Validation of four automatic devices for self-measurement of blood pressure according to the international protocol of the European Society of Hypertension
Open accessThe DOI matches its Crossref record; 6 of the 6 references it cites were found in a registry or an index.
Abstract
BACKGROUND: Four electronic devices for self-measurement of brachial blood pressure (BP): the Omron M1 Plus, the Omron M6 Comfort, the Spengler KP7500 D, and the Microlife BP A100 Plus, were evaluated in four separate studies according to the International Protocol of the European Society of Hypertension (ESH). DESIGN: The International Validation Protocol is divided into 2 phases: the first phase is performed on 15 selected subjects (45 pairs of BP measurements); if the device passes this phase, 18 supplementary subjects are included (54 pairs of BP measurements) making a total number of 33 subjects (99 pairs of BP measurements) on which the final validation is performed. METHODS: The same methodology recommended by the ESH protocol was applied for the 4 studies. In each study and for each subject, 4 BP measurements were performed simultaneously by 2 trained observers using mercury sphygmomanometers alternately with 3 measurements by the tested device. The difference between the BP value given by the device and that obtained by the two observers (mean of the two observers) was calculated for each measure. The 99 pairs of BP differences were classified into 3 categories (< or =5, 10, < or =15 mmHg). The number of differences in each category was compared with the number required by the International Protocol. An individual analysis was then done to determine for each subject the number of comparisons < or =5 mmHg. At least 22 of the 33 subjects should have 2 of their 3 comparisons < or =5 mmHg. RESULTS: All 4 tested devices passed the first and the second phase of the validation process. The average differences between the device and mercury sphygmomanometer readings were -1.4 +/- 5.5 and -0.4 +/- 4.8 mmHg for SBP and DBP respectively for the Omron M1 Plus device, -2.1 +/- 7.4 and 0.1 +/- 4.9 mmHg for SBP and DBP respectively for the Omron M6 Comfort device, -1.4 +/- 8.6 and -0.1 +/- 3.5 mmHg for SBP and DBP respectively for the Spengler KP7500 D device, and 1.6 +/- 4.2 mmHg and 0.54 +/- 2.8 mmHg for SBP and DBP respectively for the Microlife BP A100 Plus device. For all devices, readings differing by less than 5, 10, and 15 mmHg for SBP and DBP values fulfill the recommendation criteria of the International Protocol as well as the individual analysis. CONCLUSIONS: Omron M1 Plus (HEM-401 1C-E), Omron M6 Comfort (HEM 7000-E), Spengler KP7500 D, and Microlife BP A100 Plus devices fulfilled the validation recommendations of the International Protocol.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Chobanian, A. V., Bakris, G. L., Black, H. R., Cushman, W. C., Green, L. A., Izzo, J. L., Jones, D. W., Materson, B. J., Oparil, S., Wright, J. T., et al. (2003) · Hypertension · doi:10.1161/01.hyp.0000107251.49515.c2 · Cited by 13,271
Pickering, T. G., Hall, J. E., Appel, L. J., Falkner, B. E., Graves, J., Hill, M. N., Jones, D. W., Kurtz, T., Sheps, S. G., & Roccella, E. J. (2005) · Circulation · doi:10.1161/01.cir.0000154900.76284.f6 · Cited by 3,985
Pickering, T. G., Hall, J. E., Appel, L. J., Falkner, B. E., Graves, J., Hill, M. N., Jones, D. W., Kurtz, T., Sheps, S. G., & Roccella, E. J. (2005) · Hypertension · doi:10.1161/01.hyp.0000150859.47929.8e · Cited by 2,051
Parati, G., Stergiou, G. S., Asmar, R., Bilo, G., de Leeuw, P., Imai, Y., Kario, K., Lurbe, E., Manolis, A., Mengden, T., et al. (2008) · Journal of Hypertension · doi:10.1097/hjh.0b013e328308da66 · Cited by 837
Self-citationCharacterization of the oscillometric method for measuring indirect blood pressure
Geddes, L. A., Voelz, M., Combs, C., Reiner, D., & Babbs, C. F. (1982) · Annals of Biomedical Engineering · doi:10.1007/bf02367308 · Cited by 340
Reference Values for Self-recorded Blood Pressure: A Meta-analysis of Summary Data
Thijs, L., Staessen, J. A., Celis, H., de Gaudemaris, R., Imai, Y., Julius, S., & Fagard, R. (1998) · Archives of Internal Medicine · doi:10.1001/archinte.158.5.481 · Cited by 143
Cited by
McManus, R. J., Mant, J., Franssen, M., Nickless, A., Schwartz, C., Hodgkinson, J., Bradburn, P., Farmer, A., Grant, S., Greenfield, S. M., et al. (2018) · The Lancet · doi:10.1016/s0140-6736(18)30309-x · Cited by 419
Glynn, L. G., Hayes, P. S., Casey, M., Glynn, F., Alvarez-Iglesias, A., Newell, J., ÓLaighin, G., Heaney, D., O’Donnell, M., & Murphy, A. W. (2014) · British Journal of General Practice · doi:10.3399/bjgp14x680461 · Cited by 280
How consumer physical activity monitors could transform human physiology research
Wright, S. P., Hall Brown, T. S., Collier, S. R., & Sandberg, K. (2017) · American Journal of Physiology-Regulatory, Integrative and Comparative Physiology · doi:10.1152/ajpregu.00349.2016 · Cited by 234
Automated blood pressure measurement in atrial fibrillation: a systematic review and meta-analysis
Stergiou, G. S., Kollias, A., Destounis, A., & Tzamouranis, D. (2012) · Journal of Hypertension · doi:10.1097/hjh.0b013e32835850d7 · Cited by 120
Picone, D. S., Deshpande, R. A., Schultz, M. G., Fonseca, R., Campbell, N. R. C., Delles, C., Hecht Olsen, M., Schutte, A. E., Stergiou, G., Padwal, R., et al. (2020) · Hypertension · doi:10.1161/hypertensionaha.120.14719 · Cited by 118
Noninvasive Continuous Monitoring of Vital Signs With Wearables: Fit for Medical Use?
Jacobsen, M., Dembek, T. A., Kobbe, G., Gaidzik, P. W., & Heinemann, L. (2021) · Journal of Diabetes Science and Technology · doi:10.1177/1932296820904947 · Cited by 49
Simonsson, E., Levik Sandström, S., Hedlund, M., Holmberg, H., Johansson, B., Lindelöf, N., Boraxbekk, C.-J., & Rosendahl, E. (2023) · The Journals of Gerontology: Series A · doi:10.1093/gerona/glad070 · Cited by 35
Mapesi, H., Gupta, R., Wilson, H. I., Lukau, B., Amstutz, A., Lyimo, A., Muhairwe, J., Senkoro, E., Byakuzana, T., Mphunyane, M., et al. (2021) · Trials · doi:10.1186/s13063-021-05023-z · Cited by 8
Metadata
- Authors
- Asmar, R.
- Venue
- Vascular Health and Risk Management
- Publisher
- Informa UK Limited
- Published
- 2011-11
- Volume
- 3
- Issue
- 4
- Pages
- 709
- ISSN
- 1176-6344 (ISSN-L)
- 1178-2048
- OpenAlex
- W1546895422
- Cited by
- 163
- References
- 6
- FWCI
- 4.72
- Type
- journal article
- Language
- en
- Open access
- gold
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.