Lead-I ECG for detecting atrial fibrillation in patients attending primary care with an irregular pulse using single-time point testing: A systematic review and economic evaluation

Open access
Duarte, Rui, Stainthorpe, Angela, Mahon, James, Greenhalgh, Janette, Richardson, Marty, Nevitt, Sarah, Kotas, Eleanor, Boland, Angela +4 more·2019·PLOS ONE

Abstract

BACKGROUND: Atrial fibrillation (AF) is the most common type of cardiac arrhythmia and is associated with increased risk of stroke and congestive heart failure. Lead-I electrocardiogram (ECG) devices are handheld instruments that can detect AF at a single-time point. PURPOSE: To assess the diagnostic test accuracy, clinical impact and cost effectiveness of single-time point lead-I ECG devices compared with manual pulse palpation (MPP) followed by a 12-lead ECG for the detection of AF in symptomatic primary care patients with an irregular pulse. METHODS: Electronic databases (MEDLINE, MEDLINE Epub Ahead of Print and MEDLINE In-Process, EMBASE, PubMed and Cochrane Databases of Systematic Reviews, Cochrane Central Database of Controlled Trials, Database of Abstracts of Reviews of Effects, Health Technology Assessment Database) were searched to March 2018. Two reviewers screened the search results, extracted data and assessed study quality. Summary estimates of diagnostic accuracy were calculated using bivariate models. Cost-effectiveness was evaluated using an economic model consisting of a decision tree and two cohort Markov models. RESULTS: Diagnostic accuracy The diagnostic accuracy (13 publications reporting on nine studies) and clinical impact (24 publications reporting on 19 studies) results are derived from an asymptomatic population (used as a proxy for people with signs or symptoms of AF). The summary sensitivity of lead-I ECG devices was 93.9% (95% confidence interval [CI]: 86.2% to 97.4%) and summary specificity was 96.5% (95% CI: 90.4% to 98.8%). Cost effectiveness The de novo economic model yielded incremental cost effectiveness ratios (ICERs) per quality adjusted life year (QALY) gained. The results of the pairwise analysis show that all lead-I ECG devices generate ICERs per QALY gained below the £20,000-£30,000 threshold. Kardia Mobile is the most cost effective option in a full incremental analysis. Lead-I ECG tests may identify more AF cases than the standard diagnostic pathway. This comes at a higher cost but with greater patient benefit in terms of mortality and quality of life. LIMITATIONS: No published data evaluating the diagnostic accuracy, clinical impact or cost effectiveness of lead-I ECG devices for the target population are available. CONCLUSIONS: The use of single-time point lead-I ECG devices in primary care for the detection of AF in people with signs or symptoms of AF and an irregular pulse appears to be a cost effective use of NHS resources compared with MPP followed by a 12-lead ECG, given the assumptions used in the base case model. REGISTRATION: The protocol for this review is registered on PROSPERO as CRD42018090375.

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.1371/journal.pone.0226671

Cites

8 of 8 references verified0 citation errors · 0 not checkable

Statistics

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Cites

Flagged first, then by citations

Research works printed but not linked: 38

  1. [2]Kirchhof P, Benussi S, Kotecha D, Ahlsson A, Atar D, Casadei B, et al 2016 ESC guidelines for the management of atrial fibrillation developed in collaboration with EACTS. Eur Heart J. 2016; 37: 2893– 62. 10.1093/eurheartj/ehw210 27567408 · DOI · Paper page
  2. [5]National Institute for Health and Care Excellence (NICE). Atrial fibrillation: management. clinical guideline CG180. 2014. Available from: https://www.nice.org.uk/guidance/cg180/chapter/Introduction [accessed January 2018]. 31841281
  3. [7]National Institute for Health and Care Excellence (NICE). Diagnostic assessment programme manual [Internet] 2011. Available from: http://www.nice.org.uk/media/A0B/97/DAPManualFINAL.pdf [accessed January 2018]. 27466648
  4. [9]McInnes MDF, Moher D, Thombs BD, McGrath TA, Bossuyt PM, Clifford T, et al Preferred Reporting Items for a Systematic Review and Meta-analysis of Diagnostic Test Accuracy Studies: The PRISMA-DTA Statement. Jama. 2018; 319(4): 388– 96. Epub 2018/01/25. 10.1001/jama.2017.19163. 29362800 · DOI · Paper page
  5. [10]Whiting PF, Rutjes AW, Westwood ME, Mallett S, Deeks JJ, Reitsma JB, et al QUADAS-2: a revised tool for the quality assessment of diagnostic accuracy studies. Annals of Internal Medicine. 2011; 155: 529– 36. 10.7326/0003-4819-155-8-201110180-00009 22007046 · DOI · Paper page
  6. [13]Chu H, Cole SR. Bivariate meta-analysis of sensitivity and specificity with sparse data: a generalized linear mixed model approach. J Clin Epidemiol. 2006; 59 (12): 1331– 2; author reply 2–3. Epub 2006/11/14. 10.1016/j.jclinepi.2006.06.011. 17098577 · DOI
  7. [14]Takwoingi Y, Guo B, Riley RD, Deeks JJ. Performance of methods for meta-analysis of diagnostic test accuracy with few studies or sparse data. Stat Methods Med Res. 2017; 26(4): 1896– 911. Epub 2015/06/28. 10.1177/0962280215592269 26116616 PMC5564999 · DOI
  8. [15]Adderley NJ, Ryan R, Nirantharakumar K, Marshall T. Prevalence and treatment of atrial fibrillation in UK general practice from 2000 to 2016. Heart. 2018 10.1136/heartjnl-2018-312977 29991504 · DOI
  9. [16]Piccini JP, Simon DN, Steinberg BA, et al Differences in clinical and functional outcomes of atrial fibrillation in women and men: two-year results from the orbit-af registry. JAMA Cardiology. 2016; 1(3): 282– 91. 10.1001/jamacardio.2016.0529 27438106 · DOI
  10. [17]Welton NJ, McAleenan A, Thom HH, Davies P, Hollingworth W, Higgins JP, et al Screening strategies for atrial fibrillation: a systematic review and cost-effectiveness analysis. Health Technol Assess. 2017; 21(29): 1– 236. Epub 2017/06/21. 10.3310/hta21290. 28629510 · DOI
  11. [18]Engdahl J, Andersson L, Mirskaya M, Rosenqvist M. Stepwise screening of atrial fibrillation in a 75-year-old population: implications for stroke prevention. Circulation. 2013; 127(8): 930– 7. Epub 2013/01/25. 10.1161/CIRCULATIONAHA.112.126656. 23343564 · DOI · Paper page
  12. [22]Mathisen SM, Dalen I, Larsen JP, Kurz M. Long-Term Mortality and Its Risk Factors in Stroke Survivors. Journal of stroke and cerebrovascular diseases: the official journal of National Stroke Association. 2016; 25(3): 635– 41. Epub 2016/01/08. 10.1016/j.jstrokecerebrovasdis.2015.11.039. 26738815 · DOI
  13. [23]Berg J, Lindgren P, Nieuwlaat R, Bouin O, Crijns H. Factors determining utility measured with the EQ-5D in patients with atrial fibrillation. Quality of life research: an international journal of quality of life aspects of treatment, care and rehabilitation. 2010; 19(3): 381– 90. Epub 2010/01/29. 10.1007/s11136-010-9591-y. 20108048 · DOI
  14. [24]The EuroQol Group. EuroQol-a new facility for the measurement of health-related quality of life. Health Policy. 1990; 16(3): 199– 208. Epub 1990/11/05. 10.1016/0168-8510(90)90421-9. 10109801 · DOI · Paper page
  15. [25]Moher D, Liberati A, Tetzlaff J, Altman DG. Preferred reporting items for systematic reviews and meta-analyses: the PRISMA statement. BMJ. 2009; 339 10.1136/bmj.b2535 PMC3090117 21603045 · DOI · Paper page
  16. [26]Desteghe L, Raymaekers Z, Vijgen J, Dilling-Boer D, Koopman P, Schurmans J, et al Accuracy and cost-effectiveness of two handheld electrocardiogram recorders to screen for atrial fibrillation in a hospital setting. Eur Heart J. 2016; 37 (Supplement 1): 1265 10.1093/eurheartj/ehw434.. · DOI
  17. [27]Desteghe L, Raymaekers Z, Vijgen J, Dilling-Boer D, Koopman P, Schurmans J, et al Accuracy and usability of handheld electrocardiogram recorders to detectatrial fibrillation in hospitalised patients. Europace. 2016; 18 (Supplement 1): i177 10.1093/europace/euw158.. · DOI
  18. [28]Desteghe L, Raymaekers Z, Lutin M, Vijgen J, Dilling-Boer D, Koopman P, et al Performance of handheld electrocardiogram devices to detect atrial fibrillation in a cardiology and geriatric ward setting. Europace. 2017; 19(1): 29– 39. 10.1093/europace/euw025. 26893496 · DOI
  19. [29]Doliwa PS, Frykman V, Rosenqvist M. Short-term ECG for out of hospital detection of silent atrial fibrillation episodes. Scand Cardiovasc J. 2009; 43(3): 163– 8. Epub 2008/12/20. 10.1080/14017430802593435. 19096977 · DOI · Paper page
  20. [30]Haberman ZC, Jahn RT, Bose R, Tun H, Shinbane JS, Doshi RN, et al Wireless smart phone equipped ECG enables large scale screening in diverse populations. Heart Rhythm. 2014; 1): S312 10.1016/j.hrthm.2014.03.031.. 25651872 · DOI
  21. [32]Lau J, Lowres N, Neubeck L, Brieger DB, Sy RW, Galloway C, et al Validation of an iphone ECG application suitable for community screening for silent atrial fibrillation: a novel way to prevent stroke. Circulation Conference: American Heart Association. 2012; 126(21 SUPPL. 1)..
  22. [36]Williams J, Pearce K, Benett I. The effectiveness of a mobile ECG device in identifying AF: sensitivity, specificity and predictive value. Br J Cardiol. 2015; 22: 70– 2.
  23. [37]Crockford CJ, Ahmed O, Kaba R, Berry R. An analysis of the applicability of lead1 screening for cardiac arrhythmia in primary care settings using novel sensing technology & multiple commercial algorithms for automating detection to increase PPV of referrals for further investigation. Europace. 2013; 4): iv20 10.1093/europace/eut317.. · DOI
  24. [39]Battipaglia I, Gilbert K, Hogarth AJ, Tayebjee MH. Screening for atrial fibrillation in the community using a novel ECG recorder. J Atr Fibrillation. 2016; 9(2): 1433 Epub 2016/12/03. 10.4022/jafib.1433 27909536 PMC5129689 · DOI
  25. [40]Orchard J, Freedman SB, Lowres N, Peiris D, Neubeck L. iPhone ECG screening by practice nurses and receptionists for atrial fibrillation in general practice: the GP-SEARCH qualitative pilot study. Australian family physician. 2014; 43(5): 315– 9. 24791776
  26. [41]Chan LL, Chan SC, Yan BP. Feasibility and acceptability of atrial fibrillation screening using a hand-held ECG device in general practice setting in Hong Kong. Value in Health. 2017; 20 (9): A599 10.1016/j.jval.2017.08.1136.. · DOI
  27. [42]Chan NY, Choy CC. Community screening for atrial fibrillation in a Chinese population using a smartphone-based wireless single-lead ECG. J Am Coll Cardiol. 2015; 1: A467.
  28. [43]Chan NY, Choy CC. Screening for atrial fibrillation in 13 122 Hong Kong citizens with smartphone electrocardiogram. Heart. 2016a; 12 10.1136/heartjnl-2016-309993.. 27733533 · DOI · Paper page
  29. [45]Gibson J, Hanjari M, Watkins C, Chauhan U. Opportunistic detection of atrial fibrillation in primary care: a mixed methods evaluation of the introduction of new healthcare technology. Eur Stroke J. 2017; 2 (1 Supplement 1):112. 10.1177/2396987317705242.. · DOI
  30. [46]Hussain W, Thakrar D. The use of a handheld device in identifying atrial fibrillation patients during flu vaccination clinics. Europace. 2016; 18 (Supplement 2): ii19 10.1093/europace/euw268.. · DOI
  31. [47]Kaasenbrood F, Hollander M, Rutten FH, Gerhards LJ, Hoes AW, Tieleman RG. Yield of screening for atrial fibrillation in primary care with a hand-held, single-lead electrocardiogram device during influenza vaccination. Europace. 2016; 18(10): 1514– 20. 10.1093/europace/euv426. 26851813 PMC5072135 · DOI
  32. [48]Lowres N, Neubeck L, Salkeld G, Krass I, McLachlan AJ, Redfern J, et al Feasibility and cost-effectiveness of stroke prevention through community screening for atrial fibrillation using iPhone ECG in pharmacies. the SEARCH-AF study. Thromb Haemost. 2014; 111(6): 1167– 76. Epub 2014/04/02. 10.1160/TH14-03-0231. 24687081 · DOI · Paper page
  33. [49]Orchard J, Lowres N, Freedman SB, Ladak L, Lee W, Zwar N, et al Screening for atrial fibrillation during influenza vaccinations by primary care nurses using a smartphone electrocardiograph (iECG): a feasibility study. Eur J Prev Cardiol. 2016; 23(2_suppl): 13– 20. 10.1177/2047487316670255.. 27892421 · DOI
  34. [51]Waring O, Davidson N, Stout M, Pearce K. Detection of atrial fibrillation in community locations using novel technology's as a method of stroke prevention in the over 65's asymptomatic population—should it become standard practise? Europace Conference: heart rhythm congress 2016 United kingdom [Internet]. 2016; 18:[ii39 p.]. Available from: http://onlinelibrary.wiley.com/o/cochrane/clcentral/articles/744/CN-01406744/frame.html [Accessed June 2018].
  35. [52]Yan BPY, Chan LLY, Lee VWY, Freedman B. Medical outpatient clinics an ideal setting for atrial fibrillation screening using a handheld single-lead ECG with automated diagnosis. Eur Heart J. 2016; 37 (Supplement 1): 888 10.1093/eurheartj/ehw433.. · DOI
  36. [53]Duarte R, Stainthorpe A, Greenhalgh J, Richardson M, Nevitt S, Mahon J, et al Lead-I ECG for detecting atrial fibrillation in patients with an irregular pulse using single time point testing: a systematic review and economic evaluation. Health Technol Assess. [in press]. PMC6983912 31933471
  37. [54]Taggar JS, Coleman T, Lewis S, Heneghan C, Jones M. Accuracy of methods for detecting an irregular pulse and suspected atrial fibrillation: A systematic review and meta-analysis. Eur J Prev Cardiol. 2016; 23(12): 1330– 8. Epub 2015/10/16. 10.1177/2047487315611347 26464292 PMC4952027 · DOI
  38. [55]Jacobs MS, Kaasenbrood F, Postma MJ, Van Hulst M, Tieleman RG. Cost-effectiveness of screening for atrial fibrillation in primary care with a handheld, single-lead electrocardiogram device in the Netherlands. Europace. 2018; 20(1): 12– 8. 10.1093/europace/euw285. 27733465 · DOI

Cited by

Sorted by citations

Authors
Duarte, R., Stainthorpe, A., Mahon, J., Greenhalgh, J., Richardson, M., Nevitt, S., Kotas, E., Boland, A., Thom, H., Marshall, T., Hall, M., & Takwoingi, Y.
Affiliations
  • University of Liverpool (GB)
  • University of Bristol (GB)
  • University of Birmingham (GB)
  • Liverpool Heart and Chest Hospital (GB)
  • University Hospitals Birmingham NHS Foundation Trust (GB)
Corresponding author
Duarte, Rui
Venue
PLOS ONE
Published
2019-12-23
Volume
14
Issue
12
Pages
e0226671
ISSN
1932-6203 (ISSN-L)
OpenAlex
W2996595048
Cited by
28
References
8
Citations by year
2026: 3 · 2025: 3 · 2024: 7 · 2023: 5 · 2022: 3 · 2021: 4 · 2020: 3
FWCI
0.45
Type
journal article
Language
en
Open access
Locations
  • PLoS ONE (publishedVersion, cc-by)
  • Europe PMC (PubMed Central) (publishedVersion, cc-by)
  • PubMed Central (publishedVersion, cc-by)
  • PubMed (publishedVersion)
  • University of Birmingham Research Portal (University of Birmingham) (submittedVersion, cc-by)
  • RePEc: Research Papers in Economics (submittedVersion)
  • DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
  • Figshare (submittedVersion, cc-by)
  • Explore Bristol Research (submittedVersion, cc-by)
Funders
Health Technology Assessment Programme (16/30/05)
Topics
  • Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Atrial Fibrillation Management and Outcomes
  • Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › ECG Monitoring and Analysis
  • Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Cardiac pacing and defibrillation studies
Keywords
atrial fibrillation, irregular pulse, pulse palpation, 12-lead ECG, diagnostic accuracy, cost-effectiveness, primary care, quality-adjusted life years, arrhythmia detection, stroke risk, mortality, quality of life, decision tree, Markov model, economic evaluation, sensitivity and specificity
MeSH terms
Aged, Aged, 80 and over, Atrial Fibrillation, Cost-Benefit Analysis, Electrocardiography, Female, Humans, Male, Markov Chains, Middle Aged, Primary Health Care, Pulse

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review