Investigation of the Impact of Infrared Sensors on Core Body Temperature Monitoring by Comparing Measurement Sites
Open accessAbstract
Many types of thermometers have been developed to measure body temperature. Infrared thermometers (IRT) are fast, convenient and ease to use. Two types of infrared thermometers are uses to measure body temperature: tympanic and forehead. With the spread of COVID-19 coronavirus, forehead temperature measurement is used widely to screen people for the illness. The performance of this type of device and the criteria for screening are worth studying. This study evaluated the performance of two types of tympanic infrared thermometers and an industrial infrared thermometer. The results showed that these infrared thermometers provide good precision. A fixed offset between tympanic and forehead temperature were found. The measurement values for wrist temperature show significant offsets with the tympanic temperature and cannot be used to screen fevers. The standard operating procedure (SOP) for the measurement of body temperature using an infrared thermometer was proposed. The suggestion threshold for the forehead temperature is 36 °C for screening of fever. The body temperature of a person who is possibly ill is then measured using a tympanic infrared thermometer for the purpose of a double check.
Overview
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Figures





Verification
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“However, the determination of core body temperature by indirect measurement can be accomplished using several signal processing methods [144,145].”
From The Current State of Optical Sensors in Medical Wearables (2022) · cites this paper as [144]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Citations
Cites
Infrared ear thermometry compared with rectal thermometry in children: a systematic review
Craig, J. V., Lancaster, G., Taylor, S., Williamson, P. R., & Smyth, R. L. (2002) · The Lancet · doi:10.1016/s0140-6736(02)09783-0 · Cited by 337
Validity and Reliability of Devices That Assess Body Temperature During Indoor Exercise in the Heat
Ganio, M. S., Brown, C. M., Casa, D. J., Becker, S. M., Yeargin, S. W., McDermott, B. P., Boots, L. M., Boyd, P. W., Armstrong, L. E., & Maresh, C. M. (2009) · Journal of Athletic Training · doi:10.4085/1062-6050-44.2.124 · Cited by 201
Comparison of ear-based, bladder, oral, and axillary methods for core temperature measurement
Erickson, R. S., & Kirklin, S. K. (1993) · Critical Care Medicine · doi:10.1097/00003246-199310000-00022 · Cited by 190
Infrared camera assessment of skin surface temperature—Effect of emissivity
Bernard, V., Staffa, E., Mornstein, V., & Bourek, A. (2013) · Physica Medica · doi:10.1016/j.ejmp.2012.09.003 · Cited by 177
Accuracy and Precision of Noninvasive Temperature Measurement in Adult Intensive Care Patients
Lawson, L. L., Bridges, E. J., Ballou, I., Eraker, R., Greco, S. A., Shively, J., & Sochulak, V. (2007) · American Journal of Critical Care · doi:10.4037/ajcc2007.16.5.485 · Cited by 154
Yao, Y., Lian, Z., Liu, W., & Shen, Q. (2007) · Indoor and Built Environment · doi:10.1177/1420326x07084291 · Cited by 110
Insufficiency in a New Temporal-Artery Thermometer for Adult and Pediatric Patients
Suleman, M.-I., Doufas, A. G., Akça, O., Ducharme, M. B., & Sessler, D. I. (2002) · Anesthesia & Analgesia · doi:10.1097/00000539-200207000-00012 · Cited by 105
Liu, C., Chang, R., & Chang, W.-C. (2004) · Infection Control and Hospital Epidemiology · doi:10.1086/502351 · Cited by 94
Clinical accuracy of a non‐contact infrared skin thermometer in paediatric practice
Terán, C. G., Torrez‐Llanos, J., Teran‐Miranda, T. E., Balderrama, C. R., Shah, N. S., & Villarroel, P. (2011) · Child Care Health and Development · doi:10.1111/j.1365-2214.2011.01264.x · Cited by 90
Reliability of an infrared forehead skin thermometer for core temperature measurements
Kistemaker, J. A., DenHartog, E., & Daanen, H. A. M. (2006) · Journal of Medical Engineering & Technology · doi:10.1080/03091900600711381 · Cited by 87
Novel approach to assess the emissivity of the human skin
Sánchez‐Marin, F. J., Calixto-Carrera, S., & Villaseñor-Mora, C. (2009) · Journal of Biomedical Optics · doi:10.1117/1.3086612 · Cited by 87
Chiappini, E., Sollai, S., Longhi, R. C., Morandini, L., Laghi, A., Osio, C. E., Persiani, M., Lonati, S., Picchi, R., Bonsignori, F., et al. (2011) · Journal of Clinical Nursing · doi:10.1111/j.1365-2702.2010.03565.x · Cited by 79
Koçoğlu, H., Göksu, S., Işık, M., Aktürk, Z., & Bayazit, Y. A. (2002) · International Journal of Pediatric Otorhinolaryngology · doi:10.1016/s0165-5876(02)00129-5 · Cited by 76
Noninvasive Temperature Monitoring in Postanesthesia Care Units
Langham, G., Maheshwari, A., Contrera, K. J., You, J., Mascha, E. J., & Sessler, D. I. (2009) · Anesthesiology · doi:10.1097/aln.0b013e3181a864ca · Cited by 75
Hsiao, S. H., Tc, C., Chien, H.-C., Yang, C., & Chen, Y. (2020) · Journal of Hospital Infection · doi:10.1016/j.jhin.2020.04.004 · Cited by 68
Which thermometer? Factors influencing best choice for intermittent clinical temperature assessment
Crawford, D. C., Hicks, B. N., & Thompson, M. J. (2006) · Journal of Medical Engineering & Technology · doi:10.1080/03091900600711464 · Cited by 67
Re-visiting the tympanic membrane vicinity as core body temperature measurement site
Yeoh, W. K., LEE, J. K. W., Lim, H. Y., Gan, C. W., Liang, W., & Tan, K. K. (2017) · PLoS ONE · doi:10.1371/journal.pone.0174120 · Cited by 66
Sund‐Levander, M., Grodzinsky, E., Loyd, D., & Wahren, L. K. (2004) · International Journal of Nursing Practice · doi:10.1111/j.1440-172x.2004.00483.x · Cited by 63
Non-contact infrared thermometry temperature measurement for screening fever in children
Ng, D. K., Chan, C., Lee, R. S., & Leung, L. C. K. (2005) · Annals of Tropical Paediatrics · doi:10.1179/146532805x72412 · Cited by 58
Measurement of Core Body Temperature Using Graphene-Inked Infrared Thermopile Sensor
E., J. S. C., Celik, N., & Balachandran, W. (2018) · Sensors · doi:10.3390/s18103315 · Cited by 56
Clinical temperature. New physiological basis
Benzinger, T. (1969) · JAMA · doi:10.1001/jama.209.8.1200 · Cited by 56
Ng, D. K., Chan, C.-H., Chan, E. Y.-T., Kwok, K., Chow, P.-Y., Lau, W.-F., & Ho, J. C. (2005) · American Journal of Infection Control · doi:10.1016/j.ajic.2005.01.003 · Cited by 54
Mogensen, C. B., Wittenhoff, L., Fruerhøj, G., & Hansen, S. (2018) · BMC Pediatrics · doi:10.1186/s12887-018-0994-1 · Cited by 54
Comparison of esophageal, tympanic, and forehead skin temperatures in adult patients
Patel, N., Smith, C. E., Pinchak, A. C., & Hagen, J. F. (1996) · Journal of Clinical Anesthesia · doi:10.1016/0952-8180(96)00103-1 · Cited by 49
Zeiner, A., Klewer, J., Sterz, F. R., Haugk, M., Krizanac, D., Testori, C., Losert, H., Ayati, S., & Holzer, M. S. (2010) · Resuscitation · doi:10.1016/j.resuscitation.2010.03.018 · Cited by 49
Abdi, A., Asadian, S., Khatony, A., Moradi, G., & Rezaei, M. (2016) · Medical Devices Evidence and Research · doi:10.2147/mder.s109904 · Cited by 48
Tay, M. R. J., Low, Y. L., Zhao, X., Cook, A. R., & Lee, V. J. (2015) · Public Health · doi:10.1016/j.puhe.2015.07.023 · Cited by 46
“Deep-forehead” temperature correlates well with blood temperature
HARIOKA, T., Matsukawa, T., Ozaki, M., Nomura, K., Sone, T., Kakuyama, M., & Toda, H. (2000) · Canadian Journal of Anesthesia/Journal canadien d anesthésie · doi:10.1007/bf03024869 · Cited by 41
Duran, R., Vatansever, Ü., Acunaş, B., & Süt, N. (2009) · Journal of Paediatrics and Child Health · doi:10.1111/j.1440-1754.2009.01526.x · Cited by 40
Continuous Monitoring of Skin Temperature Using a Liquid-Crystal Thermometer During Anesthesia
Burgess, G. E., Cooper, J. R., Marino, R. J., & Peuler, M. J. (1978) · Southern Medical Journal · doi:10.1097/00007611-197805000-00012 · Cited by 39
Infrared ear thermometers—parameters influencing their reading and accuracy
Pušnik, I., & Drnovšek, J. (2005) · Physiological Measurement · doi:10.1088/0967-3334/26/6/016 · Cited by 39
Benzinger, T. H. (1969) · JAMA · doi:10.1001/jama.1969.03160210032008 · Cited by 35
Axillary and Tympanic Temperature Measurement in Children and Normal Values for Ages
Oğuz, F. S., Yıldız, İ., Varkal, M. A., Demirkale, Z. H., Toprak, S., Kaymakci, K., Saygılı, S., Kılıç, A., & Ünüvar, E. (2016) · Pediatric Emergency Care · doi:10.1097/pec.0000000000000693 · Cited by 34
Comparison of distal oesophageal temperature with “deep” and tracheal temperatures
Matsukawa, T., Sessler, D. I., Ozaki, M., Hanagata, K., Iwashita, H., & Kumazawa, T. (1997) · Canadian Journal of Anesthesia/Journal canadien d anesthésie · doi:10.1007/bf03014466 · Cited by 31
Traceability and calibration in temperature measurement: a clinical necessity
Simpson, R., Machin, G., McEvoy, H. C., & Rusby, R. L. (2006) · Journal of Medical Engineering & Technology · doi:10.1080/03091900600711530 · Cited by 29
An assessment of infrared tympanic thermometers for body temperature measurement
Betta, V., Cascetta, F., & Sepe, D. (1997) · Physiological Measurement · doi:10.1088/0967-3334/18/3/006 · Cited by 28
Study of facial skin and aural temperature
Ng, E. Y. K., Muljo, W., & Wong, B. S. (2006) · IEEE Engineering in Medicine and Biology Magazine · doi:10.1109/memb.2006.1636353 · Cited by 28
An Evaluation of Liquid-Crystal Thermometry as a Screening Device for Intraoperative Hyperthermia
Lees, D. E., SCHUETTE, W., Bull, J. M. C., Whang‐Peng, J., Atkinson, E., & Macnamara, T. E. (1979) · Survey of Anesthesiology · doi:10.1097/00132586-197912000-00023 · Cited by 26
An Evaluation of a Plastic Strip Thermometer
Lewit, E. M. (1982) · JAMA · doi:10.1001/jama.1982.03320280041027 · Cited by 25
IR ear thermometers: what do they measure and how do they comply with the EU technical regulation?
nik, I. P., Ham, E., & ek, J. D. (2004) · Physiological Measurement · doi:10.1088/0967-3334/25/3/010 · Cited by 25
Does forehead liquid crystal temperature accurately reflect “core” temperature?
Allen, G. C., Horrow, J. C., & Rosenberg, H. R. (1990) · Canadian Journal of Anesthesia/Journal canadien d anesthésie · doi:10.1007/bf03006486 · Cited by 25
Evaluation of Performance and Uncertainty of Infrared Tympanic Thermometers
Chung, W., & Chen, C. (2010) · Sensors · doi:10.3390/s100403073 · Cited by 25
Self-citationAn Evaluation of Liquid-Crystal Thermometry as a Screening Device for Intraoperative Hyperthermia
Lees, D. E., SCHUETTE, W., Bull, J. M., WHANG-PENG, J., ATKINSON, E. R., & Macnamara, T. E. (1978) · Anesthesia & Analgesia · doi:10.1213/00000539-197811000-00014 · Cited by 25
Dynamics and topography of human temperature regulation in response to thermal and work load
Werner, J., Heising, M., Rautenberg, W., & Leimann, K. (1985) · European Journal of Applied Physiology · doi:10.1007/bf00422853 · Cited by 23
Improved probe for a deep body thermometer
Nemoto, T., & Togawa, T. (1988) · Medical & Biological Engineering & Computing · doi:10.1007/bf02442312 · Cited by 23
Infrared ear thermometry in the critically ill patient
León, C., Rodríguez, A. H., Fernández-Palacín, A., & Flores, L. (2005) · Journal of Critical Care · doi:10.1016/j.jcrc.2004.08.011 · Cited by 19
Villaseñor-Mora, C., Sánchez‐Marin, F. J., & Calixto-Carrera, S. (2009) · Redalyc (Universidad Autónoma del Estado de México) · Cited by 19
Duncan, A. L., Bell, A., Chu, K., & Greenslade, J. H. (2008) · Australasian Emergency Nursing Journal · doi:10.1016/j.aenj.2008.05.003 · Cited by 17
Non-Invasive Forehead Segmentation in Thermographic Imaging
Rodriguez-Lozano, F. J., León‐García, F., de Adana, M. R., Palomares, J. M., & Olivares, J. (2019) · Sensors · doi:10.3390/s19194096 · Cited by 17
IR ear thermometers: what do they measure and how do they comply with the EU technical regulation?
Pušnik, I., Ham, E., & Drnovšek, J. (2004) · Physiological Measurement · doi:10.1088/0967-3334/25/5/c01 · Cited by 17
Thermal screening of facial skin arterial hot spots using non-contact infrared radiometry
Williams, E. M., Heusch, A. I., & McCarthy, P. W. (2008) · Physiological Measurement · doi:10.1088/0967-3334/29/3/005 · Cited by 16
Image processing based body temperature estimation using thermal video sequence
Sharma, A., & Yadav, A. R. (2017) · International Conference on Computing Methodologies and Communication (ICCMC) · doi:10.1109/iccmc.2017.8282585 · Cited by 14
Thermistor Recording of Forehead Skin Temperature as an Index of Carotid Artery Disease
Conrad, M. C., Toole, J. F., & Janeway, R. (1969) · Circulation · doi:10.1161/01.cir.39.1.126 · Cited by 13
Clinical evaluation of liquid crystal skin thermometers
MACKENZIE, R., & Asbury, A. J. (1994) · British Journal of Anaesthesia · doi:10.1093/bja/72.2.246 · Cited by 13
Dante, A., Franconi, I., Marucci, A. R., Alfes, C. M., & Lancia, L. (2020) · Journal of Pediatric Nursing · doi:10.1016/j.pedn.2019.11.014 · Cited by 13
An evaluation of the performance of an infrared tympanic thermometer
Cascetta, F. (1995) · Measurement · doi:10.1016/0263-2241(95)00037-2 · Cited by 12
A Comparative Study of Infrared Tympanic Thermometry and Rectal Mercury Thermometry
Valle, P. C. (1999) · Scandinavian Journal of Infectious Diseases · doi:10.1080/00365549950162014 · Cited by 12
Body temperature measurement methods and targets in Australian and New Zealand intensive care units
Cutuli, S. L., Osawa, E., Glassford, N. J., Marshall, D. C., Eyeington, C. T., Eastwood, G. M., Young, P. J., & Bellomo, R. (2018) · Critical Care and Resuscitation · doi:10.1016/s1441-2772(23)00696-8 · Cited by 9
Performance Evaluation of an Infrared Thermocouple
Chen, C., Weng, Y.-K., & Shen, T.-C. (2010) · Sensors · doi:10.3390/s101110081 · Cited by 7
Self-citationSharma, P., Pande, B., Chandrakar, P., & Pati, A. K. (2014) · Biological Rhythm Research · doi:10.1080/09291016.2014.950091 · Cited by 6
Edling, L., Carlsson, R., Magnuson, A., & Holmberg, H. (2010) · PubMed · Cited by 4
US increases Ebola screening at five airports
McCarthy, M. (2014) · BMJ · doi:10.1136/bmj.g6147 · Cited by 3
Research works printed but not linked: 2
- Berksoy, 2018, Comparison of infrared tympanic, non-contact infrared skin, and axillary thermometer to rectal temperature measurements in a pediatric emergency observation unit, Int. J. Clin. Exp. Med., 11, 567
- Myers, R.H. (1990). Classical and Modern Regression with Applications, Duxbury Press. [2nd ed.].
Cited by
Sorted by citations
Significance of sensors for industry 4.0: Roles, capabilities, and applications
Javaid, M., Haleem, A., Singh, R. P., Rab, S., & Suman, R. (2021) · Sensors International · doi:10.1016/j.sintl.2021.100110 · Cited by 373
Feasibility of continuous fever monitoring using wearable devices
Smarr, B. L., Aschbacher, K., Fisher, S. M., Chowdhary, A., Dilchert, S., Puldon, K., Rao, A., Hecht, F. M., & Mason, A. E. (2020) · Scientific Reports · doi:10.1038/s41598-020-78355-6 · Cited by 133
Review of Stress Detection Methods Using Wearable Sensors
Taskasaplidis, G. V., Fotiadis, D. A., & Bamidis, P. D. (2024) · IEEE Access · doi:10.1109/access.2024.3373010 · Cited by 131
The Current State of Optical Sensors in Medical Wearables
Vavrinsky, E., Esfahani, N. E., Hausner, M., Kuzma, A., Rezo, V., Donoval, M., & Kosnacova, H. (2022) · Biosensors · doi:10.3390/bios12040217 · Cited by 103
Bioelectronic Implantable Devices for Physiological Signal Recording and Closed‐Loop Neuromodulation
Oh, S., Jekal, J., Jia Liu, B., Kim, J., Park, J., Lee, T., & Jang, K. (2024) · Advanced Functional Materials · doi:10.1002/adfm.202403562 · Cited by 104
Metadata
- Authors
- Chen, H.-Y., Chen, A., & Chen, C.
- Affiliations
- University of California San Diego (US)
- National Chung Hsing University (TW)
- Corresponding author
- Chen, Chiachung
- Venue
- Sensors
- Publisher
- MDPI AG
- Published
- 2020-05-19
- Volume
- 20
- Issue
- 10
- Pages
- 2885
- ISSN
- 1424-8220 (ISSN-L)
- PMC
- PMC7284737
- OpenAlex
- W3025566666
- Cited by
- 139
- References
- 63
- Citations by year
- 2026: 6 · 2025: 11 · 2024: 17 · 2023: 20 · 2022: 28 · 2021: 42 · 2020: 15
- FWCI
- 12.78
- Type
- journal article
- Language
- en
- Open access
- gold
- dx.doi.org
- Licence
- Locations
- Sensors (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
- Funders
- Ministry of Science and Technology, Taiwan (MOST107-2313-B005-012)
- Topics
- Health Sciences › Medicine › Radiology, Nuclear Medicine and Imaging › Infrared Thermography in Medicine
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Health Sciences › Medicine › Critical Care and Intensive Care Medicine › Thermal Regulation in Medicine
- Keywords
- infrared thermometer, core body temperature, tympanic temperature, forehead temperature, wrist temperature, fever screening, COVID-19 pandemic, measurement site comparison, standard operating procedures, temperature measurement accuracy
- MeSH terms
- COVID-19, Body Temperature, Fever, Forehead, Humans, Infrared Rays, Pneumonia, Viral, Thermometers, Coronavirus Infections, Pandemics
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review