Optical Fibre-Based Pulse Oximetry Sensor with Contact Force Detection
Open accessAbstract
A novel optical sensor probe combining monitoring of blood oxygen saturation (SpO2) with contact pressure is presented. This is beneficial as contact pressure is known to affect SpO2 measurement. The sensor consists of three plastic optical fibres (POF) used to deliver and collect light for pulse oximetry, and a fibre Bragg grating (FBG) sensor to measure contact pressure. All optical fibres are housed in a biocompatible epoxy patch which serves two purposes: (i) to reduce motion artefacts in the photoplethysmogram (PPG), and (ii) to transduce transverse loading into an axial strain in the FBG. Test results show that using a combination of pressure measuring FBG with a reference FBG, reliable results are possible with low hysteresis which are relatively immune to the effects of temperature. The sensor is used to measure the SpO2 of ten volunteers under different contact pressures with perfusion and skewness indices applied to assess the quality of the PPG. The study revealed that the contact force ranging from 5 to 15 kPa provides errors of <2%. The combined probe has the potential to improve the reliability of reflectance oximeters. In particular, in wearable technology, the probe should find use in optimising the fitting of garments incorporating this technology.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
When pressures are kept within the range of 5-15 kPa, there is less than 2% error in the SpO2 measurements.
“Utilizing this modality, they demonstrated that, when pressures are kept within the range of 5–15 kPa, there is less than 2% error in the SpO2 measurements.”
From Sources of inaccuracy in photoplethysmography for continuous cardiovascular monitoring (2021) · cites this paper as [215]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
Photoplethysmography and its application in clinical physiological measurement
Allen, J. (2007) · Physiological Measurement · doi:10.1088/0967-3334/28/3/r01 · Cited by 3,967 · see citing works
Wearable photoplethysmographic sensors—past and present
Tamura, T., Maeda, Y., Sekine, M., & Yoshida, M. (2014) · Electronics · doi:10.3390/electronics3020282 · Cited by 912 · see citing works
E-broidery: Design and fabrication of textile-based computing
Post, E. R., Orth, M., Russo, P. R., & Gershenfeld, N. (2000) · IBM Systems Journal · doi:10.1147/sj.393.0840 · Cited by 623 · see citing works
(1997) · Institute of Physics eBooks · doi:10.1887/0750304677 · Cited by 455 · see citing works
A Novel Approach for Motion Artifact Reduction in PPG Signals Based on AS-LMS Adaptive Filter
Ram, M. R., Madhav, K. V., Krishna, E. H., Komalla, N. R., & Reddy, K. A. (2011) · IEEE Transactions on Instrumentation and Measurement · doi:10.1109/tim.2011.2175832 · Cited by 345 · see citing works
Optimal Signal Quality Index for Photoplethysmogram Signals
Elgendi, M. (2016) · Bioengineering · doi:10.3390/bioengineering3040021 · Cited by 338 · see citing works
A Review of Approaches to Mobility Telemonitoring of the Elderly in Their Living Environment
Scanaill, C. N., Carew, S., Barralon, P., Noury, N., Lyons, D., & Lyons, G. M. (2006) · Annals of Biomedical Engineering · doi:10.1007/s10439-005-9068-2 · Cited by 265 · see citing works
Two-Stage Approach for Detection and Reduction of Motion Artifacts in Photoplethysmographic Data
Krishnan, R., Natarajan, B., & Warren, S. (2010) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2009.2039568 · Cited by 263 · see citing works
Non-contact detection of oxygen saturation based on visible light imaging device using ambient light
Kong, L., Zhao, Y., Dong, L., Jian, Y., Jin, X., Li, B., Feng, Y., Liu, M., Liu, X., & Wu, H. R. (2013) · Optics Express · doi:10.1364/oe.21.017464 · Cited by 187 · see citing works
The effect of contacting force on photoplethysmographic signals
Teng, X. F., & Zhang, Y. T. (2004) · Physiological Measurement · doi:10.1088/0967-3334/25/5/020 · Cited by 171 · see citing works
Taillefer, M.-C., & Denault, A. Y. (2005) · Canadian Journal of Anesthesia/Journal canadien d anesthésie · doi:10.1007/bf03018586 · Cited by 151 · see citing works
Textile Electrodes in ECG Measurement
Pola, T., & Vanhala, J. J. (2007) · International Conference on Intelligent Sensors, Sensor Networks and Information Processing · doi:10.1109/issnip.2007.4496917 · Cited by 100 · see citing works
Reflectance pulse oximetry: Practical issues and limitations
Lee, H., Ko, H., & Lee, J. (2016) · ICT Express · doi:10.1016/j.icte.2016.10.004 · Cited by 95 · see citing works
Polymeric Optical Fiber Fabrics for Illumination and Sensorial Applications in Textiles
Selm, B., Gürel, E. A., Rothmaier, M., Rossi, R. M., & Scherer, L. J. (2010) · Journal of Intelligent Material Systems and Structures · doi:10.1177/1045389x10377676 · Cited by 75 · see citing works
Development of a luminous textile for reflective pulse oximetry measurements
Krehel, M., Wolf, M., Boesel, L. F., Rossi, R. M., Bona, G. L., & Scherer, L. J. (2014) · Biomedical Optics Express · doi:10.1364/boe.5.002537 · Cited by 66 · see citing works
Ability of the Masimo pulse CO-Oximeter to detect changes in hemoglobin
Colquhoun, D. A., Forkin, K. T., Durieux, M. E., & Thiele, R. H. (2012) · Journal of Clinical Monitoring and Computing · doi:10.1007/s10877-012-9335-3 · Cited by 61 · see citing works
Artifacts in fetal pulse oximetry: Incomplete sensor-to-skin contact
Gardosi, J. O., Damianou, D., & Schram, C. M. H. (1994) · American Journal of Obstetrics and Gynecology · doi:10.1016/s0002-9378(94)70116-4 · Cited by 49 · see citing works
Invisible Metallic Microfiber in Clothing Presents Unrecognized MRI Risk for Cutaneous Burn
Pietryga, J., Fonder, M. A., Rogg, J. M., North, D. L., & Bercovitch, L. G. (2011) · American Journal of Neuroradiology · doi:10.3174/ajnr.a2827 · Cited by 45 · see citing works
Reflectance pulse oximetry at the forehead improves by pressure on the probe
Dassel, A. C. M., Graaff, R., Sikkema, M., Meijer, A., Zijlstra, W. G., & Aarnoudse, J. G. (1995) · The Journal of Clinical Monitoring · doi:10.1007/bf01617518 · Cited by 40 · see citing works
Reflectance Forehead Pulse Oximetry: Effects of Contact Pressure During Walking
Dresher, R. P., & Mendelson, Y. (2006) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/iembs.2006.260136 · Cited by 31 · see citing works
A Single-Chip CMOS Pulse Oximeter with On-Chip Lock-In Detection
He, D., Morgan, S. P., Trachanis, D., Hese, J. V., Drogoudis, D., Fummi, F., Stefanni, F., Guarnieri, V., & Hayes‐Gill, B. (2015) · Sensors · doi:10.3390/s150717076 · Cited by 20 · see citing works
Self-citationMethod for producing angled optical fiber tips in the laboratory
Davenport, J. J., Hickey, M., Phillips, J. P., & Kyriacou, P. (2016) · Optical Engineering · doi:10.1117/1.oe.55.2.026120 · Cited by 15 · see citing works
Highly sensitive contact pressure measurements using FBG patch in endotracheal tube cuff
Correia, R. G., Blackman, O. R., Hernandez, F. U., Korposh, S., Morgan, S. P., Hayes‐Gill, B., James, S. W., Evans, D., & Norris, A. M. (2016) · Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE · doi:10.1117/12.2236859 · Cited by 12 · see citing works
Self-citationStrojnik, M., & Паез, Г. (2013) · Journal of Biomedical Optics · doi:10.1117/1.jbo.18.10.108001 · Cited by 11 · see citing works
Hickey, M., Samuels, N., Randive, N., Langford, R. M., & Kyriacou, P. A. (2011) · Computer Methods and Programs in Biomedicine · doi:10.1016/j.cmpb.2011.03.019 · Cited by 10 · see citing works
TAMURA, T. (2017) · Springer eBooks · doi:10.1007/978-3-319-69362-0_8 · Cited by 9 · see citing works
Photoplethysmographic measurements from the esophagus using a new fiber-optic reflectance sensor
Phillips, J. P., Kyriacou, P. A., Jones, D. P., Langford, M. B. R. M., & Chang, S. H. (2011) · Journal of Biomedical Optics · doi:10.1117/1.3598858 · Cited by 6 · see citing works
Suzuki, A., 若園, 芳., Suzuki, S., Tamura, M., Masuda, H., Ishikawa, T., Hashimoto, Y., Suzuki, T.-I., Kikuchi, K., Nakagawa, H., et al. (2008) · Electronics Letters · doi:10.1049/el:20081001 · Cited by 5 · see citing works
Oxygen Level Measurement Techniques: Pulse Oximetry
Babikir, S. F., & Ismail, R. A. (2015) · Cited by 5 · see citing works
Development and optimization of a miniaturized fiber-optic photoplethysmographic sensor
Morley, A., Davenport, J. J., Hickey, M., & Phillips, J. P. (2017) · Optical Engineering · doi:10.1117/1.oe.56.11.117111 · Cited by 3 · see citing works
FBG contact pressure sensitivity enhancement technology
Zhang, J., Korposh, S., Correia, R. G., & Zhang, Y. (2017) · International Conference on Transparent Optical Networks · doi:10.1109/icton.2017.8024983 · Cited by 1 · see citing works
Self-citation
Flagged first, then by citations
Research works printed but not linked: 3
- [2]Webster J.G. Design of Pulse Qximeters CRC Press Boca Raton, FL, USA 1997
- [17]Dresher R. Mendelson Y. Attachment of a Wearable Skin Reflectance Pulse Oximeter Proceedings of the 2005 BMES Annual Fall Meeting Baltimore, MD, USA 28 September 2005
- [30]Tamura T. Chen W. Seamless Healthcare Monitoring: Advancements in Wearable, Attachable, and Invisible Devices Springer Berlin, Germany 2017
Cited by
Sorted by citations
Sources of inaccuracy in photoplethysmography for continuous cardiovascular monitoring
Fine, J., Branan, K. L., Rodriguez, A. J., Boonya-ananta, T., Ajmal, Ramella-Roman, J. C., McShane, M. J., & Coté, G. L. (2021) · Biosensors · doi:10.3390/bios11040126 · Cited by 348 · see citing works
Zhao, J. C., Zhang, S., Sun, Y., Zhou, N., Yu, H., Zhang, H., & Jia, D. (2022) · ACS Photonics · doi:10.1021/acsphotonics.2c00898 · Cited by 62 · see citing works
Noninvasive hemoglobin sensing and imaging: optical tools for disease diagnosis
Taylor-Williams, M., Spicer, G. L. C., Bale, G. M., & Bohndiek, S. E. (2022) · Journal of Biomedical Optics · doi:10.1117/1.jbo.27.8.080901 · Cited by 59 · see citing works
Huang, Z., Xu, Y., Cheng, Y., Xue, M., Deng, M., Jaffrezic-Renault, N. J., & Guo, Z. (2022) · Sensors & Diagnostics · doi:10.1039/d2sd00037g · Cited by 51 · see citing works
Noninvasive Continuous Monitoring of Vital Signs With Wearables: Fit for Medical Use?
Jacobsen, M., Dembek, T. A., Kobbe, G., Gaidzik, P. W., & Heinemann, L. (2021) · Journal of Diabetes Science and Technology · doi:10.1177/1932296820904947 · Cited by 49 · see citing works
The origin of photoplethysmography
Kyriacou, P. A., & Chatterjee, S. (2021) · Elsevier eBooks · doi:10.1016/b978-0-12-823374-0.00004-9 · Cited by 47 · see citing works
Ivanoska-Dacikj, A., & Stachewicz, U. (2020) · REVIEWS ON ADVANCED MATERIALS SCIENCE · doi:10.1515/rams-2020-0048 · Cited by 46 · see citing works
Wearable and Multifunctional Self-Mixing Microfiber Sensor for Human Health Monitoring
Liang, H., Wang, Y., Kan, L., Xu, K., Dong, T., Wang, W., Gao, B., & Jiang, C. (2022) · IEEE Sensors Journal · doi:10.1109/jsen.2022.3225196 · Cited by 42 · see citing works
Pant, S., Umesh, S., & Asokan, S. (2020) · IEEE Sensors Journal · doi:10.1109/jsen.2020.2973342 · Cited by 42 · see citing works
A Comprehensive Review on the Optical Micro-Electromechanical Sensors for the Biomedical Application
Upadhyaya, A. M., Hasan, M. K., Abdel‐Khalek, S., Hassan, R. B., Srivastava, M. C., Sharan, P., Islam, S., Saad, A. M. E., & Vo, N. (2021) · Frontiers in Public Health · doi:10.3389/fpubh.2021.759032 · Cited by 42 · see citing works
Flexible Optical Fiber Sensor for Non‐Invasive Continuous Monitoring of Human Physiological Signals
Yu, Q. X., Zhang, Y., Jiang, L., Li, L., Li, X., & Zhao, J. H. (2025) · Small Methods · doi:10.1002/smtd.202401368 · Cited by 39 · see citing works
Investigating the impact of contact pressure on photoplethysmograms
Sirkiä, J., Panula, T., & Kaisti, M. (2024) · Biomedical Engineering Advances · doi:10.1016/j.bea.2024.100123 · Cited by 22 · see citing works
Qu, W., Chen, Y., Ma, C., Peng, D., Bai, X., Zhao, J., Liu, S., & Luo, L. (2024) · Coatings · doi:10.3390/coatings14020173 · Cited by 18 · see citing works
Ya, N., Zhang, D., Wang, Y., Zheng, Y., Yang, M., Wu, H., & Oudeng, G. (2024) · Nanoscale · doi:10.1039/d4nr01719f · Cited by 18 · see citing works
Design and application of flexible wearable sensors based on optical fibers
Hou, Z., Liu, J., Liao, Y., Gong, J., Li, C., Li, M., Liu, H., & Huang, Q. (2025) · Talanta · doi:10.1016/j.talanta.2025.128576 · Cited by 17 · see citing works
Ballaji, H. K., Correia, R. G., Korposh, S., Hayes‐Gill, B., Hernandez, F. U., Salisbury, B., & Morgan, S. P. (2020) · Sensors · doi:10.3390/s20226568 · Cited by 17 · see citing works
Self-citationPulse oximetry in pediatric care: Balancing advantages and limitations
Al‐Beltagi, M., Saeed, N. K., Bediwy, A. S., & Elbeltagi, R. M. (2024) · World Journal of Clinical Pediatrics · doi:10.5409/wjcp.v13.i3.96950 · Cited by 19 · see citing works
The Impact of Contact Force on Signal Quality Indices in Photoplethysmography Measurements
Cause, J. L., Morillo, Á. S., Naranjo, J. C. G., Stiens, J. H., & da Silva, B. (2024) · Applied Sciences · doi:10.3390/app14135704 · Cited by 14 · see citing works
Fibre Bragg Grating Based Interface Pressure Sensor for Compression Therapy
Bradbury, J. A., Zhang, Q., Hernandez, F. U., Correia, R. G., Korposh, S., Hayes‐Gill, B., Tamoué, F., Parnham, A., McMaster, S. A., & Morgan, S. P. (2022) · Sensors · doi:10.3390/s22051798 · Cited by 11 · see citing works
Self-citationCause, J. L., Morillo, Á. S., da Silva, B., Naranjo, J. C. G., & Stiens, J. H. (2023) · Sensors · doi:10.3390/s23146628 · Cited by 9 · see citing works
Optical Fibre Sensor for Simultaneous Measurement of Capillary Refill Time and Contact Pressure
Liu, C., Correia, R. G., Ballaji, H. K., Korposh, S., Hayes‐Gill, B., & Morgan, S. P. (2020) · Sensors · doi:10.3390/s20051388 · Cited by 9 · see citing works
Self-citationMachine-Learning Classification of Pulse Waveform Quality
OuYoung, T., Weng, W.-L., Hu, T.-Y., Lee, C.-C., Wu, L., & Hsiu, H. (2022) · Sensors · doi:10.3390/s22228607 · Cited by 8 · see citing works
Intra-tracheal multiplexed sensing of contact pressure and perfusion
Correia, R. G., Gadsby, B., Korposh, S., Norris, A. M., Hayes‐Gill, B., Sinha, R., Hardman, J. G., Gardner, D. S., Talbot, S., Harvey, D., et al. (2021) · Biomedical Optics Express · doi:10.1364/boe.442165 · Cited by 7 · see citing works
Self-citationHilaluddin, F., Ong, K. K., Norrrahim, M. N. F., Jamari, N. L., Jamal, S. H., Noor, S. A. M., Kasim, N. A. M., Yunus, W. M. Z. W., Abdullah, N., Osman, M. J., et al. (2024) · Arab Journal of Basic and Applied Sciences · doi:10.1080/25765299.2024.2386734 · Cited by 5 · see citing works
Application of BAS-BP model in 3D force decoupling of flexible tactile sensors
Jiang, X., Cheng, Y., Bu, J., & Wei, D. (2019) · AIP conference proceedings · doi:10.1063/1.5125346 · Cited by 3 · see citing works
Thull, R., Goedicke‐Fritz, S., Schmiech, D., Marnach, A., Müller, S. N., Körbel, C., Laschke, M. W., Tutdibi, E., Nourkami‐Tutdibi, N., Kaiser, E., et al. (2024) · Biosensors · doi:10.3390/bios14090437 · Cited by 2 · see citing works
Caruso, J. F., & Shefflette, A. C. S. (2021) · Isokinetics and Exercise Science · doi:10.3233/ies-200237 · Cited by 1 · see citing works
Development and Translation of intra-Tracheal Multiplexed Sensing Endotracheal Tubes (iTraXS)
Morgan, S. P., Correia, R. G., Korposh, S., Al-Badri, M., Hayes‐Gill, B., Norris, A. M., Sinha, R., Hardman, J. G., Gardner, D. S., & Talbot, S. (2020) · Optics in the Life Sciences · doi:10.1364/translational.2020.ttu2b.6
Self-citationImprovement of Force Sensor Accuracy Using Redundant Outputs for Biomedical Applications
Higa, M., HAYAMI, R., Ono, M., Nakamura, S., & Matsuda, S. (2021) · World Congress on Medical Physics and Biomedical Engineering, September 7 - 12, 2009, Munich, Germany · doi:10.1007/978-3-030-66169-4_39
OuYoung, T., Weng, W.-L., Hu, T.-Y., Lee, C.-C., Wu, L.-W., & Hsiu, H. (2021) · Research Square · doi:10.21203/rs.3.rs-870679/v1
Llera, M., Laux, E., Flahaut, F., Jeandupeux, L., Seoane, E. A., Zerbib, M., Beugnot, J., & Potty, P. (2024) · Biomedical Optics Express · doi:10.1364/boe.549790
Mechanical Contact Conditions in Wearable Reflectance Photoplethysmography: Scoping Review
Yang, C., Xie, J., He, Z., Li, J., & Liu, C. (2026) · JMIR mhealth and uhealth · doi:10.2196/99333
Self-citation
- Authors
- Liu, C., Correia, R., Ballaji, H., Korposh, S., Hayes-Gill, B., & Morgan, S.
- Affiliations
- University of Nottingham (GB)
- Optics and Photonics Group, Faculty of Engineering, University Park, Nottingham NG7 2RD, UK
- Corresponding author
- Morgan, Stephen
- Venue
- Sensors
- Publisher
- MDPI AG
- Published
- 2018-10-26
- Volume
- 18
- Issue
- 11
- Pages
- 3632
- ISSN
- 1424-8220 (ISSN-L)
- PMC
- PMC6263952
- OpenAlex
- W2898098650
- Cited by
- 32
- References
- 31
- Citations by year
- 2026: 1 · 2025: 2 · 2024: 8 · 2023: 1 · 2022: 6 · 2021: 7 · 2020: 6 · 2019: 1
- FWCI
- 1.40
- Type
- journal article
- Language
- en
- Open access
- gold
- dx.doi.org
- Licence
- Locations
- Sensors (publishedVersion, cc-by)
- Europe PMC (PubMed Central) (publishedVersion, other-oa)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- Repository@Nottingham (University of Nottingham) (submittedVersion, cc-by)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
- Funders
- Engineering and Physical Sciences Research Council (EP/K503800/1)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Physical Sciences › Engineering › Electrical and Electronic Engineering › Advanced Fiber Optic Sensors
- Physical Sciences › Chemical Engineering › Bioengineering › Analytical Chemistry and Sensors
- Keywords
- pulse oximetry, oxygen saturation, contact pressure sensing, fiber Bragg grating, plastic optical fiber, photoplethysmogram, motion artifacts, transverse loading, perfusion index, wearable sensors, sensor probe design
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review