Multiwavelength Photoacoustic Doppler Flowmetry of Living Microalgae Cells
Open accessThe DOI matches its Crossref record; 58 of the 58 references it cites were found in a registry or an index; 11 of the 35 sources its claims cite were checked for support.
Abstract
Photoacoustics can provide a direct measurement of light absorption by microalgae depending on the photosynthesis pigment within them. In this study, we have performed photoacoustic flowmetry on living microalgae cells to measure their flow characteristics, which include flow speed, flow angle, flow direction, and, more importantly, the photoacoustic absorption spectrum, all by observing the photoacoustic Doppler power spectra during their flowing state. A supercontinuum pulsed laser with a high repetition frequency is used as the light source: through intensity modulation at a specified frequency, it can provide wavelength-selectable excitation of a photoacoustic signal centered around this frequency. Our approach can be useful to simultaneously measure the flow characteristics of microalgae and easily discriminate their different species with high accuracy in both static and dynamic states, thus facilitating the study of their cultivation and their role in our ecosystem.
Overview
CoChat Summary
TL;DR
The paper demonstrates photoacoustic Doppler flowmetry (PADF) on living microalgae using an intensity-modulated supercontinuum laser, extracting flow speed, direction, angle, and wavelength-dependent absorption from Doppler-shifted power spectra. It distinguishes Chlorella and Chrysophyta by their photoacoustic absorption spectra in both stationary and flowing states.
Key findings
- Stationary photoacoustic absorption spectra showed green microalgae (Chlorella) with two distinct peaks at 480 nm and 680 nm, while golden-brown microalgae (Chrysophyta) showed a single broad absorption window from 500 nm to 700 nm, attributed to chlorophyll a/c and fucoxanthin (580–600 nm).
- Increasing flow speed broadened the PSD curve while keeping the enclosing area constant; the Doppler broadening width at 2.0 mm/s was about ten times that at 0.2 mm/s, and at 2.0 mm/s the FWHM was 3.68 Hz.
- Flow direction was determined from positive versus negative Doppler shifts around −5 Hz at 70° and 0.5 mm/s; flow angle measurements from 90° to 50° showed no net shift at 90° and an increasing net shift reaching 0.55 Hz at 50°, agreeing with the cosθ theoretical relation.
- Dynamic PAD-shift absorption spectra at 0.5 mm/s and 70° reproduced the static absorption curves for both Chlorella and Chrysophyta, using nine wavelengths at 40 nm intervals and 50 nm bandwidth.
Methods
- Precultured microalgae samples were grown in F/2 medium for 72 h, centrifuged at 3000 rpm for 5 min, concentrated, and loaded into a glass syringe on a computer-controlled microfluidic pump; Chlorella and Chrysophyta were selected for multiwavelength experiments, and Chlorella was used for flow-parameter demonstrations.
- The setup used a supercontinuum fiber laser (tunable 400–840 nm, 80 MHz repetition rate, 250 ps pulses, 92 nJ pulse energy), electro-optic intensity modulation at 2.25 MHz, a lock-in amplifier with reference set 5 Hz higher than the modulation frequency, a 2.25 MHz ultrasonic transducer perpendicular to flow, a 60 dB preamplifier, a 0.5 mm inner-diameter silicon tube, and a rotatable mount for flow angles.
- Power spectral density was calculated as the squared magnitude of the FFT normalized by effective noise bandwidth with a Hamming window; Gaussian fitting via the Levenberg–Marquardt algorithm extracted area, width, center, and FWHM, using 2300 data points, 145 s per measurement, 15.0 Hz spectral range, and 6.862 mHz resolution.
Limitations
- The paper states that flow speeds above 2.0 mm/s may require a higher frequency offset, and that at high flow speeds the noise ratio increases while signal amplitude decreases with spectral broadening, so flow speed measurement may not be very accurate above 2.0 mm/s; parameters were optimized for 0.1–1 mm/s.
- The paper states that flow-angle measurements were limited to 50° and above because the photoacoustic signal became weak or blocked below 50°, and that dynamic absorption spectra used a 40 nm wavelength interval, with finer 10 nm resolution being time-consuming; the ideal flow speed and angle were 0.5 mm/s and 70°, with other settings potentially reducing spectral accuracy due to lower SNR.
- The paper states that background acoustic signal was detected even without laser light, attributed to electromagnetic interference, electronic component noise, and environmental factors, and was reduced tenfold by separating electronics; evidently, multiwavelength species discrimination was demonstrated only for two selected species after initial screening of five microalgae types.
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Figures







Verification
Claims the paper makes
6 Valid · 1 Not valid · 10 Not checked · 57 Own results and interpretations
The cited sources support only part of this claim — read each source’s verdict below.
In the paper
“In a stationary position, the photoacoustic signal in the time domain is directly proportional to the optical absorption spectrum of the microalgae [3,37].”
§ Multiwavelength Photoacoustic Doppler Flowmetry of Living Microalgae Cells > 1. Introduction
What the sources say
- Partly supportedcites [3] Wang, H., Liao, R., Xiong, Z., Wang, Z., Li, J., Zhou, Q., Tao, Y., and Ma, H. (2022). Simultaneously Acquiring Optical and Acoustic Properties of Individual Microalgae Cells Suspended in Water. Biosensors, 12. sourceExperimental — accuracy not yet measured
Methods §2.3 links greater light absorption to a stronger thermal effect and possibly a higher acoustic frequency. It does not establish that a stationary microalga’s time-domain photoacoustic signal is directly proportional to its optical absorption spectrum; the measured cells pass through the scattering volume (Methods §2.2; Discussions §4). Page numbers are unavailable in our copy. (LLM, conf=0.82)
- Not matchedcites [37] Tuchin, V.V., Leahy, M.J., and Wang, R.K. (2021, January 5). In Situ Fluorescence/Photoacoustic Monitoring of Diatom Algae. Proceedings of the Dynamics and Fluctuations in Biomedical Photonics XVIII, Online.
Couldn’t be matched to a paper in Crossref, so it can’t be checked.
plainstates an association- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Partly supported
The cited sources support only part of this claim — read each source’s verdict below.
In the paper
“In this way, from flow state measurements, we obtain not only the photoacoustic absorption spectral information from the Doppler signal but also other important parameters, such as flow speed, flow direction, and flow angle simultaneously [40,41].”
§ Multiwavelength Photoacoustic Doppler Flowmetry of Living Microalgae Cells > 1. Introduction
What the sources say
- Partly supportedcites [40] Daoudi, 2015, Review of Photoacoustic Flow Imaging: Its Current State and Its Promises, Photoacoustics sourceExperimental — accuracy not yet measured
The Introduction (‘The case for PA flow imaging’; page unavailable) says photoacoustic imaging can provide spectroscopic information. The ‘Density tracking’ section reports a method that combines measurements to estimate total flow velocity and Doppler angle. However, the review does not establish that absorption spectra are obtained from the Doppler signal alongside all the listed flow parameters simultaneously; its ‘Doppler shift’ section notes that some methods provide no flow directionality. (LLM, conf=0.87)
- Not checked — no full text of the sourcecites [41] Sheinfeld, 2010, Photoacoustic Doppler Measurement of Flow Using Tone Burst Excitation, Opt. Express source
plainstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Partly supported
A result the paper states — it cites no source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“Each species shows different photoacoustic absorption depending on the photosynthesis pigment inside them.”
§ Multiwavelength Photoacoustic Doppler Flowmetry of Living Microalgae Cells > 2. Methods > 2.1. Sample Preparation
plainstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Rests on an assumption not shown here.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“while golden-brown (Chrysophyta) microalgae showed a single peak in a very broad window of visible wavelength ranging from 500 nm to 700 nm.”
§ Multiwavelength Photoacoustic Doppler Flowmetry of Living Microalgae Cells > 3. Results > 3.1. Absorption Spectra of Algae Samples in the Stationary State
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Rests on an assumption not shown here.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“From Figure 2, it is observable that green microalgae (Chlorella) exhibited two distinct peaks at 480 nm and 680 nm, respectively”
§ Multiwavelength Photoacoustic Doppler Flowmetry of Living Microalgae Cells > 3. Results > 3.1. Absorption Spectra of Algae Samples in the Stationary State
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Rests on an assumption not shown here.
37 more claims
Create a free account to see all 74 claims and every source verdict.
- Probably not a substantial claim. 13
- It may state two separate claims as one. 24
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
Each contradiction between two passages listed here was confirmed by an independent check.
Rated minor: it would not touch a result — a value no result rests on, a p-value that stays on the same side of the significance level, or a conflict that is clerical, resolved elsewhere in the paper, or outside its own findings.
In the paper
§ Multiwavelength Photoacoustic Doppler Flowmetry of Living Microalgae Cells > 3. Results > 3.4. Flow Angle Measurement
“a series of experiments were conducted at different flow angles ranging from θ = 0° to 50°, 60°, 70°, 80°, and 90°”
§ Multiwavelength Photoacoustic Doppler Flowmetry of Living Microalgae Cells > 4. Discussion
“Therefore, we have limited our measurements to angles of 50° and above.”
Rated minor: it would not touch a result — a value no result rests on, a p-value that stays on the same side of the significance level, or a conflict that is clerical, resolved elsewhere in the paper, or outside its own findings.
In the paper
§ Multiwavelength Photoacoustic Doppler Flowmetry of Living Microalgae Cells > 3. Results > 3.5. Photoacoustic Doppler Shift Absorption Spectrum
“we took green Chlorella and found their PAD absorption spectrum at different wavelengths ranging from 440 nm to 760 nm by the spectral interval of 40 nm”
§ Multiwavelength Photoacoustic Doppler Flowmetry of Living Microalgae Cells > 4. Discussion
“Currently, we have constructed the absorption spectral graphs in dynamic conditions using a wavelength interval of 40 nm from 415 to 850 nm. The nine power spectra shown in Figure 6a and Figure 7a are analyzed”
Citations
Cites
[34] Chlorophylls and carotenoids: Pigments of photosynthetic biomembranes
Lichtenthaler, H. K. (1987) · Methods in Enzymology · doi:10.1016/0076-6879(87)48036-1 · Cited by 13,616
Microalgae-based products: Food and public health
Ferreira de Oliveira, A. P., & Bragotto, A. P. A. (2022) · Future Foods · doi:10.1016/j.fufo.2022.100157 · Cited by 190
Dilek (Yalcin) Duygu (2012) · AFRICAN JOURNAL OF BIOTECHNOLOGY · doi:10.5897/ajb11.1863 · Cited by 182
Photoacoustic Doppler Effect from Flowing Small Light-Absorbing Particles
Fang, H., Maslov, K., & Wang, L. V. (2007) · Physical Review Letters · doi:10.1103/physrevlett.99.184501 · Cited by 171
Self-citationGalanzha, E. I., & Zharov, V. P. (2012) · Methods · doi:10.1016/j.ymeth.2012.06.009 · Cited by 159 · cited 1×
“Photoacoustics has been implemented in many versatile applications, such as the characterization of photoacoustic and fluorescence properties of living cells [3], environmental monitoring [4], and flow cytometry [5–9].”
MethodRolinger, L., Rüdt, M., & Hubbuch, J. (2020) · Analytical and Bioanalytical Chemistry · doi:10.1007/s00216-020-02407-z · Cited by 126 · cited 1×
“Photoacoustics is gaining much interest in imaging technique development [1] because of the involvement of acoustics, making it suffer less scattering [1] and achieve deeper penetration with higher image resolution over other optical spectroscopy techniques [2].”
BackgroundSahoo, D., & Seckbach, J. (2015) · Cellular origin, life in extreme habitats and astrobiology · doi:10.1007/978-94-017-7321-8 · Cited by 90
Review of Photoacoustic Flow Imaging: Its Current State and Its Promises
van den Berg, P. J., Daoudi, K., & Steenbergen, W. (2015) · Photoacoustics · doi:10.1016/j.pacs.2015.08.001 · Cited by 84
In vivo photoacoustic flow cytometry for early malaria diagnosis
Cai, C., Carey, K. A., Nedosekin, D. A., Menyaev, Y. A., Sarimollaoglu, M., Galanzha, E. I., Stumhofer, J. S., & Zharov, V. P. (2016) · Cytometry Part A · doi:10.1002/cyto.a.22854 · Cited by 78
Transverse flow imaging based on photoacoustic Doppler bandwidth broadening
Yao, J., & Wang, L. V. (2010) · Journal of Biomedical Optics · doi:10.1117/1.3339953 · Cited by 75 · cited 2×
“Furthermore, Gaussian fi tt ing was then applied to each FFT signal to clearly observe the PSD broadening It is obvious in Figure 3 that increasing flow speed results in broadening the width of the PSD curve while keeping the enclosing area constant (maintaining the total power) [32]. e ff ect.”
MethodRecent Advances of Natural Pigments from Algae
Chen, Z., Wu, W., Wen, Y., Zhang, L., Wu, Y., Farid, M. S., El-Seedi, H. R., Capanoglu, E., & Zhao, C. (2023) · Food Production, Processing and Nutrition · doi:10.1186/s43014-023-00155-y · Cited by 67 · cited 1×
“Their absorption spectrum [16] depends upon the type and concentration of photosynthesis pigments [17] inside them, such as chlorophyll, fucoxanthin, phycoerythrin, and phycocyanin [18].”
BackgroundPhotoacoustic Doppler Flow Measurement in Optically Scattering Media
Fang, H., Maslov, K., & Wang, L. V. (2007) · Applied Physics Letters · doi:10.1063/1.2825569 · Cited by 64 · cited 1×
“On the other hand, the absorption spectrum in the flow state can be extracted from the Doppler-shifted power spectra [27,38,39].”
BackgroundSelf-citationRecent application of spectroscopy for the detection of microalgae life information: A review
Liu, J.-Y., Zeng, L.-H., & Ren, Z.-H. (2020) · Applied Spectroscopy Reviews · doi:10.1080/05704928.2018.1509345 · Cited by 59 · cited 1×
“Several other spectroscopic techniques to study microalgae exist [19], including ultraviolet–visible spectroscopy [20], fluorescence spectroscopy [21], Raman spectroscopy [22], Fourier-transform infrared (FTIR) spectroscopy [23], and nuclear magnetic resonance (NMR) spectroscopy [24].”
BackgroundRaman Spectroscopy of Algae: A Review
Tomar, V. (2012) · Journal of Nanomedicine & Nanotechnology · doi:10.4172/2157-7439.1000131 · Cited by 56 · cited 1×
“Several other spectroscopic techniques to study microalgae exist [19], including ultraviolet–visible spectroscopy [20], fluorescence spectroscopy [21], Raman spectroscopy [22], Fourier-transform infrared (FTIR) spectroscopy [23], and nuclear magnetic resonance (NMR) spectroscopy [24].”
BackgroundA review on current and future advancements for commercialized microalgae species
Wong, J. F., Hong, H. J., Foo, S. C., Yap, M. K. K., & Tan, J. W. (2022) · Food Science and Human Wellness · doi:10.1016/j.fshw.2022.04.007 · Cited by 56 · cited 1×
“We experimented on different microalgae species [35,36] and observed their characteristics, such as flow speed, flow direction, flow angle, and, most importantly, their photoacoustic wavelength-dependent absorption spectrum in both stationary and moving states.”
MethodIkeda, Y., Yamagishi, A., Komura, M., Suzuki, T., Dohmae, N., Shibata, Y., Itoh, S., Koike, H., & Satoh, K. (2013) · Biochimica et Biophysica Acta (BBA) - Bioenergetics · doi:10.1016/j.bbabio.2013.02.003 · Cited by 46 · cited 2×
“The major reason is because of the improvement in the rotatable angular mount structure on which sample flow has been made.”
BackgroundPhotoacoustic imaging plus X: a review
Jiang, D., Zhu, L., Tong, S., et al. (2023) · Journal of Biomedical Optics · doi:10.1117/1.jbo.29.s1.s11513 · Cited by 42 · cited 1×
“Photoacoustics is gaining much interest in imaging technique development [1] because of the involvement of acoustics, making it suffer less scattering [1] and achieve deeper penetration with higher image resolution over other optical spectroscopy techniques [2].”
BackgroundPhotoacoustic Doppler Measurement of Flow Using Tone Burst Excitation
Sheinfeld, A., Gilead, S., & Eyal, A. (2010) · Optics Express · doi:10.1364/oe.18.004212 · Cited by 41 · cited 1×
“In this way, from flow state measurements, we obtain not only the photoacoustic absorption spectral information from the Doppler signal but also other important parameters, such as flow speed, flow direction, and flow angle simultaneously [40,41].”
MethodShu, X., Bondu, M., Dong, B., et al. (2016) · Optics Letters · doi:10.1364/ol.41.002743 · Cited by 41 · cited 1×
“Usually, supercontinuum sources with ns pulses have been used in previous photoacoustic studies [46,47]; however, we have developed a method using a picosecond pulse source with a high repetition frequency of 80 MHz and external intensity modulation around 2.25 MHz [48].”
MethodInfluentialVelocity measurements in whole blood using acoustic resolution photoacoustic Doppler
Brunker, J., & Beard, P. C. (2016) · Biomedical Optics Express · doi:10.1364/boe.7.002789 · Cited by 38 · cited 1×
“Although research on photoacoustic flow cytometry of human blood has been well established [10,11], very limited data exist on such flowmetry of living botanical cells such as microalgae [12].”
BackgroundBest practices in plant cytometry
Galbraith, D., Loureiro, J., Antoniadi, I., Bainard, J., Bureš, P., Cápal, P., Castro, M., Castro, S., Čertner, M., Čertnerová, D., et al. (2021) · Cytometry Part A · doi:10.1002/cyto.a.24295 · Cited by 35
Light-Induced Shifts in the Absorption Spectrum of Carotenoids in Red and Brown Algae
Fork, D. C., & Amesz, J. (1967) · Photochemistry and Photobiology · doi:10.1111/j.1751-1097.1967.tb09656.x · Cited by 33 · cited 2×
“Notably, we observed distinct absorption spectra of different microalgae species corresponding to light-absorbing pigments such as chlorophyll and carotenoids [18,49].”
BackgroundTzang, O., & Piestun, R. (2016) · Optics Express · doi:10.1364/oe.24.028122 · Cited by 32 · cited 1×
“The lock-in amplifier also required a reference signal at input #1, which was achieved from the function waveform The lock-in ampli fi er extracts signal information by de-modulating the received signal to isolate the speci fi c frequency component of interest from the noise [52,53]. generator.”
BackgroundReal-Time Label-Free Embolus Detection Using In Vivo Photoacoustic Flow Cytometry
Juratli, M. A., Menyaev, Y. A., Sarimollaoglu, M., Siegel, E. R., Nedosekin, D. A., Suen, J. Y., Melerzanov, A., Juratli, T. A., Galanzha, E. I., & Zharov, V. P. (2016) · PLoS ONE · doi:10.1371/journal.pone.0156269 · Cited by 32
Acoustic resolution photoacoustic Doppler velocimetry in blood-mimicking fluids
Brunker, J., & Beard, P. (2016) · Scientific Reports · doi:10.1038/srep20902 · Cited by 32 · cited 1×
“This study builds upon the foundational research on PADF, where the speed of flowing particles having parabolic flow distribution in a cylindrical tube or vessel is measured from their Doppler shift of frequencies [26–32].”
MethodLababpour, A., & Lee, C. (2006) · Journal of Bioscience and Bioengineering · doi:10.1263/jbb.101.104 · Cited by 31
Perspectives of fluorescence spectroscopy for online monitoring in microalgae industry
Sá, M., Ferrer‐Ledo, N., Gao, F., Bertinetto, C. G., Jansen, J., Crespo, J. G., Wijffels, R. H., Barbosa, M., & Galinha, C. F. (2022) · Microbial Biotechnology · doi:10.1111/1751-7915.14013 · Cited by 29
Best practices in the flow cytometry of microalgae
Čertnerová, D., & Galbraith, D. W. (2021) · Cytometry Part A · doi:10.1002/cyto.a.24328 · Cited by 25 · cited 1×
“Although research on photoacoustic flow cytometry of human blood has been well established [10,11], very limited data exist on such flowmetry of living botanical cells such as microalgae [12].”
BackgroundPhotoacoustic flow cytometry for nanomaterial research
Nedosekin, D. A., Fahmi, T., Nima, Z. A., et al. (2017) · Photoacoustics · doi:10.1016/j.pacs.2017.03.002 · Cited by 25
Kosumi, D., Fujii, R., Sugisaki, M., Oka, N., Iha, M., & Hashimoto, H. (2014) · Photosynthesis Research · doi:10.1007/s11120-014-9995-6 · Cited by 22 · cited 1×
“Notably, we observed distinct absorption spectra of different microalgae species corresponding to light-absorbing pigments such as chlorophyll and carotenoids [18,49].”
BackgroundLight-Harvesting Complexes of Diatoms: Fucoxanthin-Chlorophyll Proteins
Büchel, C. (2020) · Advances in Photosynthesis and Respiration · doi:10.1007/978-3-030-33397-3_16 · Cited by 20
A Low-Cost Optoacoustic Sensor for Environmental Monitoring
Stylogiannis, A., Κούσιας, Ν., Κοντσές, Α., Ntziachristos, L. D., & Ntziachristos, V. (2021) · Sensors · doi:10.3390/s21041379 · Cited by 19 · cited 1×
“Photoacoustics has been implemented in many versatile applications, such as the characterization of photoacoustic and fluorescence properties of living cells [3], environmental monitoring [4], and flow cytometry [5–9].”
MethodPhotoacoustic and fluorescence lifetime imaging of diatoms
Cvjetinovic, J., Salimon, A. I., Novoselova, M. V., Sapozhnikov, P. V., Shirshin, E. A., Yashchenok, A. M., Kalinina, O. Y., Korsunsky, A. M., & Gorin, D. A. (2020) · Photoacoustics · doi:10.1016/j.pacs.2020.100171 · Cited by 18 · cited 1×
“Pho-toacoustic lifetime imaging has also recently been performed using a 532 nm laser source for algae diatoms [33].”
MethodZhang, R., Yao, J., Maslov, K. I., & Wang, L. V. (2013) · Applied Physics Letters · doi:10.1063/1.4819735 · Cited by 17
Baránková, B., Lazár, D., Nauš, J., Solovchenko, A., Gorelova, O., Baulina, O., Huber, G., & Nedbal, L. (2020) · Algal Research · doi:10.1016/j.algal.2020.101881 · Cited by 17
Lange, F., Giannoni, L., & Tachtsidis, I. (2021) · Applied Sciences · doi:10.3390/app11104616 · Cited by 13 · cited 1×
“We use a supercontinuum laser source [42–45] for the excitation, which provides a narrow bandwidth, high spatial coherence, and collimated light with selectable wave-lengths in the visible region.”
MethodDiatoms: Yellow or Golden Brown Algae
Kumar, S., Baweja, P., & Sahoo, D. (2015) · Cellular origin, life in extreme habitats and astrobiology · doi:10.1007/978-94-017-7321-8_8 · Cited by 11
Wang, H., Liao, R., Xiong, Z., Wang, Z., Li, J., Zhou, Q., Tao, Y., & Ma, H. (2022) · Biosensors · doi:10.3390/bios12030176 · Cited by 10 · cited 3×
“Photoacoustics has been implemented in many versatile applications, such as the characterization of photoacoustic and fluorescence properties of living cells [3], environmental monitoring [4], and flow cytometry [5–9].”
BackgroundMethodBrunker, J., & Beard, P. C. (2014) · Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE · doi:10.1117/12.2039619 · Cited by 9 · cited 1×
“Although research on photoacoustic flow cytometry of human blood has been well established [10,11], very limited data exist on such flowmetry of living botanical cells such as microalgae [12].”
BackgroundIn situ fluorescence/photoacoustic monitoring of diatom algae
Cvjetinovic, J., Bedoshvili, Y. D., Nozdriukhin, D. V., et al. (2021) · doi:10.1117/12.2588254 · Cited by 8 · cited 1×
“In a stationary position, the photoacoustic signal in the time domain is directly proportional to the optical absorption spectrum of the microalgae [3,37].”
BackgroundDuan, X., Zheng, X., Liu, Z., Dong, T., Luo, Y., Yan, W., Wang, C., & Song, C. (2024) · Analytical Chemistry · doi:10.1021/acs.analchem.3c04665 · Cited by 8 · cited 1×
“Other interesting advancements include the implementation of photoacoustic on-chip flow cytometry of living microalgae cells [34].”
BackgroundUnit operations applied for microalgae-based solid–liquid separation
Rao, N. R. H., & Henderson, R. K. (2022) · Elsevier eBooks · doi:10.1016/b978-0-323-90971-6.00015-2 · Cited by 8
Photoacoustic Flow Cytometry Using Functionalized Microspheres for Selective Detection of Bacteria
Edgar, R. H., Samson, A., Kocsis, T., et al. (2023) · Micromachines · doi:10.3390/mi14030573 · Cited by 6
The Algae Foundation® and Algae Technology Educational Consortium
Levine, I. A., Gerk, C. L., Gómez, S. M., Nalley, J. O., & Nalley, M. M. (2021) · Journal of the World Aquaculture Society · doi:10.1111/jwas.12817 · Cited by 6
Hotos, G. N., & Bekiari, V. (2023) · International Journal of Plant Biology · doi:10.3390/ijpb14040065 · Cited by 6 · cited 2×
“A series of experiments were carried out to observe the photoacoustic effect on different microalgae species such as Green Puff ( Chrysophyta ), Golden Brown ( Chrysophyta ), Triangu-lar ( Phaeodactylum tricornutum ), Diatom ( phytoplankton ), and Green Algae ( Chlorella ) [16].”
BackgroundAntimicrobial Study of Algal Enzymes Extracted from Microalgae by UltrasonicationPreprint
Bhuyar, P., Yusoff, M. M., Rahim, M. H. A., et al. (2019) · SSRN Electronic Journal · doi:10.2139/ssrn.3356472 · Cited by 5
Tong, Y., Zhao, H., Fang, H., Zhao, Y., & Yuan, X. (2016) · AIP Advances · doi:10.1063/1.4942097 · Cited by 4 · cited 3×
“Besides the microalgae pigment study, the technique demonstrated can be further extended to other biological and environmental studies and may offer a new dimension to photoacoustic spectroscopy applications.”
BackgroundInfluentialSelf-citationAngle dependence and transit time effect in Doppler flow measurement
Klünemann, H.-H., Arnolds, B. J., & von Reutern, G.-M. (1996) · European Journal of Ultrasound · doi:10.1016/0929-8266(95)00158-n · Cited by 3
Wavelength agile photoacoustic microscopy with a pulsed supercontinuum source
Liu, M., & Buma, T. (2010) · doi:10.1109/ultsym.2010.5936015 · Cited by 3
Photoacoustics — A Novel Tool for the Study of Aquatic Photosynthesis
Pinchasov-Grinblat, Y., & Dubinsky, Z. (2013) · Photosynthesis · doi:10.5772/56600 · Cited by 3 · cited 1×
“In contrast, photoacoustic spectroscopy can offer high sensitivity, depth profiling, non-destructive label-free spectral information, low cost, and real-time monitoring capabilities, making it a robust and versatile alternative for the analysis of microalgae absorption spectra [25].”
BackgroundPhotoacoustic imaging using lock-in amplification and pulsed fiber lasers
Shi, W., Reza, P. H., & Zemp, R. J. (2016) · Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE · doi:10.1117/12.2211109 · Cited by 2 · cited 1×
“The lock-in amplifier also required a reference signal at input #1, which was achieved from the function waveform The lock-in ampli fi er extracts signal information by de-modulating the received signal to isolate the speci fi c frequency component of interest from the noise [52,53]. generator.”
BackgroundLiu, Y., Fang, H., FAROOQ, T., & Yuan, X. (2020) · 2019 International Conference on Optical Instruments and Technology: Optical Sensors and Applications · doi:10.1117/12.2549505 · Cited by 2 · cited 1×
“On the other hand, the absorption spectrum in the flow state can be extracted from the Doppler-shifted power spectra [27,38,39].”
BackgroundSelf-citationPhotoacoustic Doppler Effect and Flow Sensing
Fang, H. (2009) · Optical Science and Engineering · doi:10.1201/9781420059922.ch2 · Cited by 2
Self-citationPhotoacoustic Doppler axial flow measurement of homogenous media using structured illumination
Zhang, R., Yao, J., Maslov, K., & Wang, L. V. (2014) · Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE · doi:10.1117/12.2038356 · Cited by 1
All-fiber nanosecond laser system generating supercontinuum spectrum for photoacoustic imaging
Yavaş, S., Kipergil, E. A., Akçaalan, Ö., et al. (2013) · doi:10.1109/cleoe-iqec.2013.6801525 · Cited by 1 · cited 1×
“Usually, supercontinuum sources with ns pulses have been used in previous photoacoustic studies [46,47]; however, we have developed a method using a picosecond pulse source with a high repetition frequency of 80 MHz and external intensity modulation around 2.25 MHz [48].”
MethodInfluentialPhotoacoustic microscopy using supercontinuum light for imaging of chromoproteins in living cells
Ishihara, M., Hirasawa, T., Le, Z., et al. (2022) · doi:10.1117/12.2610957 · Cited by 1
Wu, X., Farooq, T., Li, L., Yan, S., & Fang, H. (2023) · Tenth International Conference on Information Optics and Photonics · doi:10.1117/12.3005850 · Cited by 1
Self-citationSupercontinuum Generation - an encyclopedia article
Paschotta, R. (2021) · doi:10.61835/bxg · Cited by 1
Research works printed but not linked: 1
- Oraevsky, A.A., and Wang, L.V. (2008, January 7). Photoacoustic Doppler Flowmetry. Proceedings of the SPIE BiOS, San Francisco, CA, USA.
Cited by
Farooq, T., Ali, F., & Fang, H. (2025) · Journal of Physics Conference Series · doi:10.1088/1742-6596/2966/1/012018 · Cited by 1
Self-citation
Metadata
- Authors
- Farooq, T., Wu, X., Yan, S., & Fang, H.
- Affiliations
- Shenzhen University (CN)
- Nanophotonic Research Center, Shenzhen Key Laboratory of Micro-Scale Optical Information Technology, Institute of Microscale Optoelectronics, Shenzhen 518060, China
- Corresponding author
- Yan, Sheng
- Fang, Hui
- ORCID
- Venue
- Biosensors
- Publisher
- MDPI AG
- Published
- 2024-08-16
- Volume
- 14
- Issue
- 8
- Pages
- 397
- ISSN
- 2079-6374 (ISSN-L)
- PMC
- PMC11353084
- OpenAlex
- W4401634253
- Cited by
- 1
- References
- 58
- Citations by year
- 2025: 1
- FWCI
- 0.11
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Licence
- Locations
- Biosensors (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion)
- Funders
- National Natural Science Foundation of China (12074268, 2023YQ024)
- Scientific Instrument Developing Project of Shenzhen University (12074268, 2023YQ024)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Photoacoustic and Ultrasonic Imaging
- Physical Sciences › Engineering › Biomedical Engineering › Microfluidic and Bio-sensing Technologies
- Life Sciences › Biochemistry, Genetics and Molecular Biology › Biophysics › Spectroscopy Techniques in Biomedical and Chemical Research
- Keywords
- microalgae, flow velocity measurement, flow direction, species discrimination, photosynthetic pigments, microalgae cultivation
- MeSH terms
- Photosynthesis, Microalgae, Photoacoustic Techniques
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
The claim verdicts this report counts are experimental: their accuracy is not yet measured.
A language model drew the claims from the paper on 2026-10-01, under version 3 of the claim definition (model deepseek-v4.1-flash, prompt extract-v3).
A second model labelled each claim on 2026-10-01 (model jev-1.13-20260917, definition revision gold-guideline-v0.2).
Cited sources are checked with a language model’s help (gpt-6-sol).
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).