A Chest‐Laminated Ultrathin and Stretchable E‐Tattoo for the Measurement of Electrocardiogram, Seismocardiogram, and Cardiac Time Intervals
Open accessThe DOI matches its Crossref record; 63 of the 63 references it cites were found in a registry or an index; 5 of the 50 sources its claims cite were checked for support.
Abstract
Seismocardiography (SCG) is a measure of chest vibration associated with heartbeats. While skin soft electronic tattoos (e‐tattoos) have been widely reported for electrocardiogram (ECG) sensing, wearable SCG sensors are still based on either rigid accelerometers or non‐stretchable piezoelectric membranes. This work reports an ultrathin and stretchable SCG sensing e‐tattoo based on the filamentary serpentine mesh of 28‐µm‐thick piezoelectric polymer, polyvinylidene fluoride (PVDF). 3D digital image correlation (DIC) is used to map chest vibration to identify the best location to mount the e‐tattoo and to investigate the effects of substrate stiffness. As piezoelectric sensors easily suffer from motion artifacts, motion artifacts are effectively reduced by performing subtraction between a pair of identical SCG tattoos placed adjacent to each other. Integrating the soft SCG sensor with a pair of soft gold electrodes on a single e‐tattoo platform forms a soft electro‐mechano‐acoustic cardiovascular (EMAC) sensing tattoo, which can perform synchronous ECG and SCG measurements and extract various cardiac time intervals including systolic time interval (STI). Using the EMAC tattoo, strong correlations between STI and the systolic/diastolic blood pressures, are found, which may provide a simple way to estimate blood pressure continuously and noninvasively using one chest‐mounted e‐tattoo.
Overview
CoChat Summary
TL;DR
This paper reports an ultrathin, stretchable chest-laminated e-tattoo using filamentary serpentine 28-µm PVDF for seismocardiogram (SCG) sensing and integrated Au electrodes for synchronous ECG/SCG (EMAC) sensing. It demonstrates high-fidelity SCG/ECG, motion-artifact reduction via dual PVDF subtraction, and a strong negative correlation between the systolic time interval (RAC) and blood pressure across four subjects.
Key findings
- The FS PVDF e-tattoo is 28 µm thick, has 112.9% stretchability and 0.4 mV per µε sensitivity, and an effective modulus of 8.5 MPa; the integrated EMAC tattoo measures 63.5 mm × 38.1 mm × 0.122 mm, weighs 150 mg, and has about 100% stretchability.
- 3D DIC mapped chest vibration: S1 was ≈5 µm around the left nipple area and S2 was ≈3 µm at a higher location; the middle e-tattoo was best for measuring both S1 and S2. FEM estimated substrate-dependent outputs of 2.74 mV (no substrate), 2.20 mV (Tegaderm), and 0.86 mV (PET).
- Dual PVDF subtraction reduced motion artifacts, making motion-period SNRs comparable to rest-period SNRs; a single PET-substrate sensor had negative SNR under motion.
- In four subjects, RAC negatively correlated with blood pressure: average Pearson R was −0.772 for SBP–RAC, −0.858 for DBP–RAC, and −0.538 for PBP–RAC; individual calibration was necessary.
Methods
- The device was fabricated by a cut-and-paste method to pattern FS PVDF and Au/PET electrodes on Tegaderm; characterization used DMA, electrical measurements, and ABAQUS FEM. 3D DIC used two high-speed cameras at 500 Hz for 16 s to map chest deformation.
- Human experiments were IRB-approved. Synchronous ECG and SCG were measured by the EMAC tattoo and compared with an accelerometer; blood pressure was measured by SOMNOtouch NIBP during Valsalva maneuvers in four subjects. Signals were filtered with 2–40 Hz (ECG) and 12–40 Hz (SCG) Butterworth filters, and RAC intervals were extracted from R and AC peaks.
- Analysis included Pearson correlations between BP and RAC, plus FEM modeling of substrate effects and electromechanical output.
Limitations
- Stated: fitting parameters and correlation coefficients varied by individual and posture; no master fitting equations could be derived, so per-individual calibration is necessary. Valsalva data were omitted because both SOMNO and SCG signals suffered excessive noise from body shivering, and mismatches were likely from motion or slight posture change.
- Stated: wireless NFC or Bluetooth integration remains future work for practical ambulatory applications.
- Evident from the text: BP validation involved only four subjects, and the main DIC mapping was performed on an adult male; no long-term ambulatory validation is reported.
Model-written summary. Check it against the paper before relying on it.
Integrity signals
Registry
Cites
Statistics
Figures





Verification
Claims the paper makes
4 Valid · 4 Not valid · 31 Not checked · 114 Own results and interpretations
The cited sources that could be checked couldn’t confirm or dispute this claim.
In the paper
“In contrast, there has been little progress in the development of soft SCG sensors.”
§ A Chest‐Laminated Ultrathin and Stretchable E‐Tattoo for the Measurement of Electrocardiogram, Seismocardiogram, and Cardiac Time Intervals > Introduction
What the sources say
- Could not verifycites [9] D.‐H. Kim, N. Lu, R. Ma, Y.‐S. Kim, R.‐H. Kim, S. Wang, J. Wu, S. M. Won, H. Tao, A. Islam, K. J. Yu, T.‐i. Kim, R. Chowdhury, M. Ying, L. Xu, M. Li, H.‐J. Chung, H. Keum, M. McCormick, P. Liu, Y.‐W. Zhang, F. G. Omenetto, Y. Huang, T. Coleman, J. A. Rogers, Science 2011, 333, 838. 21836009 sourceExperimental — accuracy not yet measured
The Abstract says that many skin-mounted physiological devices still use conventional designs, but it does not assess progress in soft SCG sensors. Searches for SCG and seismocardiography found passages about ECG and other sensors, not the claimed lack of progress in SCG sensors. (LLM, conf=0.75)
- Not checked — no full text of the sourcecites [10] S. X. Yang, Y. C. Chen, L. Nicolini, P. Pasupathy, J. Sacks, B. Su, R. Yang, D. Sanchez, Y. F. Chang, P. L. Wang, D. Schnyer, D. Neikirk, N. S. Lu, Adv. Mater. 2015, 27, 6423. 26398335 source
- Not checked — no full text of the sourcecites [11] S. Kabiri Ameri, R. Ho, H. Jang, L. Tao, Y. Wang, L. Wang, D. M. Schnyer, D. Akinwande, N. Lu, ACS Nano 2017, 11, 7634. 28719739 source
plain- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Could not verify
The cited sources support only part of this claim — read each source’s verdict below.
In the paper
“Physical strapping or taping is also required to reduce the acoustic mismatch caused by the air gap between the rigid accelerometer and the skin.7”
§ A Chest‐Laminated Ultrathin and Stretchable E‐Tattoo for the Measurement of Electrocardiogram, Seismocardiogram, and Cardiac Time Intervals > Introduction
What the sources say
- Partly supportedcites [7] Y. Liu, J. J. Norton, R. Qazi, Z. Zou, K. R. Ammann, H. Liu, L. Yan, P. L. Tran, K.‐I. Jang, J. W. Lee, Sci. Adv. 2016, 2, e1601185. 28138529 PMC5262452 sourceExperimental — accuracy not yet measured
Materials and Methods (Device characterization; page not provided) describes attaching a sensor in a rigid acrylic box to the throat with adhesive and medical tape. Results (Device characterization; page not provided) says mechanical impedance matching with skin matters. The paper does not say that strapping or taping is generally required to reduce an acoustic mismatch caused by an air gap. (LLM, conf=0.80)
plainstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Partly supported
The cited sources support only part of this claim — read each source’s verdict below.
In the paper
“Several wearable sensors and energy harvesters have employed electrospun PVDF fibers, but electrodes could only be placed at the ends of the fiber bundles instead of along the full length of the fiber,22 which compromises the electrical output and, subsequently, the sensitivity.”
§ A Chest‐Laminated Ultrathin and Stretchable E‐Tattoo for the Measurement of Electrocardiogram, Seismocardiogram, and Cardiac Time Intervals > Introduction
What the sources say
- Partly supportedcites [22] a) L. Persano, C. Dagdeviren, Y. W. Su, Y. H. Zhang, S. Girardo, D. Pisignano, Y. G. Huang, J. A. Rogers, Nat. Commun. 2013, 4, 1633; 23535654 b) Y. Q. Duan, Y. A. Huang, Z. P. Yin, N. B. Bu, W. T. Dong, Nanoscale 2014, 6, 3289; 24509570 c) C. E. Chang, V. H. Tran, J. B. Wang, Y. K. Fuh, L. W. Lin, Nano Lett. 2010, 10, 726. 20099876 sourceExperimental — accuracy not yet measured
The Results (page unavailable) describe electrospun P(VDF-TrFe) sensors with electrical contacts at the ends of fibre arrays. The paper does not say electrodes could only be placed there or that end contacts compromise output and sensitivity; it reports a large response to minute pressures from those devices. (LLM, conf=0.86)
hedgedstates a cause- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Partly supported
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“In Figure 5d, we find that the R peak of the ECG and the MC peak of the SCG are almost aligned, but the R peak is much easier to extract as it is the highest peak in ECG whereas the MC peak is neither aligned with S1 nor S2.”
§ A Chest‐Laminated Ultrathin and Stretchable E‐Tattoo for the Measurement of Electrocardiogram, Seismocardiogram, and Cardiac Time Intervals > Results > EMAC Sensing E‐Tattoo
plainstates an associationSingle case- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: unclear — It is unclear whether the paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
A result the paper reports from its own data — it cites no other source for it, so there is none to check. The quote shows it is in the paper.
In the paper
“Here the result of the dual sensor noise cancellation with PET substrate was provided instead of with Tegaderm substrate to prove the dual sensor system can restore the SCG signal even for the severe motion noise due to the stiff substrate.”
§ A Chest‐Laminated Ultrathin and Stretchable E‐Tattoo for the Measurement of Electrocardiogram, Seismocardiogram, and Cardiac Time Intervals > Results > Dual‐PVDF‐Sensor for Motion Artifact Cancellation
plainstates a causeSingle case- Asserted: passes — The authors put it forward as true.
- Checkable: passes — It can be checked against evidence.
- Load-bearing: passes — The paper’s conclusions depend on it.
- Based: passes — The paper gives a reason why it holds.
- Rests on 3 assumptions not shown here.
90 more claims
Create a free account to see all 153 claims and to verify the 7 references still open.
- It may state two separate claims as one. 40
- Probably not a substantial claim. 20
Claims other papers make about this one
Each is checked against this paper’s own text: does it say what the citing paper says it does?
The cited source says what the claim attributes to it.
In the citing paper
“There have been multiple previous works that have delved into using dual-mode ECG and SCG sensing for cardiac time intervals measurements. [26-29]”
From A Chest‐Conformable, Wireless Electro‐Mechanical E‐Tattoo for Measuring Multiple Cardiac Time Intervals (2023) · cites this paper as [29]
In the citing paper
“Analogue and digital signal processing methods are often applied to the captured waveforms to remove motion artefacts and improve the SNR; multisensor configurations can also be used for active noise-cancelling and localization by beamforming118,119.”
From End-to-end design of wearable sensors (2022) · cites this paper as [119]
Self-check
Percentages, sums and reported differences re-computed from the paper’s own numbers.
Each contradiction between two passages listed here was confirmed by an independent check.
Rated minor: it would not touch a result — a value no result rests on, a p-value that stays on the same side of the significance level, or a conflict that is clerical, resolved elsewhere in the paper, or outside its own findings.
In the paper
§ A Chest‐Laminated Ultrathin and Stretchable E‐Tattoo for the Measurement of Electrocardiogram, Seismocardiogram, and Cardiac Time Intervals > Results > EMAC Sensing E‐Tattoo
“f) Continuous SBP and DBP estimated by SOMNOtough (black) and EMAC tattoo (red) during Valsalva maneuver.”
§ A Chest‐Laminated Ultrathin and Stretchable E‐Tattoo for the Measurement of Electrocardiogram, Seismocardiogram, and Cardiac Time Intervals > Results > Correlation between Hemodynamic Parameters and BP
“Data during the Valsalva maneuver was omitted as both the SOMNO and SCG signals suffered from excessive noise induced by body shivering, as evident in Figure S11 (Supporting Information).”
Citations
Cites
Global status report on noncommunicable diseases 2010.
Alwan, A. (2011) · Cited by 8,494
Kim, D.-H., Lu, N., Ma, R., Kim, Y.-S., Kim, R.-H., Wang, S., Wu, J., Won, S. M., Tao, H., Islam, A., et al. (2011) · Science · doi:10.1126/science.1206157 · Cited by 4,836
Self-citationAge-predicted maximal heart rate revisited
Tanaka, H., Monahan, K. D., & Seals, D. R. (2001) · Journal of the American College of Cardiology · doi:10.1016/s0735-1097(00)01054-8 · Cited by 3,775
Direct-Write Piezoelectric Polymeric Nanogenerator with High Energy Conversion Efficiency
Chang, C., Tran, V. H., Wang, J., Fuh, Y.-K., & Lin, L. (2010) · Nano Letters · doi:10.1021/nl9040719 · Cited by 1,379
Persano, L., Dagdeviren, C., Su, Y., Zhang, Y., Girardo, S., Pisignano, D., Huang, Y., & Rogers, J. A. (2013) · Nature Communications · doi:10.1038/ncomms2639 · Cited by 1,271
Monitoring of the central blood pressure waveform via a conformal ultrasonic device
Wang, C., Li, X., Hu, H., Zhang, L., Huang, Z., Lin, M., Zhang, Z., Yin, Z., Huang, B., Gong, H., et al. (2018) · Nature Biomedical Engineering · doi:10.1038/s41551-018-0287-x · Cited by 895
Fundamental Understanding of Piezoelectric Strain Sensors
Sirohi, J., & Chopra, I. (2000) · Journal of Intelligent Material Systems and Structures · doi:10.1106/8bfb-gc8p-xq47-ycq0 · Cited by 777
Self-citationBallistocardiography and Seismocardiography: A Review of Recent Advances
Inan, O. T., Migeotte, P.-F., Park, K.-S., Etemadi, M., Tavakolian, K., Casanella, R., Zanetti, J., Tank, J., Funtova, I., Prisk, G. K., et al. (2015) · IEEE Journal of Biomedical and Health Informatics · doi:10.1109/jbhi.2014.2361732 · Cited by 775
Flexible and Stretchable Electronics for Biointegrated Devices
Kim, D.-H., Ghaffari, R., Lu, N., & Rogers, J. A. (2012) · Annual Review of Biomedical Engineering · doi:10.1146/annurev-bioeng-071811-150018 · Cited by 749
Self-citationGraphene Electronic Tattoo Sensors
Kabiri Ameri, S., Ho, R., Jang, H., Tao, L., Wang, Y., Wang, L., Schnyer, D. M., Akinwande, D., & Lu, N. (2017) · ACS Nano · doi:10.1021/acsnano.7b02182 · Cited by 669
Self-citation
Research works printed but not linked: 15
- Castiglioni P., 2007, Proc. of the Annual Int. IEEE EMBS, 3954
- Hendriks F. M., 2005, Technische Universiteit Eindhoven, 106
- Specialties M., 2013, Measurement, 3
- Lionel W. Y. L., 2016, IEEE Eng. Med. Biol., 38, 4816
- Ameri S. K., 2016, Int. Electron Devices Meeting
- Paukkunen M., 2013, J. Bioeng. Biomed. Sci., 3, 1
- Singh A., 2013, J. Dental Mater. Tech., 4, 682
- Obropta E. W., 2016, Aerospace Conf. Proc.
- Obropta E. W., 2015, IEEE Aerospace Conf.
- 10.1063/1.2817234 · DOI · Paper page
- Pandia K., 2010, 2010 IEEE Int. Conf. on Acoustics Speech and Signal Processing (ICASSP)
- Di Rienzo M., 2012, Annual Int. Conf. of the IEEE on Engineering in Medicine and Biology Society (EMBC), 5050
- Korzeniowska‐Kubacka I., 2006, Folia Cardiol., 13, 319
- Imtiaz M. S., 2013, IEEE Canadian Conf. on Electrical and Computer Engineering (CCECE), 501
- Aubert X. L., 2007, Proc. of the Annual Int. IEEE EMBS, 1707
Cited by
Items 201–218 of 218
Development and Characterization of Compliant Bioelectronic Devices for Gastrointestinal Stimulation
Chitrakar, C. (2023) · University of North Texas Digital Library (University of North Texas) · doi:10.12794/metadc2257713
Strain-Resilient Thin Film Conductor for Multifunctional Electronic SkinPreprint
Ya, C., Qi, X., Wang, L., Li, C., & Wang, Y. (2024) · SSRN Electronic Journal · doi:10.2139/ssrn.4866463
Díaz, Y., Salas, P., González, C. C., Mejía-Muñoz, J.-M., & González-Landaeta, R. (2025) · World Congress on Medical Physics and Biomedical Engineering, September 7 - 12, 2009, Munich, Germany · doi:10.1007/978-3-031-82126-4_1
Wang, J. (2024) · Highlights in Science Engineering and Technology · doi:10.54097/0sn5r384
Uribe, J., Uy, A., Alvarez, K., et al. (2025) · IEEE Dallas Circuits and Systems Conference (DCAS) · doi:10.1109/dcas65331.2025.11045473
“In contrast to the explored small dimension, a 2019 study used a 28 µm thick sensor made from metalized PVDF sheets patterned into a filamentary serpentine network (PVDF tattoo) [4].”
Zhang, J., Shen, S. Q., Zhang, N., Wang, P., Liu, S., Han, Y., Sui, M., Pu, C., Lin, R., Xu, F., et al. (2025) · Advanced Functional Materials · doi:10.1002/adfm.202511531 · Cited by 1
Self-citationA review of MEMS microphone capabilities
Sharma, R. K., & Prasad, M. (2025) · Journal of Computational Electronics · doi:10.1007/s10825-025-02468-2
Fu, K., Zhao, J., Deng, J., Zhao, C., Wang, X., & Qiu, L. (2026) · IEEE Sensors Journal · doi:10.1109/jsen.2026.3652555
“…wearable sensors often suffer from limitations that hinder their practical usability, including complex device architectures, reliance on external fixatives due to poor self-adhesion, and unstable skin–device interfaces that introduce motion artifacts [14], [15], [16], [17], [18], [19].”
Fabrication of P(VDF-TrFE) films by Screen Printing and Application to Flexible Ultrasonic Sensor
Omori, M., Tanaka, T., & Uno, M. O. (2026) · IEEJ Transactions on Sensors and Micromachines · doi:10.1541/ieejsmas.146.52
Flexible and stretchable organic materials and devices for application in emerging optoelectronics
Dauzon, E. (2020) · theses.fr (ABES) · doi:10.70675/07a65d04z7ef3z4cdaz861bzd8744fff7395
Kim, J. H., Park, G., Kim, D., Jin, H., Choi, J. R., Kim, M., Lee, S., Lee, M., Son, Y., Jon, S., et al. (2026) · npj Flexible Electronics · doi:10.1038/s41528-026-00567-z
Sahoo, S., & Jain, P. K. (2026) · Physical and Engineering Sciences in Medicine · doi:10.1007/s13246-026-01741-3
“Ha et al. [28] developed a stretchable SCG sensing e-tattoo using a piezoelectric polymer PVDF, employing adjacent SCG signal subtraction to reduce motion artefacts.”
Alzahrani, M., Kabbara, I., Peng, S., Phan, H., & Kumeria, T. (2026) · Small Methods · doi:10.1002/smtd.70728
“Among the available techniques, ECG is widely used because it provides direct information on the electrical activity of the heart and supports the diagnosis, management, and prevention of CVDs [ 54, 55 ].”
Sun, X., Zhang, C., Sun, G., Wang, P., Meng, C., & Li, Z. J. (2026) · Advanced Functional Materials · doi:10.1002/adfm.76294
Self-citationTian, G., Deng, W., Zhou, L., et al. (2026) · Advanced Materials · doi:10.1002/adma.74367
Twiddy, J. S., Wilkins, M. D., Ismail, I., Shwetar, Y. J., Maddocks, G. M., Peterson, K., Lobaton, E. J., & Daniele, M. A.-A. (2026) · Biomedical Materials & Devices · doi:10.1007/s44174-026-00737-7
Pectin-Based Flexible and Wearable Bioelectrodes for EMG Signal Recording
Sayyad, P. W., Alex, M., Al–Othman, A., Al-Nashash, H. A., & Al-Sayah, M. H. (2026) · Macromol—A Journal of Macromolecular Research · doi:10.3390/macromol6030064
Resonant‐Coupled 3D Corrugated Triboelectric Sensor for Intelligent Health Monitoring
Tang, L., Yan, S., Hui, X., Liu, Y., Ren, Y., Wang, Z., Wang, Z. L., Dong, W., Wu, F., & Guo, H. (2026) · Advanced Functional Materials · doi:10.1002/adfm.78625
Metadata
- Authors
- Ha, T., Tran, J., Liu, S., Jang, H., Jeong, H., Mitbander, R., Huh, H., Qiu, Y., Duong, J., Wang, R. L., Wang, P., Tandon, A., Sirohi, J., & Lu, N.
- Affiliations
- The University of Texas at Austin (US)
- Texas Materials Institute (US)
- Children's Medical Center (US)
- Corresponding author
- Lu, Nanshu
- ORCID
- Venue
- Advanced Science
- Publisher
- Wiley
- Published
- 2019-05-21
- Volume
- 6
- Issue
- 14
- Pages
- 1900290-1900290
- ISSN
- 2198-3844 (ISSN-L)
- PMC
- PMC6662084
- OpenAlex
- W2945869729
- Cited by
- 220
- References
- 63
- Citations by year
- 2026: 15 · 2025: 28 · 2024: 33 · 2023: 34 · 2022: 44 · 2021: 34 · 2020: 26 · 2019: 6
- FWCI
- 9.68
- Type
- journal article
- Language
- en
- Open access
- gold
- doi.org
- Locations
- Advanced Science (publishedVersion, cc-by)
- Europe PMC (PubMed Central) (publishedVersion, cc-by)
- PubMed Central (publishedVersion, cc-by)
- PubMed (publishedVersion)
- DOAJ (DOAJ: Directory of Open Access Journals) (submittedVersion, cc-by-sa)
- eScholarship (California Digital Library) (submittedVersion, cc-by)
- Funders
- Office of Naval Research (N00014‐16‐1‐2044)
- Air Force Office of Scientific Research (FA9550‐15‐1‐0112)
- National Science Foundation (CNS‐1738293, UL1TR001105)
- Topics
- Physical Sciences › Engineering › Biomedical Engineering › Advanced Sensor and Energy Harvesting Materials
- Physical Sciences › Engineering › Biomedical Engineering › Non-Invasive Vital Sign Monitoring
- Health Sciences › Medicine › Cardiology and Cardiovascular Medicine › ECG Monitoring and Analysis
- Keywords
- seismocardiography, stretchable sensors, polyvinylidene fluoride, piezoelectric polymers, electrocardiogram, chest vibration, digital image correlation, motion artifact reduction, systolic time intervals, non-invasive blood pressure estimation
About these checks
Registry, retraction, Cites and statistics results are computed, not judged, with one exception: a printed reference that no index lists is read once by a model before it counts as a citation error. Claims are extracted by a model and kept only when their quote appears verbatim in the text; each cited source is then checked, with a model’s help, for whether it says what the claim relies on. Those verdicts can be wrong — read the claim’s quote and the cited source before relying on one.
The claim verdicts this report counts are experimental: their accuracy is not yet measured.
A language model drew the claims from the paper on 2026-10-10, under version 3 of the claim definition (model deepseek-v4.1-flash, prompt extract-v3).
A second model labelled each claim on 2026-10-10 (model jev-1.13-20260917, definition revision gold-guideline-v0.2).
Cited sources are checked with a language model’s help (gpt-6-sol).
Statistics and arithmetic are recomputed, not judged.
Confidence bands are shown only for fields calibrated on our annotated test set. None is calibrated yet, so no band is shown.
What counts as a claim ›How we measure confidence ›How we check this ›
Are you an author, or do you see a problem with this page? Request a review
Data sources
Bibliographic data: Semantic Scholar, OpenAlex, Crossref.
Contains information from the Semantic Scholar Academic Graph Datasets, made available under the ODC Attribution License (ODC-BY 1.0).
Semantic Scholar’s data is described in: Kinney, R., et al. (2023). The Semantic Scholar Open Data Platform. arXiv:2301.10140.