Insights into the Problem of Alarm Fatigue with Physiologic Monitor Devices: A Comprehensive Observational Study of Consecutive Intensive Care Unit Patients
Open accessThe DOI matches its Crossref record; 23 of the 23 references it cites were found in a registry or an index.
Abstract
PURPOSE: Physiologic monitors are plagued with alarms that create a cacophony of sounds and visual alerts causing "alarm fatigue" which creates an unsafe patient environment because a life-threatening event may be missed in this milieu of sensory overload. Using a state-of-the-art technology acquisition infrastructure, all monitor data including 7 ECG leads, all pressure, SpO(2), and respiration waveforms as well as user settings and alarms were stored on 461 adults treated in intensive care units. Using a well-defined alarm annotation protocol, nurse scientists with 95% inter-rater reliability annotated 12,671 arrhythmia alarms. RESULTS: A total of 2,558,760 unique alarms occurred in the 31-day study period: arrhythmia, 1,154,201; parameter, 612,927; technical, 791,632. There were 381,560 audible alarms for an audible alarm burden of 187/bed/day. 88.8% of the 12,671 annotated arrhythmia alarms were false positives. Conditions causing excessive alarms included inappropriate alarm settings, persistent atrial fibrillation, and non-actionable events such as PVC's and brief spikes in ST segments. Low amplitude QRS complexes in some, but not all available ECG leads caused undercounting and false arrhythmia alarms. Wide QRS complexes due to bundle branch block or ventricular pacemaker rhythm caused false alarms. 93% of the 168 true ventricular tachycardia alarms were not sustained long enough to warrant treatment. DISCUSSION: The excessive number of physiologic monitor alarms is a complex interplay of inappropriate user settings, patient conditions, and algorithm deficiencies. Device solutions should focus on use of all available ECG leads to identify non-artifact leads and leads with adequate QRS amplitude. Devices should provide prompts to aide in more appropriate tailoring of alarm settings to individual patients. Atrial fibrillation alarms should be limited to new onset and termination of the arrhythmia and delays for ST-segment and other parameter alarms should be configurable. Because computer devices are more reliable than humans, an opportunity exists to improve physiologic monitoring and reduce alarm fatigue.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Figures











![Figure 12 True accelerated ventricular rhythm alarm showing why this arrhythmia is not considered an “actionable” alarm condition. Accelerated ventricular rhythm at a rate of 56 for the first 5 beats followed by 2 fusion beats; the last 2 beats are normal sinus rhythm. The invasive arterial pressure waveform (bottom tracing) shows no change between normal rhythm and accelerated ventricular rhythm which confirms the rationale for the published guidelines stating no treatment is indicated for this arrhythmia in hospital settings [5].](https://figures.cochat.ai/paper-figures/440dfa9937e3409b986af268dec77e1a/insights-into-the-problem-of-alarm-fatigue-with-physiologic-monitor-devices-a-figure-12.jpg)




Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The source backs part of the claim, not all of it — or backs it for a different population, condition or comparison.
In the citing paper
“One study conducted a comprehensive observational analysis over 31 days across five ICUs, involving 461 adults and monitoring over 2.5 million alarm instances. Of 1,154,201 unique alarms, 88.8% were for arrhythmias with a majority being false positive98.”
From Bias recognition and mitigation strategies in artificial intelligence healthcare applications (2025) · cites this paper as [98]
Experimental — accuracy not yet measured
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
The Cardiac Arrhythmia Suppression Trial (CAST) Investigators (1989) · New England Journal of Medicine · doi:10.1056/nejm198908103210629 · Cited by 3,300
Third Universal Definition of Myocardial Infarction
Thygesen, K., Alpert, J. S., Jaffe, A. S., Simoons, M. L., Chaitman, B. R., & White, H. D. (2012) · Circulation · doi:10.1161/cir.0b013e31826e1058 · Cited by 2,985
Third universal definition of myocardial infarction
the Writing Group on behalf of the Joint ESC/ACCF/AHA/WHF Task Force for the Universal Definition of Myocardial Infarction, Thygesen, K., Alpert, J. S., Jaffe, A. S., Simoons, M. L., Chaitman, B. R., & White, H. D. (2012) · Nature Reviews Cardiology · doi:10.1038/nrcardio.2012.122 · Cited by 2,053
Zipes, D. P., Camm, A. J., Borggrefe, M., Buxton, A. E., Chaitman, B., Fromer, M., Gregoratos, G., Klein, G., Myerburg, R. J., Quinones, M. A., et al. (2006) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2006.07.010 · Cited by 1,973
Weir, M. (1990) · Cardiopulmonary Physical Therapy Journal · doi:10.1097/01823246-199001020-00009 · Cited by 1,058
The association for the advancement of medical instrumentation
(1974) · Ultrasound in Medicine & Biology · doi:10.1016/0301-5629(74)90035-0 · Cited by 1,031
Drew, B. J., Ackerman, M. J., Funk, M., Gibler, W. B., Kligfield, P., Menon, V., Philippides, G. J., Roden, D. M., & Zareba, W. (2010) · Circulation · doi:10.1161/circulationaha.109.192704 · Cited by 821
Self-citationAuthors/Task Force Members (2006) · European Heart Journal · doi:10.1093/eurheartj/ehl199 · Cited by 523
ACC/AHA guidelines for ambulatory electrocardiography
Crawford, M. H., Bernstein, S. J., Deedwania, P. C., DiMarco, J. P., Ferrick, K. J., Garson, A., Green, L. A., Greene, H. L., Silka, M. J., Stone, P. H., et al. (1999) · Journal of the American College of Cardiology · doi:10.1016/s0735-1097(99)00354-x · Cited by 309
Improving Alarm Performance in the Medical Intensive Care Unit Using Delays and Clinical Context
Görges, M., Markewitz, B. A., & Westenskow, D. R. (2009) · Anesthesia & Analgesia · doi:10.1213/ane.0b013e31819bdfbb · Cited by 211
Cited by
Scalable and accurate deep learning with electronic health records
Rajkomar, A., Oren, E., Chen, K., Dai, A. M., Hajaj, N., Hardt, M., Liu, P. J., Liu, X., Marcus, J., Sun, M., et al. (2018) · npj Digital Medicine · doi:10.1038/s41746-018-0029-1 · Cited by 2,648
“This is problematic because the resulting models may produce imprecise predictions: false-positive predictions can overwhelm physicians, nurses, and other providers with false alarms and subsequent alert fatigue [14], which the Joint Commission identified as a national patient safety priority in 2014 [10].”
BackgroundScalable and accurate deep learning with electronic health records.
Rajkomar, A., Oren, E., Chen, K., Dai, A. M., Hajaj, N., Hardt, M., Liu, P. J., Liu, X., Marcus, J., Sun, M., et al. (2018) · eScholarship (California Digital Library) · Cited by 2,242
Ensuring fairness in machine learning to advance health equity.
Rajkomar, A., Hardt, M., Howell, M. D., Corrado, G., & Chin, M. H. (2018) · Annals of Internal Medicine · doi:10.7326/m18-1990 · Cited by 1,341
“If the model had a lower positive predictive value for African Americans, it might also disproportionately harm them through dismissal bias—a generalization of alert fatigue in which clinicians may learn to discount or “dismiss” alerts for African Americans because they are more likely to be false-positive (17).”
BackgroundBig Data Analytics in Healthcare
Belle, A., Thiagarajan, R., Soroushmehr, S. M. R., Navidi, F., Beard, D. A., & Najarian, K. (2015) · BioMed Research International · doi:10.1155/2015/370194 · Cited by 487
“However, such uncompounded approaches towards development and implementation of alarm systems tend to be unreliable and their sheer numbers could cause “alarm fatigue” for both care givers and patients [10–12].”
BackgroundMachine Learning and Decision Support in Critical Care
Johnson, A. E. W., Ghassemi, M. M., Nemati, S., Niehaus, K. E., Clifton, D., & Clifford, G. D. (2016) · Proceedings of the IEEE · doi:10.1109/jproc.2015.2501978 · Cited by 432
“[65] reviewed the issue of alarm fatigue associated with false alarms and suggested alarm algorithms should focus on: using all available ECG leads and extracting at least one lead with high quality data if available, providing contextual alarms based upon multiple features (e.”
BackgroundSandau, K. E., Funk, M., Auerbach, A., Barsness, G. W., Blum, K., Cvach, M., Lampert, R., May, J. L., McDaniel, G. M., Perez, M. V., et al. (2017) · Circulation · doi:10.1161/cir.0000000000000527 · Cited by 342
Bias recognition and mitigation strategies in artificial intelligence healthcare applications
Hasanzadeh, F., Josephson, C. B., Waters, G., Adedinsewo, D., Azizi, Z., & White, J. A. (2025) · npj Digital Medicine · doi:10.1038/s41746-025-01503-7 · Cited by 320
“Of 1,154,201 unique alarms, 88.8% were for arrhythmias with a majority being false positive 98 .”
InfluentialLeveraging big data analytics in healthcare enhancement: trends, challenges and opportunities
Rehman, A., Naz, S., & Razzak, I. (2022) · Multimedia Systems · doi:10.1007/s00530-020-00736-8 · Cited by 264
Impact of Alarm Fatigue on the Work of Nurses in an Intensive Care Environment—A Systematic Review
Lewandowska, K., Weisbrot, M., Cieloszyk, A., Mędrzycka-Dąbrowska, W., Krupa, S., & Ozga, D. (2020) · International Journal of Environmental Research and Public Health · doi:10.3390/ijerph17228409 · Cited by 244
“monitoring center, without directly checking the patient’s condition [23].”
BackgroundAlarm fatigue: impacts on patient safety
Ruskin, K. J., & Hueske-Kraus, D. (2015) · Current Opinion in Anaesthesiology · doi:10.1097/aco.0000000000000260 · Cited by 226
Pathinarupothi, R. K., Durga, P., & Rangan, E. S. (2019) · IEEE Internet of Things Journal · doi:10.1109/jiot.2018.2870068 · Cited by 219
“This would have caused fatigue [31] to the attending physicians who would have either set more stringent thresholds or completely ignored them, both of which defeat the very purpose of having alerts.”
BackgroundExplainable AI for clinical and remote health applications: A survey on tabular and time series data
Di Martino, F., & Delmastro, F. (2023) · Artificial Intelligence Review · doi:10.1007/s10462-022-10304-3 · Cited by 216
“[124] (2021) 8 Signal quality assessment PPG CNN SHAP, integrated gradients, guided saliency Training: ICU dataset [126]; Test: stroke dataset [127]”
MethodJiang, L., Wu, Z., Xu, X., Zhan, Y., Jin, X., Wang, L., & Qiu, Y. (2021) · Journal of International Medical Research · doi:10.1177/03000605211000157 · Cited by 207
“Actually, the low positive rate was caused by the small number of African Americans who participated in the initial experiment, and false-positive results were more likely.(26) In addition, injustices may be encountered in the design, deployment, and evaluation of models.”
BackgroundPaine, C. W., Goel, V. V., Ely, E., Stave, C. D., Stemler, S., Zander, M., & Bonafide, C. P. (2016) · Journal of Hospital Medicine · doi:10.1002/jhm.2520 · Cited by 183
“Finally, though not necessarily a limitation, several studies were conducted by authors involved in or funded by the medical device industry.(11,15,19,31,32) This has the potential to introduce bias, although we have no indication that the quality of the science was adversely impacted.”
BackgroundSevakula, R. K., Au‐Yeung, W. M., Singh, J. P., Heist, E. K., Isselbacher, E. M., & Armoundas, A. A. (2020) · Journal of the American Heart Association · doi:10.1161/jaha.119.013924 · Cited by 173
“8% of annotated arrhythmic alarms were false positives.(69) Major causes of these false arrhythmic alarms were found to be noise and artifacts in the physiological signals, which again were probably caused by patient motion or loose electrodes.”
BackgroundWinters, B. D., Cvach, M. M., Bonafide, C. P., Hu, X., Konkani, A., O’Connor, M. F., Rothschild, J. M., Selby, N. M., Pelter, M. M., McLean, B., et al. (2018) · Critical Care Medicine · doi:10.1097/ccm.0000000000002803 · Cited by 170
Self-citationMapping sources of noise in an intensive care unit
Darbyshire, J. L., Müller-Trapet, M., Cheer, J., Fazi, F. M., & Young, D. (2019) · Anaesthesia · doi:10.1111/anae.14690 · Cited by 124
“However, a significant proportion of the noise coming from speech is not required for patient care [1, 2], and almost 90% of alarms from physiological monitors are ‘false positives’, with nopatient benefit [35].”
BackgroundAssociation of electronic health record use with physician fatigue and efficiency
Khairat, S., Coleman, C., Ottmar, P., Jayachander, D. I., Bice, T., & Carson, S. S. (2020) · JAMA Network Open · doi:10.1001/jamanetworkopen.2020.7385 · Cited by 118
A call to alarms: Current state and future directions in the battle against alarm fatigue
Hravnak, M., Pellathy, T., Chen, L., Dubrawski, A. W., Wertz, A., Clermont, G., & Pinsky, M. R. (2018) · Journal of Electrocardiology · doi:10.1016/j.jelectrocard.2018.07.024 · Cited by 122
Pal, J., Taywade, M., Pal, R., & Sethi, D. (2022) · Noise and Health · doi:10.4103/nah.nah_79_21 · Cited by 149
“[12] The shocking results of the above-stated studies are that most of the noise originates from specific areas such as the staff cafeteria near the ICU where conversations occur and nursing stations where handing and taking over the patients arise.”
E-health-IoT Universe: A Review
Scarpato, N., Pieroni, A., Nunzio, L. D., & Fallucchi, F. (2017) · International Journal on Advanced Science Engineering and Information Technology · doi:10.18517/ijaseit.7.6.4467 · Cited by 130
“Similar to medical images, medical signals also pose volume and velocity obstacles especially during continuous, high-resolution acquisition and storage from a multitude of monitors connected to each patient, “alarm fatigue” for both care givers and patients [100], [101].”
BackgroundManaging alarm systems for quality and safety in the hospital setting
Bach, T. A., Berglund, L.-M., & Turk, E. (2018) · BMJ Open Quality · doi:10.1136/bmjoq-2017-000202 · Cited by 119
“Research suggests that 80%–99% of alarms are false.(7) 13–18 Alarms are designed by default to be high in sensitivity but low in specificity ‘because a manufacturer is much more likely to be held liable for an alarm system that fails to annunciate a clinically significant alarm condition than for any problems caused b…”
BackgroundNoise Levels in Surgical ICUs Are Consistently Above Recommended Standards
Tainter, C. “., Levine, A. R., Quraishi, S. A., Butterly, A. D., Stahl, D. L., Eikermann, M., Kaafarani, H. M. A., & Lee, J. (2015) · Critical Care Medicine · doi:10.1097/ccm.0000000000001378 · Cited by 100
“Modern ICUs are a cacophony of raucous devices (1).”
BackgroundEffects of monitor alarm management training on nurses’ alarm fatigue: A randomised controlled trial
Bi, J., Yin, X., Li, H., Gao, R., Zhang, Q., Zhong, T., Zan, T., Guan, B., & Li, Z. (2020) · Journal of Clinical Nursing · doi:10.1111/jocn.15452 · Cited by 104
Areia, C. M., Biggs, C. A., Santos, M., Thurley, N., Gerry, S., Tarassenko, L., Watkinson, P., & Vollam, S. (2021) · Critical Care · doi:10.1186/s13054-021-03766-4 · Cited by 98
Shashikumar, S. P., Shah, A. J., Clifford, G. D., & Nemati, S. (2018) · Proceedings of the 24th ACM SIGKDD International Conference on Knowledge Discovery & Data Mining · doi:10.1145/3219819.3219912 · Cited by 89
“Although, the two-dimensional embedding of the features only provides an approximate picture of the position of FP and FN cases, it sheds light on where the algorithm is most likely to disagree with the Holter monitor labels, which are known to be noise-prone [18].”
BackgroundApplying machine learning to continuously monitored physiological data
Rush, B., Celi, L. A., & Stone, D. J. (2019) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-018-0219-z · Cited by 93
(2016) · PubMed Central · Cited by 92
Boriani, G., Fauchier, L., Aguinaga, L., Beattie, J. M., Lundqvist, C. B., Cohen, A. A., Dan, G., Genovesi, S., Israel, C. W., Joung, B., et al. (2018) · EP Europace · doi:10.1093/europace/euy110 · Cited by 102
Weller, R. S., Foard, K. L., & Harwood, T. N. (2017) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-017-0085-0 · Cited by 86
A Supervised Approach to Robust Photoplethysmography Quality Assessment
Pereira, T., Gadhoumi, K., Ma, M., Liu, X., Xiao, R., Colorado, R. A., Keenan, K. J., Meisel, K., & Hu, X. (2020) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2019.2909065 · Cited by 79
“of the five general intensive care units between March 2013 and December 2016 as described in a previous study [22].”
MethodSelf-citationLiu, G., Han, X., Lan, T., Zhou, W., & Liu, H. (2021) · Computer Methods and Programs in Biomedicine · doi:10.1016/j.cmpb.2021.106269 · Cited by 91
Chaparro, J. D., Beus, J. M., Dziorny, A. C., Hagedorn, P. A., Hernandez, S., Kandaswamy, S., Kirkendall, E. S., McCoy, A. B., Muthu, N., & Orenstein, E. W. (2022) · Applied Clinical Informatics · doi:10.1055/s-0042-1748856 · Cited by 76
Chen, L., Dubrawski, A., Wang, D., Fiterau, M., Guillame-Bert, M., Bose, E., Kaynar, A. M., Wallace, D. J., Guttendorf, J., Clermont, G., et al. (2016) · Critical Care Medicine · doi:10.1097/ccm.0000000000001660 · Cited by 75
“Although “alarm hazards” remain at or near the top 10 technology safety concerns (36, 37), it is vital that mechanisms to improve monitoring safety are undertaken.”
BackgroundA novel attentional deep neural network-based assessment method for ECG quality
Jin, Y., Li, Z., Qin, C., Liu, J., Liu, Y., Zhao, L., & Liu, C. (2022) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2022.104064 · Cited by 86
The Imperative of Solving Nurses’ Usability Problems With Health Information Technology
Staggers, N., Elias, B. L., Makar, E., & Alexander, G. L. (2018) · JONA: The Journal of Nursing Administration · doi:10.1097/nna.0000000000000598 · Cited by 70
Electrocardiogram Signal Quality Assessment Based on Structural Image Similarity Metric
Shahriari, Y., Fidler, R., Pelter, M. M., Bai, Y., Villaroman, A., & Hu, X. (2017) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2017.2717876 · Cited by 79
“FALSE Electrocardiogram (ECG) arrhythmia alarms are frequent among the patient monitor alarms contributing to “alarm fatigue” which has been listed as one of the top technology threats that may compromise patient safety in hospitals [1]–[3].”
InfluentialSelf-citationStop the Noise: A Quality Improvement Project to Decrease Electrocardiographic Nuisance Alarms
Sendelbach, S., Wahl, S., Anthony, A., & Shotts, P. (2015) · Critical Care Nurse · doi:10.4037/ccn2015858 · Cited by 64
Shuzan, M. N. I., Chowdhury, M. H., Hossain, M. S., Chowdhury, M. E. H., Reaz, M. B. I., Uddin, M. M., Khandakar, A. A., Mahbub, Z. B., & Ali, S. H. M. (2021) · IEEE Access · doi:10.1109/access.2021.3095380 · Cited by 68
“However, it has been observed that respiratory signals extracted from ECG appeared flat in ICU patients even though they were breathing sufficiently [21].”
BackgroundAreia, C. M., King, E., Ede, J., Young, L., Tarassenko, L., Watkinson, P., & Vollam, S. (2021) · Journal of Advanced Nursing · doi:10.1111/jan.15055 · Cited by 76
“Conversely, patients felt nurses subconsciously disregard the alarms from the noise pollution and developed alarm fatigue, a known concept previously described in other studies (Bonafide et al., 2015; Drew et al., 2014; Görges et al., 2009; de Man et al., 2013).”
ResultAssessment of Clinical Alarms Influencing Nurses’ Perceptions of Alarm Fatigue
Petersen, E. M., & Costanzo, C. L. (2016) · Dimensions of Critical Care Nursing · doi:10.1097/dcc.0000000000000220 · Cited by 62
“The primary findings from the literature review include the effect of excessive alarms on staff, nurses’ response to alarms, alarm sounds and audibility, technological advancements reducing nuisance alarms, and alarm notification systems.(5,16,19) To reduce adverse events from alarm systems, false alarms, and alarm fa…”
BackgroundFernandes, C. O., Miles, S., de Lucena, C. J. P., & Cowan, D. (2019) · Journal of Medical Internet Research · doi:10.2196/15406 · Cited by 61
“Alarm fatigue involves a lack of response because of an excessive number of noncritical alarms being received by health care personnel, resulting in sensory overload and desensitization [1-4].”
Nurses' Perspectives on Clinical Alarms
Honan, L., Funk, M., Maynard, M. A., Fahs, D. B., Clark, J. T., & David, Y. B. (2015) · American Journal of Critical Care · doi:10.4037/ajcc2015552 · Cited by 74
“Drew et al 20 proposed a number of solutions that manufacturers could incorporate into monitors that would be helpful for nurses, including use of all available electrocardiographic leads to identify leads without artifact and those with adequate QRS amplitude, prompts to help in tailoring alarm settings, and delays f…”
BackgroundInfluentialAu-Yeung, W.-T. M., Sahani, A. K., Isselbacher, E. M., & Armoundas, A. A. (2019) · npj Digital Medicine · doi:10.1038/s41746-019-0160-7 · Cited by 56
Rossetti, S. C., Dykes, P. C., Knaplund, C., Cho, S., Withall, J., Lowenthal, G., Albers, D., Lee, R. Y., Jia, H., Bakken, S., et al. (2025) · Nature Medicine · doi:10.1038/s41591-025-03609-7 · Cited by 54
Ruppel, H., Vaux, L. D., Cooper, D. A., Kunz, S., Duller, B., & Funk, M. (2018) · PLoS ONE · doi:10.1371/journal.pone.0205901 · Cited by 61
Chromik, J., Klopfenstein, S. A. I., Pfitzner, B., Sinno, Z.-C., Arnrich, B., Balzer, F., & Poncette, A.-S. (2022) · Frontiers in Digital Health · doi:10.3389/fdgth.2022.843747 · Cited by 80
“This is reasonable since (11) found that these alarms represent the majority of all audible alarms.”
False Alarms and Overmonitoring: Major Factors in Alarm Fatigue Among Labor Nurses
Simpson, K. R., & Lyndon, A. (2018) · Journal of Nursing Care Quality · doi:10.1097/ncq.0000000000000335 · Cited by 55
“Caregivers can be exposed to hundreds of alarms over the course of their shift.(1,2) As physiologic monitoring has become more common, the technology is no longer limited to perioperative and intensive care units (ICUs).”
Improving Prediction Performance Using Hierarchical Analysis of Real-Time Data: A Sepsis Case Study
van Wyk, F., Khojandi, A., & Kamaleswaran, R. (2019) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2019.2894570 · Cited by 52
“Studies have reported false positive rates between 80% and 99% in hospitals [36], [37].”
BackgroundDeasy, J., Liò, P., & Ercole, A. (2020) · Scientific Reports · doi:10.1038/s41598-020-79142-z · Cited by 52
Pereira, T., Ding, C., Gadhoumi, K., Tran, N., Colorado, R. A., Meisel, K., & Hu, X. (2019) · Physiological Measurement · doi:10.1088/1361-6579/ab5b84 · Cited by 51
“In the first cohort of 3764 patients admitted to the ICUs between March 2013 and December 2016, referred to hereafter as group A, we randomly sampled 25 PPG segments per patient, each lasting 30 seconds as described in our previous study [26], for a total of 78278 PPG segments.”
MethodSelf-citationChen, E., Prakash, S., Reddi, V. J., Kim, D. A., & Rajpurkar, P. (2023) · Nature Biomedical Engineering · doi:10.1038/s41551-023-01115-0 · Cited by 83
“…recommendations in the clinician–AI loop, particularly by minimizing false positives (especially if arising from artefacts or aberrancies in continuous monitoring) in the recommendation of follow-up visits, will be essential to avoid adding to the burden of false alarms faced by clinicians 40,41 .”
Creation of a Robust and Generalizable Machine Learning Classifier for Patient Ventilator Asynchrony
Rehm, G. B., Han, J., Kuhn, B. T., Delplanque, J., Anderson, N., Adams, J. Y., & Chuah, C. (2018) · Methods of Information in Medicine · doi:10.3414/me17-02-0012 · Cited by 53
“A low alarm specificity has been shown to cause alarm fatigue, giving rise to the potential for clinicians to miss truly life-threatening changes in physiology, such that patients may physiologically deteriorate until a major problem is finally recognized [7].”
BackgroundWilliams, G. W., George, C. A., Harvey, B. C., & Freeman, J. E. (2016) · Anesthesia & Analgesia · doi:10.1213/ane.0000000000001395 · Cited by 48
Measurement of Physiological Monitor Alarm Accuracy and Clinical Relevance in Intensive Care Units
Ruppel, H., Funk, M., & Whittemore, R. (2018) · American Journal of Critical Care · doi:10.4037/ajcc2018385 · Cited by 58
Redesigning the allergy module of the electronic health record.
Blumenthal, K., Park, M., & Macy, E. (2016) · Annals of Allergy, Asthma & Immunology · doi:10.1016/j.anai.2016.05.017 · Cited by 43
Despins, L. A. (2017) · Intensive and Critical Care Nursing · doi:10.1016/j.iccn.2017.04.003 · Cited by 49
Mechanical Ventilation Alarms and Alarm Fatigue
Scott, J. B., Vaux, L. D., Dills, C. C., & Strickland, S. L. (2019) · Respiratory Care · doi:10.4187/respcare.06878 · Cited by 47
Shiyovich, A., Axelrod, M., Gilutz, H., & Plakht, Y. (2019) · Angiology · doi:10.1177/0003319719867542 · Cited by 45
Critical care nurses' perceptions and practices towards clinical alarms
Jeong, Y. J., & Kim, H. (2022) · Nursing in Critical Care · doi:10.1111/nicc.12751 · Cited by 44
Mousavi, S., Fotoohinasab, A., & Afghah, F. (2020) · PLoS ONE · doi:10.1371/journal.pone.0226990 · Cited by 44
“As reported by Aboukhalil et al. [1] and Drew et al. [2], the rate of false alarms in ICUs reaches as high as almost 90%.”
Sowan, A. K., Vera, A., Fonseca, E., Reed, C. C., Tarriela, A., & Berndt, A. (2017) · The Open Medical Informatics Journal · doi:10.2174/1874431101711010001 · Cited by 46
Adams, J. Y., Lieng, M. K., Kuhn, B. T., Rehm, G. B., Guo, E. C., Taylor, S. L., Delplanque, J., & Anderson, N. (2017) · Scientific Reports · doi:10.1038/s41598-017-15052-x · Cited by 47
Patient characteristics associated with false arrhythmia alarms in intensive care
Harris, P., Zègre‐Hemsey, J. K., Schindler, D. M., Bai, Y., Pelter, M. M., & Hu, X. (2017) · Therapeutics and Clinical Risk Management · doi:10.2147/tcrm.s126191 · Cited by 47
“4; GE Healthcare, Milwaukee, WI, USA) associated with beds in five adult ICUs over a 31-day period.(2) One of the specific”
BackgroundMethodInfluentialSelf-citationThe impact of workload on the ability to localize audible alarms
Edworthy, J. R., Reid, S., Peel, K., Lock, S., Williams, J., Newbury, C., Foster, J., & Farrington, M. (2018) · Applied Ergonomics · doi:10.1016/j.apergo.2018.05.006 · Cited by 44
Krasteva, V., Jekova, I. I., Leber, R., Schmid, R., & Abächerli, R. (2016) · Physiological Measurement · doi:10.1088/0967-3334/37/8/1273 · Cited by 41
Data-Driven Implementation of Alarm Reduction Interventions in a Cardiovascular Surgical ICU
Allan, S. H., Doyle, P. A., Sapirstein, A., & Cvach, M. M. (2016) · The Joint Commission Journal on Quality and Patient Safety · doi:10.1016/j.jcjq.2016.11.004 · Cited by 47
Reducing medical device alarms by an order of magnitude: A human factors approach
Koomen, E., Webster, C. S., Konrad, D., van der Hoeven, J. G., Best, T., Kesecioğlu, J., Gommers, D. A. M. P. J., de Vries, W. B., & Kappen, T. H. (2021) · Anaesthesia and Intensive Care · doi:10.1177/0310057x20968840 · Cited by 42
“A recent study of this problem reported from a single hospital in the USA, with 77 intensive care beds, recorded the occurrence of an astonishing 2,558,760 unique physiological alarms during intensive care in a single month.(15) Both the Joint Commission and the Emergency Care Research Institute (ECRI) have repeatedly…”
BackgroundMulti-parameter vital sign database to assist in alarm optimization for general care units
Welch, J., Kanter, B., Skora, B., McCombie, S., Henry, I., McCombie, D. B., Kennedy, R., & Soller, B. R. (2015) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-015-9790-8 · Cited by 37
“& Babs Soller babs.soller@soterawireless.com 1 Sotera Wireless Inc., 10020 Huennekens St., San Diego, CA 92121, USA been well documented in the operating room [5] and ICU [6, 7] and the concern about alarm fatigue is real [7].”
BackgroundInfluentialSeifert, M., Tola, D. H., Thompson, J. A., McGugan, L., & Smallheer, B. (2021) · Intensive and Critical Care Nursing · doi:10.1016/j.iccn.2021.103098 · Cited by 39
Weiniger, C. F., Carvalho, B., Stocki, D., & Einav, S. (2016) · Anesthesia & Analgesia · doi:10.1213/ane.0000000000001644 · Cited by 36
Kebapçı, A., & Güner, P. (2020) · Intensive and Critical Care Nursing · doi:10.1016/j.iccn.2020.102975 · Cited by 36
“certain period of time; then, they became normalized and desensitized to the sound (Bach, Berglund, & Turk, 2018; Cho, Kim, Lee, & Cho, 2016; Drew et al., 2014; Ergezen & Kol, 2019).”
BackgroundKeim‐Malpass, J., & Moorman, L. P. (2021) · International Journal of Nursing Studies Advances · doi:10.1016/j.ijnsa.2021.100019 · Cited by 46
“…ICU, the bedside cardiorespiratory monitor alone generates, on average, 187 audible alarms per day averaging 1 per every 7.7 min with upwards of 90% of them being non-actionable ( Drew et al., 2014 ; Aboukhalil et al., 2008 ; Joshi et al., 2017 ; Schondelmeyer et al., 2019 ; McClure et al., 2016 ).”
InfluentialChecklists, cognitive aids, and the future of patient safety
Webster, C. S. (2017) · British Journal of Anaesthesia · doi:10.1093/bja/aex193 · Cited by 36
Disrupting Electronic Health Records Systems: The Next Generation
Celi, L. A., Marshall, J. D., Lai, Y., & Stone, D. J. (2015) · JMIR Medical Informatics · doi:10.2196/medinform.4192 · Cited by 35
“Such personalized alarm limits would provide a smarter range of acceptable values for each patient and perhaps also act to reduce the unacceptable number of false positive alarms that currently plague bedside XSL • FO RenderX caregivers (and patients) [17].”
Pattern discovery in critical alarms originating from neonates under intensive care
Joshi, R., van Pul, C., Atallah, L., Feijs, L. M. G., Huffel, S. V., & Andriessen, P. (2016) · Physiological Measurement · doi:10.1088/0967-3334/37/4/564 · Cited by 35
“…the Healthcare Technology Foundation, the Association for the Advancement of Medical Instrumentation and the American Association of Critical-Care Nurses have recognized the hazards originating from alarm fatigue (Graham and Cvach 2010, Keller 2012, Mitka 2013, Drew et al 2014, Funk et al 2014).”
Supervised and unsupervised language modelling in Chest X-Ray radiological reports
Drozdov, I. A., Forbes, D., Szubert, B., Hall, M., Carlin, C. M., & Lowe, D. J. (2020) · PLoS ONE · doi:10.1371/journal.pone.0229963 · Cited by 42
“reduced cognitive overload) benefits [21, 22].”
Cardiac Monitoring in the Emergency Department
Zègre‐Hemsey, J. K., Garvey, J. L., & Carey, M. G. (2016) · Critical Care Nursing Clinics of North America · doi:10.1016/j.cnc.2016.04.009 · Cited by 44
Self-citationIntensive care unit nurses' perceptions and practices regarding clinical alarms: A descriptive study
Wang, L., He, W., Chen, Y., Wu, Q., Du, X., Li, Q., & Song, C. (2023) · Nursing Open · doi:10.1002/nop2.1792 · Cited by 36
“Drew et al. (2014) collected approximately 2.55 million physiological monitor alarms from 5 adult ICUs in 31 days using automatic storage devices.”
BackgroundInfluentialDaluwatte, C., Johannesen, L., Galeotti, L., Vicente, J., Strauss, D. G., & Scully, C. G. (2016) · Physiological Measurement · doi:10.1088/0967-3334/37/8/1370 · Cited by 38
Optimisation of clinical workflow and monitor settings safely reduces alarms in the NICU
Varisco, G., van de Mortel, H., Cabrera-Quirós, L., Atallah, L., Hüske-Kraus, D., Long, X., Cottaar, E. J. E., Zhan, Z. Z., Andriessen, P., & van Pul, C. (2020) · Acta Paediatrica · doi:10.1111/apa.15615 · Cited by 39
Keim‐Malpass, J., Clark, M., Lake, D. E., & Moorman, J. R. (2019) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-019-00361-5 · Cited by 33
Real‐Time Arrhythmia Detection Using Hybrid Convolutional Neural Networks
Bollepalli, S. C., Sevakula, R. K., Au-Yeung, W.-T. M., Kassab, M. B., Merchant, F. M., Bazoukis, G., Boyer, R., Isselbacher, E. M., & Armoundas, A. A. (2021) · Journal of the American Heart Association · doi:10.1161/jaha.121.023222 · Cited by 41
Frequency of passive EHR alerts in the ICU: another form of alert fatigue?
Kizzier-Carnahan, V., Artis, K. A., Mohan, V., & Gold, J. A. (2019) · Journal of Patient Safety · doi:10.1097/pts.0000000000000270 · Cited by 32
“Both alarm and alert fatigue are increasingly recognized patient safety issue in the intensive care unit (ICU).(1,2)With increasing number of alerts, clinicians often become “desensitized” to them.”
BackgroundMcFarlane, D. C., Doig, A. K., Agutter, J. A., Brewer, L., Syroid, N. D., & Mittu, R. (2018) · PLoS ONE · doi:10.1371/journal.pone.0197157 · Cited by 38
Machine learning applied to multi-sensor information to reduce false alarm rate in the ICU
Hever, G., Cohen-Lavi, L., O’Connor, M. F., Matot, I., Lerner, B., & Bitan, Y. (2019) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-019-00307-x · Cited by 34
McKie, P. M., Kor, D. J., Cook, D. A., Kessler, M., Carter, R. E., Wilson, P. M., Pencille, L. J., Hickey, B. C., & Chaudhry, R. (2019) · The American Journal of Medicine · doi:10.1016/j.amjmed.2019.10.039 · Cited by 32
A grounded theory study of alarm fatigue among nurses in intensive care units
Movahedi, A., Sadooghiasl, A., Ahmadi, F., & Vaismoradi, M. (2023) · Australian Critical Care · doi:10.1016/j.aucc.2022.12.004 · Cited by 42
Noise and sound in the intensive care unit: a cohort study
TAHVILI, A., Waite, A., Hampton, T., Welters, I. D., & Lee, P. J. (2025) · Scientific Reports · doi:10.1038/s41598-025-94365-8 · Cited by 35
“72 .”
Monitoring significant ST changes through deep learning
Xiao, R., Xu, Y. X., Pelter, M. M., Fidler, R., Badilini, F., MORTARA, D. W., & Hu, X. (2018) · Journal of Electrocardiology · doi:10.1016/j.jelectrocard.2018.07.026 · Cited by 31
“, and introducing a delay in monitoring algorithms can effectively reduce the number of alarms and mitigate alarm fatigue [3,20,21].”
BackgroundSelf-citationExplainability Metrics of Deep Convolutional Networks for Photoplethysmography Quality Assessment
Zhang, O. R., Ding, C., Pereira, T., Xiao, R., Gadhoumi, K., Meisel, K. M., Lee, R. J., Chen, Y., & Hu, X. (2021) · IEEE Access · doi:10.1109/access.2021.3054613 · Cited by 32
“Biomedical signals were acquired from 3764 patients in the intensive care unit (ICU) from 3/2013 to 12/2016 [32].”
BackgroundSelf-citationManaging Alarms in Acute Care Across the Life Span: Electrocardiography and Pulse Oximetry
(2018) · Critical Care Nurse · doi:10.4037/ccn2018468 · Cited by 36
Evaluating the Alarm Fatigue and its Associated Factors among Clinicians in Critical Care Units
Bourji, H., Sabbah, H., Al’Jamil, A., Khamis, R., Sabbah, S., Droubi, N., & Sabbah, I. (2020) · European Journal of Clinical Medicine · doi:10.24018/clinicmed.2020.1.1.8 · Cited by 36
“In hospital intensive care units, the monitor devices aid clinicians hugely in taking instantaneous actions [7].”
BackgroundInfluentialMosch, L. K., Poncette, A.-S., Spies, C., Weber‐Carstens, S., Schieler, M., Krampe, H., & Balzer, F. (2021) · JMIR Formative Research · doi:10.2196/22866 · Cited by 45
“It is unclear whether the aforementioned barriers and facilitators to digital health implementation can be transferred into the ICU context, given that it is a very specific setting: multiple professional groups work together, many different technologies are already in place, and staff stress levels are also high beca…”
BackgroundHarris, P. R. E. (2016) · Critical Care Nursing Clinics of North America · doi:10.1016/j.cnc.2016.04.002 · Cited by 28
Self-citationRespiratory rate monitoring from the photoplethysmogram via sparse signal reconstruction
Zhang, X., & Ding, Q. (2016) · Physiological Measurement · doi:10.1088/0967-3334/37/7/1105 · Cited by 27
Self-citationQuantifying the Impact of Infusion Alerts and Alarms on Nursing Workflows: A Retrospective Analysis
Yu, D., Obuseh, M., & DeLaurentis, P.-C. C. (2021) · Applied Clinical Informatics · doi:10.1055/s-0041-1730031 · Cited by 35
Alarm burden and the nursing care environment: a 213-hospital cross-sectional study
Ruppel, H., Dougherty, M., Bonafide, C. P., & Lasater, K. B. (2023) · BMJ Open Quality · doi:10.1136/bmjoq-2023-002342 · Cited by 35
“Few studies focus on the system-level factors associated with nurses’ experience of alarms.”
BackgroundBawua, L. K., Miaskowski, C. A., Hu, X., Rodway, G. W., & Pelter, M. M. (2021) · Annals of Noninvasive Electrocardiology · doi:10.1111/anec.12885 · Cited by 31
“An example of this limitation is found in a study that reported 161,931 unique RR type alarms (i.e., RR parameter high/low, or apnea) from adult patients in the intensive care unit that used IP in their bedside monitors (Drew et al., 2014).”
BackgroundMethodInfluentialSelf-citationGetting Better Hospital Alarm Sounds Into a Global Standard
Edworthy, J. R., McNeer, R. R., Bennett, C., Dudaryk, R., McDougall, S. J. P., Schlesinger, J. J., Bolton, M. L., Edworthy, J., Özcan, E., Boyd, A. D., et al. (2018) · Ergonomics in Design The Quarterly of Human Factors Applications · doi:10.1177/1064804618763268 · Cited by 29
“In clinical environments, the problem of bad alarm system implementation has reached colossal proportions, where patient deaths have been attributed to alarm fatigue (Drew et al., 2014; Sendelbach & Funk, 2013).”
Background
Metadata
- Authors
- Drew, B. J., Harris, P., Zègre-Hemsey, J. K., Mammone, T., Schindler, D., Salas-Boni, R., Bai, Y., Tinoco, A., Ding, Q., & Hu, X.
- Affiliations
- University of California, San Francisco (US)
- University of California San Francisco Medical Center (US)
- Corresponding author
- Drew, Barbara J.
- Venue
- PLoS ONE
- Publisher
- Public Library of Science (PLoS)
- Published
- 2014-10-22
- Volume
- 9
- Issue
- 10
- Pages
- e110274
- ISSN
- 1932-6203 (ISSN-L)
- PMC
- PMC4206416
- OpenAlex
- W2001741247
- Cited by
- 438
- References
- 23
- Citations by year
- 2026: 51 · 2025: 31 · 2024: 25 · 2023: 37 · 2022: 26 · 2021: 58 · 2020: 29 · 2019: 38 · 2018: 42 · 2017: 36 · 2016: 43 · 2015: 21 · 2014: 1
- FWCI
- 33.15
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Locations
- PLoS ONE (publishedVersion, cc-by)
- Europe PMC (PubMed Central) (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- Carolina Digital Repository (University of North Carolina at Chapel Hill) (submittedVersion)
- Dominican Scholar (Dominican University of California) (submittedVersion, cc-by)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
- eScholarship (California Digital Library) (submittedVersion)
- UNC Libraries
- Funders
- University of California, San Francisco
- Topics
- Health Sciences › Medicine › Surgery › Healthcare Technology and Patient Monitoring
- Health Sciences › Medicine › Public Health, Environmental and Occupational Health › Patient Safety and Medication Errors
- Health Sciences › Health Professions › Health Information Management › Electronic Health Records Systems
- Keywords
- alarm fatigue, intensive care unit patients, false positive alerts, alarm settings, ECG leads, QRS amplitude, atrial fibrillation, ventricular tachycardia, bundle branch block, persistent atrial fibrillation, patient safety
- MeSH terms
- Adult, Aged, Arrhythmias, Cardiac, Atrial Fibrillation, Diagnostic Errors, Electrocardiography, Equipment Failure, False Positive Reactions, Female, Humans, Intensive Care Units, Male, Middle Aged, Monitoring, Physiologic, Critical Illness, Clinical Alarms
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.