Accuracy of Dexcom G6 continuous glucose monitoring in non-critically ill hospitalized patients with diabetes

Open access
Davis, Georgia M., Spanakis, Elias K., Migdal, Alexandra L., Singh, Lakshmi G., Albury, Bonnie, Urrutia, Maria Agustina, Zamudio-Coronado, K. Walkiria, Scott, William H. +10 more·2021·Diabetes Care

Abstract

OBJECTIVE: Advances in continuous glucose monitoring (CGM) have transformed ambulatory diabetes management. Until recently, inpatient use of CGM has remained investigational, with limited data on its accuracy in the hospital setting. RESEARCH DESIGN AND METHODS: To analyze the accuracy of Dexcom G6, we compared retrospective matched-pair CGM and capillary point-of-care (POC) glucose data from three inpatient CGM studies (two interventional and one observational) in general medicine and surgery patients with diabetes treated with insulin. Analysis of accuracy metrics included mean absolute relative difference (MARD), median absolute relative difference (ARD), and proportion of CGM values within 15, 20, and 30% or 15, 20, and 30 mg/dL of POC reference values for blood glucose >100 mg/dL or ≤100 mg/dL, respectively (% 15/15, % 20/20, % 30/30). Clinical reliability was assessed with Clarke error grid (CEG) analyses. RESULTS: A total of 218 patients were included (96% with type 2 diabetes) with a mean age of 60.6 ± 12 years. The overall MARD (n = 4,067 matched glucose pairs) was 12.8%, and median ARD was 10.1% (interquartile range 4.6, 17.6]. The proportions of readings meeting % 15/15, % 20/20, and % 30/30 criteria were 68.7, 81.7, and 93.8%, respectively. CEG analysis showed 98.7% of all values in zones A and B. MARD and median ARD were higher in the case of hypoglycemia (<70 mg/dL) and severe anemia (hemoglobin <7 g/dL). CONCLUSIONS: Our results indicate that CGM technology is a reliable tool for hospital use and may help improve glucose monitoring in non–critically ill hospitalized patients with diabetes.

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.2337/dc20-2856

Cites

18 of 18 references verified0 citation errors · 0 not checkable

Statistics

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.

A few free checks a day · papers already verified open instantly · results are public

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

  • Data on the safety and efficacy of real-time CGM use in the hospital, particularly with implementation of remote monitoring (e.g., a glucose telemetry system), has grown.

    “Data on the safety and efficacy of real-time CGM use in the hospital, particularly with implementation of remote monitoring (e.g., a glucose telemetry system), has grown (47,50-52).”
  • There are limited studies on the efficacy or safety of SGLT2 inhibitors in hospitalized patients.

    “There are limited studies assessing its efficacy or safety in hospitalized persons.1384”

    From American Association of Clinical Endocrinology Clinical Practice Guideline: Developing a Diabetes Mellitus Comprehensive Care Plan-2022 Update. (2022) · cites this paper as [1384]

    The Study Design and Conclusions discuss inpatient Dexcom G6 continuous glucose monitoring, not SGLT2 inhibitors. A search for SGLT2 inhibitors and their efficacy or safety in hospitalized patients found no passage addressing the claim; the Conclusions’ reference to limited hospital experience concerns CGM instead. (LLM, conf=0.77)

    Experimental — accuracy not yet measured
  • Preliminary studies show no adverse safety signals when SGLT2 inhibitors are started at discharge in stable patients.

    “If persons are stable and ready for discharge, starting these agents at discharge and communicating with the patient’s primary clinician may represent a good opportunity to avoid clinical inertia, with preliminary studies showing no adverse safety signals.1384”

    From American Association of Clinical Endocrinology Clinical Practice Guideline: Developing a Diabetes Mellitus Comprehensive Care Plan-2022 Update. (2022) · cites this paper as [1384]

    The Study Design and Conclusion describe an inpatient study of Dexcom G6 glucose-monitoring accuracy, not treatment initiated at discharge. Searches for SGLT2 inhibitors, discharge initiation, and adverse safety signals found no passage addressing the claim. (LLM, conf=0.77)

    Experimental — accuracy not yet measured
  • Studies assessing effectiveness of inpatient CGM may ultimately lead to approved use of CGM for monitoring hospitalized individuals.

    “Studies have been published assessing the effectiveness of this approach, which may ultimately lead to the approved use of CGM for monitoring hospitalized individuals (277,286-295).”

    From 7. Diabetes Technology: Standards of Care in Diabetes—2025 (2025) · cites this paper as [290]

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Cites

Flagged first, then by citations

Cited by

Sorted by citations

Authors
Davis, G. M., Spanakis, E. K., Migdal, A. L., Singh, L. G., Albury, B., Urrutia, M. A., Zamudio-Coronado, K. W., Scott, W. H., Doerfler, R., Lizama, S., Satyarengga, M., Munir, K., Galindo, R. J., Vellanki, P., Cardona, S., Pasquel, F. J., Peng, L., & Umpierrez, G. E.
Affiliations
  • Emory University (US)
  • University of Maryland, Baltimore (US)
  • Baltimore VA Medical Center (US)
Corresponding author
Umpierrez, Guillermo E.
Venue
Diabetes Care
Published
2021-07-01
Volume
44
Issue
7
Pages
1641-1646
ISSN
  • 0149-5992 (ISSN-L)
  • 1935-5548
OpenAlex
W3168630898
Cited by
160
References
18
Citations by year
2026: 23 · 2025: 35 · 2024: 38 · 2023: 41 · 2022: 19 · 2021: 4
FWCI
19.04
Type
journal article
Language
en
Open access
Locations
  • Diabetes Care (publishedVersion)
  • PubMed Central (publishedVersion)
  • PubMed (publishedVersion)
Funders
  • U.S. Department of Veterans Affairs (I01 CX001825)
  • National Institutes of Health (UL1TR002378)
  • National Institute of General Medical Sciences (K23GM128221)
  • National Institute of Diabetes and Digestive and Kidney Diseases (K23DK113241, K23DK122199, K23DK123384, P30DK111024)
  • National Center for Advancing Translational Sciences
Topics
  • Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
  • Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Hyperglycemia and glycemic control in critically ill and hospitalized patients
  • Health Sciences › Medicine › Physiology › Clinical Laboratory Practices and Quality Control
Keywords
Dexcom G6, continuous glucose monitoring, inpatient diabetes management, CGM accuracy, hypoglycemia, anemia, Clarke error grid, insulin-treated diabetes
MeSH terms
Aged, Blood Glucose, Diabetes Mellitus, Type 1, Diabetes Mellitus, Type 2, Humans, Middle Aged, Retrospective Studies, Blood Glucose Self-Monitoring, Reproducibility of Results

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review