. J., Deedwania, P. C., DiMarco, J. P., Ferrick, K. J., Garson, A., Green, L. A., Greene, H. L., Silka, M. J., Stone, P. H., et al. (1999) · Journal of the American College of Cardiology · doi:10.1016/s0735-1097(99)00354-x · Cited by 309
Improving Alarm Performance in the Medical Intensive Care Unit Using Delays and Clinical Context
Görges, M., Markewitz, B. A., & Westenskow, D. R. (2009) · Anesthesia & Analgesia · doi:10.1213/ane.0b013e31819bdfbb · Cited by 211
Cited by
Scalable and accurate deep learning with electronic health records
Rajkomar, A., Oren, E., Chen, K., Dai, A. M., Hajaj, N., Hardt, M., Liu, P. J., Liu, X., Marcus, J., Sun, M., et al. (2018) · npj Digital Medicine · doi:10.1038/s41746-018-0029-1 · Cited by 2,648
“This is problematic because the resulting models may produce imprecise predictions: false-positive predictions can overwhelm physicians, nurses, and other providers with false alarms and subsequent alert fatigue [14], which the Joint Commission identified as a national patient safety priority in 2014 [10].”
BackgroundScalable and accurate deep learning with electronic health records.
Rajkomar, A., Oren, E., Chen, K., Dai, A. M., Hajaj, N., Hardt, M., Liu, P. J., Liu, X., Marcus, J., Sun, M., et al. (2018) · eScholarship (California Digital Library) · Cited by 2,242
Ensuring fairness in machine learning to advance health equity.
Rajkomar, A., Hardt, M., Howell, M. D., Corrado, G., & Chin, M. H. (2018) · Annals of Internal Medicine · doi:10.7326/m18-1990 · Cited by 1,341
“If the model had a lower positive predictive value for African Americans, it might also disproportionately harm them through dismissal bias—a generalization of alert fatigue in which clinicians may learn to discount or “dismiss” alerts for African Americans because they are more likely to be false-positive (17).”
BackgroundBig Data Analytics in Healthcare
Belle, A., Thiagarajan, R., Soroushmehr, S. M. R., Navidi, F., Beard, D. A., & Najarian, K. (2015) · BioMed Research International · doi:10.1155/2015/370194 · Cited by 487
“However, such uncompounded approaches towards development and implementation of alarm systems tend to be unreliable and their sheer numbers could cause “alarm fatigue” for both care givers and patients [10–12].”
BackgroundMachine Learning and Decision Support in Critical Care
Johnson, A. E. W., Ghassemi, M. M., Nemati, S., Niehaus, K. E., Clifton, D., & Clifford, G. D. (2016) · Proceedings of the IEEE · doi:10.1109/jproc.2015.2501978 · Cited by 432
“[65] reviewed the issue of alarm fatigue associated with false alarms and suggested alarm algorithms should focus on: using all available ECG leads and extracting at least one lead with high quality data if available, providing contextual alarms based upon multiple features (e.”
Background
Metadata
- Authors
- Drew, B. J., Harris, P., Zègre-Hemsey, J. K., Mammone, T., Schindler, D., Salas-Boni, R., Bai, Y., Tinoco, A., Ding, Q., & Hu, X.
- Affiliations
- University of California, San Francisco (US)
- University of California San Francisco Medical Center (US)
- Corresponding author
- Drew, Barbara J.
- Venue
- PLoS ONE
- Publisher
- Public Library of Science (PLoS)
- Published
- 2014-10-22
- Volume
- 9
- Issue
- 10
- Pages
- e110274
- ISSN
- 1932-6203 (ISSN-L)
- PMC
- PMC4206416
- OpenAlex
- W2001741247
- Cited by
- 438
- References
- 23
- Citations by year
- 2026: 51 · 2025: 31 · 2024: 25 · 2023: 37 · 2022: 26 · 2021: 58 · 2020: 29 · 2019: 38 · 2018: 42 · 2017: 36 · 2016: 43 · 2015: 21 · 2014: 1
- FWCI
- 33.15
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Locations
- PLoS ONE (publishedVersion, cc-by)
- Europe PMC (PubMed Central) (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- Carolina Digital Repository (University of North Carolina at Chapel Hill) (submittedVersion)
- Dominican Scholar (Dominican University of California) (submittedVersion, cc-by)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
- eScholarship (California Digital Library) (submittedVersion)
- UNC Libraries
- Funders
- University of California, San Francisco
- Topics
- Health Sciences › Medicine › Surgery › Healthcare Technology and Patient Monitoring
- Health Sciences › Medicine › Public Health, Environmental and Occupational Health › Patient Safety and Medication Errors
- Health Sciences › Health Professions › Health Information Management › Electronic Health Records Systems
- Keywords
- alarm fatigue, intensive care unit patients, false positive alerts, alarm settings, ECG leads, QRS amplitude, atrial fibrillation, ventricular tachycardia, bundle branch block, persistent atrial fibrillation, patient safety
- MeSH terms
- Adult, Aged, Arrhythmias, Cardiac, Atrial Fibrillation, Diagnostic Errors, Electrocardiography, Equipment Failure, False Positive Reactions, Female, Humans, Intensive Care Units, Male, Middle Aged, Monitoring, Physiologic, Critical Illness, Clinical Alarms
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
How to cite
Cite CoChat for the checks on this page, and the paper for its own findings.
- This page
- CoChat. (2026). Citation check: Insights into the Problem of Alarm Fatigue with Physiologic Monitor Devices: A Comprehensive Observational Study of Consecutive Intensive Care Unit Patients. https://app.cochat.ai/papers/10.1371/journal.pone.0110274
- The paper
- Drew, B. J., Harris, P., Zègre-Hemsey, J. K., Mammone, T., Schindler, D., Salas-Boni, R., Bai, Y., Tinoco, A., Ding, Q., & Hu, X. (2014). Insights into the Problem of Alarm Fatigue with Physiologic Monitor Devices: A Comprehensive Observational Study of Consecutive Intensive Care Unit Patients. PLoS ONE. https://doi.org/10.1371/journal.pone.0110274
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.