Monitoring of the central blood pressure waveform via a conformal ultrasonic device
Open accessAbstract
Continuous monitoring of the central blood pressure waveform from deeply embedded vessels such as the carotid artery and jugular vein has clinical value for the prediction of all-cause cardiovascular mortality. However, existing non-invasive approaches, including photoplethysmography and tonometry, only enable access to the superficial peripheral vasculature. Although current ultrasonic technologies allow non-invasive deep tissue observation, unstable coupling with the tissue surface resulting from the bulkiness and rigidity of conventional ultrasound probes introduces usability constraints. Here, we describe the design and operation of an ultrasonic device that is conformal to the skin and capable of capturing blood pressure waveforms at deeply embedded arterial and venous sites. The wearable device is ultrathin (240 μm) and stretchable (with strains up to 60%), and enables the non-invasive, continuous and accurate monitoring of cardiovascular events from multiple body locations, which should facilitate its use in a variety of clinical environments. An ultrasonic and stretchable device conformal to the skin that captures blood pressure waveforms at deeply embedded arterial and venous sites enables the continuous monitoring of cardiovascular events.
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Integrity signals
Registry
Cites
Statistics
Claims the paper makes
The claims and statistics of this paper haven’t been checked yet.
Run the check — free, no account needed. It reads the open-access text and checks the sources it cites.
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
Other flexible health monitoring mechanisms like optoelectronic and ultrasound devices can be applied to heartbeat, pulse oximetry, blood pressure, and other on-body measurements.
“Other mechanisms of flexible health monitoring in rehabilitation, like optoelectronic[263,264] and ultrasound[265,266] devices, can also be applied to heartbeat, pulse oximetry, blood pressure, and other on-body measurements.”
From Flexible Electronics and Devices as Human–Machine Interfaces for Medical Robotics (2022) · cites this paper as [266]
The Abstract describes a wearable, skin-conformal ultrasonic device that continuously monitors blood-pressure waveforms. The Results (‘Dynamic hemodynamics monitoring’) also report heart-rate measurements. Our copy does not show this ultrasonic device measuring pulse oximetry, so it supports only part of the measurement list attributed to it. (LLM, conf=0.84)
Experimental — accuracy not yet measuredA novel spiked ultra-flexible neural interface (SUNI) collected sensory information from rat mechanoreceptors through a spike structure.
“Figure 4b displayed a novel spiked ultra-flexible neural interface (SUNI), which collected sensory information from rat mechanoreceptors through the spike structure (Wang, J. et al., 2018).”
From Artificial intelligence enhanced sensors - enabling technologies to next-generation healthcare and biomedical platform (2023) · cites this paper as Wang, J. et al., 2018
High-quality recordings from SUNI contributed to differentiating tactile and proprioceptive stimuli and provided for high spatial resolution classification of neural signals.
“High-quality recordings from SUNI contributed differentiate tactile and proprioceptive stimuli, and furtherly provided for high spatial resolution classification of neural signals.”
From Artificial intelligence enhanced sensors - enabling technologies to next-generation healthcare and biomedical platform (2023) · cites this paper as Wang, J. et al., 2018
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
The self-check needs the paper’s full text.
Cites
Fully integrated wearable sensor arrays for multiplexed in situ perspiration analysis
Gao, W., Emaminejad, S., Nyein, H. Y. Y., Challa, S., Chen, K., Peck, A., Fahad, H. M., Ota, H., Shiraki, H., Kiriya, D., et al. (2016) · Nature · doi:10.1038/nature16521 · Cited by 5,087 · see citing works
Kim, D.-H., Lu, N., Ma, R., Kim, Y.-S., Kim, R.-H., Wang, S., Wu, J., Won, S. M., Tao, H., Islam, A., et al. (2011) · Science · doi:10.1126/science.1206157 · Cited by 4,836 · see citing works
The CAFE Investigators, Williams, B., Lacy, P. S., Thom, S., Cruickshank, K. J., Stanton, A. V., Collier, D., Hughes, A. D., Thurston, H., & O’Rourke, M. F. (2006) · Circulation · doi:10.1161/circulationaha.105.595496 · Cited by 2,287 · see citing works
Schwartz, G., Tee, B. C.-K., Mei, J., Appleton, A. L., Kim, D. H., Wang, H., & Bao, Z. (2013) · Nature Communications · doi:10.1038/ncomms2832 · Cited by 2,042 · see citing works
Heart Rate, Life Expectancy and the Cardiovascular System: Therapeutic Considerations
Boudoulas, K. D., BORER, J. S., & Boudoulas, H. K. (2015) · Cardiology · doi:10.1159/000435947 · Cited by 1,952 · see citing works
Dağdeviren, C., Su, Y., Joe, P., Yona, R., Liu, Y., Kim, Y., Huang, Y., Damadoran, A. R., Xia, J., Martin, L. W., et al. (2014) · Nature Communications · doi:10.1038/ncomms5496 · Cited by 987 · see citing works
Toward ubiquitous blood pressure monitoring via pulse transit time: Theory and practice
Mukkamala, R., Hahn, J.-O., Inan, O. T., Mestha, L. K., Kim, C.-S., Toreyin, H., & Kyal, S. (2015) · IEEE Transactions on Biomedical Engineering · doi:10.1109/tbme.2015.2441951 · Cited by 935 · see citing works
Central Pulse Pressure and Mortality in End-Stage Renal Disease
Safar, M. E., Blacher, J., Pannier, B., Guérin, A. P., Marchais, S. J., Guyonvarc’h, P.-M., & London, G. M. (2002) · Hypertension · doi:10.1161/hy0202.098325 · Cited by 800 · see citing works
Kumar, A., Añel, R., Bunnell, E., Habet, K. J., Zanotti, S. L., Marshall, S., Neumann, A., Ali, A., Cheang, M., Kavinsky, C. J., et al. (2004) · Critical Care Medicine · doi:10.1097/01.ccm.0000114996.68110.c9 · Cited by 790 · see citing works
Howard, G., Sharrett, A. R., Heiss, G., Evans, G. W., Chambless, L. E., Riley, W. A., & Burke, G. L. (1993) · Stroke · doi:10.1161/01.str.24.9.1297 · Cited by 724 · see citing works
Central blood pressure: current evidence and clinical importance
McEniery, C. M., Cockcroft, J. R., Roman, M. J., Franklin, S. S., & Wilkinson, I. B. (2014) · European Heart Journal · doi:10.1093/eurheartj/eht565 · Cited by 679 · see citing works
Central Blood Pressure Measurements and Antihypertensive Therapy
Agabiti‐Rosei, E., Mancia, G., O’Rourke, M. F., Roman, M. J., Safar, M. E., Smulyan, H., Wang, J.-G., Wilkinson, I. B., Williams, B., & Vlachopoulos, C. V. (2007) · Hypertension · doi:10.1161/hypertensionaha.107.090068 · Cited by 618 · see citing works
Hemodynamic correlates of myocardial oxygen consumption during upright exercise
Kitamura, K., Jorgensen, C. R., Gobel, F. L., Taylor, H. L., & Wang, Y. (1972) · Journal of Applied Physiology · doi:10.1152/jappl.1972.32.4.516 · Cited by 575 · see citing works
Role of Pulse Pressure Amplification in Arterial Hypertension
Avolio, A. P., Bortel, L. M. V., Boutouyrie, P., Cockcroft, J. R., McEniery, C. M., Protogerou, A. D., Roman, M. J., Safar, M. E., Segers, P., & Smulyan, H. (2009) · Hypertension · doi:10.1161/hypertensionaha.109.134379 · Cited by 517 · see citing works
Williams, K. J., & Tabas, I. A. (2006) · Circulation · doi:10.1161/circulationaha.105.606962 · Cited by 486 · see citing works
Martín, A., Kim, J., Kurniawan, J. F., Sempionatto, J. R., Moreto, J. R., Tang, G., Campbell, A. S., Shin, A., Lee, M. Y., Liu, X., et al. (2017) · ACS Sensors · doi:10.1021/acssensors.7b00729 · Cited by 475 · see citing works
THE BLOOD SUPPLY OF VARIOUS SKIN AREAS AS ESTIMATED BY THE PHOTOELECTRIC PLETHYSMOGRAPH
Hertzman, A. B. (1938) · American Journal of Physiology-Legacy Content · doi:10.1152/ajplegacy.1938.124.2.328 · Cited by 455 · see citing works
Piezoelectric single crystal ultrasonic transducers for biomedical applications
Zhou, Q., Lam, K. H., Zheng, H., Qiu, W., & Shung, K. K. (2014) · Progress in Materials Science · doi:10.1016/j.pmatsci.2014.06.001 · Cited by 444 · see citing works
Self-citationPini, R., Cavallini, M. C., Palmieri, V., Marchionni, N., Bari, M. D., Devereux, R. B., Masotti, G., & Roman, M. J. (2008) · Journal of the American College of Cardiology · doi:10.1016/j.jacc.2008.03.031 · Cited by 434 · see citing works
Battery-free, stretchable optoelectronic systems for wireless optical characterization of the skin
Kim, J., Salvatore, G. A., Araki, H., Chiarelli, A. M., Xie, Z., Banks, A. R., Sheng, X., Liu, Y., Lee, J. W., Jang, K., et al. (2016) · Science Advances · doi:10.1126/sciadv.1600418 · Cited by 407 · see citing works
Self-citationArterial stiffness: a brief review
Shirwany, N. A., & Zou, M. (2010) · Acta Pharmacologica Sinica · doi:10.1038/aps.2010.123 · Cited by 387 · see citing works
Ultraflexible, large-area, physiological temperature sensors for multipoint measurements
Yokota, T., Inoue, Y., Terakawa, Y., Reeder, J., Kaltenbrunner, M., Ware, T., Yang, K., Mabuchi, K., Murakawa, T., Sekino, M., et al. (2015) · Proceedings of the National Academy of Sciences · doi:10.1073/pnas.1515650112 · Cited by 382 · see citing works
Stretchable ultrasonic transducer arrays for three-dimensional imaging on complex surfaces
Hu, H., Zhu, X., Wang, C., Zhang, L., Li, X., Lee, S., Huang, Z., Chen, R., Chen, Z., Wang, C., et al. (2018) · Science Advances · doi:10.1126/sciadv.aar3979 · Cited by 369 · see citing works
Self-citationAvolio, A. P., Butlin, M., & Walsh, A. (2009) · Physiological Measurement · doi:10.1088/0967-3334/31/1/r01 · Cited by 309 · see citing works
Materials and Designs for Wireless Epidermal Sensors of Hydration and Strain
Huang, X., Liu, Y., Cheng, H., Shin, W., Fan, J. A., Liu, Z., Lu, C., Kong, G., Chen, K., Patnaik, D., et al. (2014) · Advanced Functional Materials · doi:10.1002/adfm.201303886 · Cited by 302 · see citing works
Butman, S. M., Ewy, G. A., Standen, J. R., Kern, K. B., & Hahn, E. A. (1993) · Journal of the American College of Cardiology · doi:10.1016/0735-1097(93)90405-p · Cited by 290 · see citing works
Optical blood pressure estimation with photoplethysmography and FFT-based neural networks
Xing, X., & Sun, M. (2016) · Biomedical Optics Express · doi:10.1364/boe.7.003007 · Cited by 263 · see citing works
DeLoach, S. S., & Townsend, R. R. (2008) · Clinical Journal of the American Society of Nephrology · doi:10.2215/cjn.03340807 · Cited by 254 · see citing works
Roman, M. J., Okin, P. M., Kizer, J. R., Lee, E. T., Howard, B. V., & Devereux, R. B. (2010) · Journal of Hypertension · doi:10.1097/hjh.0b013e328333d228 · Cited by 251 · see citing works
Arterial tonometry: Review and analysis
Drzewiecki, G. M., Melbin, J., & Noordergraaf, A. (1983) · Journal of Biomechanics · doi:10.1016/0021-9290(83)90037-4 · Cited by 236 · see citing works
Maeda, Y., Sekine, M., & Tamura, T. (2011) · Journal of Medical Systems · doi:10.1007/s10916-010-9505-0 · Cited by 203 · see citing works
Ishibashi, Y., Duncker, D. J., Zhang, J. J., & Bache, R. J. (1998) · Circulation Research · doi:10.1161/01.res.82.3.346 · Cited by 197 · see citing works
Monitoring Arterial Blood Pressure: What You May Not Know
McGhee, B. H., & Bridges, E. J. (2002) · Critical Care Nurse · doi:10.4037/ccn2002.22.2.60 · Cited by 189 · see citing works
The diameter of the intact carotid artery in man and its change with pulse pressure
Arndt, J. O., Klauske, J., & Mersch, F. (1968) · Pflügers Archiv - European Journal of Physiology · doi:10.1007/bf00363770 · Cited by 172 · see citing works
Different Effects of Fosinopril and Atenolol on Wave Reflections in Hypertensive Patients
Chen, C.-H., Ting, C.-T., Lin, S., Hsu, T., Yin, F. C. P., SIU, C. O., Chou, P., Wang, S.-P., & Chang, M.-S. (1995) · Hypertension · doi:10.1161/01.hyp.25.5.1034 · Cited by 148 · see citing works
Effect of strain rate on the fracture behaviour of skin
Arumugam, V., Naresh, M. D., & Sanjeevi, R. (1994) · Journal of Biosciences · doi:10.1007/bf02716820 · Cited by 142 · see citing works
Langewouters, G. J., Settels, J. J., Roelandt, R., & Wesseling, K. H. (1998) · Journal of Medical Engineering & Technology · doi:10.3109/03091909809009997 · Cited by 123 · see citing works
Single crystal PMN-PT/Epoxy 1-3 composite for energy-harvesting application
Ren, K., Liu, Y., Geng, X., Hofmann, H. F., & Zhang, Q. M. (2006) · IEEE Transactions on Ultrasonics Ferroelectrics and Frequency Control · doi:10.1109/tuffc.2006.1610572 · Cited by 113 · see citing works
Sandberg, M., Zhang, Q., STYF, J. R., Gerdle, B. U. C., & Lindberg, L. (2005) · Acta Physiologica Scandinavica · doi:10.1111/j.1365-201x.2005.01412.x · Cited by 91 · see citing works
Ding, F., Fan, W.-X., Zhang, R.-Y., Zhang, Q., Li, Y., & Wang, J.-G. (2011) · American Journal of Hypertension · doi:10.1038/ajh.2011.145 · Cited by 83 · see citing works
Self-citationCorrelation of sonographic measurements of the internal jugular vein with central venous pressure
Donahue, S. P., Wood, J. P., Patel, B. M., et al. (2009) · The American Journal of Emergency Medicine · doi:10.1016/j.ajem.2008.06.005 · Cited by 69 · see citing works
High Frequency PMN-PT 1-3 Composite Transducer for Ultrasonic Imaging Application
Sun, P., Wang, G., Wu, D., et al. (2010) · Ferroelectrics · doi:10.1080/00150193.2010.485546 · Cited by 57 · see citing works
Bruyndonckx, L., Radtke, T., Eser, P., Vrints, C. J., Ramet, J., Wilhelm, M., & Conraads, V. M. (2013) · International Journal of Cardiology · doi:10.1016/j.ijcard.2013.07.236 · Cited by 53 · see citing works
Central Blood Pressure as an Index of Antihypertensive Control: Determinants and Potential Value
Trudeau, L. (2014) · Canadian Journal of Cardiology · doi:10.1016/j.cjca.2014.03.010 · Cited by 50 · see citing works
Accuracy of duplex ultrasound in evaluating carotid artery anatomy before endarterectomy
Wain, R. A., Lyon, R. T., Veith, F. J., et al. (1998) · Journal of Vascular Surgery · doi:10.1016/s0741-5214(98)70354-4 · Cited by 45 · see citing works
High Frequency PMN-PT 1-3 Composite Transducer for Ultrasonic Imaging Application.
Sun, P., Wang, G., Wu, D., et al. (2010) · PubMed · doi:10.1080/00150193.2010.48554 · Cited by 31 · see citing works
Camacho, F., Avolio, A. P., & Lovell, N. H. (2004) · Physiological Measurement · doi:10.1088/0967-3334/25/4/008 · Cited by 30 · see citing works
Continuous and less invasive central hemodynamic monitoring by blood pressure waveform analysis
Mukkamala, R., & Xu, D. Z. (2010) · American Journal of Physiology-Heart and Circulatory Physiology · doi:10.1152/ajpheart.00303.2010 · Cited by 30 · see citing works
Moffitt, E. A., Sethna, D. H., Gray, R. J., et al. (1984) · Canadian Journal of Anesthesia/Journal canadien d anesthésie · doi:10.1007/bf03011476 · Cited by 29 · see citing works
Bioelectronic devices: Long-lived recordings
Rotenberg, M. Y., & Tian, B. (2017) · Nature Biomedical Engineering · doi:10.1038/s41551-017-0048 · Cited by 27 · see citing works
Flagged first, then by citations
Research works printed but not linked: 1
- [43]Williams B Differential impact of blood pressure–lowering drugs on central aortic pressure and clinical outcomes: principal results of the Conduit Artery Function Evaluation (CAFE) study. Circulation 113, 1213– 1225 (2006). 16476843
Cited by
Sorted by citations
End-to-end design of wearable sensors
Ates, H. C., Nguyen, P. Q., Gonzalez-Macia, L., Morales-Narváez, E., Güder, F., Collins, J. J., & Dincer, C. (2022) · Nature Reviews Materials · doi:10.1038/s41578-022-00460-x · Cited by 1,306 · see citing works
A laser-engraved wearable sensor for sensitive detection of uric acid and tyrosine in sweat
Yang, Y., Song, Y., Bo, X., Min, J., Pak, O. S., Zhu, L., Wang, M., Tu, J., Kogan, A., Zhang, H., et al. (2020) · Nature Biotechnology · doi:10.1038/s41587-019-0321-x · Cited by 1,270 · see citing works
An epidermal patch for the simultaneous monitoring of haemodynamic and metabolic biomarkers
Sempionatto, J. R., Lin, M., Yin, L., De la paz, E., Pei, K., Sonsa-ard, T., de Loyola Silva, A. N., Khorshed, A. A., Zhang, F., Tostado, N., et al. (2021) · Nature Biomedical Engineering · doi:10.1038/s41551-021-00685-1 · Cited by 658 · see citing works
Self-citationWireless battery-free wearable sweat sensor powered by human motion
Song, Y., Min, J., Yu, Y., Wang, H., Yang, Y., Zhang, H., & Gao, W. (2020) · Science Advances · doi:10.1126/sciadv.aay9842 · Cited by 652 · see citing works
A wearable cardiac ultrasound imager
Hu, H., Huang, H., Li, M., Gao, X., Yin, L., Qi, R., Wu, R. S., Chen, X., Ma, Y., Shi, K., et al. (2023) · Nature · doi:10.1038/s41586-022-05498-z · Cited by 644 · see citing works
Self-citationFlexible Hybrid Electronics for Digital Healthcare
Ma, Y., Zhang, Y., Cai, S., Han, Z., Liu, X., Wang, F., Cao, Y., Wang, Z., Li, H., Chen, Y., et al. (2020) · Advanced Materials · doi:10.1002/adma.201902062 · Cited by 620 · see citing works
Self-citationFlexible Electronics and Devices as Human–Machine Interfaces for Medical Robotics
Heng, W., Solomon, S., & Gao, W. (2022) · Advanced Materials · doi:10.1002/adma.202107902 · Cited by 609 · see citing works
Multifunctional Skin‐Inspired Flexible Sensor Systems for Wearable Electronics
Xu, K., Lu, Y., & Takei, K. (2019) · Advanced Materials Technologies · doi:10.1002/admt.201800628 · Cited by 602 · see citing works
Hong, Y. J., Jeong, H., Cho, K. W., Lu, N., & Kim, D. (2019) · Advanced Functional Materials · doi:10.1002/adfm.201808247 · Cited by 559 · see citing works
Machine-knitted washable sensor array textile for precise epidermal physiological signal monitoring
Fan, W., He, Q., Meng, K., Tan, X., Zhou, Z., Zhang, G., Yang, J., & Wang, Z. L. (2020) · Science Advances · doi:10.1126/sciadv.aay2840 · Cited by 516 · see citing works
Wen, F., Zhang, Z., He, T., & Lee, C. (2021) · Nature Communications · doi:10.1038/s41467-021-25637-w · Cited by 494 · see citing works
Self-citationFlexible Wearable Sensors for Cardiovascular Health Monitoring
Chen, S., Qi, J., Fan, S., Qiao, Z., Yeo, J. C., & Lim, C. T. (2021) · Advanced Healthcare Materials · doi:10.1002/adhm.202100116 · Cited by 453 · see citing works
Wearable Sensing and Telehealth Technology with Potential Applications in the Coronavirus Pandemic
Ding, X., Clifton, D., Ji, N., Lovell, N. H., Bonato, P., Chen, W., Yu, X., Xue, Z., Xiang, T., Long, X., et al. (2021) · IEEE Reviews in Biomedical Engineering · doi:10.1109/rbme.2020.2992838 · Cited by 281 · see citing works
Wang, C., He, T., Zhou, H., Zhang, Z., & Lee, C. (2023) · Bioelectronic Medicine · doi:10.1186/s42234-023-00118-1 · Cited by 295 · see citing works
Self-citationContinuous monitoring of deep-tissue haemodynamics with stretchable ultrasonic phased arrays
Wang, C., Qi, B., Lin, M., Zhang, Z., Makihata, M., Liu, B., Zhou, S., Huang, Y.-H., Hu, H., Gu, Y., et al. (2021) · Nature Biomedical Engineering · doi:10.1038/s41551-021-00763-4 · Cited by 274 · see citing works
Self-citationKwon, Y.-T., Kim, Y.-S., Kwon, S., Mahmood, M., Lim, H.-R., Park, S.-W., Kang, S.-O., Choi, J. J., Herbert, R., Jang, Y. C., et al. (2020) · Nature Communications · doi:10.1038/s41467-020-17288-0 · Cited by 233 · see citing works
Ha, T., Tran, J., Liu, S., Jang, H., Jeong, H., Mitbander, R., Huh, H., Qiu, Y., Duong, J., Wang, R. L., et al. (2019) · Advanced Science · doi:10.1002/advs.201900290 · Cited by 220 · see citing works
Shi, Q., & Lee, C. (2019) · Advanced Science · doi:10.1002/advs.201900617 · Cited by 174 · see citing works
Cuffless Blood Pressure Measurement
Mukkamala, R., Stergiou, G. S., & Avolio, A. P. (2022) · Annual Review of Biomedical Engineering · doi:10.1146/annurev-bioeng-110220-014644 · Cited by 180 · see citing works
Sharman, J. E., O’Brien, E., Alpert, B., Schutte, A. E., Delles, C., Hecht Olsen, M., Asmar, R., Atkins, N., Barbosa, E., Calhoun, D., et al. (2020) · Journal of Hypertension · doi:10.1097/hjh.0000000000002246 · Cited by 139 · see citing works
Ultrasound-induced wireless energy harvesting: from materials strategies to functional applications
Jiang, L., Yang, Y., Chen, Y., & Zhou, Q. (2020) · Nano Energy · doi:10.1016/j.nanoen.2020.105131 · Cited by 131 · see citing works
Self-citationPPG2ABP: Translating photoplethysmogram (PPG) signals to arterial blood pressure (ABP) waveforms
Ibtehaz, N., Mahmud, S., Chowdhury, M. E. H., Khandakar, A., Salman Khan, M., Ayari, M. A., Tahir, A. M., & Rahman, M. S. (2022) · Bioengineering · doi:10.3390/bioengineering9110692 · Cited by 133 · see citing works
Designing wearable microgrids: towards autonomous sustainable on-body energy management
Yin, L., Kim, K. N., Trifonov, A., Podhajny, T., & Wang, J. (2022) · Energy & Environmental Science · doi:10.1039/d1ee03113a · Cited by 111 · see citing works
Modular and Reconfigurable Wireless E‐Tattoos for Personalized Sensing
Jeong, H., Wang, L., Ha, T., Mitbander, R., Yang, X., Dai, Z., Qiao, S., Shen, L., Sun, N., & Lu, N. (2019) · Advanced Materials Technologies · doi:10.1002/admt.201900117 · Cited by 111 · see citing works
A wearable echomyography system based on a single transducer
Gao, X., Chen, X., Lin, M., Yue, W., Hu, H., Qin, S., Zhang, F., Lou, Z., Yin, L., Huang, H., et al. (2024) · Nature Electronics · doi:10.1038/s41928-024-01271-4 · Cited by 95 · see citing works
Self-citationCheng, S., Gu, Z., Zhou, L., Hao, M., An, H., Song, K., Wu, X., Zhang, K., Zhao, Z., Dong, Y., et al. (2021) · Frontiers in Bioengineering and Biotechnology · doi:10.3389/fbioe.2021.765987 · Cited by 64 · see citing works
Application of telemedicine in COVID-19: a bibliometric analysis
Lan, X., Yu, H., & Cui, L. (2022) · Frontiers in Public Health · doi:10.3389/fpubh.2022.908756 · Cited by 28 · see citing works
Ultrasound powered implants: design, performance considerations and simulation results
Rosa, B. M. G., & Yang, G.-Z. (2020) · Scientific Reports · doi:10.1038/s41598-020-63097-2 · Cited by 18 · see citing works
Measurements of arterial pressure and flow in vivo
Sharman, J. E., Segers, P., & Chirinos, J. A. (2022) · Textbook of Arterial Stiffness and Pulsatile Hemodynamics in Health and Disease · doi:10.1016/b978-0-323-91391-1.00002-9 · Cited by 3 · see citing works
- Authors
- Wang, C., Li, X., Hu, H., Zhang, L., Huang, Z., Lin, M., Zhang, Z., Yin, Z., Huang, B., Gong, H., Bhaskaran, S., Gu, Y., Makihata, M., Guo, Y., Lei, Y., Chen, Y., Wang, C., Li, Y., Zhang, T., Chen, Z., Pisano, A. P., Zhang, L., Zhou, Q., & Xu, S.
- Affiliations
- University of California San Diego (US)
- Tianjin University (CN)
- University of Southern California (US)
- Corresponding author
- Xu, Sheng
- ORCID
- Venue
- Nature Biomedical Engineering
- Published
- 2018-09-11
- Volume
- 2
- Issue
- 9
- Pages
- 687-695
- ISSN
- 2157-846X (ISSN-L)
- PMC
- PMC6428206
- OpenAlex
- W2891821958
- Semantic Scholar
- 54036ec5fa00d18b5c4ccbf2436f52e6b3707b01
- Cited by
- 895
- References
- 50
- Citations by year
- 2026: 90 · 2025: 157 · 2024: 124 · 2023: 134 · 2022: 118 · 2021: 120 · 2020: 97 · 2019: 52 · 2018: 3
- FWCI
- 27.16
- Type
- journal article
- Language
- en
- Open access
- green
- europepmc.org
- Licence
- Locations
- Nature Biomedical Engineering (publishedVersion)
- PubMed (publishedVersion)
- Europe PMC (PubMed Central) (acceptedVersion)
- PubMed Central (acceptedVersion)
- eScholarship (California Digital Library) (submittedVersion)
- Funders
- National Institutes of Health (1r21eb025521-01)
- National Institute of Biomedical Imaging and Bioengineering (R21 EB025521)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Physical Sciences › Engineering › Biomedical Engineering › Advanced Sensor and Energy Harvesting Materials
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › Cardiovascular Health and Disease Prevention
- Keywords
- central blood pressure, blood pressure waveform, wearable ultrasound device, stretchable sensors, carotid artery, jugular vein, deep tissue imaging, non-invasive blood pressure monitoring, cardiovascular mortality prediction
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review