Ion Mobility Spectrometry in Food Analysis: Principles, Current Applications and Future Trends
Open accessThe DOI matches its Crossref record; 95 of the 95 references it cites were found in a registry or an index.
Abstract
In the last decade, ion mobility spectrometry (IMS) has reemerged as an analytical separation technique, especially due to the commercialization of ion mobility mass spectrometers. Its applicability has been extended beyond classical applications such as the determination of chemical warfare agents and nowadays it is widely used for the characterization of biomolecules (e.g., proteins, glycans, lipids, etc.) and, more recently, of small molecules (e.g., metabolites, xenobiotics, etc.). Following this trend, the interest in this technique is growing among researchers from different fields including food science. Several advantages are attributed to IMS when integrated in traditional liquid chromatography (LC) and gas chromatography (GC) mass spectrometry (MS) workflows: (1) it improves method selectivity by providing an additional separation dimension that allows the separation of isobaric and isomeric compounds; (2) it increases method sensitivity by isolating the compounds of interest from background noise; (3) and it provides complementary information to mass spectra and retention time, the so-called collision cross section (CCS), so compounds can be identified with more confidence, either in targeted or non-targeted approaches. In this context, the number of applications focused on food analysis has increased exponentially in the last few years. This review provides an overview of the current status of IMS technology and its applicability in different areas of food analysis (i.e., food composition, process control, authentication, adulteration and safety).
Overview
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Lanucara, F., Holman, S. W., Gray, C. J., & Eyers, C. E. (2014) · Nature Chemistry · doi:10.1038/nchem.1889 · Cited by 901 · cited 4×
“Furthermore, it is important to remark that each IMS form can improve the performance of LC–MS methods in one way or another, but they are not exempt of limitations [4,14].”
BackgroundInfluentialTheory of plasma chromatography/gaseous electrophoresis. Review
Revercomb, H. E., & Mason, E. A. (1975) · Analytical Chemistry · doi:10.1021/ac60357a043 · Cited by 755 · cited 1×
“Nevertheless, both parameters can be easily correlated according to Mason–Schamp equation (Equation (3)) when the separation occurs at low electric fields [12].”
BackgroundEiceman, G. A., & Karpas, Z. (2005) · CRC Press eBooks · doi:10.1201/9781420038972 · Cited by 569
Food authentication: Techniques, trends & emerging approaches
Danezis, G. P., Tsagkaris, A. S., Camin, F., Brusic, V., & Georgiou, C. A. (2016) · TrAC Trends in Analytical Chemistry · doi:10.1016/j.trac.2016.02.026 · Cited by 559 · cited 1×
“Chemical fingerprinting of food products and chemometrics are widely used for food authentication and ultimately to identify food fraud [76].”
BackgroundFundamentals of Traveling Wave Ion Mobility Spectrometry
Shvartsburg, A. A., & Smith, R. D. (2008) · Analytical Chemistry · doi:10.1021/ac8016295 · Cited by 534 · cited 1×
“More detailed information about the operation and physical principles of the different IMS forms can be found in specialized literature [17,19,25,27,28].”
BackgroundStow, S. M., Causon, T. J., Zheng, X., Kurulugama, R. T., Mairinger, T., May, J. C., Rennie, E. E., Baker, E. S., Smith, R. D., McLean, J. A., et al. (2017) · Analytical Chemistry · doi:10.1021/acs.analchem.7b01729 · Cited by 534 · cited 4×
“In this condition, DTIMS operates as a secondary method and CCS calibration is required to measure this molecular descriptor [22].”
BackgroundMethodInfluentialRecommendations for reporting ion mobility Mass Spectrometry measurements
Gabelica, V., Shvartsburg, A. A., Afonso, C., Barran, P., Benesch, J. L. P., Bleiholder, C., Bowers, M. T., Bilbao, A., Bush, M. F., Campbell, J. L., et al. (2019) · Mass Spectrometry Reviews · doi:10.1002/mas.21585 · Cited by 493 · cited 9×
“Since this is not always possible, CCS is a method-dependent value [20].”
BackgroundMethodInfluentialIon mobility mass spectrometry of proteins and proteinassemblies
Uetrecht, C., Rose, R. J., van Duijn, E., Lorenzen, K., & Heck, A. J. R. (2010) · Chem. Soc. Rev. · doi:10.1039/b914002f · Cited by 462
Review on ion mobility spectrometry. Part 1: current instrumentation
Cumeras, R., Figueras, E., Davis, C. E., Baumbach, J. I., & Gràcia, I. (2015) · The Analyst · doi:10.1039/c4an01100g · Cited by 459
Fundamentals of ion mobility spectrometry
Gabelica, V., & Marklund, E. (2018) · Current Opinion in Chemical Biology · doi:10.1016/j.cbpa.2017.10.022 · Cited by 440 · cited 4×
“fields (E/N > 4–10 Townsends (Td); 1 Td = 10−21 V m2) and Equation (3) is no longer applicable [10].”
BackgroundInfluentialIon Mobility-Mass Spectrometry: Time-Dispersive Instrumentation
May, J. C., & McLean, J. A. (2015) · Analytical Chemistry · doi:10.1021/ac504720m · Cited by 435 · cited 1×
“, ion confinement and release) technologies [18].”
Lapthorn, C., Pullen, F., & Chowdhry, B. Z. (2013) · Mass Spectrometry Reviews · doi:10.1002/mas.21349 · Cited by 398 · cited 1×
“In this context, the integration of IMS in LC–MS workflows introduces a third separation dimension that improves peak capacity and allows the separation of isobars and isomers [33].”
BackgroundFundamentals of ion mobility spectrometry
Gabelica, V., & Marklund, E. G. (2018) · Archive ouverte UNIGE (University of Geneva) · Cited by 375
Fundamentals of Trapped Ion Mobility Spectrometry
Michelmann, K., Silveira, J. A., Ridgeway, M. E., & Park, M. A. (2015) · Journal of the American Society for Mass Spectrometry · doi:10.1007/s13361-014-0999-4 · Cited by 374 · cited 2×
“More detailed information about the operation and physical principles of the different IMS forms can be found in specialized literature [17,19,25,27,28].”
BackgroundReview on metabolomics for food authentication
Cubero-Leon, E., Peñalver, R., & Maquet, A. (2014) · Food Research International · doi:10.1016/j.foodres.2013.11.041 · Cited by 334 · cited 1×
“In this sense, it is very efficient and has already shown clear benefits over traditional methods [77].”
MethodIon Mobility Derived Collision Cross Sections to Support Metabolomics Applications
Paglia, G., Williams, J. P., Menikarachchi, L., Thompson, J. W., Tyldesley-Worster, R., Halldórsson, S., Rolfsson, O., Moseley, A., Grant, D., Langridge, J., et al. (2014) · Analytical Chemistry · doi:10.1021/ac500405x · Cited by 319 · cited 1×
“Inter-laboratory reproducibility has barely been evaluated, and only few studies have tackled this issue [22,112,136].”
BackgroundTrapped ion mobility spectrometry: A short review
Ridgeway, M. E., Lubeck, M., Jordens, J., Mann, M., & Park, M. A. (2018) · International Journal of Mass Spectrometry · doi:10.1016/j.ijms.2018.01.006 · Cited by 308 · cited 1×
“, 45–150 Td) is applied to trap the ions and they are only dragged through the drift tube towards the detector by the gas [24].”
BackgroundCollision cross section compendium to annotate and predict multi-omic compound identities
Picache, J. A., Rose, B. S., Balinski, A., Leaptrot, K. L., Sherrod, S. D., May, J. C., & McLean, J. A. (2019) · Chemical Science · doi:10.1039/c8sc04396e · Cited by 279 · cited 4×
“In an effort to overcome this issue, several CCS databases have been reported in the last years [15], although they are not enough considering the number of molecules that remain uncharacterized.”
BackgroundInfluentialMetabolomics and lipidomics using traveling-wave ion mobility mass spectrometry
Paglia, G., & Astarita, G. (2017) · Nature Protocols · doi:10.1038/nprot.2017.013 · Cited by 248 · cited 1×
“A protocol has been reported for TWIMS calibration and operation [137], but different calibration procedures and calibrants are applied in other IMS systems.”
MethodIon mobility spectrometry: a review. Part 1. Structural analysis by mobility measurement
Creaser, C. S., Griffiths, J. R., Bramwell, C. J., Noreen, S., Hill, C. A., & Thomas, C. L. P. (2004) · The Analyst · doi:10.1039/b404531a · Cited by 235 · cited 1×
“Therefore, CCS is widely used for structural elucidation since it provides knowledge about the three dimensional (3D) conformation of ions in the gas phase [13,14].”
BackgroundDeng, L., Webb, I. K., Garimella, S. V. B., Hamid, A. M., Zheng, X., Norheim, R. V., Prost, S. A., Anderson, G. A., Sandoval, J. A., Baker, E. S., et al. (2017) · Analytical Chemistry · doi:10.1021/acs.analchem.7b00185 · Cited by 228 · cited 1×
“As a result of the improvements on Rp expected in UHRIMS technology, IMS separations will be able to provide similar peak capacities as LC at lower analysis times, which will be revolutionary from an analytical point of view and transformative for several applications [121].”
BackgroundSchneider, B. B., Nazarov, E. G., Londry, F., Vouros, P., & Covey, T. R. (2016) · Mass Spectrometry Reviews · doi:10.1002/mas.21453 · Cited by 195 · cited 3×
“More detailed information about the operation and physical principles of the different IMS forms can be found in specialized literature [17,19,25,27,28].”
BackgroundResultHines, K. M., May, J. C., McLean, J. A., & Xu, L. (2016) · Analytical Chemistry · doi:10.1021/acs.analchem.6b01728 · Cited by 188 · cited 1×
“Therefore, more studies about CCS calibration are still needed since, in the case of some classes of compounds, the accuracy of CCS measurements depends on the chemical nature of the calibrant [139].”
BackgroundAdding a new separation dimension to MS and LC–MS: What is the utility of ion mobility spectrometry?
D'Atri, V., Causon, T., Hernandez‐Alba, O., Mutabazi, A., Veuthey, J., Cianferani, S., & Guillarme, D. (2018) · Journal of Separation Science · doi:10.1002/jssc.201700919 · Cited by 184 · cited 6×
“However, LC–IM–MS platforms are currently becoming very popular and are used for numerous applications [4].”
BackgroundMethodInfluentialIon Mobility Collision Cross Section Compendium
May, J. C., Morris, C. B., & McLean, J. A. (2017) · Analytical Chemistry · doi:10.1021/acs.analchem.6b04905 · Cited by 183 · cited 4×
“accepted to indicate the type of IMS technology and drift gas used for CCS measurements [11].”
BackgroundMethodInfluentialDodds, J. N., May, J. C., & McLean, J. A. (2017) · Analytical Chemistry · doi:10.1021/acs.analchem.7b02827 · Cited by 182
Gerhardt, N., Schwolow, S., Rohn, S., Pérez-Cacho, P. R., Galán-Soldevilla, H., Arce, L., & Weller, P. (2019) · Food Chemistry · doi:10.1016/j.foodchem.2018.11.095 · Cited by 170 · cited 11×
“05 ppm, which is usually below th dor threshold of many compounds in extra virgin olive oils (EVOO) [53].”
BackgroundInfluentialGuo, Y. M., Chen, D., Dong, Y., Ju, H., Wu, C., & Lin, S. (2018) · Journal of Chromatography B · doi:10.1016/j.jchromb.2018.09.011 · Cited by 170 · cited 3×
“In addition to olive oils, HS–GC–DTIMS has also been applied to characterize the volatile fingerprint of fresh and dried Tricholoma matsutake Singer samples [50] and green tea aromas [55].”
BackgroundMethodGerhardt, N., Birkenmeier, M., Sanders, D., Rohn, S., & Weller, P. (2017) · Analytical and Bioanalytical Chemistry · doi:10.1007/s00216-017-0338-2 · Cited by 167 · cited 4×
“, 98% and 92% by PCA-linear discriminant analysis (LDA) and PCA-k-nearest neighbors (kNN), respectively) have been reached to effectively discriminate between EVOO from Italy and Spain [80].”
MethodInfluentialIon mobility spectrometry detection for gas chromatography
Kanu, A. B., & Hill, H. H. (2008) · Journal of Chromatography A · doi:10.1016/j.chroma.2007.10.110 · Cited by 161 · cited 2×
“Regarding GC–IMS methods, IMS has been useful to separate pesticides that were not resolved in the chromatographic dimension [100,101], although in general it merely acts as detector [32].”
MethodIon mobility spectrometry coupled to gas chromatography: A rapid tool to assess eggs freshness
Cavanna, D., Zanardi, S., Dall'Asta, C., & Suman, M. (2019) · Food Chemistry · doi:10.1016/j.foodchem.2018.07.204 · Cited by 160 · cited 4×
“In addition to the selective analysis of specific volatile substances, the profiling of VOCs fraction by IMS approaches can provide more complete information about the freshness and storage conditions of food products such as fish [71] and eggs [72].”
BackgroundInfluentialHigh-Resolution Ion Cyclotron Mobility Spectrometry
Merenbloom, S. I., Glaskin, R. S., Henson, Z. B., & Clemmer, D. E. (2009) · Analytical Chemistry · doi:10.1021/ac801880a · Cited by 149 · cited 1×
“In addition, a third alternative was previously explored, namely the high resolution ion cyclotron mobility spectrometry (cyclic-DTIMS) [122], which is a circular 180.”
BackgroundCyclic Ion Mobility Mass Spectrometry Distinguishes Anomers and Open-Ring Forms of Pentasaccharides
Ujma, J., Ropartz, D., Giles, K., Richardson, K., Langridge, D., Wildgoose, J. L., Green, M., & Pringle, S. D. (2019) · Journal of the American Society for Mass Spectrometry · doi:10.1007/s13361-019-02168-9 · Cited by 148 · cited 1×
“In cyclic-TWIMS, the drift cell is arranged in a circle configuration with pre- and post-store cells for ejecting the ions in and out [124,125].”
BackgroundSelf-citationIon mobility spectrometry for food quality and safety
Vautz, W., Zimmermann, D., Hartmann, M., Baumbach, J., Nolte, J., & Jung, J. (2006) · Food Additives & Contaminants · doi:10.1080/02652030600889590 · Cited by 146 · cited 3×
“GC–IMS for their determination allows decreasing the analysis time from 3 h to 10 min [5].”
BackgroundInfluentialHernández‐Mesa, M., Escourrou, A., Monteau, F., Bizec, B. L., & Dervilly, G. (2017) · TrAC Trends in Analytical Chemistry · doi:10.1016/j.trac.2017.07.006 · Cited by 145 · cited 7×
“Despite its potential to support compound identification, the lack of CCS libraries is considered to be the major drawback for reaching this goal and implementing the CCS parameter in current analytical workflows [8].”
BackgroundMethodSelf-citationHines, K. M., Ross, D. H., Davidson, K. L., Bush, M. F., & Xu, L. (2017) · Analytical Chemistry · doi:10.1021/acs.analchem.7b01709 · Cited by 135
The potential of ion mobility–mass spectrometry for non-targeted metabolomics
Mairinger, T., Causon, T. J., & Hann, S. (2017) · Current Opinion in Chemical Biology · doi:10.1016/j.cbpa.2017.10.015 · Cited by 135 · cited 1×
“, metabolomics, lipidomics) and, ultimately, for compound identification [15,16].”
Deng, L., Ibrahim, Y., Baker, E. S., Aly, N. A., Hamid, A. M., Zhang, X., Zheng, X., Garimella, S., Webb, I., Prost, S. A., et al. (2016) · ChemistrySelect · doi:10.1002/slct.201600460 · Cited by 131 · cited 2×
“it has been shown to i proved separations of lipid and peptide is mers [128] as well as glycans that”
BackgroundMethodIon mobility spectrometry fingerprints: A rapid detection technology for adulteration of sesame oil
Zhang, L., Qian, S., Li, P., Zhang, Q., Ma, F., Zhang, W., & Ding, X. (2015) · Food Chemistry · doi:10.1016/j.foodchem.2015.06.096 · Cited by 129 · cited 1×
“In this context, stand-alone in combination with chemometrics has been investigated for the rapid detection of EVOO [88], sesame oil [89], and flaxseed oil [90] adulteration with vegetable oils of lower quality.”
BackgroundHinnenkamp, V., Klein, J., Meckelmann, S. W., Balsaa, P., Schmidt, T. C., & Schmitz, O. J. (2018) · Analytical Chemistry · doi:10.1021/acs.analchem.8b02711 · Cited by 127 · cited 1×
“1% between DTCCS and TWCCS values, both values cannot be compared in all cases [135].”
BackgroundRegueiro, J., Negreira, N., & Berntssen, M. H. G. (2016) · Analytical Chemistry · doi:10.1021/acs.analchem.6b03381 · Cited by 124 · cited 3×
“Recent studies have shown high repeatability over time [30,112], and negligible impact of sample matrix on CCS values [39,113].”
BackgroundMethodResultBijlsma, L., Bade, R., Celma, A., Mullin, L. G., Cleland, G., Stead, S. L., Hernández, F. J., & Sancho, J. V. (2017) · Analytical Chemistry · doi:10.1021/acs.analchem.7b00741 · Cited by 121 · cited 1×
“For example, a machine-learning model based on artificial neural networks (ANNs) has been developed for the CCS prediction of small molecules selecting 205 compounds for model training, verification and blind test sets (ratio 68:16:16) [131].”
MethodCapabilities and limitations of ion mobility spectrometry for field screening applications
Hill, H. H., & Simpson, G. (1997) · Field Analytical Chemistry & Technology · doi:10.1002/(sici)1520-6521(1997)1:3<119::aid-fact2>3.0.co;2-s · Cited by 121 · cited 1×
“From a technical point of view, limited dynamic range and ion loss have been traditionally attributed to LC–IM–MS in comparison to LC–MS methods [36].”
ResultHamid, A. M., Ibrahim, Y., Garimella, S., Webb, I., Deng, L., Chen, T., Anderson, G., Prost, S. A., Norheim, R. V., Tolmachev, A. V., et al. (2015) · Analytical Chemistry · doi:10.1021/acs.analchem.5b02481 · Cited by 106 · cited 2×
“Ions travel through a serpentine path and are confined due to the application of DC and RF potentials [127].”
BackgroundHaynes, S. E., Polasky, D. A., Dixit, S. M., Majmudar, J. D., Neeson, K., Ruotolo, B. T., & Martin, B. R. (2017) · Analytical Chemistry · doi:10.1021/acs.analchem.7b00112 · Cited by 105 · cited 1×
“The implementation of IMS in LC–MS workflows increase peak capacity at least 2 or 3-fold [34,35], although this improvement ultimately depends on IMS resolution and the analytical application.”
BackgroundFaleh, A. B., Warnke, S., & Rizzo, T. R. (2019) · Analytical Chemistry · doi:10.1021/acs.analchem.9b00659 · Cited by 105 · cited 2×
“2% [129], which can have potential applicability in food composition analysis and food fingerprinting.”
BackgroundHybrid ion mobility and mass spectrometry as a separation tool
Ewing, M. A., Glover, M. S., & Clemmer, D. E. (2015) · Journal of Chromatography A · doi:10.1016/j.chroma.2015.10.080 · Cited by 104 · cited 5×
“Finally, FAIMS and DMS instruments do not separate ions in function of their mobility in a neutral gas as in DTIMS, TWIMS or TIMS, but rather by the ratio of low-field to high-field mobilities [17].”
BackgroundMethodInfluentialZheng, X., Wojcik, R., Zhang, X., Ibrahim, Y. M., Burnum-Johnson, K. E., Orton, D. J., Monroe, M. E., Moore, R. J., Smith, R. D., & Baker, E. S. (2017) · Annual Review of Analytical Chemistry · doi:10.1146/annurev-anchem-061516-045212 · Cited by 103 · cited 3×
“sensitivity, peak capacity, and compound identification [3,4].”
MethodKirk, A. T., Bohnhorst, A., Raddatz, C., Allers, M., & Zimmermann, S. (2019) · Analytical and Bioanalytical Chemistry · doi:10.1007/s00216-019-01807-0 · Cited by 101 · cited 6×
“Rp of DTIMS systems is around 100 whereas TWIMS instruments currently reach a maximum Rp of 40 [14,23].”
BackgroundInfluentialCollision Cross Section (CCS) Database: An Additional Measure to Characterize Steroids
Hernández‐Mesa, M., Bizec, B. L., Monteau, F., García-Campaña, A. M., & Dervilly, G. (2018) · Analytical Chemistry · doi:10.1021/acs.analchem.7b05117 · Cited by 100
Self-citationGlycan analysis by ion mobility-mass spectrometry and gas-phase spectroscopy
Manz, C., & Pagel, K. (2017) · Current Opinion in Chemical Biology · doi:10.1016/j.cbpa.2017.10.021 · Cited by 96
Hofmann, J., Struwe, W. B., Scarff, C. A., Scrivens, J. H., Harvey, D. J., & Pagel, K. (2014) · Analytical Chemistry · doi:10.1021/ac5028353 · Cited by 94 · cited 1×
“Other alternative CCS calibrants such as the dextran have also been investigated for obtaining the CCS of carbohydrates [138].”
BackgroundKurulugama, R. T., Darland, E. J., Kuhlmann, F. E., Stafford, G. C., & Fjeldsted, J. C. (2015) · The Analyst · doi:10.1039/c5an00991j · Cited by 92 · cited 1×
“Specifically, the nature of buffer gas has a great influence on the CCS of molecules [132], and CCS values measured in different drift gases cannot be directly compared, as shown for a wide variety of pesticides whose CCSs have been obtained in He, CO2, N2O and SF6 by DTIMS [133].”
BackgroundSaraji, M., Jafari, M. T., & Mossaddegh, M. (2015) · Journal of Chromatography A · doi:10.1016/j.chroma.2015.12.008 · Cited by 91 · cited 1×
“Regarding GC–IMS methods, IMS has been useful to separate pesticides that were not resolved in the chromatographic dimension [100,101], although in general it merely acts as detector [32].”
MethodDetermination of aflatoxins B1 and B2 using ion mobility spectrometry
Sheibani, A., Tabrizchi, M., & Ghaziaskar, H. S. (2007) · Talanta · doi:10.1016/j.talanta.2007.11.006 · Cited by 86 · cited 1×
“, aflatoxins B1 and B2) in pistachios [98], whereas MISPE has been use to study the uptake and translocation of a neonicotinoid pesticide (i.”
Advancing the large-scale CCS database for metabolomics and lipidomics at the machine-learning era
Zhou, Z., Tu, J., & Zhu, Z. (2017) · Current Opinion in Chemical Biology · doi:10.1016/j.cbpa.2017.10.033 · Cited by 84 · cited 2×
“In order to extend the knowledge provided by CCS databases created experimentally, other strategies such as computational modeling and, more recently, machine-learning based prediction have also been investigated [130].”
BackgroundBlaženović, I., Shen, T., Mehta, S. S., Kind, T., Ji, J., Piparo, M., Cacciola, F., Mondello, L., & Fiehn, O. (2018) · Analytical Chemistry · doi:10.1021/acs.analchem.8b01527 · Cited by 82 · cited 6×
“correlated [44,46] (Figure 3), so their coupling improves peak capacity and enables the detection of a”
BackgroundInfluentialAlves, T. D. O., D’Almeida, C. T. D. S., Scherf, K. A., & Ferreira, M. S. L. (2019) · Frontiers in Plant Science · doi:10.3389/fpls.2019.01470 · Cited by 75
Struwe, W. B., Baldauf, C., Hofmann, J., Rudd, P. M., & Pagel, K. (2016) · Chemical Communications · doi:10.1039/c6cc06247d · Cited by 75 · cited 1×
“However, both carbohydrates are baseline resolved by IMS under negative ionization conditions [57].”
BackgroundZheng, X., Dupuis, K. T., Aly, N. A., Zhou, Y., Smith, F. B., Tang, K., Smith, R. D., & Baker, E. S. (2018) · Analytica Chimica Acta · doi:10.1016/j.aca.2018.02.054 · Cited by 75 · cited 1×
“While, until a few years ago, CCS had been barely used as verification parameter in screening approaches due to the lack of databases, this situation has now recently changed and several CCS databases for pesticides [112,113], veterinary drugs [114–116], mycotoxins [117] and other contaminants [108], have been reporte…”
ResultIon Trapping for Ion Mobility Spectrometry Measurements in a Cyclical Drift Tube
Glaskin, R. S., Ewing, M. A., & Clemmer, D. E. (2013) · Analytical Chemistry · doi:10.1021/ac4015066 · Cited by 73 · cited 1×
“Successive improved versions of this technology have been shown to reach a Rp in excess of 1000 (as a consequence of 100 transits or cycles and involving a drift length of over 180 m) [123].”
MethodIon mobility-derived collision cross section database: Application to mycotoxin analysis
Righetti, L., Bergmann, A., Galaverna, G., Rolfsson, Ó., Paglia, G., & Dall’Asta, C. (2018) · Analytica Chimica Acta · doi:10.1016/j.aca.2018.01.047 · Cited by 72 · cited 2×
“Finally, recent applications of LC–ESI–TWIMS–ToF–MS methods, which include the CCS as signal filter in data processing, report the analysis of mycotoxins in cereals [117] and pesticides in green tea powder, fresh garlic, leek, fresh herb chives and rye [112], fish feed [31], and in 20 food products (e.”
MethodResultTu, J., Zhou, Z., Li, T., & Zhu, Z. (2019) · TrAC Trends in Analytical Chemistry · doi:10.1016/j.trac.2019.03.017 · Cited by 71
Venter, P. B., Muller, M., Vestner, J., Stander, M. A., Tredoux, A. G. J., Pasch, H., & de Villiers, A. J. (2018) · Analytical Chemistry · doi:10.1021/acs.analchem.8b03234 · Cited by 70 · cited 1×
“, flavonoid and non-flavonid phenolics) [58].”
Goscinny, S., Joly, L., Pauw, E. D., Hanot, V., & Eppe, G. (2015) · Journal of Chromatography A · doi:10.1016/j.chroma.2015.05.057 · Cited by 64 · cited 1×
“For example, drift time has been used to identify 100 pesticides in different vegetables and fruits by LC–TWIMS–ToF–MS [111].”
MethodOn-line analysis of coffee roasting with ion mobility spectrometry–mass spectrometry (IMS–MS)
Gloess, A. N., Yeretzian, C., Knochenmuss, R., & Groessl, M. (2017) · International Journal of Mass Spectrometry · doi:10.1016/j.ijms.2017.11.017 · Cited by 61 · cited 2×
“MS has been proposed for the on-line monitoring of VOCs formation in coffee roasting processes [75].”
BackgroundBauer, A., Luetjohann, J., Hanschen, F. S., Schreiner, M., Kuballa, J., Jantzen, E., & Rohn, S. (2017) · Food Chemistry · doi:10.1016/j.foodchem.2017.09.131 · Cited by 60 · cited 1×
“Other LC-TWIMS-MS applications in the context of food safety encompass the identification and structural characterization of residues and contaminants, either for evaluating exposure [118] or for studying their biotransformation [119].”
MethodBeach, D. G., Kerrin, E. S., & Quilliam, M. A. (2015) · Analytical and Bioanalytical Chemistry · doi:10.1007/s00216-015-9012-8 · Cited by 60 · cited 1×
“This approach has been followed for the separation of the neurotoxin β-N-methylamino-l-alanine (BMAA) and its isomer β-amino-N-methylalanine (BAMA) in mussel tissues [110].”
BackgroundHernández‐Mesa, M., Monteau, F., Bizec, B. L., & Dervilly, G. (2019) · Analytica Chimica Acta X · doi:10.1016/j.acax.2019.100006 · Cited by 58 · cited 7×
“In the context of chemical food safety, signal to noise ratio (S/N) of LC–ESI–ToF–MS methods intended to monitoring steroids and their metabolites in livestock were increased between 2 and 7 fold by the integration of TWIMS in the analytical workflow (Figure 5) [30].”
BackgroundMethodSelf-citationDetection of Adulteration in Canola Oil by Using GC-IMS and Chemometric Analysis
Chen, T., Chen, X., Lu, D., & Chen, B. (2018) · International Journal of Analytical Chemistry · doi:10.1155/2018/3160265 · Cited by 57 · cited 1×
“, sunflower, soybean, and peanut oils) [86], and crude palm oil with process byproducts of lower quality (i.”
BackgroundKermani, M., Jafari, M. T., & Saraji, M. (2019) · Microchimica Acta · doi:10.1007/s00604-018-3215-6 · Cited by 56 · cited 1×
“Regarding GC–IMS methods, IMS has been useful to separate pesticides that were not resolved in the chromatographic dimension [100,101], although in general it merely acts as detector [32].”
MethodTejada‐Casado, C., Hernández‐Mesa, M., Monteau, F., Lara, F. J. S., del Olmo‐Iruela, M., García-Campaña, A. M., Bizec, B. L., & Dervilly, G. (2018) · Analytica Chimica Acta · doi:10.1016/j.aca.2018.09.065 · Cited by 56 · cited 1×
“, mixture of small molecules such as caffeine or sulfadimethoxine, and including poly-DL-alanine) [116] are currently widely used as CCS calibrants.”
MethodSelf-citationRegueiro, J., Negreira, N., Hannisdal, R., & Berntssen, M. H. G. (2017) · Food Control · doi:10.1016/j.foodcont.2017.02.053 · Cited by 55 · cited 2×
“Finally, recent applications of LC–ESI–TWIMS–ToF–MS methods, which include the CCS as signal filter in data processing, report the analysis of mycotoxins in cereals [117] and pesticides in green tea powder, fresh garlic, leek, fresh herb chives and rye [112], fish feed [31], and in 20 food products (e.”
BackgroundMethodZhang, H., Jiang, J.-M., Zheng, D., Yuan, M., Wang, Z., Zhang, H., Zheng, C., Xiao, L., & Xu, H. (2018) · Journal of Pharmaceutical and Biomedical Analysis · doi:10.1016/j.jpba.2018.09.036 · Cited by 55 · cited 1×
“For example, 2D–ultra-high performance liquid chromatography (UHPLC)–TWIMS–ToF–MS has been applied to the characterization of ginsenosides in white and red ginsengs which are diet supplements, and from the 201 compounds identified, 10 pairs of co-eluting isobaric ginsenosides were resolved only by IMS [60].”
MethodHits and misses in research trends to monitor contaminants in foods
Lehotay, S. J., & Chen, Y. (2018) · Analytical and Bioanalytical Chemistry · doi:10.1007/s00216-018-1195-3 · Cited by 52 · cited 1×
“, toxins) or anthropogenic contaminants such as persistent organic pollutants (POPs) [1].”
BackgroundAhmed, E., Kabir, K. M. M., Wang, H., Xiao, D., Fletcher, J., & Donald, W. A. (2019) · Analytica Chimica Acta · doi:10.1016/j.aca.2019.01.038 · Cited by 52 · cited 2×
“separation of isomeric perfluoroalkyl substances (PFAS) by DMS–trap quadrupole (LTQ)–MS [104].”
BackgroundMethodKaufmann, A., & Walker, S. (2017) · Rapid Communications in Mass Spectrometry · doi:10.1002/rcm.7981 · Cited by 46 · cited 1×
“However, several technological advances have been accomplished in the last years and, for example, some recent studies indicate that hyphenation with an ion mobility device does not affect the linear dynamic range [37].”
BackgroundReal-time 2D separation by LC × differential ion mobility hyphenated to mass spectrometry
Varesio, E., Blanc, J. C. Y. L., & Gérard Hopfgartner, D. (2011) · Analytical and Bioanalytical Chemistry · doi:10.1007/s00216-011-5444-y · Cited by 45 · cited 1×
“) can be added into the buffer gas in DMS and FAIMS systems in order to improve ion mobility separation [109].”
BackgroundMultivariate modelling of fish freshness index based on ion mobility spectrometry measurements
Raatikainen, O. J., Рейникайнен, В., Minkkinen, P. O., Ritvanen, T., Muje, P., Pursiainen, J., Hiltunen, T., Hyvönen, P., von Wright, A., & Рейникайнен, С.-П. (2005) · Analytica Chimica Acta · doi:10.1016/j.aca.2005.02.029 · Cited by 45 · cited 1×
“In addition to the selective analysis of specific volatile substances, the profiling of VOCs fraction by IMS approaches can provide more complete information about the freshness and storage conditions of food products such as fish [71] and eggs [72].”
BackgroundLipidomic Response to Coffee Consumption
Kuang, A., Erlund, I., Herder, C., Westerhuis, J. A., Tuomilehto, J. O., & Cornelis, M. C. (2018) · Nutrients · doi:10.3390/nu10121851 · Cited by 43 · cited 1×
“, 853 lipid species from 14 lipid classes) have been investigated by DMS-triple quadrupole (QqQ)/linear ion trap (LIT)–MS [43].”
BackgroundGoscinny, S., McCullagh, M., Far, J., Pauw, E. D., & Eppe, G. (2019) · Rapid Communications in Mass Spectrometry · doi:10.1002/rcm.8395 · Cited by 43 · cited 6×
“Finally, recent applications of LC–ESI–TWIMS–ToF–MS methods, which include the CCS as signal filter in data processing, report the analysis of mycotoxins in cereals [117] and pesticides in green tea powder, fresh garlic, leek, fresh herb chives and rye [112], fish feed [31], and in 20 food products (e.”
BackgroundMethodResultInfluentialBauer, A., Kuballa, J., Rohn, S., Jantzen, E., & Luetjohann, J. (2018) · Journal of Separation Science · doi:10.1002/jssc.201701059 · Cited by 43 · cited 6×
“Finally, recent applications of LC–ESI–TWIMS–ToF–MS methods, which include the CCS as signal filter in data processing, report the analysis of mycotoxins in cereals [117] and pesticides in green tea powder, fresh garlic, leek, fresh herb chives and rye [112], fish feed [31], and in 20 food products (e.g., strawberries…”
BackgroundMethodInfluentialLi, J., Yuan, H., Yao, Y., Hua, J., Yang, Y., Dong, C., Deng, Y., Wang, J., Li, H., Jiang, Y., et al. (2018) · Talanta · doi:10.1016/j.talanta.2018.08.039 · Cited by 42 · cited 1×
“In addition to olive oils, HS–GC–DTIMS has also been applied to characterize the volatile fingerprint of fresh and dried Tricholoma matsutake Singer samples [50] and green tea aromas [55].”
MethodLópez‐Morales, C. A., Vázquez‐Leyva, S., Vallejo‐Castillo, L., Carballo‐Uicab, G., Muñoz‐García, L., Pucheta, J. E. H., Zepeda, L. G., Velasco-Velázquez, M. A., Pavón, L., Pérez‐Tapia, S. M., et al. (2019) · Journal of Food Science · doi:10.1111/1750-3841.14466 · Cited by 36
Lipok, C., Hippler, J., & Schmitz, O. J. (2017) · Journal of Chromatography A · doi:10.1016/j.chroma.2017.07.013 · Cited by 35
Exploring the Complexity of Steviol Glycosides Analysis Using Ion Mobility Mass Spectrometry
McCullagh, M., Douce, D. S., Hoeck, E. V., & Goscinny, S. (2018) · Analytical Chemistry · doi:10.1021/acs.analchem.7b05002 · Cited by 30 · cited 2×
“, CCS), giving a more complete overview of food composition [44,62].”
BackgroundRighetti, L., Fenclová, M., Dellafiora, L., Hajšlová, J., Stranska-Zachariasova, M., & Dall’Asta, C. (2017) · Food Chemistry · doi:10.1016/j.foodchem.2017.11.113 · Cited by 30 · cited 1×
“Other LC-TWIMS-MS applications in the context of food safety encompass the identification and structural characterization of residues and contaminants, either for evaluating exposure [118] or for studying their biotransformation [119].”
MethodCharacterization of aroma profile and characteristic aromas during lychee wine fermentation
Tang, Z., Zeng, X., Brennan, M. A., Han, Z., Niu, D., & Huo, Y. (2019) · Journal of Food Processing and Preservation · doi:10.1111/jfpp.14003 · Cited by 29 · cited 1×
“GC-I S has also been applied to the characterizatio of V s fraction during the fer e tati of lychee bevera es [73] and ine [74].”
MethodIon Mobility-Mass Spectrometry in Metabolomic, Lipidomic, and Proteomic Analyses
Chouinard, C. D., Nagy, G., Smith, R. D., & Baker, E. S. (2018) · Comprehensive analytical chemistry · doi:10.1016/bs.coac.2018.11.001 · Cited by 27 · cited 2×
“Several prototypes based on DTIMS/TWIMS–FTICR/Orbitrap–MS hyphenation have already been proposed; so further developments are expected [41].”
BackgroundMethodHalbfeld, C., Ebert, B. E., & Blank, L. M. (2014) · Metabolites · doi:10.3390/metabo4030751 · Cited by 25 · cited 3×
“GC-I S has also been applied to the characterizatio of V s fraction during the fer e tati of lychee bevera es [73] and ine [74].”
BackgroundMethodAlves, T. D. O., D’Almeida, C. T. D. S., Victorio, V. C. M., Souza, G. H. M. F., Cameron, L. C., & Ferreira, M. S. L. (2018) · Journal of Food Composition and Analysis · doi:10.1016/j.jfca.2018.07.012 · Cited by 21 · cited 2×
“including allergens [48], and samples (e.”
BackgroundOranzi, N. R., Kemperman, R. H. J., Wei, M. S., Petkovska, V. I., Granato, S. W., Rochon, B., Kaszycki, J. L., Rotta, A. L., Fouque, K. J. D., Fernandez-Lima, F. A., et al. (2019) · Analytical Chemistry · doi:10.1021/acs.analchem.8b05723 · Cited by 19 · cited 1×
“In general, similar CCS values are provided by these three IMS technologies when using the same buffer gas as observed for the sodium adduct of 25-hydroxyvitamin D3 [134].”
Detection and identification of sugar alcohol sweeteners by ion mobility spectrometry
Browne, C. A., Forbes, T. P., & Sisco, E. (2016) · Analytical Methods · doi:10.1039/c6ay01554a · Cited by 17 · cited 1×
“As an example, stand-alone IMS has been used for the analysis of seven alcohol sweeteners in chewing gum, identifying the presence of sorbitol [51].”
MethodXu, Z., Li, J., Chen, A., Ma, X., & Yang, S. (2018) · Rapid Communications in Mass Spectrometry · doi:10.1002/rcm.8154 · Cited by 14 · cited 2×
“Despite the characteristic of ion filtering is usually assigned to FAIMS/DMS technology, DTIMS and TWIMS are also able to isolate analyte signals from chemical background, improving concentration sensitivity [30,103].”
BackgroundLu, Y., Guo, J., Yu, J., Guo, J., Jia, X., Liu, W., & Tian, P. (2018) · Analytical Methods · doi:10.1039/c8ay02288g · Cited by 12 · cited 1×
“Nevertheless, the determination of phenolic acids in seedling roots by high-performance liquid chromatography (HPLC)–ESI–DTIMS has been recently reported [56], and this application can be extended to food composition analysis.”
Method
Metadata
- Authors
- Hernández-Mesa, M., Ropartz, D., García-Campaña, A. M., Rogniaux, H., Dervilly-Pinel, G., & Le Bizec, B.
- Venue
- Molecules
- Publisher
- MDPI AG
- Published
- 2019-07-25
- Volume
- 24
- Issue
- 15
- Pages
- 2706
- ISSN
- 1420-3049 (ISSN-L)
- OpenAlex
- W2963457189
- Cited by
- 202
- References
- 95
- FWCI
- 9.05
- Type
- journal article
- Language
- en
- Open access
- gold
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.