Evaluation and Optimization of a New PET Reconstruction Algorithm, Bayesian Penalized Likelihood Reconstruction, for Lung Cancer Assessment According to Lesion Size
Abstract
OBJECTIVE. The purpose of this study was to characterize the Bayesian penalized likelihood (BPL) reconstruction algorithm in comparison with an ordered subset expectation maximization (OSEM) reconstruction algorithm and to determine its optimal penalization factor (expressed as a beta value) for clinical use. MATERIALS AND METHODS. FDG PET/CT scans of 46 patients with lung cancer were reconstructed using OSEM and BPL with beta values of 200, 300, 400, 500, and 1000. The liver signal-to-noise ratio, mean standardized uptake value (SUV mean ) of the liver, and maximum standardized uptake value (SUV max ) and SUV mean of the cancers were measured. Tumors were categorized into three size groups, and the percentage difference in the tumor SUV max between OSEM and BPL with a beta value of 200 as well as the percentage difference in the SUV max between BPL with a beta value of 200 and BPL with a beta value of 1000 were calculated. Image quality was assessed by visual scoring. RESULTS. BPL showed a significantly higher liver signal-to-noise ratio than OSEM, except for BPL with a beta value of 200. The liver SUV mean showed no statistical difference among all algorithms. The SUV max and SUV mean of tumors decreased as the beta value increased. BPL with a beta value of 200 produced a significantly higher tumor SUV max than did OSEM ( p max than did OSEM ( p max between OSEM and BPL with a beta value of 200 and the percentage difference in the SUV max between BPL with a beta value of 20…
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Inter-method SUV differences may no longer be significant at beta values of 300 to 600, which have been rated as optimal for visual reading of [18F]FDG-PET images by human readers.
“However, this difference is directly dependent on the user-defined penalization factor beta during PL reconstruction, and at beta values of 300 to 600, which have been rated as optimal for visual reading of [18F]FDG-PET images by human readers [107,112,113], inter-method SUV differences may no longer be significant [41,108,111].”
From Influences on PET Quantification and Interpretation (2022) · cites this paper as [113]
PL reconstruction has the potential to systematically improve SNR compared to OSEM-based algorithms.
“Based on the potential of PL reconstruction to systematically improve SNR compared to OSEM-based algorithms [108,113,168]”
From Influences on PET Quantification and Interpretation (2022) · cites this paper as [113]
Several authors have reported improved SNR and image quality with PL reconstruction for different radiopharmaceuticals, including [18F]FDG, [18F]F-PSMA-1007, [68Ga]Ga-PSMA-11, [68Ga]Ga-DOTATATE, and 89Zr-labelled tracers.
“several authors have reported improved SNR and image quality with different radiopharmaceuticals, such as [18F]FDG [107,112,113], [18F]F-PSMA-1007 [143], [68Ga]Ga-PSMA-11 [169,170], [68Ga]Ga-DOTATATE [171] or 89Zr-labelled tracers [172].”
From Influences on PET Quantification and Interpretation (2022) · cites this paper as [113]
PL reconstruction has repeatedly been shown to increase conspicuity and detection rates of pulmonary lesions, even compared to OSEM with PSF and/or TOF.
“PL reconstruction has repeatedly been shown to increase conspicuity and detection rates of pulmonary lesions, even compared to OSEM with PSF and/or TOF [112,113,164,173,174].”
From Influences on PET Quantification and Interpretation (2022) · cites this paper as [113]
Cites
Accelerated image reconstruction using ordered subsets of projection data
Hudson, H. M., & Larkin, R. S. (1994) · IEEE Transactions on Medical Imaging · doi:10.1109/42.363108 · Cited by 3,713 · see citing works
Standards for PET Image Acquisition and Quantitative Data Analysis
Boellaard, R. (2009) · Journal of Nuclear Medicine · doi:10.2967/jnumed.108.057182 · Cited by 906 · see citing works
A Systematic Review of the Factors Affecting Accuracy of SUV Measurements
Adams, M. C., Turkington, T. G., Wilson, J. M., & Wong, T. Z. (2010) · American Journal of Roentgenology · doi:10.2214/ajr.10.4923 · Cited by 662 · see citing works
Effect of respiratory gating on quantifying PET images of lung cancer.
Nehmeh, S. A., Erdi, Y. E., Ling, C. C., Rosenzweig, K. E., Schöder, H. M., Larson, S. M., Macapinlac, H. A., Squire, O. D., & Humm, J. L. (2002) · PubMed · Cited by 406 · see citing works
18F-FDG PET and PET/CT in the Evaluation of Cancer Treatment Response
Ben‐Haim, S., & Ell, P. J. (2009) · Journal of Nuclear Medicine · doi:10.2967/jnumed.108.054205 · Cited by 354 · see citing works
Teoh, E. J., McGowan, D. R., Macpherson, R. E., Bradley, K. M., & Gleeson, F. V. (2015) · Journal of Nuclear Medicine · doi:10.2967/jnumed.115.159301 · Cited by 265 · see citing works
Nuyts, J. L., Beque, D., Dupont, P. J., & Mortelmans, L. (2002) · IEEE Transactions on Nuclear Science · doi:10.1109/tns.2002.998681 · Cited by 211 · see citing works
Influence of reconstruction iterations on 18F-FDG PET/CT standardized uptake values.
Jaskowiak, C., Bianco, J. A., Perlman, S. B., & Fine, J. P. (2005) · PubMed · Cited by 162 · see citing works
Ahn, S., Ross, S. G., Asma, E., Miao, J., Jin, X. H., Cheng, L., Wollenweber, S. D., & Manjeshwar, R. M. (2015) · Physics in Medicine and Biology · doi:10.1088/0031-9155/60/15/5733 · Cited by 154 · see citing works
Werner, M. K., Parker, J. A., Kolodny, G. M., English, J. R., & Palmer, M. R. (2009) · American Journal of Roentgenology · doi:10.2214/ajr.09.2516 · Cited by 100 · see citing works
Parvizi, N., Franklin, J. M., McGowan, D. R., Teoh, E., Bradley, K. M., & Gleeson, F. V. (2015) · European Journal of Radiology · doi:10.1016/j.ejrad.2015.06.025 · Cited by 91 · see citing works
Japanese guideline for the oncology FDG-PET/CT data acquisition protocol: synopsis of Version 1.0
Fukukita, H., Senda, M., Terauchi, T., Suzuki, K., Daisaki, H., Matsumoto, K., Ikari, Y., & Hayashi, M. (2010) · Annals of Nuclear Medicine · doi:10.1007/s12149-010-0377-7 · Cited by 83 · see citing works
Howard, B. A., Morgan, R., Thorpe, M. P., Turkington, T. G., Oldan, J., James, O. G., & Borges-Neto, S. (2017) · Annals of Nuclear Medicine · doi:10.1007/s12149-017-1192-1 · Cited by 70 · see citing works
Teoh, E., McGowan, D. R., Bradley, K. M., Belcher, E., Black, E., Moore, A., Sykes, A., & Gleeson, F. V. (2016) · European Radiology · doi:10.1007/s00330-016-4253-2 · Cited by 55 · see citing works
Teoh, E., McGowan, D. R., Schuster, D. M., Tsakok, M., Gleeson, F. V., & Bradley, K. M. (2018) · British Journal of Radiology · doi:10.1259/bjr.20170727 · Cited by 42 · see citing works
Vieira, T. S., Faria, D. B., Silva, F. D. A., Pimentel, F. L., & de Oliveira, J. P. (2017) · Clinical Nuclear Medicine · doi:10.1097/rlu.0000000000001713 · Cited by 13 · see citing works
Flagged first, then by citations
Cited by
Sorted by citations
Influences on PET Quantification and Interpretation
Rogasch, J. M. M., Hofheinz, F., van Heek, L., Voltin, C.-A., Boellaard, R., & Kobe, C. (2022) · Diagnostics · doi:10.3390/diagnostics12020451 · Cited by 50 · see citing works
Cumulative radiation doses from recurrent PET–CT examinations
Hosono, M., Takenaka, M., Monzen, H., Tamura, M., Kudo, M., & Nishimura, Y. (2021) · British Journal of Radiology · doi:10.1259/bjr.20210388 · Cited by 45 · see citing works
Self-citationWyrzykowski, M., Siminiak, N., Kaźmierczak, M., Ruchała, M., & Czepczyński, R. (2020) · EJNMMI Research · doi:10.1186/s13550-020-00690-6 · Cited by 34 · see citing works
Tian, D., Yang, H., Li, Y., Cui, B., & Lu, J. (2022) · EJNMMI Physics · doi:10.1186/s40658-021-00428-w · Cited by 30 · see citing works
Shkumat, N. A., Vali, R., & Shammas, A. (2020) · Pediatric Radiology · doi:10.1007/s00247-020-04640-1 · Cited by 29 · see citing works
Witkowska-Patena, E., Budzyńska, A., Giżewska, A., Dziuk, M., & Walęcka-Mazur, A. (2020) · Annals of Nuclear Medicine · doi:10.1007/s12149-019-01433-x · Cited by 27 · see citing works
Texte, E., Gouel, P., Thureau, S., Lequesne, J., Barres, B., Edet‐Sanson, A., Decazes, P., Véra, P., & Hapdey, S. (2020) · EJNMMI Physics · doi:10.1186/s40658-020-00300-3 · Cited by 27 · see citing works
Moving the goalposts while scoring―the dilemma posed by new PET technologies
Rogasch, J. M. M., Boellaard, R., Pike, L., Borchmann, P., Johnson, P., Wolf, J., Barrington, S. F., & Kobe, C. (2021) · European Journal of Nuclear Medicine and Molecular Imaging · doi:10.1007/s00259-021-05403-2 · Cited by 20 · see citing works
Rogasch, J. M. M., Frost, N., Bluemel, S., Michaels, L., Penzkofer, T., von Laffert, M., Temmesfeld‐Wollbrück, B., Neudecker, J., Rückert, J., Ochsenreither, S., et al. (2021) · Lung Cancer · doi:10.1016/j.lungcan.2021.05.003 · Cited by 13 · see citing works
Xu, L., Li, R.-S., Wu, R., Yang, R., You, Q., Yao, X., Xie, H.-F., Lv, Y., Dong, Y., Wang, F., et al. (2022) · EJNMMI Physics · doi:10.1186/s40658-022-00451-5 · Cited by 12 · see citing works
Liu, Y., Gao, M.-J., Zhou, J., Du, F., Chen, L., Huang, Z., Hu, J., & Lou, C. (2021) · BMC Medical Imaging · doi:10.1186/s12880-021-00664-7 · Cited by 11 · see citing works
Miwa, K., Yamao, T., Hashimoto, F., Miyaji, N., Kamitaka, Y., Masubuchi, M., Murata, T., Yoshii, T., Kobayashi, R., Fukuda, S., et al. (2025) · Annals of Nuclear Medicine · doi:10.1007/s12149-025-02088-7 · Cited by 9 · see citing works
Naghavi‐Behzad, M., Vogsen, M., Gerke, O., Dahlsgaard-Wallenius, S. E., Nissen, H. J., Jakobsen, N. M., Braad, P., Vilstrup, M. H., Deak, P. D., Hildebrandt, M. G., et al. (2023) · Journal of Imaging · doi:10.3390/jimaging9030065 · Cited by 7 · see citing works
Brambilla, C. R., Scheins, J. J., Issa, A. S. M., Tellmann, L., Herzog, H. R., Kops, E. R., Shah, N. J., Neuner, I., & Lerche, C. W. (2021) · PLoS ONE · doi:10.1371/journal.pone.0245580 · Cited by 6 · see citing works
Tatsumi, M., Soeda, F., Kamiya, T., Ueda, J., Katayama, D., Matsunaga, K., Watabe, T., Kato, H., & Tomiyama, N. (2021) · Frontiers in Oncology · doi:10.3389/fonc.2021.707023 · Cited by 4 · see citing works
Velo, P., Ismail, M. I., Mohandas, K. K., & Kasilingam, L. (2022) · Journal of medical imaging and radiation sciences · doi:10.1016/j.jmir.2022.09.010 · Cited by 3 · see citing works
Steroids May Be Associated With Extensive Skeletal Muscle Uptake of 18F-FDG
Kawabata, K., Hosono, M., Mori, Y., Tsukamoto, S., Ito, S., Ando, S., & Kanagaki, M. (2023) · Clinical Nuclear Medicine · doi:10.1097/rlu.0000000000004856 · Cited by 2 · see citing works
Self-citationValibeiglou, Z., Islamian, J. P., Soleymani, Y., Farzanehfar, S., Aghahosseini, F., Gilani, N., Rahmim, A., & Sheikhzadeh, P. (2025) · BMC Medical Imaging · doi:10.1186/s12880-025-01774-2 · Cited by 1 · see citing works
Naeimi, A., Harsini, S., Aghdam, R. A., Klein, R., & Abbaspour, F. (2025) · Frontiers in Oncology · doi:10.3389/fonc.2025.1597919 · Cited by 1 · see citing works
Clinical Evaluation of BSREM Reconstruction in Pediatric Oncology Using [ 18 F]FDG PET/CT
Shkumat, N. A., Vali, R., & Shammas, A. (2025) · Journal of Nuclear Medicine Technology · doi:10.2967/jnmt.125.269555 · Cited by 1 · see citing works
Tang, C. Y. L., Lim, G. K. Y., Chua, W. M., Ng, C. W. Q., Koo, S. X., Goh, C. X.-Y., Thang, S. P., Zaheer, S., Lam, W. W.-C., & Huang, H. L. (2023) · Nuclear Medicine Communications · doi:10.1097/mnm.0000000000001687 · Cited by 1 · see citing works
[PET] 5. Recent Advances in PET Image Reconstruction Using a Bayesian Penalized Likelihood Algorithm
Miwa, K., Miyaji, N., Yamao, T., Kamitaka, Y., Wagatsuma, K., & Murata, T. (2023) · Japanese Journal of Radiological Technology · doi:10.6009/jjrt.2023-2200 · Cited by 1 · see citing works
Hussain, A. A., Forssell-Aronsson, E., Rosholm, T., & Mehrara, E. (2024) · Brazilian Journal of Radiation Sciences · doi:10.15392/2319-0612.2024.2487 · Cited by 1 · see citing works
Digitale SPECT und PET: Klinische Konsequenzen
Rogasch, J. M. M. (2022) · Angewandte Nuklearmedizin · doi:10.1055/a-1715-5305 · Cited by 1 · see citing works
Implementation and Analysis of 3-D Tomographic Reconstructions from Space-Based Imaging Platforms
Hinton, G. W. (2021) · UA Campus Repository (The University of Arizona)
Brambilla, C. R., Scheins, J. J., Tellmann, L., Issa, A. S. M., Herzog, H. R., Shah, N. J., Neuner, I., & Lerche, C. W. (2023) · EJNMMI Research · doi:10.1186/s13550-023-00957-8
Digitale SPECT und PET: Klinische Konsequenzen
Rogasch, J. M. M. (2023) · Radiopraxis · doi:10.1055/a-1993-7420
E-Commerce Credit Evaluation Method Based on Bayesian Algorithm
Wang, J. (2023) · Lecture notes on data engineering and communications technologies · doi:10.1007/978-981-99-1157-8_76
Dahal, T., & Nepal, B. (2023) · IP Indian Journal of Anatomy and Surgery of Head Neck and Brain · doi:10.18231/j.ijashnb.2023.017
Ghafari, A., Mofrad, M. S., Kasraie, N., Ay, M. R., Seyyedi, N., & Sheikhzadeh, P. (2024) · Journal of Medical and Biological Engineering · doi:10.1007/s40846-024-00882-8
Murat, H., Zulkifli, M. A. A., Said, M. A., Kechik, M. M. A., Tahir, D., & Karim, M. K. A. (2024) · Radiography · doi:10.1016/j.radi.2024.12.011
Xu, L., Li, R., Wu, R., Yang, R., You, Q.-Q., Yao, X.-C., Xie, H.-F., Lv, Y., Dong, Y., Wang, F., et al. (2022) · Figshare · doi:10.6084/m9.figshare.19451626
Satoh, Y., Miwa, K., Takenaka, A., Inui, Y., Watanabe, M., Yamao, T., Miyaji, N., Ota, S., Leung, E. K., Quan, X., et al. (2026) · EJNMMI Physics · doi:10.1186/s40658-026-00846-8
- Authors
- Otani, T., Hosono, M., Kanagaki, M., Onishi, Y., Matsubara, N., Kawabata, K., & Kimura, H.
- Affiliations
- Kyoto University (JP)
- Kindai University (JP)
- Hyogo Prefectural Amagasaki General Medical Center (JP)
- ORCID
- Venue
- American Journal of Roentgenology
- Publisher
- American Roentgen Ray Society
- Published
- 2019-08
- Volume
- 213
- Issue
- 2
- Pages
- W50-W56
- ISSN
- 0002-9580 (ISSN-L)
- 0361-803X
- 1546-3141
- OpenAlex
- W2936311652
- Cited by
- 33
- References
- 16
- Citations by year
- 2026: 1 · 2025: 4 · 2024: 3 · 2023: 8 · 2022: 6 · 2021: 7 · 2020: 4
- FWCI
- 2.81
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- American Journal of Roentgenology (publishedVersion)
- PubMed (publishedVersion)
- Funders
- Japan Society for the Promotion of Science (18K07651)
- Topics
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Medical Imaging Techniques and Applications
- Health Sciences › Medicine › Pulmonary and Respiratory Medicine › Lung Cancer Diagnosis and Treatment
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Radiation Dose and Imaging
- Keywords
- Bayesian penalized likelihood reconstruction, PET image reconstruction, lung cancer, tumor size, image quality, quantitative image analysis, small lesions
- MeSH terms
- Positron Emission Tomography Computed Tomography, Aged, Aged, 80 and over, Algorithms, Bayes Theorem, Female, Humans, Lung Neoplasms, Male, Middle Aged, Neoplasm Staging, Radiographic Image Interpretation, Computer-Assisted, Retrospective Studies, Phantoms, Imaging, Radiopharmaceuticals, Fluorodeoxyglucose F18, Signal-To-Noise Ratio
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review