Comparison of noninvasive pulse transit time estimates as markers of blood pressure using invasive pulse transit time measurements as a reference
Open accessAbstract
Pulse transit time (PTT) measured as the time delay between invasive proximal and distal blood pressure (BP) or flow waveforms (invasive PTT [I-PTT]) tightly correlates with BP PTT estimated as the time delay between noninvasive proximal and distal arterial waveforms could therefore permit cuff-less BP monitoring. A popular noninvasive PTT estimate for this application is the time delay between ECG and photoplethysmography (PPG) waveforms (pulse arrival time [PAT]). Another estimate is the time delay between proximal and distal PPG waveforms (PPG-PTT). PAT and PPG-PTT were assessed as markers of BP over a wide physiologic range using I-PTT as a reference. Waveforms for determining I-PTT, PAT, and PPG-PTT through central arteries were measured from swine during baseline conditions and infusions of various hemodynamic drugs. Diastolic, mean, and systolic BP varied widely in each subject (group average (mean ± SE) standard deviation between 25 ± 2 and 36 ± 2 mmHg). I-PTT correlated well with all BP levels (group average R(2) values between 0.86 ± 0.03 and 0.91 ± 0.03). PPG-PTT also correlated well with all BP levels (group average R(2) values between 0.81 ± 0.03 and 0.85 ± 0.02), and its R(2) values were not significantly different from those of I-PTT PAT correlated best with systolic BP (group average R(2) value of 0.70 ± 0.04), but its R(2) values for all BP levels were significantly lower than those of I-PTT (P < 0.005) and PPG-PTT (P < 0.02). The pre-ejection period component of PAT was responsible for its inferior correlation with BP In sum, PPG-PTT was not different from I-PTT and superior to the popular PAT as a marker of BP.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Pulse arrival time serves as a surrogate measurement for blood pressure.
“The outputs of the chest and limb sensors are then capable of calculating pulse arrival time (PAT), which serves as a surrogate measurement for BP (24, 25).”
From Comprehensive pregnancy monitoring with a network of wireless, soft, and flexible sensors in high- and low-resource health settings (2021) · cites this paper as [24]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
Toward ubiquitous blood pressure monitoring via pulse transit time: Theory and practice
Mukkamala, R., Hahn, J.-O., Inan, O. T., Mestha, L. K., Kim, C.-S., Toreyin, H., & Kyal, S. (2015) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2015.2441951 · Cited by 935 · see citing works
Self-citationThe dynamic elastic properties of the arterial wall
Bergel, D. H. (1961) · The Journal of Physiology · doi:10.1113/jphysiol.1961.sp006687 · Cited by 907 · see citing works
Alterations with Age in the Viscoelastic Properties of Human Arterial Walls
Learoyd, B. M., & Taylor, M. G. (1966) · Circulation Research · doi:10.1161/01.res.18.3.278 · Cited by 684 · see citing works
Determination of pulse wave velocities with computerized algorithms
Chiu, Y. C., Arand, P. W., Shroff, S. G., Feldman, T. E., & Carroll, J. D. (1991) · American Heart Journal · doi:10.1016/0002-8703(91)90153-9 · Cited by 313 · see citing works
Convertino, V. A., Cooke, W. H., & Holcomb, J. B. (2006) · The Journal of Trauma: Injury, Infection, and Critical Care · doi:10.1097/01.ta.0000196663.34175.33 · Cited by 133 · see citing works
Chen, Y., Wen, C., Tao, G., Bi, M., & Li, G. (2009) · Annals of Biomedical Engineering · doi:10.1007/s10439-009-9759-1 · Cited by 115 · see citing works
Regional variation of series elasticity in canine arterial smooth muscles
Cox, R. H. (1978) · American Journal of Physiology-Heart and Circulatory Physiology · doi:10.1152/ajpheart.1978.234.5.h542 · Cited by 89 · see citing works
Influence of muscle length on series elasticity in arterial smooth muscle
Cox, R. H. (1978) · American Journal of Physiology-Cell Physiology · doi:10.1152/ajpcell.1978.234.5.c146 · Cited by 82 · see citing works
Pulse arrival time as a method of obtaining systolic and diastolic blood pressure indirectly
Geddes, L. A., Voelz, M., James, S., & Reiner, D. (1981) · Medical & Biological Engineering & Computing · doi:10.1007/bf02442787 · Cited by 65 · see citing works
Ochiai, R., Takeda, J., Hosaka, H., Sugo, Y., Tanaka, R., & Soma, T. (1999) · Journal of Clinical Monitoring and Computing · doi:10.1023/a:1009950731297 · Cited by 58 · see citing works
Young, C. C., Mark, J. B., White, W. D., DeBree, A., Vender, J. S., & Fleming, A. J. (1995) · The Journal of Clinical Monitoring · doi:10.1007/bf01617519 · Cited by 56 · see citing works
Measurement of pulse-wave velocity using a beat-sampling technique
Pruett, J. D., BOURLAND, J. D., & Geddes, L. A. (1988) · Annals of Biomedical Engineering · doi:10.1007/bf02364622 · Cited by 53 · see citing works
Improved Pulse Wave Velocity Estimation Using an Arterial Tube-Load Model
Gao, M., Zhang, G., Olivier, N. B., & Mukkamala, R. (2014) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2013.2291385 · Cited by 21 · see citing works
Self-citationThe velocity of the arterial pulse wave: a viscous-fluid shock wave in an elastic tube
Painter, P. R. (2008) · Theoretical Biology and Medical Modelling · doi:10.1186/1742-4682-5-15 · Cited by 18 · see citing works
Flagged first, then by citations
Cited by
Sorted by citations
El-Hajj, C., & Kyriacou, P. A. (2020) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2020.101870 · Cited by 300 · see citing works
Wearable Sensing and Telehealth Technology with Potential Applications in the Coronavirus Pandemic
Ding, X., Clifton, D., Ji, N., Lovell, N. H., Bonato, P., Chen, W., Yu, X., Xue, Z., Xiang, T., Long, X., et al. (2021) · IEEE Reviews in Biomedical Engineering · doi:10.1109/rbme.2020.2992838 · Cited by 281 · see citing works
Cuffless Blood Pressure Measurement
Mukkamala, R., Stergiou, G. S., & Avolio, A. P. (2022) · Annual Review of Biomedical Engineering · doi:10.1146/annurev-bioeng-110220-014644 · Cited by 180 · see citing works
Self-citationHuynh, T. H., Jafari, R., & Chung, W. (2018) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2018.2865751 · Cited by 181 · see citing works
Multi-Sensor Fusion Approach for Cuff-Less Blood Pressure Measurement
Miao, F., Liu, Z.-D., Liu, J., Wen, B., He, Q., & Li, Y. (2019) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2019.2901724 · Cited by 136 · see citing works
Photoplethysmography signal processing and synthesis
Mejía-Mejía, E., Allen, J., Budidha, K., El-Hajj, C., Kyriacou, P. A., & Charlton, P. H. (2022) · Photoplethysmography · doi:10.1016/b978-0-12-823374-0.00015-3 · Cited by 138 · see citing works
Martin, S. L.-O., Carek, A. M., Kim, C., Ashouri, H., Inan, O. T., Hahn, J., & Mukkamala, R. (2016) · Scientific Reports · doi:10.1038/srep39273 · Cited by 131 · see citing works
Self-citationKim, C., Carek, A. M., Inan, O. T., Mukkamala, R., & Hahn, J. (2018) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2018.2797239 · Cited by 104 · see citing works
Self-citationRyu, D., Kim, D. H., Price, J. T., Lee, J. Y., Chung, H. U., Allen, E., Walter, J. R., Jeong, H., Cao, J., Kulikova, E., et al. (2021) · Proceedings of the National Academy of Sciences · doi:10.1073/pnas.2100466118 · Cited by 105 · see citing works
Conventional pulse transit times as markers of blood pressure changes in humans
Block, R., Yavarimanesh, M., Natarajan, K., Carek, A. M., Mousavi, A. S., Chandrasekhar, A., Kim, C., Zhu, J., Schifitto, G., Mestha, L. K., et al. (2020) · Scientific Reports · doi:10.1038/s41598-020-73143-8 · Cited by 107 · see citing works
Self-citationA novel art of continuous noninvasive blood pressure measurement
Fortin, J., Rogge, D. E., Fellner, C., Flotzinger, D., Grond, J., Lerche, K., & Saugel, B. (2021) · Nature Communications · doi:10.1038/s41467-021-21271-8 · Cited by 104 · see citing works
Cuffless Blood Pressure Monitoring: Promises and Challenges
Pandit, J. A., Lores, E., & Batlle, D. (2020) · Clinical Journal of the American Society of Nephrology · doi:10.2215/cjn.03680320 · Cited by 101 · see citing works
Wearable Cuff-Less Blood Pressure Estimation at Home via Pulse Transit Time
Ganti, V. G., Carek, A. M., Nevius, B. N., Heller, J. A., Etemadi, M., & Inan, O. T. (2020) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2020.3021532 · Cited by 101 · see citing works
Cuff-Less Methods for Blood Pressure Telemonitoring
Bard, D. M., Joseph, J. I., & van Helmond, N. (2019) · Frontiers in Cardiovascular Medicine · doi:10.3389/fcvm.2019.00040 · Cited by 84 · see citing works
The Microsoft Research Aurora Project: Important Findings on Cuffless Blood Pressure Measurement
Mukkamala, R., Shroff, S. G., Landry, C., Kyriakoulis, K. G., Avolio, A. P., & Stergiou, G. S. (2022) · Hypertension · doi:10.1161/hypertensionaha.122.20410 · Cited by 82 · see citing works
Self-citationRajala, S., Lindholm, H., & Taipalus, T. (2018) · Physiological Measurement · doi:10.1088/1361-6579/aac7ac · Cited by 70 · see citing works
Pulse Wave Velocity: Methodology, Clinical Applications, and Interplay with Heart Rate Variability
Pilz, N., Heinz, V., Ax, T., Fesseler, L., Patzak, A. R., & Bothe, T. L. (2024) · Reviews in Cardiovascular Medicine · doi:10.31083/j.rcm2507266 · Cited by 70 · see citing works
Xu, Z., Liu, J., Chen, X., Wang, Y., & Zhao, Z. (2017) · Computers in Industry · doi:10.1016/j.compind.2017.04.003 · Cited by 71 · see citing works
Multimodal Wrist Biosensor for Wearable Cuff-less Blood Pressure Monitoring System
Rachim, V. P., & Chung, W. (2019) · Scientific Reports · doi:10.1038/s41598-019-44348-3 · Cited by 67 · see citing works
Liu, C., Kim, J., Kwak, S. S., Hourlier‐Fargette, A., Avila, R., Vogl, J., Tzavelis, A., Chung, H. U., Lee, J. Y., Kim, D. H., et al. (2021) · Advanced Healthcare Materials · doi:10.1002/adhm.202100383 · Cited by 57 · see citing works
Lin, W.-H., Chen, F., Geng, Y., Ji, N., Fang, P., & Li, G. (2020) · Biomedical Signal Processing and Control · doi:10.1016/j.bspc.2020.102198 · Cited by 57 · see citing works
Wearable sensors: can they benefit patients with chronic kidney disease?
Wieringa, F. P., Broers, N. J. H., Kooman, J. P., SANDE, F. M. V. D., & Hoof, C. A. V. (2017) · Expert Review of Medical Devices · doi:10.1080/17434440.2017.1342533 · Cited by 56 · see citing works
Huttunen, J. M. J., Kärkkäinen, L., & Lindholm, H. J. (2019) · PLoS Computational Biology · doi:10.1371/journal.pcbi.1007259 · Cited by 52 · see citing works
Butlin, M., Shirbani, F., Barin, E. S., Tan, I., Spronck, B., & Avolio, A. P. (2018) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2018.2823333 · Cited by 45 · see citing works
Stojanova, A., Koceski, S., & Koceska, N. (2019) · Journal of Medical Systems · doi:10.1007/s10916-018-1138-8 · Cited by 46 · see citing works
Pernice, R., Antonacci, Y., Zanetti, M., Busacca, A., Marinazzo, D., Faes, L., & Nollo, G. (2021) · Frontiers in Neuroscience · doi:10.3389/fnins.2020.602584 · Cited by 42 · see citing works
Sagirova, Z. N., Kuznetsova, N., Gogiberidze, N. A., Gognieva, D. G., Suvorov, А. Y., Chomakhidze, P. S., Omboni, S., Saner, H. E., & Kopylov, P. (2021) · Sensors · doi:10.3390/s21103525 · Cited by 42 · see citing works
Mamun, M. M. R. K., & Sherif, A. (2022) · Bioengineering · doi:10.3390/bioengineering10010027 · Cited by 37 · see citing works
Escobar, B., Villa, R. A. T., & Kyriacou, P. A. (2018) · Frontiers in Physiology · doi:10.3389/fphys.2018.01848 · Cited by 35 · see citing works
Current Status and Prospects of Health-Related Sensing Technology in Wearable Devices
Cho, J. (2019) · Journal of Healthcare Engineering · doi:10.1155/2019/3924508 · Cited by 33 · see citing works
Kao, Y.-H., Chao, P. C.-P., & Wey, C. (2018) · Microsystem Technologies · doi:10.1007/s00542-018-3877-3 · Cited by 32 · see citing works
Deep Learning Fused Wearable Pressure and PPG Data for Accurate Heart Rate Monitoring
Mehrgardt, P., Khushi, M., Poon, S. K.-S., & Withana, A. I. (2021) · IEEE Sensors Journal · doi:10.1109/jsen.2021.3123243 · Cited by 33 · see citing works
Cuff-less blood pressure measurement using pulse arrival time and a Kalman filter
Zhang, Q. M., Chen, X., Fang, Z., Xue, Y., Zhan, Q., Yang, T., & Xia, S. (2017) · Journal of Micromechanics and Microengineering · doi:10.1088/1361-6439/27/2/024002 · Cited by 31 · see citing works
Cohen, J. B., Byfield, R. L., Hardy, S. T., Juraschek, S. P., Miller, N. H., Mukkamala, R., Picone, D. S., Thiele, R. H., Yang, E., & Brady, T. M. (2025) · Hypertension · doi:10.1161/hyp.0000000000000254 · Cited by 29 · see citing works
Self-citationFreithaler, M., Chandrasekhar, A., Dhamotharan, V., Landry, C., Shroff, S. G., & Mukkamala, R. (2023) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2023.3275031 · Cited by 26 · see citing works
Self-citationSpeth, J., Vance, N., Sporrer, B., Niu, L., Flynn, P. J., & Czajka, A. (2024) · IEEE Transactions on Instrumentation and Measurement · doi:10.1109/tim.2024.3476556 · Cited by 25 · see citing works
Cuffless and Continuous Blood Pressure Monitoring Using a Single Chest-Worn Device
Park, J., Yang, S.-M., Sohn, J., Lee, J., Lee, S., Ku, Y., & Kim, H. C. (2019) · IEEE Access · doi:10.1109/access.2019.2942184 · Cited by 25 · see citing works
Multimodal Monitoring of Cardiovascular Responses to Postural Changes
Mol, A., Maier, A. B., Wezel, R. J. A. V., & Meskers, C. G. M. (2020) · Frontiers in Physiology · doi:10.3389/fphys.2020.00168 · Cited by 24 · see citing works
Carek, A. M., & Holz, C. (2018) · Proceedings of the ACM on Interactive Mobile Wearable and Ubiquitous Technologies · doi:10.1145/3264906 · Cited by 22 · see citing works
Estimation of systolic blood pressure by Random Forest using heart sounds and a ballistocardiogram
González-Landaeta, R., Ramirez, B., & Mejia, J. (2022) · Scientific Reports · doi:10.1038/s41598-022-22205-0 · Cited by 21 · see citing works
Photoplethysmography in noninvasive blood pressure monitoring
Mukkamala, R., Hahn, J., & Chandrasekhar, A. (2021) · Elsevier eBooks · doi:10.1016/b978-0-12-823374-0.00010-4 · Cited by 21 · see citing works
Self-citationPoppa, T., Witte, S. D., Vanderhasselt, M., Bechara, A., & Baeken, C. (2020) · International Journal of Psychophysiology · doi:10.1016/j.ijpsycho.2019.12.011 · Cited by 19 · see citing works
Evaluation of blood pressure estimation models based on pulse arrival time
Bote, J. M., Recas, J., & Hermida, R. (2020) · Computers & Electrical Engineering · doi:10.1016/j.compeleceng.2020.106616 · Cited by 19 · see citing works
Kim, D., Lee, E., Kim, J., Park, P., & Cho, S. (2019) · IEEE Sensors Journal · doi:10.1109/jsen.2019.2950367 · Cited by 18 · see citing works
Huttunen, J. M. J., Kärkkäinen, L., Honkala, M., & Lindholm, H. J. (2019) · International Journal for Numerical Methods in Biomedical Engineering · doi:10.1002/cnm.3303 · Cited by 17 · see citing works
Continuous and Noninvasive Blood Pressure Estimation by Two-Sensor Measurement of Pulse Transit Time
Rasool, A., Rafiq, M., Nasir, A., & Kashif, F. M. (2018) · International Conference on Emerging Technologies · doi:10.1109/icet.2018.8603612 · Cited by 13 · see citing works
Impairing Factors in Remote-PPG Pulse Transit Time Measurements on the Face
Moço, A., Stuijk, S., van Gastel, M., & de Haan, G. (2018) · IEEE/CVF Conference on Computer Vision and Pattern Recognition Workshops (CVPRW) · doi:10.1109/cvprw.2018.00184 · Cited by 11 · see citing works
Fischer, C., & Penzel, T. (2018) · Physiological Measurement · doi:10.1088/1361-6579/aaf298 · Cited by 12 · see citing works
Advancing PPG-Based Continuous Blood Pressure Monitoring from a Generative Perspective
Ji, H., & Zhou, P. (2024) · ACM Conference on Embedded Networked Sensor Systems (SenSys) · doi:10.1145/3666025.3699365 · Cited by 11 · see citing works
Xu, Y., Ping, P., Wang, D., & Zhang, W. (2018) · Journal of Healthcare Engineering · doi:10.1155/2018/5396030 · Cited by 11 · see citing works
Vasan, S., Lim, M. H., Eikelis, N., & Lambert, É. (2024) · Scientific Reports · doi:10.1038/s41598-024-65039-8 · Cited by 10 · see citing works
Relationship between mechanisms of blood pressure change and facial skin temperature distribution
Nakane, N., Oiwa, K., & Nozawa, A. (2019) · Artificial Life and Robotics · doi:10.1007/s10015-019-00565-7 · Cited by 9 · see citing works
Initialization of Pulse Transit Time-Based Blood Pressure Monitors
Mukkamala, R., & Hahn, J. (2019) · Springer eBooks · doi:10.1007/978-3-030-24701-0_10 · Cited by 8 · see citing works
Self-citationPernice, R., Parisi, A., Guarino, S., Fallica, G., Vinciguerra, V., Ferla, G., Faes, L., & Busacca, A. (2020) · IEEE Mediterranean Electrotechnical Conference (MELECON) · doi:10.1109/melecon48756.2020.9140716 · Cited by 7 · see citing works
Pernice, R., Nollo, G., Zanetti, M., Cecco, M. D., Busacca, A., & Faes, L. (2019) · IEEE EUROCON - International Conference on Computer as a Tool (EUROCON) · doi:10.1109/eurocon.2019.8861614 · Cited by 7 · see citing works
Non-invasive pulse wave velocity measurement in mice
Zakaria, H., & Hasimun, P. (2017) · International Seminar on Sensors, Instrumentation, Measurement and Metrology (ISSIMM) · doi:10.1109/issimm.2017.8124270 · Cited by 7 · see citing works
Xu, Y., Zhang, W., Wang, D., & Ping, P. (2018) · IEEE International Instrumentation and Measurement Technology Conference · doi:10.1109/i2mtc.2018.8409817 · Cited by 6 · see citing works
Ito, A., Nagumo, K., Oiwa, K., & Nozawa, A. (2024) · IEEJ Transactions on Electrical and Electronic Engineering · doi:10.1002/tee.24011 · Cited by 5 · see citing works
Practical Application of a New Cuffless Blood Pressure Measurement Method
Gogiberidze, N. A., Suvorov, А. Y., Sultygova, E. A., Sagirova, Z. N., Kuznetsova, N., Gognieva, D. G., Chomakhidze, P. S., Фролов, В. А., Aa, B., Фератовна, М. Д., et al. (2023) · Pathophysiology · doi:10.3390/pathophysiology30040042 · Cited by 5 · see citing works
Nordine, M., Pille, M., Kraemer, J. F., Berger, C., Brandhorst, P., Kaeferstein, P., Kopetsch, R., Wessel, N., Trauzeddel, R. F., & Treskatsch, S. (2023) · Sensors · doi:10.3390/s23063304 · Cited by 6 · see citing works
PTT Based Portable Cuffless Systolic Blood Pressure Estimation
Chatterjee, A., Pal, S., & Mitra, M. (2020) · IEEE Calcutta Conference (CALCON) · doi:10.1109/calcon49167.2020.9106545 · Cited by 5 · see citing works
Chen, K., Zhang, X., Zou, Q., & Wang, X. (2020) · International Conference on Intelligent Transportation, Big Data & Smart City (ICITBS) · doi:10.1109/icitbs49701.2020.00196 · Cited by 4 · see citing works
Pulse Arrival Time Is Associated With Hemorrhagic Volume in a Porcine Model: A Pilot Study
Booth, G. J., Cole, J., Geiger, P., Adams, J., Barnhill, J., & Hughey, S. B. (2021) · Military Medicine · doi:10.1093/milmed/usab069 · Cited by 4 · see citing works
Camera-based blood pressure monitoring
Natarajan, K., Yavarimanesh, M., Wang, W., & Mukkamala, R. (2021) · Elsevier eBooks · doi:10.1016/b978-0-12-822281-2.00014-7 · Cited by 4 · see citing works
Self-citationEstimation of Beat-to-Beat Interval and Systolic Time Intervals Using Phono- and Seismocardiograms
Ahmaniemi, T., Rajala, S., & Lindholm, H. J. (2019) · Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) · doi:10.1109/embc.2019.8856931 · Cited by 4 · see citing works
AI plays a crucial role in the diagnosis and prognosis cardio pregnant women
Hilal, S. (2024) · Elsevier eBooks · doi:10.1016/b978-0-443-22008-1.00010-5 · Cited by 3 · see citing works
Machine Learned Pulse Transit Time (MLPTT) Measurements from Photoplethysmography
Mehrgardt, P., Khushi, M., Withana, A. I., & Poon, S. K.-S. (2020) · Lecture notes in computer science · doi:10.1007/978-3-030-63836-8_5 · Cited by 3 · see citing works
Gogiberidze, N. A., Sagirova, Z. N., Kuznetsova, N., Gognieva, D. G., Chomakhidze, P. S., Saner, H. E., & Kopylov, P. (2021) · Sechenov Medical Journal · doi:10.47093/2218-7332.2021.12.1.39-49 · Cited by 3 · see citing works
Erkurşun, B., Uçar, A., & Tekin, A. (2022) · Sensors and Actuators A Physical · doi:10.1016/j.sna.2022.113858 · Cited by 3 · see citing works
Fingertip Plethysmography Using A Sensitive Wafer-Level-Packaged Capacitive MEMS Strain Sensor
Jiang, X., Shokouhmand, A., Ebadi, N., & Ayazi, F. (2024) · IEEE International Conference on Micro Electro Mechanical Systems (MEMS) · doi:10.1109/mems58180.2024.10439370 · Cited by 2 · see citing works
A Photoplethysmographic Monitor for Local Pulse Wave Velocity Measurement
Matheus, P., Antonio, W., Pedro, H., & Talles, M. (2020) · International Journal of Computer Applications · doi:10.5120/ijca2020919811 · Cited by 2 · see citing works
Age Dependence of Pulse Transit Time Derived from dual PPGs and PPG-ECG duo ‐ A comparative Study
Mukundan, L., & Kayalvizhi, N. (2017) · IEEE India Conference · doi:10.1109/indicon.2017.8487630 · Cited by 2 · see citing works
Evaluation of Pulse Transit Time for Different Sensing Methodologies of Arterial Waveforms
Janjua, G. M. W., Finlay, D., Guldenring, D., Haq, A. U., & McLaughlin, J. A. (2023) · IEEE Access · doi:10.1109/access.2023.3264291 · Cited by 2 · see citing works
Furudate, H., Nagumo, K., & Nozawa, A. (2025) · IEEE International Conference on Imaging Systems and Techniques (IST) · doi:10.1109/ist66504.2025.11268398 · Cited by 1 · see citing works
Sharma, P., Parveen, S., Masood, S., & Noohu, M. M. (2021) · Comparative Exercise Physiology · doi:10.3920/cep210016 · Cited by 1 · see citing works
TAMURA, T. (2017) · Springer eBooks · doi:10.1007/978-3-319-69362-0_4 · Cited by 1 · see citing works
Blood Pressure Measurement using in Wrist PPG and ECG
Kusumah, I. H., Artiyasa, M., Al-Bukhori, Khoiri, M. I., Ramadhan, A. D., & Supiyandi (2020) · IEEE International Conference on Computing, Engineering and Design (ICCED) · doi:10.1109/icced51276.2020.9415821 · Cited by 1 · see citing works
van Helmond, N., & Joseph, J. I. (2018) · Physiological Measurement · doi:10.1088/1361-6579/aadf11 · Cited by 1 · see citing works
Noninvasive, Intrusive, and Nonintrusive Measurements
Sanei, S., Jarchi, D., & Constantinides, A. G. (2020) · doi:10.1002/9781119390060.ch6 · Cited by 1 · see citing works
A Study on the Wearable and Cuffless Continuous Blood Pressure Monitoring System
박종현 (2019) · Seoul National University Open Repository (Seoul National University)
Devices based on the time difference between signals
Bolić, M. (2023) · Elsevier eBooks · doi:10.1016/b978-0-12-820947-9.00007-6
State-of-the-Art Analysis of Cuffless Methods for Blood Pressure Measurement
Burbach, S., Brück, R., & Keil, A. (2024) · World Congress on Medical Physics and Biomedical Engineering, September 7 - 12, 2009, Munich, Germany · doi:10.1007/978-3-031-62502-2_15
Chand, S., Chiu, N.-T., Chou, Y.-H., Alian, A. A., Shelley, K. H., & Wu, H. (2025) · Physiological Measurement · doi:10.1088/1361-6579/adf489
Resting blood pressure estimation from a low-resolution single thermal facial image
Furudate, H., Nagumo, K., & Nozawa, A. (2025) · Artificial Life and Robotics · doi:10.1007/s10015-025-01062-w
Ryan, L. E., Clarke, M. M., Ray, D., Glenning, J. P., Harrington, H. A., Smolich, J. J., Cheung, M. M. H., & Mynard, J. P. (2025) · Heart Lung and Circulation · doi:10.1016/j.hlc.2025.09.005
Pointwise Reliability Assessment for Continuous Non-Invasive Blood Pressure Estimation
Loureiro, J., Zhao, Z., Gil, P. S., Bresch, E., Henriques, J. M. O., & Muehlsteff, J. (2026) · World Congress on Medical Physics and Biomedical Engineering, September 7 - 12, 2009, Munich, Germany · doi:10.1007/978-3-032-39674-7_36
- Authors
- Gao, M., Olivier, N. B., & Mukkamala, R.
- Affiliations
- Michigan State University (US)
- Corresponding author
- Mukkamala, Ramakrishna
- Venue
- Physiological Reports
- Publisher
- Wiley
- Published
- 2016-05
- Volume
- 4
- Issue
- 10
- Pages
- e12768
- ISSN
- 2051-817X (ISSN-L)
- PMC
- PMC4886159
- OpenAlex
- W2406255031
- Cited by
- 86
- References
- 14
- Citations by year
- 2026: 1 · 2025: 5 · 2024: 8 · 2023: 5 · 2022: 5 · 2021: 12 · 2020: 16 · 2019: 14 · 2018: 13 · 2017: 6 · 2016: 1
- FWCI
- 5.46
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Licence
- Locations
- Physiological Reports (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- Funders
- National Institutes of Health (EB018818)
- Retia Medical
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Cardiovascular Health and Disease Prevention
- Health Sciences › Medicine › Surgery › Hemodynamic Monitoring and Therapy
- Keywords
- pulse transit time, blood pressure estimation, cuffless blood pressure measurement, photoplethysmography, pulse arrival time, non-invasive blood pressure measurement, pre-ejection period, arterial pressure waveform, pigs
- MeSH terms
- Animals, Blood Pressure, Blood Pressure Determination, Swine, Time Factors, Pulse Wave Analysis
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review