A Chest‐Laminated Ultrathin and Stretchable E‐Tattoo for the Measurement of Electrocardiogram, Seismocardiogram, and Cardiac Time Intervals

Open access

Abstract

Seismocardiography (SCG) is a measure of chest vibration associated with heartbeats. While skin soft electronic tattoos (e‐tattoos) have been widely reported for electrocardiogram (ECG) sensing, wearable SCG sensors are still based on either rigid accelerometers or non‐stretchable piezoelectric membranes. This work reports an ultrathin and stretchable SCG sensing e‐tattoo based on the filamentary serpentine mesh of 28‐µm‐thick piezoelectric polymer, polyvinylidene fluoride (PVDF). 3D digital image correlation (DIC) is used to map chest vibration to identify the best location to mount the e‐tattoo and to investigate the effects of substrate stiffness. As piezoelectric sensors easily suffer from motion artifacts, motion artifacts are effectively reduced by performing subtraction between a pair of identical SCG tattoos placed adjacent to each other. Integrating the soft SCG sensor with a pair of soft gold electrodes on a single e‐tattoo platform forms a soft electro‐mechano‐acoustic cardiovascular (EMAC) sensing tattoo, which can perform synchronous ECG and SCG measurements and extract various cardiac time intervals including systolic time interval (STI). Using the EMAC tattoo, strong correlations between STI and the systolic/diastolic blood pressures, are found, which may provide a simple way to estimate blood pressure continuously and noninvasively using one chest‐mounted e‐tattoo.

Overview

CoChat Summary

TL;DR

This paper reports an ultrathin, stretchable chest-laminated e-tattoo using filamentary serpentine 28-µm PVDF for seismocardiogram (SCG) sensing and integrated Au electrodes for synchronous ECG/SCG (EMAC) sensing. It demonstrates high-fidelity SCG/ECG, motion-artifact reduction via dual PVDF subtraction, and a strong negative correlation between the systolic time interval (RAC) and blood pressure across four subjects.

Key findings

  • The FS PVDF e-tattoo is 28 µm thick, has 112.9% stretchability and 0.4 mV per µε sensitivity, and an effective modulus of 8.5 MPa; the integrated EMAC tattoo measures 63.5 mm × 38.1 mm × 0.122 mm, weighs 150 mg, and has about 100% stretchability.
  • 3D DIC mapped chest vibration: S1 was ≈5 µm around the left nipple area and S2 was ≈3 µm at a higher location; the middle e-tattoo was best for measuring both S1 and S2. FEM estimated substrate-dependent outputs of 2.74 mV (no substrate), 2.20 mV (Tegaderm), and 0.86 mV (PET).
  • Dual PVDF subtraction reduced motion artifacts, making motion-period SNRs comparable to rest-period SNRs; a single PET-substrate sensor had negative SNR under motion.
  • In four subjects, RAC negatively correlated with blood pressure: average Pearson R was −0.772 for SBP–RAC, −0.858 for DBP–RAC, and −0.538 for PBP–RAC; individual calibration was necessary.

Methods

  • The device was fabricated by a cut-and-paste method to pattern FS PVDF and Au/PET electrodes on Tegaderm; characterization used DMA, electrical measurements, and ABAQUS FEM. 3D DIC used two high-speed cameras at 500 Hz for 16 s to map chest deformation.
  • Human experiments were IRB-approved. Synchronous ECG and SCG were measured by the EMAC tattoo and compared with an accelerometer; blood pressure was measured by SOMNOtouch NIBP during Valsalva maneuvers in four subjects. Signals were filtered with 2–40 Hz (ECG) and 12–40 Hz (SCG) Butterworth filters, and RAC intervals were extracted from R and AC peaks.
  • Analysis included Pearson correlations between BP and RAC, plus FEM modeling of substrate effects and electromechanical output.

Limitations

  • Stated: fitting parameters and correlation coefficients varied by individual and posture; no master fitting equations could be derived, so per-individual calibration is necessary. Valsalva data were omitted because both SOMNO and SCG signals suffered excessive noise from body shivering, and mismatches were likely from motion or slight posture change.
  • Stated: wireless NFC or Bluetooth integration remains future work for practical ambulatory applications.
  • Evident from the text: BP validation involved only four subjects, and the main DIC mapping was performed on an adult male; no long-term ambulatory validation is reported.

Model-written summary. Check it against the paper before relying on it.

Integrity signals

Registry

Crossref · 10.1002/advs.201900290

Cites

63 of 63 references verified0 citation errors · 0 not checkable

Statistics

Nothing to recompute.

Figures

Figure 1 Stretchable PVDF vibration sensor (i.e. PVDF e‐tattoo) and 3D DIC method for SCG measurement. a) A schematic of the stretchable PVDF e‐tattoo. b) Photographs of an undeformed and stretched PVDF e‐tattoo. c) The PVDF e‐tattoo (red boxed) and a commercial accelerometer (green arrowed) attached on human chest. d) SCG signals measured by the PVDF e‐tattoo and the accelerometer. e) A photograph of the 3D DIC setup for mapping human chest deformation. f) A photograph of a human chest mounted with three PVDF e‐tattoos and painted with a random speckle pattern. Positions of the three e‐tattoos are denoted as Top, Mid., and Bot. g,h) The out‐of‐plane displacement map averaged at S1 and S2 peak times, respectively. i) Measured signals by 3D DIC method (raw, filtered) and the accelerometer (Acc.) from the chest. j,k) SCG signals at three different positions (Top, Mid., and Bot.) captured by 3D DIC method and PVDF e‐tattoos, respectively. l) The correlation of SCG peak times (S1 and S2) measured by the 3D DIC method, PVDF e‐tattoo, and the accelerometer.
Figure 1 Stretchable PVDF vibration sensor (i.e. PVDF e‐tattoo) and 3D DIC method for SCG measurement. a) A schematic of the stretchable PVDF e‐tattoo. b) Photographs of an undeformed and stretched PVDF e‐tattoo. c) The PVDF e‐tattoo (red boxed) and a commercial accelerometer (green arrowed) attached on human chest. d) SCG signals measured by the PVDF e‐tattoo and the accelerometer. e) A photograph of the 3D DIC setup for mapping human chest deformation. f) A photograph of a human chest mounted with three PVDF e‐tattoos and painted with a random speckle pattern. Positions of the three e‐tattoos are denoted as Top, Mid., and Bot. g,h) The out‐of‐plane displacement map averaged at S1 and S2 peak times, respectively. i) Measured signals by 3D DIC method (raw, filtered) and the accelerometer (Acc.) from the chest. j,k) SCG signals at three different positions (Top, Mid., and Bot.) captured by 3D DIC method and PVDF e‐tattoos, respectively. l) The correlation of SCG peak times (S1 and S2) measured by the 3D DIC method, PVDF e‐tattoo, and the accelerometer.CC BY · PubMed Central
Figure 2 Electromechanical characterization of the PVDF e‐tattoo. a) A schematic of a PVDF e‐tattoo subjected to tensile test with in situ electrical measurements. b) Photographs of a PVDF e‐tattoo under tensile test, from 0% to 30%. c) Stress–strain curves of three different PVDF e‐tattoos, whose slope suggests the effective modulus to be 8.5 MPa. d) The measured electrical resistance versus tensile strain curves indicate the stretchability of a straight PVDF ribbon to be 5.8% (black) and a filamentary serpentine (FS) PVDF network to be 112.9% (red). e) Voltage output from the FS PVDF e‐tattoo under 5 Hz sinusoidal strain amplitude from 0.01% to 1%. f) Experimental (red) and FEM (black) results of compensated peak‐to‐peak voltage output with respect to peak‐to‐peak applied strains.
Figure 2 Electromechanical characterization of the PVDF e‐tattoo. a) A schematic of a PVDF e‐tattoo subjected to tensile test with in situ electrical measurements. b) Photographs of a PVDF e‐tattoo under tensile test, from 0% to 30%. c) Stress–strain curves of three different PVDF e‐tattoos, whose slope suggests the effective modulus to be 8.5 MPa. d) The measured electrical resistance versus tensile strain curves indicate the stretchability of a straight PVDF ribbon to be 5.8% (black) and a filamentary serpentine (FS) PVDF network to be 112.9% (red). e) Voltage output from the FS PVDF e‐tattoo under 5 Hz sinusoidal strain amplitude from 0.01% to 1%. f) Experimental (red) and FEM (black) results of compensated peak‐to‐peak voltage output with respect to peak‐to‐peak applied strains.CC BY · PubMed Central
Figure 3 Substrate effects on FS PVDF vibration sensors. a) Photographs of sensors with no substrate, 47‐µm‐thick Tegaderm, and 50‐µm‐thick PET covered over the PVDF. b) SCG signals recorded by the three different types of PVDF sensors at the Mid. location. The 12–40 Hz range, time‐averaged maximum principal strain maps of c) no substrate case, d) Tegaderm case, and e) PET case. FEM results of maximum principal strain in the FS PVDF with f) no substrate, g) Tegaderm substrate, and h) PET substrate.
Figure 3 Substrate effects on FS PVDF vibration sensors. a) Photographs of sensors with no substrate, 47‐µm‐thick Tegaderm, and 50‐µm‐thick PET covered over the PVDF. b) SCG signals recorded by the three different types of PVDF sensors at the Mid. location. The 12–40 Hz range, time‐averaged maximum principal strain maps of c) no substrate case, d) Tegaderm case, and e) PET case. FEM results of maximum principal strain in the FS PVDF with f) no substrate, g) Tegaderm substrate, and h) PET substrate.CC BY · PubMed Central
Figure 4 Dual‐sensor‐based motion artifact cancellation scheme. a) Raw and filtered SCG signals captured by one PET‐covered PVDF sensor under rest (blue shaded) and motion (yellow‐shaded) conditions. b) Periodograms of the filtered SCG signals under rest and motion conditions. c) A photograph of the dual sensor system attached on human chest. d) Filtered SCG signals and e) corresponding envelopes recorded by upper and lower PVDF sensors and their subtraction result under normal and motion conditions. f) Signal‐to‐noise ratios (SNR) of single and dual sensor systems with different substrates (PET vs Tegaderm) under normal and motion conditions.
Figure 4 Dual‐sensor‐based motion artifact cancellation scheme. a) Raw and filtered SCG signals captured by one PET‐covered PVDF sensor under rest (blue shaded) and motion (yellow‐shaded) conditions. b) Periodograms of the filtered SCG signals under rest and motion conditions. c) A photograph of the dual sensor system attached on human chest. d) Filtered SCG signals and e) corresponding envelopes recorded by upper and lower PVDF sensors and their subtraction result under normal and motion conditions. f) Signal‐to‐noise ratios (SNR) of single and dual sensor systems with different substrates (PET vs Tegaderm) under normal and motion conditions.CC BY · PubMed Central
Figure 5 Stretchable EMAC sensing tattoo for continuous BP estimation. a) A photograph of the EMAC tattoo that houses an FS PVDF vibration sensor and a pair of FS Au electrodes. b) Tattoo‐like behavior of ultrathin, stretchable EMAC on human skin. c) A photograph of an EMAC sensing tattoo on human chest with wire connections. d) Synchronously measured ECG (black) and SCG (red) by EMAC tattoo after filtering with a variety of cardiac time intervals illustrated. e) The negative correlation between SBP/DBP measured by SOMNOtouch and RAC measured by an EMAC sensing tattoo for one subject. f) Continuous SBP and DBP estimated by SOMNOtough (black) and EMAC tattoo (red) during Valsalva maneuver.
Figure 5 Stretchable EMAC sensing tattoo for continuous BP estimation. a) A photograph of the EMAC tattoo that houses an FS PVDF vibration sensor and a pair of FS Au electrodes. b) Tattoo‐like behavior of ultrathin, stretchable EMAC on human skin. c) A photograph of an EMAC sensing tattoo on human chest with wire connections. d) Synchronously measured ECG (black) and SCG (red) by EMAC tattoo after filtering with a variety of cardiac time intervals illustrated. e) The negative correlation between SBP/DBP measured by SOMNOtouch and RAC measured by an EMAC sensing tattoo for one subject. f) Continuous SBP and DBP estimated by SOMNOtough (black) and EMAC tattoo (red) during Valsalva maneuver.CC BY · PubMed Central

Verification

Claims the paper makes

4 Valid · 4 Not valid · 31 Not checked · 114 Own results and interpretations

90 more claims

Create a free account to see all 153 claims and to verify the 7 references still open.

See all claims — free account
  • It may state two separate claims as one. 40
  • Probably not a substantial claim. 20

Create a free account to see these claims.

Claims other papers make about this one

Each is checked against this paper’s own text: does it say what the citing paper says it does?

Self-check

Percentages, sums and reported differences re-computed from the paper’s own numbers.

Each contradiction between two passages listed here was confirmed by an independent check.

1 of 1

Citations

Cites

Show all 63 references

Research works printed but not linked: 15

  1. Castiglioni P., 2007, Proc. of the Annual Int. IEEE EMBS, 3954
  2. Hendriks F. M., 2005, Technische Universiteit Eindhoven, 106
  3. Specialties M., 2013, Measurement, 3
  4. Lionel W. Y. L., 2016, IEEE Eng. Med. Biol., 38, 4816
  5. Ameri S. K., 2016, Int. Electron Devices Meeting
  6. Paukkunen M., 2013, J. Bioeng. Biomed. Sci., 3, 1
  7. Singh A., 2013, J. Dental Mater. Tech., 4, 682
  8. Obropta E. W., 2016, Aerospace Conf. Proc.
  9. Obropta E. W., 2015, IEEE Aerospace Conf.
  10. 10.1063/1.2817234 · DOI · Paper page
  11. Pandia K., 2010, 2010 IEEE Int. Conf. on Acoustics Speech and Signal Processing (ICASSP)
  12. Di Rienzo M., 2012, Annual Int. Conf. of the IEEE on Engineering in Medicine and Biology Society (EMBC), 5050
  13. Korzeniowska‐Kubacka I., 2006, Folia Cardiol., 13, 319
  14. Imtiaz M. S., 2013, IEEE Canadian Conf. on Electrical and Computer Engineering (CCECE), 501
  15. Aubert X. L., 2007, Proc. of the Annual Int. IEEE EMBS, 1707

Cited by

See all 220

Metadata

Authors
Ha, T., Tran, J., Liu, S., Jang, H., Jeong, H., Mitbander, R., Huh, H., Qiu, Y., Duong, J., Wang, R. L., Wang, P., Tandon, A., Sirohi, J., & Lu, N.
Affiliations
  • The University of Texas at Austin (US)
  • Texas Materials Institute (US)
  • Children's Medical Center (US)
Corresponding author
Lu, Nanshu
Venue
Advanced Science
Publisher
Wiley
Published
2019-05-21
Volume
6
Issue
14
Pages
1900290-1900290
ISSN
2198-3844 (ISSN-L)
OpenAlex
W2945869729
Cited by
220
References
63
Citations by year
2026: 15 · 2025: 28 · 2024: 33 · 2023: 34 · 2022: 44 · 2021: 34 · 2020: 26 · 2019: 6
FWCI
9.68
Type
journal article
Language
en
Open access
Locations
  • Advanced Science (publishedVersion, cc-by)
  • Europe PMC (PubMed Central) (publishedVersion, cc-by)
  • PubMed Central (publishedVersion, cc-by)
  • PubMed (publishedVersion)
  • DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
  • eScholarship (California Digital Library) (submittedVersion, cc-by)
Funders
  • Office of Naval Research (N00014‐16‐1‐2044)
  • Air Force Office of Scientific Research (FA9550‐15‐1‐0112)
  • National Science Foundation (CNS‐1738293, UL1TR001105)
Topics
  • Physical Sciences › Engineering › Biomedical Engineering › Advanced Sensor and Energy Harvesting Materials
  • Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
  • Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › ECG Monitoring and Analysis
Keywords
seismocardiography, stretchable sensors, polyvinylidene fluoride, piezoelectric polymers, electrocardiogram, chest vibration, digital image correlation, motion artifact reduction, systolic time intervals, non-invasive blood pressure estimation

About these checks

Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.

The claim verdicts this report counts are experimental: their accuracy is not yet measured.

A language model drew the claims from the paper on 2026-10-10, under version 3 of the claim definition (model deepseek-v4.1-flash, prompt extract-v3).

A second model labelled each claim on 2026-10-10 (model jev-1.13-20260917, definition revision gold-guideline-v0.2).

Cited sources are checked with a language model’s help (gpt-6-sol).

Statistics and arithmetic are recomputed, not judged.

Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.

What counts as a claim ›How we measure confidence ›How we check this ›

Are you an author, or do you see a problem with this page? Request a review

How to cite

Cite CoChat for the checks on this page, and the paper for its own findings.

This page
CoChat. (2026). Citation check: A Chest‐Laminated Ultrathin and Stretchable E‐Tattoo for the Measurement of Electrocardiogram, Seismocardiogram, and Cardiac Time Intervals. https://app.cochat.ai/papers/10.1002/advs.201900290
The paper
Ha, T., Tran, J., Liu, S., Jang, H., Jeong, H., Mitbander, R., Huh, H., Qiu, Y., Duong, J., Wang, R. L., Wang, P., Tandon, A., Sirohi, J., & Lu, N. (2019). A Chest‐Laminated Ultrathin and Stretchable E‐Tattoo for the Measurement of Electrocardiogram, Seismocardiogram, and Cardiac Time Intervals. Advanced Science. https://doi.org/10.1002/advs.201900290

Data sources

Bibliographic data: Semantic Scholar, OpenAlex, Crossref.

Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).

Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.