Accuracy of Wristband Fitbit Models in Assessing Sleep: Systematic Review and Meta-Analysis

Open access

Published version of a preprint

The DOI matches its Crossref record; 45 of the 46 references it cites were found in a registry or an index.

Abstract

Background: Wearable sleep monitors are of high interest to consumers and researchers because of their ability to provide estimation of sleep patterns in free-living conditions in a cost-efficient way. Objective: We conducted a systematic review of publications reporting on the performance of wristband Fitbit models in assessing sleep parameters and stages. Methods: In adherence with the Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA) statement, we comprehensively searched the Cumulative Index to Nursing and Allied Health Literature (CINAHL), Cochrane, Embase, MEDLINE, PubMed, PsycINFO, and Web of Science databases using the keyword Fitbit to identify relevant publications meeting predefined inclusion and exclusion criteria. Results: The search yielded 3085 candidate articles. After eliminating duplicates and in compliance with inclusion and exclusion criteria, 22 articles qualified for systematic review, with 8 providing quantitative data for meta-analysis. In reference to polysomnography (PSG), nonsleep-staging Fitbit models tended to overestimate total sleep time (TST; range from approximately 7 to 67 mins; effect size=-0.51, P<.001; heterogenicity: I2=8.8%, P=.36) and sleep efficiency (SE; range from approximately 2% to 15%; effect size=-0.74, P<.001; heterogenicity: I2=24.0%, P=.25), and underestimate wake after sleep onset (WASO; range from approximately 6 to 44 mins; effect size=0.60, P<.001; heterogenicity: I2=0%, P=.92) and there was no significant difference in sleep onset latency (SOL; P=.37; heterogenicity: I2=0%, P=.92). In reference to PSG, nonsleep-staging Fitbit models correctly identified sleep epochs with accuracy values between 0.81 and 0.91, sensitivity values between 0.87 and 0.99, and specificity values between 0.10 and 0.52. Recent-generation Fitbit models that collectively utilize heart rate variability and body movement to assess sleep stages performed better than early-generation nonsleep-staging ones that utilize only body movement. Sleep-staging Fitbit models, in comparison to PSG, showed no significant difference in measured values of WASO (P=.25; heterogenicity: I2=0%, P=.92), TST (P=.29; heterogenicity: I2=0%, P=.98), and SE (P=.19) but they underestimated SOL (P=.03; heterogenicity: I2=0%, P=.66). Sleep-staging Fitbit models showed higher sensitivity (0.95-0.96) and specificity (0.58-0.69) values in detecting sleep epochs than nonsleep-staging models and those reported in the literature for regular wrist actigraphy. Conclusions: Sleep-staging Fitbit models showed promising performance, especially in differentiating wake from sleep. However, although these models are a convenient and economical means for consumers to obtain gross estimates of sleep parameters and time spent in sleep stages, they are of limited specificity and are not a substitute for PSG.

Overview

Integrity signals

Registry

Crossref · 10.2196/16273

Cites

45 of 45 references verified0 citation errors · 0 not checkable

Statistics

Recomputing p-values needs the paper’s full text.

Figures

Figure 1 Flow diagram adapted from Moher et al [14] describing the search strategy of databases to retrieve and qualify publications of relevance for review.
Figure 1 Flow diagram adapted from Moher et al [14] describing the search strategy of databases to retrieve and qualify publications of relevance for review.CC BY · PubMed Central
Figure 2 Forest plot of the standardized mean difference (Hedges g) between Fitbit and polysomnography for the variable of sleep onset latency (SOL). Results are shown as effect size (ES) and 95% CI. The difference in symbol size indicates the difference in weight of the respective studies. The diamond symbol shows the 95% CI of the overall effect and the tails show the 95% prediction interval of the overall effect. PLMS: periodic limb movement in sleep.
Figure 2 Forest plot of the standardized mean difference (Hedges g) between Fitbit and polysomnography for the variable of sleep onset latency (SOL). Results are shown as effect size (ES) and 95% CI. The difference in symbol size indicates the difference in weight of the respective studies. The diamond symbol shows the 95% CI of the overall effect and the tails show the 95% prediction interval of the overall effect. PLMS: periodic limb movement in sleep.CC BY · PubMed Central
Figure 3 Forest plot of the standardized mean difference (Hedges g) between Fitbit and polysomnography for the variable of wake after sleep onset (WASO). Results are shown as effect size (ES) and 95% CI. The difference in symbol size indicates the difference in the weight of the respective studies. The diamond symbol shows the 95% CI of the overall effect and the tails show the 95% prediction interval of the overall effect. PLMS: periodic limb movement in sleep.
Figure 3 Forest plot of the standardized mean difference (Hedges g) between Fitbit and polysomnography for the variable of wake after sleep onset (WASO). Results are shown as effect size (ES) and 95% CI. The difference in symbol size indicates the difference in the weight of the respective studies. The diamond symbol shows the 95% CI of the overall effect and the tails show the 95% prediction interval of the overall effect. PLMS: periodic limb movement in sleep.CC BY · PubMed Central
Figure 4 Forest plot of the standardized mean difference (Hedges g) between Fitbit and polysomnography for the variable of total sleep time (TST). Results are shown as effect size (ES) and 95% CI. The difference in symbol size indicates the difference in weight of the respective studies. The diamond symbol shows the 95% CI of the overall effect and the tails show the 95% prediction interval of the overall effect. PLMS: periodic limb movement in sleep.
Figure 4 Forest plot of the standardized mean difference (Hedges g) between Fitbit and polysomnography for the variable of total sleep time (TST). Results are shown as effect size (ES) and 95% CI. The difference in symbol size indicates the difference in weight of the respective studies. The diamond symbol shows the 95% CI of the overall effect and the tails show the 95% prediction interval of the overall effect. PLMS: periodic limb movement in sleep.CC BY · PubMed Central
Figure 5 Forest plot of the standardized mean difference (Hedges g) between Fitbit and polysomnography for the variable of sleep efficiency (SE). Results are shown as effect size (ES) and 95% CI. The difference in symbol size indicates the difference in weight of the respective studies. The diamond symbol shows the 95% CI of the overall effect and the tails show the 95% prediction interval of the overall effect.
Figure 5 Forest plot of the standardized mean difference (Hedges g) between Fitbit and polysomnography for the variable of sleep efficiency (SE). Results are shown as effect size (ES) and 95% CI. The difference in symbol size indicates the difference in weight of the respective studies. The diamond symbol shows the 95% CI of the overall effect and the tails show the 95% prediction interval of the overall effect.CC BY · PubMed Central

Verification

Claims the paper makes

We couldn’t retrieve this paper’s full text, so its claims and self-check weren’t run. The registry and citation checks still apply.

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

Citations

Cites

Show all 45 references

Research works printed but not linked: 2

  1. Wright, SP, 2017, FASEB J, 31, 1
  2. CoeRIt's the effect size, stupid: What effect size is and why it is importantProceedings of the Annual Conference of the British Educational Research Association2002Annual Conference of the British Educational Research AssociationSeptember 12-14, 2002University of Exeter, UK

Cited by

See all 470

Metadata

Authors
Haghayegh, S., Khoshnevis, S., Smolensky, M. H., Diller, K. R., & Castriotta, R. J.
Affiliations
  • The University of Texas at Austin (US)
  • University of Southern California (US)
Corresponding author
Haghayegh, Shahab
Venue
Journal of Medical Internet Research
Published
2019-11-28
Volume
21
Issue
11
Pages
e16273
ISSN
1438-8871 (ISSN-L)
OpenAlex
W2980985183
Cited by
446
References
46
Citations by year
2026: 70 · 2025: 73 · 2024: 94 · 2023: 66 · 2022: 63 · 2021: 51 · 2020: 26 · 2019: 3
FWCI
5.28
Type
journal article
Language
en
Open access
Licence
cc-by
Locations
  • Journal of Medical Internet Research (publishedVersion, cc-by)
  • Europe PMC (PubMed Central) (publishedVersion, cc-by)
  • PubMed Central (publishedVersion, cc-by)
  • PubMed (publishedVersion)
  • DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
Topics
  • Social Sciences › Psychology › Experimental and Cognitive Psychology › Sleep and related disorders
  • Physical Sciences › Computer Science › Computer Vision and Pattern Recognition › Context-Aware Activity Recognition Systems
  • Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
Keywords
Fitbit, sleep stage estimation, total sleep time, sleep efficiency, wake after sleep onset, sleep latency, polysomnography comparison, heart rate variability, human motion detection, device accuracy, consumer health technology
MeSH terms
Female, Humans, Male, Sleep, Wrist, Actigraphy

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.