Localised drug release using MRI-controlled focused ultrasound hyperthermia
Open accessThe DOI matches its Crossref record; 45 of the 45 references it cites were found in a registry or an index.
Abstract
PURPOSE: Thermosensitive liposomes provide a mechanism for triggering the local release of anticancer drugs, but this technology requires precise temperature control in targeted regions with minimal heating of surrounding tissue. The objective of this study was to evaluate the feasibility of using MRI-controlled focused ultrasound (FUS) and thermosensitive liposomes to achieve thermally mediated localised drug delivery in vivo. MATERIALS AND METHODS: Results are reported from ten rabbits, where a FUS beam was scanned in a circular trajectory to heat 10-15 mm diameter regions in normal thigh to 43°C for 20-30 min. MRI thermometry was used for closed-loop feedback control to achieve temporally and spatially uniform heating. Lyso-thermosensitive liposomal doxorubicin was infused intravenously during hyperthermia. Unabsorbed liposomes were flushed from the vasculature by saline perfusion 2 h later, and tissue samples were harvested from heated and unheated thigh regions. The fluorescence intensity of the homogenised samples was used to calculate the concentration of doxorubicin in tissue. RESULTS: Closed-loop control of FUS heating using MRI thermometry achieved temperature distributions with mean, T90 and T10 of 42.9°C, 41.0°C and 44.8°C, respectively, over a period of 20 min. Doxorubicin concentrations were significantly higher in tissues sampled from heated than unheated regions of normal thigh muscle (8.3 versus 0.5 ng/mg, mean per-animal difference = 7.8 ng/mg, P < 0.05, Wilcoxon matched pairs signed rank test). CONCLUSIONS: The results show the potential of MRI-controlled focused ultrasound hyperthermia for enhanced local drug delivery with temperature-sensitive drug carriers.
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Dose translation from animal to human studies revisited.
Reagan‐Shaw, S., Nihal, M., & Ahmad, N. (2008) · The FASEB Journal · doi:10.1096/fj.07-9574lsf · Cited by 6,524 · cited 1×
“The dose used in these experiments corresponds to a human equivalent dose of approximately 30 mg/m(2) [62], which is similar to what has been used in dogs (14–20 mg/m(2) [13]) and rats (30 mg/m(2) [15]), and consistent with what is being administered in current clinical trials.”
MethodAnalysis of Tissue and Arterial Blood Temperatures in the Resting Human Forearm
Pennes, H. H. (1948) · Journal of Applied Physiology · doi:10.1152/jappl.1948.1.2.93 · Cited by 4,646
Analysis of Tissue and Arterial Blood Temperatures in the Resting Human Forearm
Pennes, H. H. (1998) · Journal of Applied Physiology · doi:10.1152/jappl.1998.85.1.5 · Cited by 3,982 · cited 1×
“01 s time-step, the three-dimensional temperature distribution was recalculated by the Pennes bio-heat transfer equation [59] using a previously described finite-difference time-domain implementation [58].”
MethodSchenck, J. F. (1996) · Medical Physics · doi:10.1118/1.597854 · Cited by 1,522 · cited 1×
“A moving object in the bore of the MRI causes a local change in the main magnetic field related to its magnetic susceptibility [55] violating the assumption of static field inhomogeneities in the PRF shift subtraction method [30].”
A precise and fast temperature mapping using water proton chemical shift
Ishihara, Y., Calderon, A., Watanabe, H., Okamoto, K., Suzuki, Y., Kuroda, K., & Suzuki, Y. (1995) · Magnetic Resonance in Medicine · doi:10.1002/mrm.1910340606 · Cited by 1,027 · cited 1×
“Since then it has been shown that spatial temperature distributions can be non-invasively monitored in most tissues using MRI thermometry based on the proton resonance frequency shift [29, 30].”
BackgroundDesign of Liposomes for Enhanced Local Release of Drugs by Hyperthermia
Yatvin, M. B., Weinstein, J. N., Dennis, W. H., & Blumenthal, R. (1978) · Science · doi:10.1126/science.364652 · Cited by 956 · cited 1×
“[1–5].”
Needham, D., Anyarambhatla, G., Kong, G., & Dewhirst, M. W. (2000) · PubMed · Cited by 746
Hyperthermia enables tumor-specific nanoparticle delivery: effect of particle size.
Kong, G., Braun, R. D., & Dewhirst, M. W. (2000) · PubMed · Cited by 707
MR Imaging-guided Focused Ultrasound Surgery of Fibroadenomas in the Breast: A Feasibility Study
Hynynen, K., Pomeroy, O., Smith, D. N., Huber, P. E., McDannold, N. J., Kettenbach, J., Baum, J., Singer, S., & Jolesz, F. A. (2001) · Radiology · doi:10.1148/radiology.219.1.r01ap02176 · Cited by 634 · cited 1×
“Stable measurements can be made during ultrasound heating [31], and have been used clinically to guide point-by-point coagulation of tumours [32–35].”
MethodSelf-citationNeedham, D., & Dewhirst, M. W. (2001) · Advanced Drug Delivery Reviews · doi:10.1016/s0169-409x(01)00233-2 · Cited by 570 · cited 1×
“The temperature elevation and exposure duration were chosen to be sufficient to trigger drug release from temperaturesensitive liposomes [3] without causing significant tissue damage [61].”
Background
Cited by
Liposomal drug delivery systems: From concept to clinical applications
Allen, T. M., & Cullis, P. R. (2013) · Advanced Drug Delivery Reviews · doi:10.1016/j.addr.2012.09.037 · Cited by 4,839
Multifunctional, stimuli-sensitive nanoparticulate systems for drug delivery
Torchilin, V. P. (2014) · Nature Reviews Drug Discovery · doi:10.1038/nrd4333 · Cited by 1,663
Current Trends in The Use of Liposomes for Tumor Targeting
Deshpande, P. P., Biswas, S., & Torchilin, V. P. (2013) · Nanomedicine · doi:10.2217/nnm.13.118 · Cited by 657
Grüll, H., & Langereis, S. (2012) · Journal of Controlled Release · doi:10.1016/j.jconrel.2012.04.041 · Cited by 360
Thermosensitive liposomal drug delivery systems: state of the art review
Hossann, M., Kneidl, B., Peller, M., Lindner, L., & Winter, G. (2014) · International Journal of Nanomedicine · doi:10.2147/ijn.s49297 · Cited by 342
Ranjan, A., Jacobs, G. C., Woods, D. L., Negussie, A. H., Partanen, A., Yarmolenko, P. S., Gacchina, C. E., Sharma, K. V., Frenkel, V., Wood, B. J., et al. (2012) · Journal of Controlled Release · doi:10.1016/j.jconrel.2011.12.011 · Cited by 267
Phase‐shift, stimuli‐responsive perfluorocarbon nanodroplets for drug delivery to cancer
Rapoport, N. (2012) · WIREs Nanomedicine and Nanobiotechnology · doi:10.1002/wnan.1176 · Cited by 246
First noninvasive thermal ablation of a brain tumor with MR-guided focusedultrasound
Coluccia, D., Fandino, J., Schwyzer, L., O’Gorman, R., Remonda, L., Anon, J., Martin, E., & Werner, B. (2014) · Journal of Therapeutic Ultrasound · doi:10.1186/2050-5736-2-17 · Cited by 234
“Beyond thermal ablation, HIFU has notably been shown to allow safe, nondestructive, and transient focal blood-brain barrier disruption to facilitate drug delivery [24,25] and is being evaluated as a tool to induce hyperthermia to enhance the therapeutic effect of radiotherapy and chemotherapy [26-28].”
BackgroundLyon, P. C., Gray, M. D., Mannaris, C., Folkes, L. K., Stratford, M., Campo, L., Chung, D. Y. F., Scott, S., Anderson, M., Goldin, R., et al. (2018) · The Lancet Oncology · doi:10.1016/s1470-2045(18)30332-2 · Cited by 233
Hyperthermia-induced drug targeting
May, J. P., & Li, S.-D. (2013) · Expert Opinion on Drug Delivery · doi:10.1517/17425247.2013.758631 · Cited by 228
“Focused ultrasound with MR is currently being investigated by several groups [1,45,93-95] and is in clinical trials for thermal ablation of prostate cancer and uterine fibroids (ID: NCT01092988 and ID: NCT01226576).”
BackgroundMethodTagami, T., Foltz, W. D., Ernsting, M. J., Lee, C. M., Tannock, I. F., May, J. P., & Li, S.-D. (2011) · Biomaterials · doi:10.1016/j.biomaterials.2011.05.029 · Cited by 213
Chemical Components for the Design of Temperature-Responsive Vesicles as Cancer Therapeutics
Al-Ahmady, Z., & Kostarelos, K. (2016) · Chemical Reviews · doi:10.1021/acs.chemrev.5b00578 · Cited by 164
Phenix, C. P., Togtema, M., Pichardo, S., Zehbe, I., & Curiel, L. (2014) · Journal of Pharmacy & Pharmaceutical Sciences · doi:10.18433/j3zp5f · Cited by 161
“HIFU to deliver ThermoDox was examined in rabbit muscle demonstrating increased Dox uptake in the area of hyperthermia (145).”
BackgroundMolecular Theranostics: A Primer for the Imaging Professional
Lee, D. Y., & Li, K. C. P. (2011) · American Journal of Roentgenology · doi:10.2214/ajr.11.6797 · Cited by 95
“Other approaches use exogenous methods such as light, heat, and ultrasound to trigger drug release [45–57].”
MethodUltrasonically triggered drug delivery: Breaking the barrier
Husseini, G. A., Pitt, W. G., & Martins, A. M. (2014) · Colloids and Surfaces B: Biointerfaces · doi:10.1016/j.colsurfb.2014.07.051 · Cited by 87
“muscle [286,288,289] and even bone marrow [286,287].”
Metadata
- Authors
- Staruch, R., Chopra, R., & Hynynen, K.
- Venue
- International Journal of Hyperthermia
- Publisher
- Informa UK Limited
- Published
- 2011-03
- Volume
- 27
- Issue
- 2
- Pages
- 156-171
- ISSN
- 0265-6736 (ISSN-L)
- 1464-5157
- OpenAlex
- W2045151743
- Cited by
- 159
- References
- 45
- FWCI
- 9.10
- Type
- journal article
- Language
- en
- Open access
- bronze
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.