Effects of Different Hematocrit Levels on Glucose Measurements With Handheld Meters for Point-of-Care Testing
The DOI matches its Crossref record; 23 of the 23 references it cites were found in a registry or an index.
Abstract
Objectives. —To determine the effects of low, normal, and high hematocrit levels on glucose meter measurements and to assess the clinical risks of hematocrit errors. Design. —Changes in glucose measurements between low and high hematocrit levels were calculated to determine hematocrit effects. The differences between glucose measured with meters and with a plasma glucose method (YSI 2300) also were compared. Setting. —Six handheld glucose meters were assessed in vitro at low (19.1%), normal (38.5%), and high (58.3%) hematocrit levels, and at 6 glucose concentrations ranging from 2.06 mmol/L (37.1 mg/dL) to 30.24 mmol/L (544.7 mg/dL). Results. —Most systems, regardless of the reference to which they were calibrated, demonstrated positive bias at lower hematocrit levels and negative bias at higher hematocrit levels. Low, normal, and high hematocrit levels progressively lowered Precision G and Precision QID glucose measurements. Hematocrit effects on the other systems were more dependent on the glucose concentration. Overall, Accu-Chek Comfort Curve showed the least sensitivity to hematocrit changes, except at the lowest glucose concentration. Conclusions. —We strongly recommend that clinical professionals choose glucose systems carefully and interpret glucose measurements with extreme caution when the patient's hematocrit value changes, particularly if there is a simultaneous change in glucose level.
Overview
Integrity signals
Registry
Cites
Statistics
Verification
Claims the paper makes
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
In the citing paper
“In strip-based glucose assays, high hematocrit values, due to increased blood viscosity, can reduce the diffusion of plasma and determine underestimated results, whereas low hematocrit levels, like in anemia, are associated with overestimated readings 101,102.”
From Glucose biosensors in clinical practice: principles, limits and perspectives of currently used devices (2022) · cites this paper as [102]
In the citing paper
“Glucose extraction is approximately in linear relation with the density and duration of iontophoretic current (<0.5 mA/cm2) 102.”
From Glucose biosensors in clinical practice: principles, limits and perspectives of currently used devices (2022) · cites this paper as [102]
In the citing paper
“Important variables that may influence the results of bedside glucose monitoring include changes in hematocrit (114), altitude, environmental temperature or humidity, hypotension, hypoxia and high triglyceride concentrations (115), and various drugs.”
From Guidelines and recommendations for laboratory analysis in the diagnosis and management of diabetes mellitus (2011) · cites this paper as [114]
In the citing paper
“Quite often the relevance of factors such as HCT is not considered by the treating physician and, subsequently, by the PwD [28].”
From Self-Monitoring of Blood Glucose as an Integral Part in the Management of People with Type 2 Diabetes Mellitus (2022) · cites this paper as [28]
Citations
Cites
A Text-Book of Physical Chemistry
(1936) · Nature · doi:10.1038/138864b0 · Cited by 1,146
A textbook of physical chemistry
Adamson, A. W. (1973) · Cited by 343
Tang, Z., Du, X., Louie, R. F., & Kost, G. J. (2000) · American Journal of Clinical Pathology · doi:10.1309/qaw1-x5xw-bvrq-5lkq · Cited by 207
Self-citationFingerstick Glucose Determination in Shock
Atkin, S. H., Dasmahapatra, A., Jaker, M. A., Chorost, M. I., & Reddy, S. (1991) · Annals of Internal Medicine · doi:10.7326/0003-4819-114-12-1020 · Cited by 189
Louie, R. F., Tang, Z., Sutton, D. V., Lee, J. H., & Kost, G. J. (2000) · Archives of Pathology & Laboratory Medicine · doi:10.5858/2000-124-0257-pocgt · Cited by 180
Self-citationBurrin, J. M., & Price, C. P. (1985) · Annals of Clinical Biochemistry International Journal of Laboratory Medicine · doi:10.1177/000456328502200401 · Cited by 150
Kost, G. J., Vu, H.-T., Lee, J. H., Bourgeois, P., Kiechle, F. L., Martin, C., Miller, S. S., Okorodudu, A. O., Podczasy, J. J., Webster, R., et al. (1998) · Critical Care Medicine · doi:10.1097/00003246-199803000-00036 · Cited by 120
Self-citationCommercial biosensors : applications to clinical, bioprocess, and environmental samples
Ramsay, G. (1998) · Wiley eBooks · Cited by 109
Accuracy of fingerstick glucose values in shock patients
Sylvain, H., Pokorny, M., English, S., Benson, N., Whitley, T., Ferenczy, C., & Harrison, J. (1995) · American Journal of Critical Care · doi:10.4037/ajcc1995.4.1.44 · Cited by 89
Neonatal Hyperviscosity: I. Incidence
Wirth, F. H., Goldberg, K. E., & Lubchenco, L. O. (1979) · PEDIATRICS · doi:10.1542/peds.63.6.833 · Cited by 79
Research works printed but not linked: 4
- Hills, L., G.Azurin, and X.Wang. et al. Glutathione as an interferent in near patient whole blood glucose devices.In: Proceedings of the 17th International Symposium of the International Federation of Clinical Chemistry. Madison, Wis: Omnipress. 1998. 207–219.
- Ramsay, G. Commercial Biosensors: Applications to Clinical, Bioprocess, and Environmental Samples. New York, NY: John Wiley & Sons Inc. 1998. 23.
- Adamson, A. W. A Textbook of Physical Chemistry. New York, NY: Academic Press. 1973. 339–437.
- Hills, L., G.Azurin, and X.Wang. et al. Glutathione as an interferent in near patient whole blood glucose devices.In: Proceedings of the 17th International Symposium of the International Federation of Clinical Chemistry. Madison, Wis: Omnipress. 1998. 207–219.
Cited by
Sacks, D. B., Arnold, M., Bakris, G. L., Bruns, D. E., Horvath, A. R., Kirkman, M. S., Lernmark, A., Metzger, B. E., & Nathan, D. M. (2011) · Clinical Chemistry · doi:10.1373/clinchem.2010.161596 · Cited by 1,360
Management of diabetes and hyperglycemia in hospitals.
Clement, S., Braithwaite, S. S., Magee, M. F., Ahmann, A., Smith, E. P., Schafer, R. G., Hirsch, I. B., & on behalf of the Diabetes in Hospitals Writing Committee (2004) · Diabetes Care · doi:10.2337/diacare.27.2.553 · Cited by 1,266
Glucose Biosensors: An Overview of Use in Clinical Practice
Yoo, E.-H., & Lee, S.-Y. (2010) · Sensors · doi:10.3390/s100504558 · Cited by 1,137
Sacks, D. B., Bruns, D. E., Goldstein, D. E., Maclaren, N. K., McDonald, J. M., & Parrott, M. (2002) · Clinical Chemistry · doi:10.1093/clinchem/48.3.436 · Cited by 1,130
Sacks, D. B., Arnold, M., Bakris, G. L., Bruns, D. E., Horvath, A. R., Kirkman, M. S., Lernmark, A., Metzger, B. E., & Nathan, D. M. (2011) · Diabetes Care · doi:10.2337/dc11-9998 · Cited by 718
Jacobi, J., Bircher, N., Krinsley, J., Agus, M., Braithwaite, S. S., Deutschman, C., Freire, A. X., Geehan, D., Kohl, B., Nasraway, S. A., et al. (2012) · Critical Care Medicine · doi:10.1097/ccm.0b013e3182653269 · Cited by 571
A1c versus glucose testing: a comparison
Sacks, D. B. (2011) · Diabetes Care · doi:10.2337/dc10-1546 · Cited by 461
Reliability of point-of-care testing for glucose measurement in critically ill adults*
Kanji, S., Buffie, J., Hutton, B., Bunting, P. S., Singh, A., McDonald, K., Fergusson, D., McIntyre, L. A., & Hebert, P. C. (2005) · Critical Care Medicine · doi:10.1097/01.ccm.0000189939.10881.60 · Cited by 395
Glucose Measurement: Confounding Issues in Setting Targets for Inpatient Management
Dungan, K., Chapman, J., Braithwaite, S. S., & Buse, J. (2007) · Diabetes Care · doi:10.2337/dc06-1679 · Cited by 292
Zhou, H., & Guo, Z. (2019) · Journal of Materials Chemistry A · doi:10.1039/c9ta02682g · Cited by 227
Electrochemical detection of glucose from whole blood using paper-based microfluidic devices
Noiphung, J., Songjaroen, T., Dungchai, W., Henry, C. S., Chailapakul, O., & Laiwattanapaisal, W. (2013) · Analytica Chimica Acta · doi:10.1016/j.aca.2013.06.021 · Cited by 227
Pullano, S. A., Greco, M., Bianco, M. G., Foti, D., Brunetti, A., & Fiorillo, A. S. (2022) · Theranostics · doi:10.7150/thno.64035 · Cited by 210
A Paper-Based “Pop-up” Electrochemical Device for Analysis of Beta-Hydroxybutyrate
Wang, C.-C., Hennek, J. W., Ainla, A., Kumar, A. A., Lan, W.-J., Im, J., Smith, B. S., Zhao, M., & Whitesides, G. M. (2016) · Analytical Chemistry · doi:10.1021/acs.analchem.6b00568 · Cited by 180
Louie, R. F., Tang, Z., Sutton, D. V., Lee, J. H., & Kost, G. J. (2000) · Archives of Pathology & Laboratory Medicine · doi:10.5858/2000-124-0257-pocgt · Cited by 180
Self-citationESPGHAN/ESPEN/ESPR/CSPEN guidelines on pediatric parenteral nutrition: Carbohydrates
Mesotten, D., Joosten, K., van Kempen, A., Verbruggen, S., Braegger, C., Bronsky, J., Cai, W., Campoy, C., Carnielli, V., Darmaun, D., et al. (2018) · Clinical Nutrition · doi:10.1016/j.clnu.2018.06.947 · Cited by 161
Measurement of glucose levels in the newborn
Beardsall, K. (2010) · Early Human Development · doi:10.1016/j.earlhumdev.2010.05.005 · Cited by 153
Khan, A. I., Vasquez, Y., Gray, J., Wians Jr, F. H., & Kroll, M. H. (2006) · Archives of Pathology & Laboratory Medicine · doi:10.5858/2006-130-1527-tvorbp · Cited by 133
S. Karon, B., Griesmann, L., Scott, R., Bryant, S. C., Dubois, J. A., Shirey, T. L., Presti, S., & Santrach, P. J. (2008) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2007.0257 · Cited by 132
Comparative accuracy of 17 point-of-care glucose meters
Ekhlaspour, L., Mondesir, D., Lautsch, N., Balliro, C., Hillard, M., Magyar, K., Radocchia, L. G., Esmaeili, A., Sinha, M., & Russell, S. J. (2017) · Journal of Diabetes Science and Technology · doi:10.1177/1932296816672237 · Cited by 127
Glucose Homeostasis During Short-term and Prolonged Exposure to High Altitudes
Woolcott, O. O., Ader, M., & Bergman, R. N. (2015) · Endocrine Reviews · doi:10.1210/er.2014-1063 · Cited by 133
Hirsch, I. B., Bode, B. W., Childs, B. P., Close, K. L., Fisher, W. A., Gavin, J. R., Ginsberg, B. H., Raine, C. H., & Verderese, C. A. (2008) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2008.0104 · Cited by 120
Overview on self-monitoring of blood glucose
Montagnana, M., Caputo, M., Giavarina, D., & Lippi, G. (2009) · Clinica Chimica Acta · doi:10.1016/j.cca.2009.01.002 · Cited by 116
Interferences and Limitations in Blood Glucose Self-Testing: An Overview of the Current Knowledge
Erbach, M., Freckmann, G., Hinzmann, R., Kulzer, B., Ziegler, R., Heinemann, L., & Schnell, O. (2016) · Journal of Diabetes Science and Technology · doi:10.1177/1932296816641433 · Cited by 115
Laboratory Results: Timeliness as a Quality Attribute and Strategy
Howanitz, J. H., & Howanitz, P. J. (2001) · American Journal of Clinical Pathology · doi:10.1309/h0dy-6vtw-nb36-u3l6 · Cited by 108
Karon, B. S., Gandhi, G. Y., Nuttall, G. A., Bryant, S. C., Schaff, H. V., McMahon, M. M., & Santrach, P. J. (2007) · American Journal of Clinical Pathology · doi:10.1309/6rfqckaajgkwb8m4 · Cited by 98
Performance Evaluation of Blood Glucose Monitoring Devices
Chen, E. T., Nichols, J. H., Duh, S.-H., & Hortin, G. (2003) · Diabetes Technology & Therapeutics · doi:10.1089/152091503322526969 · Cited by 96
Glucose Measurement in the Operating Room: More Complicated than It Seems
Rice, M. J., Pitkin, A. D., & Coursin, D. B. (2010) · Anesthesia & Analgesia · doi:10.1213/ane.0b013e3181cc07de · Cited by 96
Kulkarni, A. P., Saxena, M. K., Price, G., O’Leary, M. J., Jacques, T. C., & Myburgh, J. (2004) · Intensive Care Medicine · doi:10.1007/s00134-004-2500-5 · Cited by 88
An Overview of Hypoglycemia in the Critically Ill
Lacherade, J.-C., Jacqueminet, S., & Preiser, J.-C. (2009) · Journal of Diabetes Science and Technology · doi:10.1177/193229680900300603 · Cited by 86
Use of a hand-held meter for detecting subclinical ketosis in dairy cows
Voyvoda, H., & Erdoğan, H. (2010) · Research in Veterinary Science · doi:10.1016/j.rvsc.2010.04.007 · Cited by 95
Clinical Chemistry: Challenges for Analytical Chemistry and the Nanosciences from Medicine
Durner, J. (2010) · Angewandte Chemie International Edition · doi:10.1002/anie.200903363 · Cited by 88
Accuracy evaluation of a new glucometer with automated hematocrit measurement and correction
Rao, L. V., Jakubiak, F., Sidwell, J. S., Winkelman, J. W., & Snyder, M. (2005) · Clinica Chimica Acta · doi:10.1016/j.cccn.2005.01.027 · Cited by 81
Evaluation of a Veterinary Glucometer for Use in Horses
Hackett, E. S., & McCue, P. M. (2010) · Journal of Veterinary Internal Medicine · doi:10.1111/j.1939-1676.2010.0481.x · Cited by 86
PANOUSIS, N. K., Brozos, C., Karagiannis, I., Giadinis, N. D., Lafi, S. Q., & Kritsepi‐Konstantinou, M. (2011) · Research in Veterinary Science · doi:10.1016/j.rvsc.2011.06.019 · Cited by 83
Sensitivity enhancement for colorimetric glucose assays on whole blood by on-chip beam-guidance
Grumann, M., Steigert, J., Riegger, L., Moser, I., Enderle, B., Riebeseel, K., Urban, G. A., Zengerle, R., & Ducrée, J. (2006) · Biomedical Microdevices · doi:10.1007/s10544-006-8172-x · Cited by 77
Comparison of Point-of-Care and Laboratory Glucose Analysis in Critically Ill Patients
Lacara, T., Domagtoy, C., Lickliter, D., Quattrocchi, K., Snipes, L., Kuszaj, J. M., & Prasnikar, M. (2007) · American Journal of Critical Care · doi:10.4037/ajcc2007.16.4.336 · Cited by 82
Hematocrit Interference of Blood Glucose Meters for Patient Self-Measurement
Ramljak, S., Lock, J. P., Schipper, C., Musholt, P. B., Forst, T., Lyon, M., & Pfützner, A. (2013) · Journal of Diabetes Science and Technology · doi:10.1177/193229681300700123 · Cited by 72
Brazg, R. L., Klaff, L. J., & Parkin, C. G. (2013) · Journal of Diabetes Science and Technology · doi:10.1177/193229681300700117 · Cited by 68
Review-Enzymatic and Non-Enzymatic Electrochemical Sensor for Lactate Detection in Human Biofluids
Shakhih, M. F. M., Rosslan, A. S., Noor, A. M., Ramanathan, S., Lazim, А. M., & Wahab, A. A. (2021) · Journal of The Electrochemical Society · doi:10.1149/1945-7111/ac0360 · Cited by 69
Clinical assessment of the accuracy of blood glucose measurement devices
Pfützner, A., Mitri, M., Musholt, P. B., Sachsenheimer, D., Borchert, M., Yap, A., & Forst, T. (2012) · Current Medical Research and Opinion · doi:10.1185/03007995.2012.673479 · Cited by 64
Hollis, A. R., Schaer, B. L. D., Boston, R. C., & Wilkins, P. A. (2008) · Journal of Veterinary Internal Medicine · doi:10.1111/j.1939-1676.2008.0148.x · Cited by 63
Oxygen Effects on Glucose Measurements with a Reference Analyzer and Three Handheld Meters
Tang, Z., Louie, R. F., Payes, M., Chang, K.-C. J., & Kost, G. J. (2000) · Diabetes Technology & Therapeutics · doi:10.1089/15209150050194215 · Cited by 63
Self-citationPoint-of-Care Testing: Millennium Technology for Critical Care
Louie, R. F., Tang, Z., Shelby, D. G., & Kost, G. J. (2000) · Laboratory Medicine · doi:10.1309/0y5f-b7np-5y67-gw7t · Cited by 62
Self-citationEvaluation of four portable blood glucose meters in diabetic and non-diabetic dogs and cats
Kang, M., Kim, D., Jeong, I., Choi, G.-C., & Park, H. (2015) · Veterinary Quarterly · doi:10.1080/01652176.2015.1092617 · Cited by 70
Standardized Evaluation of Nine Instruments for Self-Monitoring of Blood Glucose
Kristensen, G. B. B., Monsen, G., Skeie, S., & Sandberg, S. (2008) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2008.0034 · Cited by 56
Pleus, S., Freckmann, G., Schauer, S., Heinemann, L., Ziegler, R., Ji, L., Mohan, V., Calliari, L. E., & Hinzmann, R. (2022) · Diabetes Therapy · doi:10.1007/s13300-022-01254-8 · Cited by 55
Diagnostic Modalities in Critical Care: Point-of-Care Approach
Rajšić, S., Breitkopf, R., Bachler, M., & Treml, B. (2021) · Diagnostics · doi:10.3390/diagnostics11122202 · Cited by 66
Cook, A., Laughlin, D., Moore, M., North, D., Wilkins, K., Wong, G., Wallace-Scroggs, A., & Halvorsen, L. (2009) · American Journal of Critical Care · doi:10.4037/ajcc2009626 · Cited by 58
Koufakis, T., Karras, S. Ν., Mustafa, O. G., Zebekakis, P. E., & Kotsa, K. (2018) · High Altitude Medicine & Biology · doi:10.1089/ham.2018.0076 · Cited by 60
Barriers to Glucose Control in the Intensive Care Unit
Anger, K. E., & Szumita, P. M. (2006) · Pharmacotherapy The Journal of Human Pharmacology and Drug Therapy · doi:10.1592/phco.26.2.214 · Cited by 55
Mann, E. A., Salinas, J., Pidcoke, H. F., Wolf, S. E., Holcomb, J. B., & Wade, C. E. (2008) · The Journal of Trauma: Injury, Infection, and Critical Care · doi:10.1097/ta.0b013e318160b9e4 · Cited by 52
Gerber, K. L., & Freeman, K. P. (2016) · Veterinary Clinical Pathology · doi:10.1111/vcp.12310 · Cited by 62
Effect of hematocrit on accuracy of two point-of-care glucometers for use in dogs
Paul, A. E. H., Shiel, R. E., Juvet, F., Mooney, C. T., & Mansfield, C. (2011) · American Journal of Veterinary Research · doi:10.2460/ajvr.72.9.1204 · Cited by 51
Automation and Computerization of (Bio)sensing Systems
Raju, C. M., Elpa, D. P., & Urban, P. L. (2024) · ACS Sensors · doi:10.1021/acssensors.3c01887 · Cited by 54
Glucose monitoring in neonates: need for accurate and non-invasive methods
Woo, H. C., Tolosa, L. M., El‐Metwally, D., & Viscardi, R. M. (2013) · Archives of Disease in Childhood Fetal & Neonatal · doi:10.1136/archdischild-2013-304682 · Cited by 49
Fram, R. Y., Cree‐Green, M., Wolfe, R. R., Mlcak, R. P., Qian, T., Chinkes, D. L., & Herndon, D. N. (2010) · Critical Care Medicine · doi:10.1097/ccm.0b013e3181de8b9e · Cited by 53
Assessing the quality of glucose monitor studies: a critical evaluation of published reports
Mahoney, J., & Ellison, J. (2007) · Clinical Chemistry · doi:10.1373/clinchem.2006.083493 · Cited by 48
Musholt, P. B., Schipper, C., Thomé, N., Ramljak, S., Schmidt, M., Forst, T., & Pfützner, A. (2011) · Journal of Diabetes Science and Technology · doi:10.1177/193229681100500520 · Cited by 46
Kost, G. J., Tran, N. K., Abad, V. J., & Louie, R. F. (2007) · Clinica Chimica Acta · doi:10.1016/j.cca.2007.11.019 · Cited by 46
Self-citationBlood Glucose Measurement in the Intensive Care Unit: What is the Best Method?
Le, H. T., Harris, N., Estilong, A. J., Olson, A., & Rice, M. J. (2013) · Journal of Diabetes Science and Technology · doi:10.1177/193229681300700226 · Cited by 44
Evaluation of 4 Handheld Blood Glucose Monitors for Use in Seabird Rehabilitation
Lieske, C. L., Ziccardi, M. H., Mazet, J. A. K., Newman, S. H., & Gardner, I. A. (2002) · Journal of Avian Medicine and Surgery · doi:10.1647/1082-6742(2002)016[0277:eohbgm]2.0.co;2 · Cited by 49
Ghys, T., Goedhuys, W., Spincemaille, K., Gorus, F., & Gerlo, E. (2007) · Clinica Chimica Acta · doi:10.1016/j.cca.2007.07.025 · Cited by 45
Performance of a new interference-resistant glucose meter
Vanavanan, S., Santanirand, P., Chaichanajarernkul, U., Chittamma, A., DuBois, J. A., Shirey, T., & Heinz, M. (2010) · Clinical Biochemistry · doi:10.1016/j.clinbiochem.2009.09.010 · Cited by 39
Dector, A., Escalona-Villalpando, R. A., Dector, D., Vallejo‐Becerra, V., Chávez-Ramírez, A. U., Arriaga, L. G., & García, J. L. (2015) · Journal of Power Sources · doi:10.1016/j.jpowsour.2015.04.089 · Cited by 41
Comparison of Glucose Point-of-Care Values With Laboratory Values in Critically Ill Patients
Shearer, A., Boehmer, M., Closs, M., Rosa, R. D., Hamilton, J., Horton, K. M., McGrath, R., & Schulman, C. S. (2009) · American Journal of Critical Care · doi:10.4037/ajcc2009448 · Cited by 42
Louie, R. F., Sumner, S. L., Belcher, S., Mathew, R., Tran, N. K., & Kost, G. J. (2009) · Disaster Medicine and Public Health Preparedness · doi:10.1097/dmp.0b013e3181979a06 · Cited by 39
Self-citationPetritz, O. A., Antinoff, N., Chen, S., Kass, P. H., & Paul‐Murphy, J. (2013) · Journal of the American Veterinary Medical Association · doi:10.2460/javma.242.3.350 · Cited by 43
Subcutaneous Insulin Therapy in the Hospital Setting: Issues, Concerns, and Implementation
Magee, M. F., & Clement, S. C. (2004) · Endocrine Practice · doi:10.4158/ep.10.s2.81 · Cited by 37
Lane, S. L., Koenig, A., & Brainard, B. M. (2015) · Journal of the American Veterinary Medical Association · doi:10.2460/javma.246.3.307 · Cited by 38
Evaluating the Accuracy of Modern Glucose Meters
Bergenstal, R. M. (2008) · Insulin · doi:10.1016/s1557-0843(08)80004-8 · Cited by 41
Selleri, P., Girolamo, N. D., & Novari, G. (2014) · Journal of the American Veterinary Medical Association · doi:10.2460/javma.245.1.87 · Cited by 40
Forrow, N. J., & Bayliff, S. W. (2005) · Biosensors and Bioelectronics · doi:10.1016/j.bios.2004.12.011 · Cited by 35
Hematocrit compensation in electrochemical blood glucose monitoring systems
Teodorczyk, M., Cardosi, M., & Setford, S. (2012) · Journal of Diabetes Science and Technology · doi:10.1177/193229681200600320 · Cited by 33
Assessing the Performance of Handheld Glucose Testing for Critical Care
Kost, G. J., Tran, N. K., Louie, R. F., Gentile, N. L., & Abad, V. J. (2008) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2008.0049 · Cited by 32
Self-citationInterference Studies with Two Hospital-Grade and Two Home-Grade Glucose Meters
Lyon, M. E., Baskin, L. B., Braakman, S., Presti, S., DuBois, J. A., & Shirey, T. (2009) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2009.0035 · Cited by 31
Pereira, A. J., Corrêa, T. D., de Almeida, F. P., Deliberato, R. O., Lobato, M. D. S., Akamine, N., Silva, E., & Cavalcanti, A. B. (2015) · PLoS ONE · doi:10.1371/journal.pone.0129568 · Cited by 33
Evaluation of a point-of-care glucose meter for general use in complex tertiary care facilities
Chan, P. C., Rozmanc, M., Seiden‐Long, I., & Kwan, J. (2009) · Clinical Biochemistry · doi:10.1016/j.clinbiochem.2009.03.023 · Cited by 32
Russell, C., Palmer, J., Boston, R. C., & Wilkins, P. A. (2007) · Journal of Veterinary Emergency and Critical Care · doi:10.1111/j.1476-4431.2007.00236.x · Cited by 27
Holtzinger, C., Szelag, E., DuBois, J. A., Shirey, T., & Presti, S. (2008) · Point of Care The Journal of Near-Patient Testing & Technology · doi:10.1097/poc.0b013e318162f4da · Cited by 29
Zhang, W., Hu, L., Chen, H., Gao, S., Zhang, X., & Jin, J. (2017) · Journal of Materials Chemistry B · doi:10.1039/c7tb00644f · Cited by 28
Kermani, S. K., Khatony, A., Jalali, R., Rezaei, M., & Abdi, A. (2017) · JOURNAL OF CLINICAL AND DIAGNOSTIC RESEARCH · doi:10.7860/jcdr/2017/23926.9613 · Cited by 37
Point‐of‐Care Testing: Principles, Practice, and Critical‐Emergency‐Disaster Medicine
Kost, G. J., Tran, N. K., & Louie, R. F. (2008) · Wiley eBooks · doi:10.1002/9780470027318.a0540.pub2 · Cited by 38
Self-citationFrancescato, M. P., Geat, M., Stel, G., & Cauci, S. (2011) · Clinica Chimica Acta · doi:10.1016/j.cca.2011.10.012 · Cited by 29
Gao, R., Wu, Y., Huang, J., Song, L., Qian, H., Song, X., Cheng, L., Wang, R., Luo, L., Zhao, G., et al. (2019) · Sensors and Actuators B Chemical · doi:10.1016/j.snb.2019.01.065 · Cited by 27
How accurately do we measure blood glucose levels in intensive care unit (ICU) patients?
Wahl, H. G. (2009) · Best practice & research. Clinical anaesthesiology · doi:10.1016/j.bpa.2009.09.003 · Cited by 32
Rosenthal, K. L., & Wyre, N. R. (2011) · Elsevier eBooks · doi:10.1016/b978-1-4160-6621-7.00007-5 · Cited by 27
Endocrine Assessments During Critical Illness
Sakharova, O. V., & Inzucchi, S. E. (2007) · Critical Care Clinics · doi:10.1016/j.ccc.2007.05.007 · Cited by 29
Acierno, M. J., Mitchell, M. A., Schuster, P. J., Freeman, D. E., Guzman, D. S.-M., & Tully, T. N. (2009) · American Journal of Veterinary Research · doi:10.2460/ajvr.70.2.172 · Cited by 29
Evaluation of different POCT devices for glucose measurement in a clinical neonatal setting
Roth‐Kleiner, M., Stadelmann, C., Urfer, J., Ruffieux, C., & Werner, D. (2010) · European Journal of Pediatrics · doi:10.1007/s00431-010-1243-2 · Cited by 25
Connectivity. The millennium challenge for point-of-care testing.
Kost, G. J. (2000) · PubMed · doi:10.5858/2000-124-1108-c · Cited by 24
Self-citationMohsenzadeh, M. S., Zaeemi, M., Razmyar, J., & Azizzadeh, M. (2015) · Journal of Avian Medicine and Surgery · doi:10.1647/2014-020 · Cited by 23
Evaluation of hematocrit interference with MyStar extra and seven competitive devices
Demircik, F., Ramljak, S., Hermanns, I., Pfützner, A., & Pfützner, A. (2015) · Journal of Diabetes Science and Technology · doi:10.1177/1932296814565790 · Cited by 22
IMPORTANCE OF THE PRE-ANALYTICAL PHASE IN BLOOD GLUCOSE ANALYSIS
Janssen, K., & Delanghe, J. (2010) · Acta Clinica Belgica · doi:10.1179/acb.2010.068 · Cited by 22
Accuracy of different methods for blood glucose measurement in critically ill patients
Júnior, S. A. P., de Assunção, M. S. C., Mazza, B. F., Fernandes, H. D. S., Jackiu, M., Freitas, F. G. R., & Machado, F. R. (2009) · Sao Paulo Medical Journal · doi:10.1590/s1516-31802009000500003 · Cited by 22
Pathophysiology and aetiology of hypoglycaemic crises
Morgan, R. K., Cortes, Y., & Murphy, L. A. (2018) · Journal of Small Animal Practice · doi:10.1111/jsap.12911 · Cited by 26
Kavsak, P. A., Zielinski, N., Li, D., McNamara, P. J., & Adeli, K. (2004) · Clinical Biochemistry · doi:10.1016/j.clinbiochem.2004.06.003 · Cited by 22
Accuracy of the HemoCue Portable Glucose Analyzer in a Large Nonhomogeneous Population
Torjman, M. C., Jahn, L., Joseph, J. I., Crothall, K. D., & Goldstein, B. J. (2001) · Diabetes Technology & Therapeutics · doi:10.1089/15209150152811216 · Cited by 22
Tran, N. K., Godwin, Z. R., Bockhold, J. C., Passerini, A. G., Cheng, J., & Ingemason, M. (2014) · Journal of Burn Care & Research · doi:10.1097/bcr.0b013e31829b3700 · Cited by 21
Point‐of‐care glucose monitoring on the neonatal unit
Makaya, T., Memmott, A., & Bustani, P. (2011) · Journal of Paediatrics and Child Health · doi:10.1111/j.1440-1754.2011.02253.x · Cited by 21
Warner, J. V., Wu, J. Y., Buckingham, N., McLeod, D. S. A., Mottram, B., & Carter, A. C. (2011) · Diabetes Technology & Therapeutics · doi:10.1089/dia.2010.0129 · Cited by 20
Metadata
- Authors
- Tang, Z., Lee, J. H., Louie, R. F., & Kost, G. J.
- Affiliations
- University of California, Davis (US)
- ORCID
- Kost, Gerald J.
- Venue
- Archives of Pathology & Laboratory Medicine
- Published
- 2000-08
- Volume
- 124
- Issue
- 8
- Pages
- 1135-1140
- ISSN
- 0003-9985 (ISSN-L)
- 1543-2165
- OpenAlex
- W1883317613
- Cited by
- 293
- References
- 23
- Citations by year
- 2026: 10 · 2025: 6 · 2024: 1 · 2023: 5 · 2022: 10 · 2021: 8 · 2020: 9 · 2019: 10 · 2018: 13 · 2017: 10 · 2016: 10 · 2015: 14 · 2014: 14 · 2013: 21 · 2012: 11
- FWCI
- 3.06
- Type
- journal article
- Language
- en
- Open access
- closed
- Locations
- Archives of Pathology & Laboratory Medicine (publishedVersion)
- PubMed (publishedVersion)
- Topics
- Health Sciences › Health Professions › Medical Laboratory Technology › Clinical Laboratory Practices and Quality Control
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Diabetes Management and Research
- Health Sciences › Medicine › Endocrinology, Diabetes and Metabolism › Hyperglycemia and glycemic control in critically ill and hospitalized patients
- Keywords
- hematocrit, glucose measurement, point-of-care testing, measurement bias, plasma glucose, meter calibration
- MeSH terms
- Blood Glucose, California, Diagnostic Techniques, Endocrine, Evaluation Studies as Topic, Hematocrit, Humans, Predictive Value of Tests, Reproducibility of Results, Biosensing Techniques, Point-of-Care Systems, Diagnostic Equipment
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.