Wide Field Spectral Imaging with Shifted Excitation Raman Difference Spectroscopy Using the Nod and Shuffle Technique

Open access
Korinth, Florian, Schmälzlin, Elmar, Stiebing, Clara, Urrutia, Tanya, Micheva, Genoveva, Sandin, Christer, Müller, André, Maiwald, Martin, Sumpf, Bernd, Krafft, Christoph, Tränkle, Günther, Roth, Martin M., Popp, Jürgen · 2020 · Sensors

The DOI matches its Crossref record; 59 of the 60 references it cites were found in a registry or an index.

Abstract

Wide field Raman imaging using the integral field spectroscopy approach was used as a fast, one shot imaging method for the simultaneous collection of all spectra composing a Raman image. For the suppression of autofluorescence and background signals such as room light, shifted excitation Raman difference spectroscopy (SERDS) was applied to remove background artifacts in Raman spectra. To reduce acquisition times in wide field SERDS imaging, we adapted the nod and shuffle technique from astrophysics and implemented it into a wide field SERDS imaging setup. In our adapted version, the nod corresponds to the change in excitation wavelength, whereas the shuffle corresponds to the shifting of charges up and down on a Charge-Coupled Device (CCD) chip synchronous to the change in excitation wavelength. We coupled this improved wide field SERDS imaging setup to diode lasers with 784.4/785.5 and 457.7/458.9 nm excitation and applied it to samples such as paracetamol and aspirin tablets, polystyrene and polymethyl methacrylate beads, as well as pork meat using multiple accumulations with acquisition times in the range of 50 to 200 ms. The results tackle two main challenges of SERDS imaging: gradual photobleaching changes the autofluorescence background, and multiple readouts of CCD detector prolong the acquisition time.

Overview

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Integrity signals

Registry

Crossref · 10.3390/s20236723

Cites

59 of 59 references verified0 citation errors · 0 not checkable

Statistics

Figures

Figure 1 Emission spectra of the dual-wavelength GaN external cavity diode laser (ECDL) measured at an operating temperature of 40°C and an optical output power of 0.4 W.
Figure 1 Emission spectra of the dual-wavelength GaN external cavity diode laser (ECDL) measured at an operating temperature of 40°C and an optical output power of 0.4 W.CC BY · PubMed Central
Figure 2 Wide field shifted excitation Raman difference spectroscopy (SERDS) imaging setup based on two micro lens arrays.
Figure 2 Wide field shifted excitation Raman difference spectroscopy (SERDS) imaging setup based on two micro lens arrays.CC BY · PubMed Central
Figure 3 Illustration of the nod and shuffle technique by schematically showing the Charge-Coupled Device (CCD) chip (one square denotes one pixel on the CCD) and the shifting of charges (red arrows) after each acquisition and accumulation. The black lines symbolize the spectral signal traces with the corresponding excitation wavelengths to the left of them. The white spots inside the black lines display increasing signal strength, i.e., Raman bands that get more pronounced during the accumulation. The red boxes show the respective regions on the CCD, which are illuminated during the acquisition, where each box corresponds to the same spaxel (spatial Position P1, P2, P3, and P4) of the Raman image. For simplicity, only four spaxels of the Raman image are shown.
Figure 3 Illustration of the nod and shuffle technique by schematically showing the Charge-Coupled Device (CCD) chip (one square denotes one pixel on the CCD) and the shifting of charges (red arrows) after each acquisition and accumulation. The black lines symbolize the spectral signal traces with the corresponding excitation wavelengths to the left of them. The white spots inside the black lines display increasing signal strength, i.e., Raman bands that get more pronounced during the accumulation. The red boxes show the respective regions on the CCD, which are illuminated during the acquisition, where each box corresponds to the same spaxel (spatial Position P1, P2, P3, and P4) of the Raman image. For simplicity, only four spaxels of the Raman image are shown.CC BY · PubMed Central
Figure 4 Illustration of the raw, wide field Raman data measured with two different excitation wavelengths (λ1 = 784.4 nm, λ2 = 785.5 nm) and the nod and shuffle technique: (a) zoomed view on a part of the CCD chip after acquisition for a closer look on the signal traces. Twenty spectra of each wavelength are always arranged in a block of traces (black square bracket), corresponding to one row of pixels in the Raman images. The upper blue box marks two intensity traces (Position 1) that correspond to a spaxel in the Raman image without a sample. The lower orange box marks two intensity traces (Position 2) that correspond to a spaxel on the Raman image with a paracetamol pill; (b) signal intensities of the two signal traces (in dark red and dark green) marked by the upper blue box in the CCD image showing only the room light spectra. In red (λ1) and green (λ2), the signal intensities of the two signal traces marked by the lower orange box in the CCD image are shown, which correspond to the paracetamol Raman spectra superposed by room light spectra. The dotted, vertical black lines indicate the position of the fluorescent lamp bands and above their spectral position in nm.
Figure 4 Illustration of the raw, wide field Raman data measured with two different excitation wavelengths (λ1 = 784.4 nm, λ2 = 785.5 nm) and the nod and shuffle technique: (a) zoomed view on a part of the CCD chip after acquisition for a closer look on the signal traces. Twenty spectra of each wavelength are always arranged in a block of traces (black square bracket), corresponding to one row of pixels in the Raman images. The upper blue box marks two intensity traces (Position 1) that correspond to a spaxel in the Raman image without a sample. The lower orange box marks two intensity traces (Position 2) that correspond to a spaxel on the Raman image with a paracetamol pill; (b) signal intensities of the two signal traces (in dark red and dark green) marked by the upper blue box in the CCD image showing only the room light spectra. In red (λ1) and green (λ2), the signal intensities of the two signal traces marked by the lower orange box in the CCD image are shown, which correspond to the paracetamol Raman spectra superposed by room light spectra. The dotted, vertical black lines indicate the position of the fluorescent lamp bands and above their spectral position in nm.CC BY · PubMed Central
Figure 5 SERDS imaging using nod and shuffle of a paracetamol (P) and an aspirin (A) pill with an acquisition of 200 times 200 ms. (a) Picture of the sample on the probe head, the blue square corresponds to the field of view; (b) cluster map of the SERDS image; (c) clustered mean SERDS spectra and their standard deviation, cluster 1 = paracetamol (black), cluster 2 = no sample/residual room light (green), cluster 3 = aspirin (red); the dotted, vertical black lines indicate the position of the room light bands.
Figure 5 SERDS imaging using nod and shuffle of a paracetamol (P) and an aspirin (A) pill with an acquisition of 200 times 200 ms. (a) Picture of the sample on the probe head, the blue square corresponds to the field of view; (b) cluster map of the SERDS image; (c) clustered mean SERDS spectra and their standard deviation, cluster 1 = paracetamol (black), cluster 2 = no sample/residual room light (green), cluster 3 = aspirin (red); the dotted, vertical black lines indicate the position of the room light bands.CC BY · PubMed Central
Figure 6 SERDS imaging of polystyrene and polymethyl methacrylate beads: (a) bright field image of the sample, red box shows a region of larger PMMA beads; (b) PS intensity map based on the intensity of three marker bands in the baseline corrected, reconstructed Raman spectra; (c) PMMA intensity map based on the intensity of three marker bands in the baseline corrected, reconstructed Raman spectra; (d) mean SERDS spectra and their standard deviation for nine spectra with the highest intensities for PMMA and PS marked by boxes (green for PS, blue for PMMA) in the corresponding intensity maps; (e) baseline corrected and reconstructed mean Raman spectra with standard deviation of the SERDS spectra shown in (d); The vertical lines in (d,e) indicate the marker bands (green for PS; blue for PMMA) that were used to create the intensity maps (b,c).
Figure 6 SERDS imaging of polystyrene and polymethyl methacrylate beads: (a) bright field image of the sample, red box shows a region of larger PMMA beads; (b) PS intensity map based on the intensity of three marker bands in the baseline corrected, reconstructed Raman spectra; (c) PMMA intensity map based on the intensity of three marker bands in the baseline corrected, reconstructed Raman spectra; (d) mean SERDS spectra and their standard deviation for nine spectra with the highest intensities for PMMA and PS marked by boxes (green for PS, blue for PMMA) in the corresponding intensity maps; (e) baseline corrected and reconstructed mean Raman spectra with standard deviation of the SERDS spectra shown in (d); The vertical lines in (d,e) indicate the marker bands (green for PS; blue for PMMA) that were used to create the intensity maps (b,c).CC BY · PubMed Central
Figure 7 Compensation of photobleaching in SERDS imaging of the fat region of a pork chop using nod and shuffle: (a) image of the pork chop sample with measurement area marked by a blue square; (b) SERDS spectra of the SERDS image (red) with an acquisition time of 50 ms and 40 accumulations, SERDS spectra of the SERDS image (green) with an acquisition time of 2000 ms in a single accumulation. The darker lines (dark green and dark red, respectively) show the mean SERDS spectra of the respective SERDS image. The total exposure time is 4 s for each difference spectrum (2 s for each excitation wavelength).
Figure 7 Compensation of photobleaching in SERDS imaging of the fat region of a pork chop using nod and shuffle: (a) image of the pork chop sample with measurement area marked by a blue square; (b) SERDS spectra of the SERDS image (red) with an acquisition time of 50 ms and 40 accumulations, SERDS spectra of the SERDS image (green) with an acquisition time of 2000 ms in a single accumulation. The darker lines (dark green and dark red, respectively) show the mean SERDS spectra of the respective SERDS image. The total exposure time is 4 s for each difference spectrum (2 s for each excitation wavelength).CC BY · PubMed Central
Figure 8 SERDS imaging of a piece of pork meat with protein and lipid-rich areas: (a) intensity map of the SERDS spectra at 2900 cm−1; (b) cluster map of the SERDS spectra; (c) clustered normalized, mean SERDS spectra and their standard deviation; (d) reconstructed and baseline corrected mean Raman spectra of the two clusters shown in (c). The high wavenumber regions were multiplied by 0.2 for better visualization.
Figure 8 SERDS imaging of a piece of pork meat with protein and lipid-rich areas: (a) intensity map of the SERDS spectra at 2900 cm−1; (b) cluster map of the SERDS spectra; (c) clustered normalized, mean SERDS spectra and their standard deviation; (d) reconstructed and baseline corrected mean Raman spectra of the two clusters shown in (c). The high wavenumber regions were multiplied by 0.2 for better visualization.CC BY · PubMed Central

Verification

Claims the paper makes

The claims and statistics of this paper haven’t been checked yet.

Free: it reads the open-access text and checks the first 5 claims that cite a source we can check. Create a free account to check every claim.

A few free checks a day · papers already verified open instantly · results are public

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

The self-check needs the paper’s full text.

Citations

Cites

Show all 59 references

Research works printed but not linked: 1

  1. Heming R. Herzog H. Deckert V. Optical CCD lock-in device for Raman difference spectroscopy DGaO Proc. 2008 109 33

Cited by

Metadata

Authors
Korinth, F., Schmälzlin, E., Stiebing, C., Urrutia, T., Micheva, G., Sandin, C., Müller, A., Maiwald, M., Sumpf, B., Krafft, C., Tränkle, G., Roth, M. M., & Popp, J.
Affiliations
  • Leibniz Institute of Photonic Technology (DE)
  • Leibniz Institute for Astrophysics Potsdam (DE)
  • Sandin Advanced Visualization, Tylögränd 14, 12156 Johanneshov, Sweden
  • Ferdinand-Braun-Institut (DE)
Corresponding author
Krafft, Christoph
Venue
Sensors
Publisher
MDPI AG
Published
2020-11-24
Volume
20
Issue
23
Pages
6723
ISSN
1424-8220 (ISSN-L)
OpenAlex
W3107090361
Cited by
13
References
60
Citations by year
2026: 1 · 2025: 2 · 2024: 4 · 2022: 3 · 2021: 3
FWCI
0.71
Type
journal article
Language
en
Open access
Locations
  • Sensors (publishedVersion, cc-by)
  • arXiv (Cornell University) (publishedVersion)
  • PubMed Central (publishedVersion, cc-by)
  • PubMed (publishedVersion)
  • DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
  • RENATE (submittedVersion, cc-by-nc-nd)
  • arXiv (Cornell University) (cc-by)
  • RENATE (cc-by)
Funders
Senate Competition Committee of the Leibniz Association (SAW-2016-IPHT-2)
Topics
  • Life Sciences › Biochemistry, Genetics and Molecular Biology › Biophysics › Spectroscopy Techniques in Biomedical and Chemical Research
  • Physical Sciences › Chemistry › Analytical Chemistry › Spectroscopy and Chemometric Analyses
  • Social Sciences › Arts and Humanities › Archeology (arts and humanities) › Cultural Heritage Materials Analysis
Keywords
shifted excitation Raman difference spectroscopy, integral field spectroscopy, autofluorescence suppression, photobleaching, acquisition time reduction, laser diode excitation, tablets, pork

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.