Accuracy of Dexcom G6 continuous glucose monitoring in non-critically ill hospitalized patients with diabetes
Open accessAbstract
OBJECTIVE: Advances in continuous glucose monitoring (CGM) have transformed ambulatory diabetes management. Until recently, inpatient use of CGM has remained investigational, with limited data on its accuracy in the hospital setting. RESEARCH DESIGN AND METHODS: To analyze the accuracy of Dexcom G6, we compared retrospective matched-pair CGM and capillary point-of-care (POC) glucose data from three inpatient CGM studies (two interventional and one observational) in general medicine and surgery patients with diabetes treated with insulin. Analysis of accuracy metrics included mean absolute relative difference (MARD), median absolute relative difference (ARD), and proportion of CGM values within 15, 20, and 30% or 15, 20, and 30 mg/dL of POC reference values for blood glucose >100 mg/dL or ≤100 mg/dL, respectively (% 15/15, % 20/20, % 30/30). Clinical reliability was assessed with Clarke error grid (CEG) analyses. RESULTS: A total of 218 patients were included (96% with type 2 diabetes) with a mean age of 60.6 ± 12 years. The overall MARD (n = 4,067 matched glucose pairs) was 12.8%, and median ARD was 10.1% (interquartile range 4.6, 17.6]. The proportions of readings meeting % 15/15, % 20/20, and % 30/30 criteria were 68.7, 81.7, and 93.8%, respectively. CEG analysis showed 98.7% of all values in zones A and B. MARD and median ARD were higher in the case of hypoglycemia (<70 mg/dL) and severe anemia (hemoglobin <7 g/dL). CONCLUSIONS: Our results indicate that CGM technology is a reliable tool for hospital use and may help improve glucose monitoring in non–critically ill hospitalized patients with diabetes.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Data on the safety and efficacy of real-time CGM use in the hospital, particularly with implementation of remote monitoring (e.g., a glucose telemetry system), has grown.
“Data on the safety and efficacy of real-time CGM use in the hospital, particularly with implementation of remote monitoring (e.g., a glucose telemetry system), has grown (47,50-52).”
From 16. Diabetes Care in the Hospital: Standards of Care in Diabetes—2026 (2026) · cites this paper as [52]
There are limited studies on the efficacy or safety of SGLT2 inhibitors in hospitalized patients.
“There are limited studies assessing its efficacy or safety in hospitalized persons.1384”
From American Association of Clinical Endocrinology Clinical Practice Guideline: Developing a Diabetes Mellitus Comprehensive Care Plan-2022 Update. (2022) · cites this paper as [1384]
The Study Design and Conclusions discuss inpatient Dexcom G6 continuous glucose monitoring, not SGLT2 inhibitors. A search for SGLT2 inhibitors and their efficacy or safety in hospitalized patients found no passage addressing the claim; the Conclusions’ reference to limited hospital experience concerns CGM instead. (LLM, conf=0.77)
Experimental — accuracy not yet measuredPreliminary studies show no adverse safety signals when SGLT2 inhibitors are started at discharge in stable patients.
“If persons are stable and ready for discharge, starting these agents at discharge and communicating with the patient’s primary clinician may represent a good opportunity to avoid clinical inertia, with preliminary studies showing no adverse safety signals.1384”
From American Association of Clinical Endocrinology Clinical Practice Guideline: Developing a Diabetes Mellitus Comprehensive Care Plan-2022 Update. (2022) · cites this paper as [1384]
The Study Design and Conclusion describe an inpatient study of Dexcom G6 glucose-monitoring accuracy, not treatment initiated at discharge. Searches for SGLT2 inhibitors, discharge initiation, and adverse safety signals found no passage addressing the claim. (LLM, conf=0.77)
Experimental — accuracy not yet measuredStudies assessing effectiveness of inpatient CGM may ultimately lead to approved use of CGM for monitoring hospitalized individuals.
“Studies have been published assessing the effectiveness of this approach, which may ultimately lead to the approved use of CGM for monitoring hospitalized individuals (277,286-295).”
From 7. Diabetes Technology: Standards of Care in Diabetes—2025 (2025) · cites this paper as [290]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
Battelino, T., Danne, T., Bergenstal, R. M., Amiel, S. A., Beck, R., Biester, T., Bosi, E., Buckingham, B. A., Cefalu, W. T., Close, K. L., et al. (2019) · Diabetes Care · doi:10.2337/dci19-0028 · Cited by 4,032 · cited 1× · see citing works
“use, and standardization of metrics for CGM data reporting (1).”
MethodStatistical Tools to Analyze Continuous Glucose Monitor Data
Clarke, W. L., & Kovatchev, B. P. (2009) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2008.0138 · Cited by 248 · see citing works
Time Lag of Glucose From Intravascular to Interstitial Compartment in Humans
Basu, A., Dube, S., Slama, M. Q., Errazuriz, I., Amezcua, J. C., Kudva, Y. C., Peyser, T. A., Carter, R. E., Cobelli, C., & Basu, R. (2013) · Diabetes · doi:10.2337/db13-1132 · Cited by 208 · see citing works
Measures of Accuracy for Continuous Glucose Monitoring and Blood Glucose Monitoring Devices
Freckmann, G., Pleus, S., Grady, M., Setford, S., & Levy, B. (2019) · Journal of Diabetes Science and Technology · doi:10.1177/1932296818812062 · Cited by 202 · see citing works
Shah, V. N., Laffel, L. M., Wadwa, R. P., & Garg, S. K. (2018) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2018.0143 · Cited by 197 · see citing works
Galindo, R. J., Migdal, A. L., Davis, G. M., Urrutia, M. A., Albury, B., Zambrano, C., Vellanki, P., Pasquel, F. J., Fayfman, M., Peng, L., et al. (2020) · Diabetes Care · doi:10.2337/dc19-2073 · Cited by 181 · see citing works
Self-citationContinuous glucose monitors and automated insulin dosing systems in the hospital consensus guideline
Galindo, R. J., Umpierrez, G. E., Rushakoff, R. J., Basu, A., Lohnes, S., Nichols, J. H., Spanakis, E. K., Espinoza, J., Palermo, N. E., Awadjie, D. G., et al. (2020) · Journal of Diabetes Science and Technology · doi:10.1177/1932296820954163 · Cited by 170 · see citing works
Self-citationGalindo, R. J., Aleppo, G., Klonoff, D. C., Spanakis, E. K., Agarwal, S., Vellanki, P., Olson, D. E., Umpierrez, G. E., Davis, G. M., & Pasquel, F. J. (2020) · Journal of Diabetes Science and Technology · doi:10.1177/1932296820932903 · Cited by 155 · see citing works
Self-citationSingh, L. G., Satyarengga, M., Marcano, I., Scott, W. H., Pinault, L. F., Feng, Z., Sorkin, J. D., Umpierrez, G. E., & Spanakis, E. K. (2020) · Diabetes Care · doi:10.2337/dc20-0840 · Cited by 154 · see citing works
Self-citationRacial Disparities in Clinical Trials
King, T. E. (2002) · New England Journal of Medicine · doi:10.1056/nejm200205023461812 · Cited by 100 · see citing works
Gómez, A. M., Umpierrez, G. E., Múñoz, Ó., Herrera, F., Rubio, C., Aschner, P. J., & Buendía, R. (2015) · Journal of Diabetes Science and Technology · doi:10.1177/1932296815602905 · Cited by 100 · see citing works
Self-citationDavis, G. M., Faulds, E., Walker, T., Vigliotti, D., Rabinovich, M., Hester, J., Peng, L., McLean, B., Hannon, P., Poindexter, N., et al. (2021) · Diabetes Care · doi:10.2337/dc20-2085 · Cited by 94 · see citing works
Self-citationDiabetes technology in the inpatient setting for management of hyperglycemia
Davis, G. M., Galindo, R. J., Migdal, A. L., & Umpierrez, G. E. (2020) · Endocrinology and Metabolism Clinics of North America · doi:10.1016/j.ecl.2019.11.002 · Cited by 62 · see citing works
Self-citationRemote Glucose Monitoring of Hospitalized, Quarantined Patients With Diabetes and COVID-19
Shehav-Zaltzman, G., Segal, G., Konvalina, N., & Tirosh, A. (2020) · Diabetes Care · doi:10.2337/dc20-0696 · Cited by 62 · see citing works
Reutrakul, S., Genco, M. T., Salinas, H., Sargis, R. M., Paul, C., Eisenberg, Y., Fang, J., Caskey, R. N., Henkle, S., Fatoorehchi, S., et al. (2020) · Diabetes Care · doi:10.2337/dc20-1503 · Cited by 58 · see citing works
Bamashmus, M. A., Gunaid, A. A., & Khandekar, R. B. (2009) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2007.0299 · Cited by 48 · see citing works
Garg, S. K., & Aktürk, H. K. (2018) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2018.0142 · Cited by 37 · see citing works
The Impact of COVID-19 on CGM Use in the Hospital
Ehrhardt, N. M., & Hirsch, I. B. (2020) · Diabetes Care · doi:10.2337/dci20-0046 · Cited by 29 · see citing works
Flagged first, then by citations
Cited by
Sorted by citations
Blonde, L., Umpierrez, G. E., Reddy, S. S., McGill, J. B., Berga, S. L., Bush, M., Chandrasekaran, S., DeFronzo, R. A., Einhorn, D., Galindo, R. J., et al. (2022) · Endocrine Practice · doi:10.1016/j.eprac.2022.08.002 · Cited by 507 · see citing works
Self-citation7. Diabetes Technology: Standards of Care in Diabetes—2025
American Diabetes Association Professional Practice Committee, ElSayed, N. A., McCoy, R. G., Aleppo, G., Balapattabi, K., Beverly, E. A., Briggs Early, K., Bruemmer, D., Echouffo-Tcheugui, J. B., Ekhlaspour, L., et al. (2025) · Diabetes Care · doi:10.2337/dc25-s007 · Cited by 348 · see citing works
7. Diabetes Technology: <i>Standards of Care in Diabetes</i> — <i>2023</i>
ElSayed, N. A., Aleppo, G., Aroda, V. R., Bannuru, R. R., Brown, F. M., Bruemmer, D., Collins, B. S., Hilliard, M. E., Isaacs, D., Johnson, E. L., et al. (2023) · Diabetes Care · doi:10.2337/dc23-s007 · Cited by 301 · see citing works
7. Diabetes Technology: <i>Standards of Care in Diabetes—2024</i>
American Diabetes Association Professional Practice Committee, ElSayed, N. A., Aleppo, G., Bannuru, R. R., Bruemmer, D., Collins, B. S., Ekhlaspour, L., Hilliard, M. E., Johnson, E. L., Khunti, K., et al. (2024) · Diabetes Care · doi:10.2337/dc24-s007 · Cited by 281 · see citing works
Spanakis, E. K., Urrutia, A., Galindo, R. J., Vellanki, P., Migdal, A. L., Davis, G., Fayfman, M., Idrees, T., Pasquel, F. J., Coronado, W. Z., et al. (2022) · Diabetes Care · doi:10.2337/dc22-0716 · Cited by 153 · see citing works
Self-citation16. Diabetes Care in the Hospital: Standards of Care in Diabetes—2026
American Diabetes Association Professional Practice Committee for Diabetes*, Bajaj, M., McCoy, R. G., Balapattabi, K., Bannuru, R. R., Bellini, N. J., Bennett, A. K., Beverly, E. A., Briggs Early, K., ChallaSivaKanaka, S., et al. (2026) · Diabetes Care · doi:10.2337/dc26-s016 · Cited by 80 · see citing works
Davis, G. M., Hughes, M. S., Brown, S. A., Sibayan, J., Perez-Guzman, M. C., Stumpf, M., Thompson, Z., Basina, M., Patel, R. M., Hester, J., et al. (2023) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2023.0304 · Cited by 41 · see citing works
Self-citationBaker, M., Musselman, M. E., Rogers, R., & Hellman, R. (2022) · American Journal of Health-System Pharmacy · doi:10.1093/ajhp/zxab456 · Cited by 27 · see citing works
Vellanki, P., Cardona, S., Galindo, R. J., Urrutia, M. A., Pasquel, F. J., Davis, G. M., Fayfman, M., Migdal, A., Peng, L., & Umpierrez, G. E. (2022) · Diabetes Care · doi:10.2337/dc21-1606 · Cited by 10 · see citing works
Self-citation
- Authors
- Davis, G. M., Spanakis, E. K., Migdal, A. L., Singh, L. G., Albury, B., Urrutia, M. A., Zamudio-Coronado, K. W., Scott, W. H., Doerfler, R., Lizama, S., Satyarengga, M., Munir, K., Galindo, R. J., Vellanki, P., Cardona, S., Pasquel, F. J., Peng, L., & Umpierrez, G. E.
- Affiliations
- Emory University (US)
- University of Maryland, Baltimore (US)
- Baltimore VA Medical Center (US)
- Corresponding author
- Umpierrez, Guillermo E.
- ORCID
- Venue
- Diabetes Care
- Publisher
- American Diabetes Association
- Published
- 2021-07-01
- Volume
- 44
- Issue
- 7
- Pages
- 1641-1646
- ISSN
- 0149-5992 (ISSN-L)
- 1935-5548
- PMC
- PMC8323182
- OpenAlex
- W3168630898
- Cited by
- 160
- References
- 18
- Citations by year
- 2026: 23 · 2025: 35 · 2024: 38 · 2023: 41 · 2022: 19 · 2021: 4
- FWCI
- 19.04
- Type
- journal article
- Language
- en
- Open access
- bronze
- doi.org
- Locations
- Diabetes Care (publishedVersion)
- PubMed Central (publishedVersion)
- PubMed (publishedVersion)
- Funders
- U.S. Department of Veterans Affairs (I01 CX001825)
- National Institutes of Health (UL1TR002378)
- National Institute of General Medical Sciences (K23GM128221)
- National Institute of Diabetes and Digestive and Kidney Diseases (K23DK113241, K23DK122199, K23DK123384, P30DK111024)
- National Center for Advancing Translational Sciences
- Topics
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Hyperglycemia and glycemic control in critically ill and hospitalized patients
- Health Sciences › Medicine › Physiology › Clinical Laboratory Practices and Quality Control
- Keywords
- Dexcom G6, continuous glucose monitoring, inpatient diabetes management, CGM accuracy, hypoglycemia, anemia, Clarke error grid, insulin-treated diabetes
- MeSH terms
- Aged, Blood Glucose, Diabetes Mellitus, Type 1, Diabetes Mellitus, Type 2, Humans, Middle Aged, Retrospective Studies, Blood Glucose Self-Monitoring, Reproducibility of Results
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review