Development of a Miniaturized Glucose Monitoring System by Combining a Needle-Type Glucose Sensor With Microdialysis Sampling Method: Long-term subcutaneous tissue glucose monitoring in ambulatory diabetic patients
The DOI matches its Crossref record.
Abstract
OBJECTIVE: To develop a reliable and practical glucose monitoring system by combining a needle-type glucose sensor with a microdialysis sampling technique for long-term subcutaneous tissue glucose measurements. RESEARCH DESIGN AND METHODS: A microdialysis Cuprophan hollowfiber probe (inner diameter, 0.20 mm; length, 15 mm) was perfused with isotonic saline solution (120 μl/h) and glucose concentrations in the dialysate were measured by a needle-type glucose sensor extracorporeally. This system was tested both in vitro and in vivo. Subcutaneous tissue glucose concentrations were then monitored continuously in 5 healthy and 8 diabetic volunteers for 7 to 8 days. A hollow-fiber probe was inserted into the abdominal subcutaneous tissue. RESULTS: This monitoring system achieved excellent results in vitro. Subcutaneous tissue glucose concentrations were measured in a wide range from 1.7 to >027.8 mM glucose, with a time delay of 6.9 ±1.2 min associated with a rise in glucose and 8.8 ±1.6 min with a fall in the glucose level (means ± SE). The overall correlation between subcutaneous tissue (Y) and blood (X) glucose concentration was Y = 1.08X ± 0.19 (r = 0.99). The subcutaneous tissue glucose concentration could be monitored precisely for 4 days without any in vivo calibrations and for 7 days by introducing in vivo calibrations. CONCLUSIONS: Glycemic excursions could be monitored precisely in the subcutaneous tissue by this microdialysis sampling method with a needle-type glucose sensor in ambulatory diabetic patients.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cited by
Electrochemical Glucose Sensors and Their Applications in Diabetes Management
Heller, A., & Feldman, B. (2008) · Chemical Reviews · doi:10.1021/cr068069y · Cited by 1,707
Sacks, D. B., Arnold, M., Bakris, G. L., Bruns, D. E., Horvath, A. R., Kirkman, M. S., Lernmark, A., Metzger, B. E., & Nathan, D. M. (2011) · Clinical Chemistry · doi:10.1373/clinchem.2010.161596 · Cited by 1,360
Glucose Biosensors: An Overview of Use in Clinical Practice
Yoo, E.-H., & Lee, S.-Y. (2010) · Sensors · doi:10.3390/s100504558 · Cited by 1,137
“Continuous subcutaneous glucose monitoring can also be achieved without direct contact between the interstitial fluid and transducer by using the microdialysis technique [80,81].”
MethodSacks, D. B., Bruns, D. E., Goldstein, D. E., Maclaren, N. K., McDonald, J. M., & Parrott, M. (2002) · Clinical Chemistry · doi:10.1093/clinchem/48.3.436 · Cited by 1,130
Artificial Pancreas: Past, Present, Future
Cobelli, C., Renard, E., & Kovatchev, B. (2011) · Diabetes · doi:10.2337/db11-0654 · Cited by 600
Methods for reducing biosensor membrane biofouling
Wisniewski, N., & Reichert, M. (2000) · Colloids and Surfaces B: Biointerfaces · doi:10.1016/s0927-7765(99)00148-4 · Cited by 543
Biosensors in clinical chemistry
D'Orazio, P. (2003) · Clinica Chimica Acta · doi:10.1016/s0009-8981(03)00241-9 · Cited by 524
Continuous glucose monitoring and closed‐loop systems
Hovorka, R. (2006) · Diabetic Medicine · doi:10.1111/j.1464-5491.2005.01672.x · Cited by 443
Sensors for glucose monitoring: technical and clinical aspects
Koschinsky, T., & Heinemann, L. (2001) · Diabetes/Metabolism Research and Reviews · doi:10.1002/dmrr.188 · Cited by 288
Implanted Electrochemical Glucose Sensors for the Management of Diabetes
Heller, A. (1999) · Annual Review of Biomedical Engineering · doi:10.1146/annurev.bioeng.1.1.153 · Cited by 215
Glucose monitoring: state of the art and future possibilities
Wilkins, E., & Atanasov, P. (1996) · Medical Engineering & Physics · doi:10.1016/1350-4533(95)00046-1 · Cited by 180
A critical review of mathematical models and data used in diabetology
Boutayeb, A., & Chetouani, A. (2006) · BioMedical Engineering OnLine · doi:10.1186/1475-925x-5-43 · Cited by 126
In vivo glucose sensing for diabetes management: progress towards non-invasive monitoring
Pickup, J., McCartney, L., Rolinski, O., & Birch, D. (1999) · BMJ · doi:10.1136/bmj.319.7220.1289 · Cited by 66
Evolution of Devices in Diabetes Management
Penfornis, A., Personeni, E., & Borot, S. (2011) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2011.0058 · Cited by 49
“43”
Metadata
- Authors
- Hashiguchi, Y., Sakakida, M., Nishida, K., Uemura, T., Kajiwara, K.-I., & Shichiri, M.
- Venue
- Diabetes Care
- Publisher
- American Diabetes Association
- Published
- 1994-05-01
- Volume
- 17
- Issue
- 5
- Pages
- 387-396
- ISSN
- 0149-5992 (ISSN-L)
- 1935-5548
- OpenAlex
- W2097943039
- Cited by
- 82
- References
- 0
- FWCI
- 1.95
- Type
- journal article
- Language
- en
- Open access
- closed
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.