Clinical Validation of Heart Rate Apps: Mixed-Methods Evaluation Study
Open accessThe DOI matches its Crossref record; 6 of the 6 references it cites were found in a registry or an index.
Abstract
BACKGROUND: Photoplethysmography (PPG) is a proven way to measure heart rate (HR). This technology is already available in smartphones, which allows measuring HR only by using the smartphone. Given the widespread availability of smartphones, this creates a scalable way to enable mobile HR monitoring. An essential precondition is that these technologies are as reliable and accurate as the current clinical (gold) standards. At this moment, there is no consensus on a gold standard method for the validation of HR apps. This results in different validation processes that do not always reflect the veracious outcome of comparison. OBJECTIVE: The aim of this paper was to investigate and describe the necessary elements in validating and comparing HR apps versus standard technology. METHODS: The FibriCheck (Qompium) app was used in two separate prospective nonrandomized studies. In the first study, the HR of the FibriCheck app was consecutively compared with 2 different Food and Drug Administration (FDA)-cleared HR devices: the Nonin oximeter and the AliveCor Mobile ECG. In the second study, a next step in validation was performed by comparing the beat-to-beat intervals of the FibriCheck app to a synchronized ECG recording. RESULTS: In the first study, the HR (BPM, beats per minute) of 88 random subjects consecutively measured with the 3 devices showed a correlation coefficient of .834 between FibriCheck and Nonin, .88 between FibriCheck and AliveCor, and .897 between Nonin and AliveCor. A single way analysis of variance (ANOVA; P=.61 was executed to test the hypothesis that there were no significant differences between the HRs as measured by the 3 devices. In the second study, 20,298 (ms) R-R intervals (RRI)-peak-to-peak intervals (PPI) from 229 subjects were analyzed. This resulted in a positive correlation (rs=.993, root mean square deviation [RMSE]=23.04 ms, and normalized root mean square error [NRMSE]=0.012) between the PPI from FibriCheck and the RRI from the wearable ECG. There was no significant difference (P=.92) between these intervals. CONCLUSIONS: Our findings suggest that the most suitable method for the validation of an HR app is a simultaneous measurement of the HR by the smartphone app and an ECG system, compared on the basis of beat-to-beat analysis. This approach could lead to more correct assessments of the accuracy of HR apps.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
The impact of the MIT-BIH Arrhythmia Database
Moody, G. B., & Mark, R. G. (2001) · IEEE Engineering in Medicine and Biology Magazine · doi:10.1109/51.932724 · Cited by 4,812
On the Analysis of Fingertip Photoplethysmogram Signals
Elgendi, M. (2012) · Current Cardiology Reviews · doi:10.2174/157340312801215782 · Cited by 1,206 · cited 2×
“The amount of backscattered light corresponds with the variation in blood volume [10].”
Mobile Health Technology Evaluation
Kumar, S., Nilsen, W. J., Abernethy, A., Atienza, A., Patrick, K., Pavel, M., Riley, W. T., Shar, A., Spring, B., Spruijt-Metz, D., et al. (2013) · American Journal of Preventive Medicine · doi:10.1016/j.amepre.2013.03.017 · Cited by 900 · cited 1×
“This enables the use of a smartphone as a wireless HR monitor [3].”
Kroll, R. R., Boyd, J. G., & Maslove, D. M. (2016) · Journal of Medical Internet Research · doi:10.2196/jmir.6025 · Cited by 123 · cited 2×
“Validation can be done in two ways: (1) by comparing the HR [4,14] or (2) by comparing the ECG-derived R-R intervals (RRI) [15,16] and the PPG-derived peak-to-peak intervals (PPI) [17] as shown in Figure 3.”
Friberg, L., Engdahl, J., Frykman, V., Svennberg, E., Levin, L.-Å., & Rosenqvist, M. (2013) · EP Europace · doi:10.1093/europace/eus217 · Cited by 100 · cited 1×
“PPG is already used in the clinic to measure oxygen saturation and pulse rate [7].”
Lee, Q. Y., Redmond, S. J., Chan, G. S., Middleton, P. M., Steel, E., Malouf, P., Critoph, C., Flynn, G., O’Lone, E., & Lovell, N. H. (2013) · BioMedical Engineering OnLine · doi:10.1186/1475-925x-12-19 · Cited by 34
Cited by
Wearables and the Medical Revolution
Dunn, J., Runge, R., & Snyder, M. (2018) · Personalized Medicine · doi:10.2217/pme-2018-0044 · Cited by 654
Self-powered cardiovascular electronic devices and systems
Zheng, Q., Tang, Q., Wang, Z. L., & Li, Z. (2021) · Nature Reviews Cardiology · doi:10.1038/s41569-020-0426-4 · Cited by 348
Béres, S., Holczer, L., & Hejjel, L. (2019) · Measurement Science Review · doi:10.2478/msr-2019-0030 · Cited by 33
Metadata
- Authors
- Vandenberk, T., Stans, J., Mortelmans, C., Van Haelst, R., Van Schelvergem, G., Pelckmans, C., Smeets, C. J., Lanssens, D., De Cannière, H., Storms, V., Thijs, I. M., Vaes, B., & Vandervoort, P. M.
- Venue
- JMIR mHealth and uHealth
- Publisher
- JMIR Publications Inc.
- Published
- 2017-08-25
- Volume
- 5
- Issue
- 8
- Pages
- e129
- ISSN
- 2291-5222 (ISSN-L)
- OpenAlex
- W2747585374
- Cited by
- 58
- References
- 6
- FWCI
- 2.44
- Type
- journal article
- Language
- en
- Open access
- gold
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.